Substantial signal dropout was observed in arterial spin label (ASL) images, acquired at 3T in a young female patient three-months after mild traumatic.

Slides:



Advertisements
Similar presentations
Magnetic Resonance Imaging Lorenz Mitschang Physikalisch-Technische Bundesanstalt, 23 rd February 2009 I. Basic Concepts.
Advertisements

Chapter 7: Gradient Echo Imaging Methods
MRI Phillip W Patton, Ph.D..
Richard Wise FMRI Director +44(0)
Basis of the BOLD signal
Institute for Biomedical Engineering EXCITE Afternoon Hands-On MRI Sessions: fMRI & DTI.
The importance of MRI, a few numbers  MRI units worldwide in 2003  75 millions scans per year performed  Constant need for over 1000 MRI technologists.
By: Tracy Ssali Medical Biophysics 3970z April 4 th 2012 Supervisors: Keith St. Lawrence, PhD Udunna Anazodo, PhD candidate.
Chapter 14: Artifacts Mark D. Herbst, MD, PhD. Artifact Something on the image that does not represent something in the patient –Many causes Equipment.
Magnetic Resonance Imaging Magnetic field gradients.
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
Principles of MRI. Some terms: –Nuclear Magnetic Resonance (NMR) quantum property of protons energy absorbed when precession frequency matches radio frequency.
Functional Brain Signal Processing: EEG & fMRI Lesson 12 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Diffusion Tensor MRI And Fiber Tacking Presented By: Eng. Inas Yassine.
Realigning and Unwarping MfD
Relaxation Exponential time constants T1 T2 T2*
FMRI: Biological Basis and Experiment Design Lecture 15: CBF and Localization II CBF techniques Big veins and big voxels 1 light year = 5,913,000,000,000.
Psy 8960, Fall ‘06 Spin Echo, high field1 High Field and Spin Echo: the Minnesota story Preview: fMRI decision tree Why image at 7T?
Psy 8960, Fall ‘06 Fieldmaps1 Fieldmaps and distortion What is a fieldmap? How can we predict distortion? How can we correct distortion?
Arterial Spin Labeling at 7T - Double Edged Sword
FMRI: Biological Basis and Experiment Design Lecture 7: Gradients and k-space FFT examples –Sampling and aliasing Gradient Gradient echo K-space
Perfusion Imaging S. Lalith Talagala, Ph. D
Perfusion Imaging Lalith Talagala, Ph. D
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
Psy 8960, Spring ‘07 Gradient and Spin Echoes1 Echoes, gradients and dephasing Gradient echoSpin echo.
QUIPPS II 1.Saturate volume of interest RF G SS. QUIPPS II 1.Saturate volume of interest 2.Tag (invert) incoming blood RF G SS.
MRI Image Formation Karla Miller FMRIB Physics Group.
Introduction of Medical Imaging Chun Yuan. Organization of the Course 8 Lectures (1.5 hours per lecture) – Introduction of medical imaging and MRI – Basic.
What are we measuring in fMRI? Caroline Catmur Jack Kelly.
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
Magnetic Resonance Imaging
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Medical Imaging Systems: MRI Image Formation
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Pulse Sequences Types of Pulse Sequences: Functional Techniques
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
In Burst experiments, the read gradient is much larger then the other imaging gradients, hence, only translation along this gradient direction will cause.
Topics in Magnetic Resonance Angiography Research and Some Potential New Directions Dennis L. Parker et al.
MR Compatibility Torques Artifacts Not Compatible Yes Yes MR Safe No
Contrast Mechanisms in MRI Introduction to Cardiovascular Engineering Michael Jay Schillaci, PhD Managing Director, Physicist Tuesday, September 16 th,
Perfusion MRI in GSK Study
Contrast Mechanism and Pulse Sequences
Statistical Parametric Mapping
An Investigation of Perfusion in Stroke Survivors Kathleen Brumm 1, Thomas Liu 2, Joanna Perthen 2, Frank Haist 2, Tracy Love 1, 2, 3 1 SDSU/ UCSD Joint.
FMRI – Week 4 – Contrast Scott Huettel, Duke University MR Contrast FMRI Graduate Course (NBIO 381, PSY 362) Dr. Scott Huettel, Course Director.
Conclusions Simulated fMRI phantoms with real motion and realistic susceptibility artifacts have been generated and tested using SPM2. Image distortion.
Basic MRI Chapter 1 Lecture. Introduction MRI uses radio waves and a magnetic field to make images MRI uses radio waves and a magnetic field to make images.
1. Introduction Magnetic resonance spectroscopy (NMR) and imaging (MRI) are safe and highly versatile modalities, based on fundamental concepts in quantum.
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
Origin of Negative BOLD fMRI Signals
MRI: Contrast Mechanisms and Pulse Sequences
Introduction Material and Methods Result 2: Evolution of the lesion Result 1: DTI can map the injured spinal cord Conclusion J. Cohen-Adad 1, H. Leblond.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Charged particle. Moving charge = current Associated magnetic field - B.
Receive Coil Arrays and Parallel Imaging for fMRI of the Human Brain
Thursday Case of the Day What is the source of the artifact in the upper right-hand quadrant of the image? History: An AP view of the pelvis was acquired.
Parameters which can be “optimized” Functional Contrast Image signal to noise Hemodynamic Specificity Image quality (warping, dropout) Speed Resolution.
FMRI data acquisition.
Sunday Case of the Day Physics
Jody Culham Brain and Mind Institute Department of Psychology
MRI Physics in a Nutshell Christian Schwarzbauer
Monday Case of the Day Physics
Sunday Case of the Day Physics (Case 1: MR)
Monday Case of the Day Physics
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
Tuesday Case of the Day Physics
Pulse sequence diagrams show the benefits of parallel imaging for DWI
Presentation transcript:

Substantial signal dropout was observed in arterial spin label (ASL) images, acquired at 3T in a young female patient three-months after mild traumatic brain injury Tuesday Case of the Day The likely cause of the artifact is: A. Malfunctioning receive coil B. Transmit (B 1 ) inhomogeneity C. Susceptibility-related T 2 * dephasing D. Magnetic field (B 0 ) inhomogeneity History: Authors : Dustin K. Ragan, PhD 1 and Charles E. Willis, PhD, DABR 2 1 Department of Pediatrics, Washington University, St. Louis, MO 2 U.T. M. D. Anderson Cancer Center, Houston, TX Physics Figure 1. Examples of the artifact (red arrows) which was only observed in this one sequence and this one patient Figure 2. No artifacts are displayed on the DTI images

Findings: Anatomical and DTI images did not display signal loss Figure 4: DTI image without artifact 1. The same coil was used to acquire all images, but signal loss is only visible in one sequence 2. DTI images are also typically acquired with an EPI-readout, which would also be sensitivity signal dropout from magnetic susceptibility in the brain 3. Artifact is located at the periphery of the brain Figure 3: ASL image with artifact Figure 5: T1-weighted anatomical image

Diagnosis: D. Magnetic field inhomogeneity

Coil problems would manifest across all of the acquired images. Therefore, choice A. Malfunctioning receive coil is incorrect. Transmit inhomogeneity typically produces shading on the interior of the images, not at the periphery. Therefore, choice B. Transmit (B 1 ) inhomogeneity is incorrect. The patient was wearing a shirt with iron-based glitter up to the shoulder. This distorted the position of the labeling pulses, causing them to saturate brain tissue instead of blood. (Magnetic field inhomogeneties in the brain can distort both ASL and DTI, as in the figure, but only ASL is sensitive to the neck) Susceptibility-related dephasing is common in EPI acquisitions, which is used in both ASL and DTI acquisitions. Because only the ASL images were affected, choice C. Susceptibility-related dephasing is incorrect. Also, large susceptibility artifacts are relatively rare in mild TBI, so it is inconsistent with the patient’s presentation. Discussion: Pseudocontinous ASL uses a preparatory train of RF pulses positioned at the neck along with (usually) an EPI readout to acquire images of cerebral blood flow Figure 6: Susceptibility artifacts present on both ASL and DTI images

Discussion: The gradient strengths used in ASL labeling are relatively weak, which amplifies the distorting effect of field inhomogeneities Physical position Magnetic field offset Strong gradient Weak gradient ΔB=G  x A magnetic field shift produces a much larger distortion in the presence of a weak gradient than a strong one Gradient strengths used slice selection are around ~40 mT/m; those used in labeling are ~10 mT/m As a result, the ASL labeling pulses are highly susceptible to position shifts RF pulses are not localized and can unintentionally affect the entire sensitive volume of the coil

References/Bibliography: Jahanian H, et al. B 0 field inhomogeneity considerations in pseudo-continuous arterial spin labelng (pCASL): effects on tagging efficiency and correction strategy. NMR in Biomedicine 24: Haacke M., et al. Magnetic Field Inhomogeneity Effects and T 2 * Dephasing, In Magnetic Resonance Imaging: Physical Principles and Sequence Design. p New York, John Wiley & Sons.