CERN – 7 Oct 04 1 LHCb Muon Grounding Anatoli Katchouk CERN Alessandro Balla, Paolo Ciambrone, Maurizio Carletti, Giovanni Corradi, Giulietto Felici, Rosario Lenci INFN - Laboratori Nazionali di Frascati Maurizio Gatta INFN - RomaII
CERN – 7 Oct 04 2 Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct 04 3 Reference ground Safety ground Signal ground Grounding, shielding and power distribution must be handled together to avoid EMI pickup and preserve signal integrity Signal ground : on-detector electronics ground reference Reference ground : off-detector electronics ground reference Safety ground : connection to earth How many grounds ?
CERN – 7 Oct 04 4 Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct 04 5 Z1 Safety connections Detector must be connected to earth for safety reasons Chamb 1Chamb 2 Chamb 3 Chamb … Copper braid (sitting on the edge of the wall) Z2 Z3 Z4 Zstrip
CERN – 7 Oct 04 6 Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct 04 7 HV supply voltage is distributed using shielded cables Shield can not be connected to ground to both sides to avoid ground loops A resistor (~ 1-10 k) is used on the detector side to break the loop Some current will flow through the low resistance safety earth network (unavoidable) HV connections HV PS Load 1-10 k
CERN – 7 Oct 04 8 Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct 04 9 Front end on-chamber connections (I) SPBCARDIACOPB LVPS regulator board Safety ground must be connected to FC through a low impedance path (copper braid) Faraday cage is the chamber signal ground Ground connection LVPS connector CARDIAC connector Faraday cage
CERN – 7 Oct Front end on-chamber connections (II)
CERN – 7 Oct Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct Data transmission between front end electronics and DAQ system is implemented via differential electrical connections (LVDS) good immunity to common mode noise As we will use shielded cables one end is connected directly to ground while the other end could be connected to ground through a capacitor to avoid low frequency loops Front End Off-chamber connections Shielded Twisted cable Exceeding the common mode range can damage LVDS driver/receiver circuits |V A -V B | must be less than the LVDS common mode range A B
CERN – 7 Oct Grounding components Grounding HV connections Front End Off-chamber connections Safety connections LVPS connections Front End On-chamber connections
CERN – 7 Oct LVPS connections (I) Number of Chambers FEE boards x Chamber Icc x Chamber (A) Icc (A) x Station Total (A) Current M1 R419231,38264,96 927,36 R348125,52264,96 R224 11,04264,96 R ,04132,48 M2/M3 R ,38264,96 850,08 R ,52264,96 R ,12242,88 R ,4477,28 M4/M5 R ,38264,96 529,92 R348 62,76132,48 R224 62,7666,24 R112 5,5266,24 Total4217,28 Front end current requirements ( LHCb internal note) CARIOCA V DIALOG V
CERN – 7 Oct LVPS connections (II) 40 A Module N A M1R3 M2/M3-R3 M4/M5-R1 N A M4/M5-R3 N A M1R4 M2/M3-R4 M4/M5-R4 LVPS Patch panel & fuses Chamber 1 Chamber 2 Chamber n
CERN – 7 Oct Conclusions The on-chamber shielding and grounding scheme has been studied and successfully tested in several TB The proposed grounding scheme requires : FC as detector signal ground reference Floating low voltage power supply Chamber connected only to the safety ground High voltage ground broken by resistor The maximum ground voltage difference between on- detector and off-detector electronics must be less than LVDS allowed common voltage