Comparison of T1 FLAIR and T1 FSE Images Andrew Allmendinger, DO Sylvie Destian, MD
Neither I nor my immediate family members have a financial relationship with commercial organizations that may have a direct or indirect interest in the content of this presentation.
In most institutions, a sagittal T1 FLAIR sequence has replaced the T1 FSE in routine MR imaging of the lumbar spine.
Faster Provides better conspicuity of abnormalities in the cord, CSF, disc, and bone 1,2,3. 1 Erdem LO, Erdem CZ, Acikgoz B, Gundogdu S. Degenerative disc disease of the lumbar spine: a prospective comparison of fast T1-weighted fluid- attenuated inversion recovery and T1-weighted turbo spin echo MR imaging. Eur J Radiol, 2005 Aug; 55(2): Lavdas E, Vlychou M, Arikidis N, Kapsalaki E, Roka V, Fezoulidis IV. Comparison of T1 weighted fluid-attenuated inversion recovery images of the lumbar spine at 3.0 Tesla. Acta Radiol 2010 Apr; 51(3): Melhem ER, Israel DA, Eustace S, Jara H. MR of the spine with a fast T1-weighted fluid-attenuated inversion recovery sequence. AJNR 1997 Mar;18:
T1 FLAIR = T1 FSE? To determine whether there is a difference in the appearance of the marrow signal on T1 FLAIR and T1 FSE images in normal individuals To determine whether the difference is related to age.
Sagittal T1 FLAIR TE 22, TI 500, TR 2000, Echo Train 12, Freq 448, Phase 224, NEX 2, Slice thickness 4/1, Freq encoding A/P, imaging options: NPW, scan time 2:16 Sagittal T1 FSE TE minimum full, TR 650, Echo Train 2, Freq 256, Phase 192, NEX 2, Slice thickness 4/1, Freq encoding A/P, imaging options: FC, NPW, TRF, FAST, scan time 3:14
1.5T GE Signa /Excite Prospective study All patients having non contrast MRI of the lumbar spine during a 3 month period in 2009 (n=217).
Exclusion criteria: older than 75 years prior lumbar surgery predisposing medical condition HIV disease multiple myeloma osteoporosis metastatic disease renal failure severe degenerative disease Final n=93
A midline sagittal T1 FLAIR and T1 FSE image from each exam, randomly labeled “A” or “B”, were placed side by side in a power point presentation. B A
All case pairs were reviewed by 3 board certified radiologists, each with more than 5 years experience in MR imaging. The radiologists compared the two images and were asked whether marrow signal intensity = or difference perceived? T1 FLAIR or T1 FSE darker?
The marrow signal intensity appears lower on the T1 FLAIR image (A), compared to the T1 FSE image (B) on the right. A B
The marrow signal intensity appears lower on the T1 FSE image (A) compared to the T1 FLAIR image (B) on the right. B A
The marrow signal intensity on the T1 FSE image (A) and the T1 FLAIR image (B) appears equal. AB
AgeT1 FLAIR Darker T1 FSE Darker 0 to to to to to to to to Total (n=70)5416 Where a difference was perceived, the marrow signal intensity was judged darker on T1 FLAIR 77% of the time (p <.0001).
AgeT1 FLAIR Darker T1 FSE Darker 0 to to to to to to to to 7540 Total (n=35)323 Where a difference was perceived, the marrow signal intensity was judged darker on T1 FLAIR 91% of the time (p < ).
AgeT1 FLAIR Darker T1 FSE Darker 0 to to to to to to to to Total (n=87)852 Where a difference was perceived, the marrow signal intensity was judged darker on T1 FLAIR 98% of the time (p <.0001).
T1 FLAIR and T1 FSE were judged equal in 31% of the cases. When a difference was perceived, the T1 FLAIR image was judged darker in 88% of the cases. T1 FLAIR was judged darker in all age groups from
1 patient < 20 years One radiologist T1 FLAIR darker Two radiologists T1 FLAIR = T1 FSE More data needed 0-20 age group
Regardless of age, marrow signal intensity on T1 FLAIR images may appear artificially low in normal individuals. It is important to recognize this potential pitfall in order to avoid misdiagnoses and/or additional unnecessary exams.