Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact.

Slides:



Advertisements
Similar presentations
ARTIFACTS IN COMPUTED TOMOGRAPHY
Advertisements

IMAGE GENERATION IN CT.
Computed Tomography Principles
Image Reconstruction.
Image Quality Bushong Chapter 7.
Computed Tomography Physics, Instrumentation, and Imaging
Historical Development
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Image Artifacts Chapter 8 Bushong.
Computed Tomography II
Seeram Chapter 13: Single Slice Spiral - Helical CT
RAD309 Patient Dose.
Module H Computed Tomography Physics, Instrumentation, and Imaging.
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
CT Physics V.G.Wimalasena Principal School of radiography.
Computed Tomography III
Computed Tomography RAD309
tomos = slice, graphein = to write
Seeram Chapter 5: Data Acquisition in CT
Chapter 2 Stewart C. Bushong
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
LEC ( 2 ) RAD 323. Reconstruction techniques dates back to (1917), when scientist (Radon) developed mathematical solutions to the problem of reconstructing.
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Computer Tomography By Moustafa M. Mohamed. Introduction to Medical Imaging Uses of medical imaging Obtain information about internal body organs or the.
Computed Tomography Q & A
Medical Image Analysis Image Reconstruction Figures come from the textbook: Medical Image Analysis, by Atam P. Dhawan, IEEE Press, 2003.
Fan Yang, M.D., Ph.D. Ichiro Ikuta, M.D., M.M.Sc. Head Holder Image Artifact: Shadows in the midst of Light, and How We Fixed It. 34 Maple Street Norwalk,
Computed Tomography Physics, Instrumentation, and Imaging
CT Instrumentation and X-ray system
Fig. 3 shows how a dose distribution that is initially relatively sharp becomes blurred with time, until after 16 hours, virtually no useful information.
Module A Computed Tomography Physics, Instrumentation, and Imaging.
Module C Computed Tomography Physics, Instrumentation, and Imaging.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
Part No...., Module No....Lesson No
Quality Assurance.
Monday Case of the Day History: During acceptance testing of a new general radiographic room, tube output measurements differed between the small and large.
History: A 61-year-old man with suspected thoracic aneurysm underwent ECG-gated chest CT angiogram. He was instructed to hold his breath during the exam.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Thursday Case of the Day The arrow indicates a region of high apparent FDG uptake posterior to the bladder. What is the cause of this artifact? a) attenuation.
Thursday Case of the Day The likely cause of the artifact is: A. Patient Motion B. Improper Cupping Correction C.Off-Focus Radiation combined with heel.
Optimizing The Image RVT: Chapter 7
Computed Tomography The images in the following presentation follow the “fair use” rules of the U.S. Copyright law.
Tuesday Case of the Day What is the best solution to mitigate the artifact? A. There is no artifact, patient is wearing compression garment(s) which are.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
Sunday Case of the Day How does the presence of the object impact patient dose? A. Increases by 20% - 50% B. Increases by 3% - 5% C. No Change D. Decreases.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Page 1 Physics Case of the Day - Thursday During routine QA testing of a computed tomography scanner, the physicist recorded the data below for the tube.
Single Slice Spiral - Helical CT
Computed Tomography Basics
Computed Tomography Data Acquisition
Learning Objectives By the end of this lesson you should…
CT PRINCIPLES AYMAN OSAMA.
Sunday Case of the Day Physics
Computed Tomography Image Reconstruction
Sunday Case of the Day Physics
Monday Case of the Day Physics
Strategies for Reducing Radiation Exposure From Multidetector Computed Tomography in the Acute Care Setting  Aaron Sodickson, MD, PhD  Canadian Association.
Monday Case of the Day Physics
Tuesday Physics Case of the Day
Wednesday Case of the Day
Tuesday Case of the Day Physics
David J. Brenner, Maria A. Georgsson  Gastroenterology 
Thursday Case of the Day
Intrinsic Uniformity QC Setup
Type 1 pedicle marrow signal intensity changes associated with degenerative facet disease. Type 1 pedicle marrow signal intensity changes associated with.
Pál Maurovich-Horvat et al. JIMG 2010;3:
Presentation transcript:

Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact which mimic disease in brain matter. Authors: Mark Supanich PhD, Donovan Bakalyar PhD, Donald Peck PhD Henry Ford Health System, Detroit MI Physics Figure 1: Images from a standard helical head study at left shows hypointense artifactual regions (arrows) that mimic disease. At right is a scan of the Quality Assurance phantom using the standard protocol showing the diffuse, hypointense, blurry artifact. The Source of the Artifact is: A. Debris in the CT scanner gantry in front of the detectors B. Air bubbles in the cooling oil of the X-ray tube assembly C. Faulty calibration of a block of detectors D. Deposits on the filter in the X-ray tube assembly

Findings: A Physicist was able to reproduce the diffuse, blurry artifact when scanning a Quality Assurance (QA) phantom using the standard head protocols (Figure 2). The artifact was also observed in previous tech QA scans after windowing and leveling. No foreign objects or stains were noted on the gantry or on the patient bed. Hypointense regions were noted not to be entirely contiguous from slice to slice (Figure 2). Figure 2: Adjacent slices of QA phantom showing diffuse hypointense regions that differ in appearance in each slice.

Diagnosis: B. Air bubbles in the cooling oil of the X-ray tube assembly

Discussion: As photons exit the tube assembly, they pass through cooling oil. If the photons pass through air bubbles located in the cooling oil, they are less likely to be attenuated than if they passed only through oil. The decreased attenuation in each CT view leads to a reconstructed pattern of hypointense regions. Figure 3: Diagram from reference 1 showing that photons pass through cooling oil as they exit the x-ray tube housing.

Discussion: Small air bubbles in the cooling oil are hypoattenuating and are geometrically magnified in an attenuation profile as shown in Figure 4. Figure 4. Cutaway magnified view of anode with air bubbles in surrounding oil shown at top. The photons exit the tube assembly and encounter an attenuating object. The combination of the hypoattenuating air bubbles and the object result in an attenuation profile with a decrease in attenuation at certain projections due to air bubbles. Air bubbles move through the liquid as the gantry and the tube assembly rotate, resulting in different attenuation profiles at each view. (Stationary bubbles would result in thick, fuzzy circles.) This view-to-view movement results in the non-circular appearance of the blotchy patterns which, in turn, masks the behavior normally expected from CT artifacts.

Discussion: The vendor’s service engineer was contacted to perform maintenance on the tube to remove the air bubbles. Figure 5 shows the QC phantom imaged with the air bubbles present (left) and after servicing of the tube (right). Figure 5: Quality Assurance phantom imaged and displayed at the same WW/WL with air bubbles present in cooling oil (left) and with no bubbles present (right).

Discussion: Choice A: Debris in front of detector would result in artifacts in the images that would be smaller and sharper than observed due to minimal geometric magnification. Choice C: Faulty calibration of a block of detectors could explain the existence of one circular hypointense region at isocenter as shown in Figure 6 from reference 2, but the multiple areas of hypointensity and their non-circular shape rule out this answer. Choice D: Deposits on the filter in the assembly would show the effects of geometric blurring, but would lead to regions of hyperintensity due to the increased attenuation of the photons encountering the deposits. The resulting artifacts would be circular, as the location of the attenuation change would be fixed in the detector frame of reference. Figure 6: Regions of hypointensity due to the faulty calibration of a block of detectors due to a foreign substance on the collimator assembly.. The location of the artifact varies from scan to scan due to patient positioning. Adapted from reference 2.

References/Bibliography: 1.Schardt et al. “New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology,” Med Phys 2004; 31(9): Cody, D. et al. “Multi-detector row CT artifacts that mimic disease,” Radiology 2005;236: