Tuesday Case of the Day - Physics Zheng Feng Lu, Ph.D., Stephen Thomas, M.D. Department of Radiology, University of Chicago, Chicago, IL Question: What.

Slides:



Advertisements
Similar presentations
Chapter3 Pulse-Echo Ultrasound Instrumentation
Advertisements

Foundations of Medical Ultrasonic Imaging

01: Introduction to Ultrasound George David, M.S. Associate Professor of Radiology.
Volume I Companion Presentation Frank R. Miele Pegasus Lectures, Inc.
Resident Physics Lectures
Hospital Physics Group
Ultrasound Physics Have no fear Presentation by Alexis Palley MD
Basic Physics of Ultrasound Beth Baughman DuPree M.D. FACS Medical Director Breast Health Program Holy Redeemer Health System 2011.
Basic Physics of Ultrasound
ECE 501 Introduction to BME
Seismic reflection Seismic reflection profiling basically same principle as echo sounding But lower frequency used for greater subbottom penetration Trade.
Sonar Chapter 9. History Sound Navigation And Ranging (SONAR) developed during WW II –Sound pulses emitted reflected off metal objects with characteristic.
Echoes. Reflection  The reflection of sound waves applies to frequencies above human hearing. 20 kHz is human maximum20 kHz is human maximum Bats can.
Ultrasound Medical Imaging Imaging Science Fundamentals.
Ultrasound Imaging Atam Dhawan.
Ultrasound.
 Some animals such as bats, use ultrasound waves to detect obstacles and objects around them.  Ultrasounds are reflected of surfaces or objects and.
Welcome.
GENERATING AND DETECTING OF ULTRASOUND
Ultrasound – Physics & Advances
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Development of an Ultrasound Lab Laura Wade April 4 th Z.
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Ultrasound Physics Sound is a mechanical, longitudinal wave that travels in a straight line Sound requires a medium through which to travel Ultrasound.
Ultrasonography.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Principles of Magnetic Resonance
Doppler echocardiography & Magnetic Resonance Imaging.
Display of Motion & Doppler Ultrasound
George David Associate Professor Ultrasound Physics 04: Scanner ‘97.
1 March 2013 PGI Chandigarh Dr Sarvesh Natani.
Building Three-Dimensional Images Using a Time-Reversal Chaotic Cavity
US FUNDAMENTALS: B-MODE AND DOPPLER MACHINE SETTINGS.
Medical Imaging Ultrasound Edwin L. Dove 1412 SC
Display of Motion & Doppler Ultrasound
Review of Ultrasonic Imaging
Sarah Gillies Ultrasound Sarah Gillies
Ultrasound Imaging Basic Principle: sound wave pulse emitted at >20 kHZ, some reflected, and some transmitted. Reflection from single signal – A-scan Reproduced.
Doppler Ultrasound Dr Mohamed El Safwany, MD.. Introduction The Doppler Effect refers to the change in frequency that results when either the detector/observer.
Resident Physics Lectures
The Universal Wave Equation
Saudi Board of Radiology: Physics Refresher Course Kostas Chantziantoniou, MSc 2, DABR Head, Imaging Physics Section King Faisal Specialist Hospital &
Ultrasound Physics Reflections & Attenuation ‘97.
Ultrasonic İmaging.
BIOE 220/RAD 220 REVIEW SESSION 3 February 6, 2011.
Ultrasound Basis Michel Slama Amiens.
Piezoelectric Effect  Sound waves striking a PZ material produce an electrical signal  Can be used to detect sound (and echoes)!
Artifacts Ultrasound Physics George David, M.S.
Supervisor : Evgeniy Kuksin Hanna Abo Hanna Alaa Mozalbat 32CH BEAM-FORMER FOR MEDICAL ULTRASOUND SCANNER.
Performance Issues in Doppler Ultrasound 1. 2 Fundamental Tradeoffs In pulsed modes (PW and color), maximum velocity without aliasing is In pulsed modes,
TISSUE HARMONIC IMAGING (THI)
TISSUE HARMONIC IMAGING (THI)
TISSUE HARMONIC IMAGING NUR FASHIHA BINTI AZMAN A DIAGNOSTIC IMAGING AND RADIOTHRAPY /2.
NUR INANI BINTI MOHD RUZLAN A Tissue harmonic imaging is a technique in ultrasonography that provides images of better quality as compared to conventional.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Waves (I) S3 Physics
Ultrasound Physics Image Formation ‘97.
Sound in medicine Lect.10.
Reflections & Attenuation
Visit for more Learning Resources
Ultrasound Physics Image Formation ‘97.
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Basic Physics of Ultrasound
PARAMETRIC SUB-BOTTOM PROFILER: A NEW APPROACH FOR AN OLD PROBLEM
What we will do today: Carry out calculations involving ultrasound.
Medical Ultrasound: Image Optimisation
ArtIfacts in usg Artifacts are part n parcel of any imagin modality .Us artifact can be avoidable and may be due to improper scanning technique. Recognition.
Fundamentals of Sonographic Wave Propagation and New Technologies
Chapter 12 Image Artifacts
Presentation transcript:

Tuesday Case of the Day - Physics Zheng Feng Lu, Ph.D., Stephen Thomas, M.D. Department of Radiology, University of Chicago, Chicago, IL Question: What is the cause of the echogenic band (Arrow) obscuring part of the urinary bladder (image A)? a)Reverberation caused by multiple reflections between two strong reflectors. b)Range ambiguity caused by concurrent high pulse repetition frequency and large field depth. c)Speed error that occurs when sound wave travels through substances of different sound speed. d)Mirror image that duplicates a structure on the opposite side of a specular reflector. Image A: Frame rate 3Hz with multiple focal zones Image B: Frame rate 26Hz with single focal zone History: 50-year-old male underwent sonography of the urinary bladder for hematuria. Technique: Curvilinear array transducer C1-5 Mhz with a nominal frequency 4.0 MHz and a 15cm depth of field. Images A & B: Transverse images of the urinary bladder.

Diagnosis: The correct answer is (b), range ambiguity (artifact) caused by concurrent high pulse repetition frequency and large field depth.

Discussion A B-mode ultrasound image is formed by emitting a pulse and collecting the resultant echo signal along the insonified path. After each transmit pulse, sufficient time must be allocated for the echoes to return using the following range equation: D = ct/2 Where D is the depth, t is the delay time between pulse emission and echo reception and c is the sound propagation speed. Echoes from a deeper depth takes a longer time to return. When echoes from deep structure from the previous transmit pulses are received simultaneously as the echoes from shallow structure insonified by the present pulse, the “range ambiguity” artifact occurs. This typically happens when multiple focal zones are set for a large field-of-view because multiple pulses are needed along the same beam line while each pulse can only focus at one particular depth. Consequently, very high pulse repetition frequency is typically employed for multiple focal zone settings, which causes the “range ambiguity” artifact especially when the field-of-view is large. This may become more obvious when the object has low attenuation such as a urinary bladder. By reducing the number of focal zones to one, the pulse repetition frequency is reduced; thus the range ambiguity artifact is removed. Clinically, range ambiguity artifact is important to recognize and correct as it can obscure important features such as tissue planes and organ pathology.

References RT O’Brien, JA Zagzebski and FA Delaney, “Ultrasound corner: range ambiguity artifact”, Veterinary Radiology & Ultrasound, Vol. 42, No. 6, 2001, pp 542 – 545. A Goldstein, “Range ambiguities in real-time ultrasound”, J Clin Ultrasound, Vol 9, No. 2, 1981, pp83-90.