F. Pepe Observatoire de Genève Optical astronomical spectroscopy at the VLT (Part 2)

Slides:



Advertisements
Similar presentations
HIRES Technical concept and design E. Oliva, HIRES meeting, Brera (Milan, Italy)1.
Advertisements

1 ATST Imager and Slit Viewer Optics Ming Liang. 2 Optical layout of the telescope, relay optics, beam reducer and imager. Optical Layouts.
Copyright © 2009 Pearson Education, Inc. Chapter 35 Diffraction and Polarization.
The waves spread out from the opening!
Wavelength Selection Devices
Spectral Resolution and Spectrometers
Echelle Spectroscopy Dr Ray Stathakis, AAO. What is it? n Echelle spectroscopy is used to observe single objects at high spectral detail. n The spectrum.
Diffraction of Light Waves
Spectral Resolution and Spectrometers A Brief Guide to Understanding and Obtaining the Proper Resolution of the 785 Raman System.
Spectroscopic Data ASTR 3010 Lecture 16 Textbook Ch. 11.
UVES The Ultraviolet & Visible Echelle Spectrograph John Pritchard ESO p2pp3 Workshop: UVES Overview | | John Pritchard.
Instrumentation Concepts Ground-based Optical Telescopes
AURA New Initiatives Office S.C. Barden, M. Liang, K.H. Hinkle, C.F.W. Harmer, R.R. Joyce (NOAO/NIO) September 17, 2001 Instrumentation Concepts for the.
Grism Spectroscopy with FLITECAM Erin C. Smith (UCLA) Ian S. McLean (UCLA)
Wide-field, triple spectrograph with R=5000 for a fast 22 m telescope Roger Angel, Steward Observatory 1 st draft, December 4, 2002 Summary This wide-field,
Interference and Diffraction
Spectrographs. Spectral Resolution d 1 2 Consider two monochromatic beams They will just be resolved when they have a wavelength separation of d Resolving.
Astronomical Spectroscopy
WAVES MEDIUM VIBRATES PERPENDICULARLY TO THE WAVE DIRECTION IF f IS THE WAVE FREQUENCE AND λ IS THE WAVELEGTH THEN c, THE WAVE VELOCITY, IS GIVEN BY: c.
CEE 540 Spring Term 2012 go over syllabus and course requirements project spectra lab demo: parts of a spectrograph image in focal plane hg spectral emission.
4-1 Chap. 7 (Optical Instruments), Chap. 8 (Optical Atomic Spectroscopy) General design of optical instruments Sources of radiation Selection of wavelength.
Grazing-incidence design and others L. Poletto Istituto Nazionale per la Fisica della Materia (INFM) Department of Electronics and Informatics - Padova.
Optical characteristics of the EUV spectrometer (EUS) for SOLO L. Poletto, G. Tondello Istituto Nazionale per la Fisica della Materia (INFM) Department.
1 Components of Optical Instruments, Cont… Lecture 7.
Astronomical Instrumentation Often, astronomers use additional optics between the telescope optics and their detectors. This is called the instrumentation.
THE LHIRES-III SPECTROGRAPH © C2PU, Observatoire de la Cote d’Azur, Université de Nice Sophia-Antipolis Jean-Pierre Rivet CNRS, OCA, Dept. Lagrange
5 Components Common to All Optical Spectrometers Source Transparent Sample Holder Wavelength Selector Radiation Detector Signal Processor and Readout.
Visual Angle How large an object appears, and how much detail we can see on it, depends on the size of the image it makes on the retina. This, in turns,
Spectroscopy Spectral analysis is probably the most important method for learning about the physics of the astronomical sources Simplest method to get.
B.Delabre November 2003ANGRA DOS REIS - BRAZIL ESO 2 nd GENERATION INSTRUMENTATION – OPTICAL DESIGNS ESO VLT SECOND GENERATION INSTRUMENTATION Optical.
Spectrographs. Literature: Astronomical Optics, Daniel Schneider Astronomical Observations, Gordon Walker Stellar Photospheres, David Gray.
2009 Aug 20 — SAC update WFOS/MOBIE1 WFOS/Multi-Object Broadband Imaging Echellette MOBIE Team, to date: PI / optical designer: Rebecca Bernstein Project.
Spectroscopy principles
14 October Observational Astronomy SPECTROSCOPY and spectrometers Kitchin, pp
SPECTROSCOPY From Science to Science System view Fundamentals of Echelle Spectroscopy and other components End - to - End operations Verification Management.
Spectroscopic Observations (Massey & Hanson 2011, arXiv v2.pdf) Examples of Spectrographs Spectroscopy with CCDs Data Reduction and Calibration.
High Resolution Echelle Spectrograph for Chinese Weihai 1m Telescope. Leiwang, Yongtian Zhu, Zhongwen Hu Nanjing institute of Astronomical Optics Technology.
Near Infrared Spectro-polarimeter (NIRSP) Conceptual Design Don Mickey Jeff Kuhn Haosheng Lin.
18 October Observational Astronomy SPECTROSCOPY and spectrometers Kitchin, pp
Spectrographs. Literature: Astronomical Optics, Daniel Schroeder Astronomical Observations, Gordon Walker Stellar Photospheres, David Gray.
Astronomical Spectroscopy Notes from Richard Gray, Appalachian State, and D. J. Schroeder 1974 in “Methods of Experimental Physics, Vol. 12-Part A Optical.
Multi-slit spectroscopy In sky-noise dominated conditions (most interesting!) the use of slits is essential: eg: Faint object, extra-galactic, surveys:
An IFU for IFOSC on IUCAA 2m Telescope
JGR 19 Apr Basics of Spectroscopy Gordon Robertson (University of Sydney)
NORDFORSK Summer School, La Palma, June-July 2006 NOT: Telescope and Instrumentation Michal I. Andersen & Heidi Korhonen Astrophysikalisches Institut Potsdam.
The Study of IFU for the Li Jiang 2.4m Telescope ZHANG Jujia 张居甲 Yun Nan Astronomical Observatory. CAS Sino-French IFU Workshop Nov Li Jiang.
The waves spread out from the opening!
Neon school 2006 ESO-Garching M. Dennefeld NEON Archive School 2006 Introduction to Spectroscopic Techniques (low dispersion) M. Dennefeld (IAP-Paris)
Diffraction by N-slits. Optical disturbance due to N slits.
1.Stable radiation source 2.Wavelength selector 3.Transparent sample holder: cells/curvettes made of suitable material (Table 7- 2) 4.Radiation detector.
Optical characteristics of the EUV spectrometer for the grazing-incidence region L. Poletto, G. Tondello Istituto Nazionale per la Fisica della Materia.
Spectrographs. Literature: Astronomical Optics, Daniel Schroeder Astronomical Observations, Gordon Walker Stellar Photospheres, David Gray.
Science with Giant Telescopes - Jun 15-18, Instrument Concepts InstrumentFunction range (microns) ResolutionFOV GMACSOptical Multi-Object Spectrometer.
Visible Spectro-polarimeter (ViSP) Conceptual Design David Elmore HAO/NCAR
Laser Spectroscopy/SJX Chap. 4 Components of Spectroscopic Instruments 1 In this chapter we discuss basic spectroscopic instruments and techniques employed.
Optical astronomical spectroscopy at the VLT (Part 1) F. Pepe Observatoire de Genève.
2. Diffraction Grating Basic principles: Diffraction
Spectrometer The instrument used for the astronomers MinGyu Kim
Astronomical Spectroscopic Techniques. Contents 1.Optics (1): Stops, Pupils, Field Optics and Cameras 2.Basic Electromagnetics –Math –Maxwell's equations.
Overview: Astronomical Spectroscopy Basic Spectrograph Optics Objective Prism Spectrographs Slit Spectrographs.
Sept 16, 2005Ast 920 Meeting 21 Astronomy 920 Commissioning of the Prime Focus Imaging Spectrograph Meeting 2 Intro; VPH Grating Spectroscopy Ken Nordsieck.
Astronomical Spectroscopic Techniques
Diffraction gratings By M. Ravi Kiran.
ESAC 2017 JWST Workshop JWST User Documentation Hands on experience
THE LHIRES-III SPECTROGRAPH
An IFU slicer spectrometer for SNAP
Introduction to Spectroscopy
Chapter 35-Diffraction Chapter 35 opener. Parallel coherent light from a laser, which acts as nearly a point source, illuminates these shears. Instead.
PRISMS – one of the e.g. of optical instrumentation
The waves spread out from the opening!
Presentation transcript:

F. Pepe Observatoire de Genève Optical astronomical spectroscopy at the VLT (Part 2)

Course outline PART 1 - Principles of spectroscopy – Fundamental parameters – Overview of spectrometry methods PART 2 – ‘Modern’ spectrographs – ‘Simple’ spectroimager FORS – Echelle spectrographs UVES CRIRES PART 3 - Spectroscopy on the VLTs – Multi-Object spectrographs (MOS) and Intergral-Field Units (IFU) and spectro-imagers Giraffe VIMOS Sinfoni – Future instruments X-shooter MUSE ESPRESSO

Resolving power and resolution The objective translates angles into positions on the detector. Each position (pixel) of the detector ‘sees’ a given angle of the parallel (collimated) beam The collimated beam is never perfectly parallel, because either of the limited diameter of the beam, which produces diffraction  = 1.22  /D 1, or because of the finite slit, which produces and angular divergence  = s/f 1. The angular divergence is translated into a distance  = f 2  or  = f 2  on the CCD. This means that over this distances the wavelengths are mixed (cannot be separated angularly.

Resolving power and resolution Resolving power is the maximum spectral resolution which can be reached if the slit s = 0 and the angular divergence is limited by diffraction arising from the limited beam diameter. For a given Dispersion D we get the resolving power: Spectral resolution is the effective spectral resolution which is finally reached when assuming a finite slit s. For a given Dispersion D we get the spectral resolution:

Example of spectrographs Basic Parameters Telescope diameter: D T Source/seeing/slit: s sky Collimated beam of the spectrograph: D 1 Other parameters: Telescope focal length: f T Telescope F-number (focal ratio): F T = F = f T /D T Physical slit/fiber width: s = f T x s sky Collimator focal length: f 1 Objective focal length: f 2

Grating spectrograph Camera Collimator Focal ratios defined as F i = f i / D i

Example of simple spectrographs Spectral resolution: where

Example of ‘simple’ spectrographs Telescope diameter: D T = 1.2 m Source/seeing/slit: s sky = 1 arcsec Collimated beam of the spectrograph: D 1 = 75 mm

Example of ‘simple’ spectrographs With prism: BK7 (normal glass), t = 50 mm With grism: m = 1,  = 150 gr/mm, cos 

FORS: Filter mode

FORS: Grism mode Grisms from 150 gr/mm to 1400 gr/mm Spectroscopic modes – ‘slitless’ MOS mode (R is given by seeing) – Mask with up to 9 ‘long slits’ of 0.3” x 6.8’ – Up to 19 movable ‘slitlets’ of 0.3” x 22.5” – MOS-MXU: Laser-cut masks (any format) Spectral resolution depends on grism dispersion and slit width in dispersion direction.

FORS: Grism mode Movable slilets (localized sourcess in ‘crowded fields) Long slit (extended sources, 1-D image + spectrum)

FORS: MOS mode

This image shows the first spectrum obtained of the comet 1995 Q1. The spectral coverage extends from about 3650 Angstrom (left) to 4100 Angstrom (right) in the violet region. The spectrograph slit was oriented along the main tail at position angle ~ 145 degrees. The total slit length was 5.6 arcminutes. The spectrum is typical for a comet at Comet Bradfield's current distance from the Sun (0.5 A.U., or about 75 million kilometres).

FORS: Spectral resolution Telescope diameter: D T = 8.2 m Source/seeing/slit: s sky = 0.3 arcsec Collimated beam : D 1 = 125 mm With grism: m = 1,  = 1400 gr/mm, cos 

Example of ‘simple’ spectrographs With single prisms resolution remains small Possibility to increase dispersion by adding up several prisms With gratings or grisms, the spectral resolution is up to several 1000 but still modest. Other ‘tricks’ required

Echelle grating Increase dispersion of the ordinary greating by increasing the difference between entrance and exit angle The use in Littrow condition will make the mounting symmetric and maximize  + , since a is set equal to  :  =  Use in blaze condition for maximum efficiency: Groove angle  = 

Echelle grating From the grating equation we derive the dispersion law of an echelle grating: Example: R = 4 grating:

Characteristics of Echelle grating High dispersion -> high resolution Spectrum becomes VERY long and efficiency drops because of blaze function Alternative: use at high order, e.g. m = 100 (  = 30 gr/mm) -> Orders overlap spacially and must be separated (filtering, pre-dispersion or cross-dispersion) Free spectral range: Numbers of orders for full ‘octave’:

Multiple orders Many orders to cover desired  Free spectral range F  /m Orders lie on top of each other: (m) = (n)  (n/m) Solution: – use narrow passband filter to isolate one order at a time – cross-disperse to fill detector with many orders at once Cross dispersion may use prisms or low dispersion grating Blaze condition fulfilled

Monochromatic image of the slit Main dispersion (echelle grating) -> Cross dispersion (prism, grism, grating) ->

UVES: Spectral format

UVES: Order separation

UVES: Spectral resolution Telescope diameter: D T = 8.2 m Source/seeing/slit: s sky = 0.3 arcsec Collimated beam of the spectrograph: D 1 = 200 mm Echelle grating: R4 -> tan  = 4, D 1 = 200 mm

UVES: Sampling and resolution

UVES: Image slicer ‘trick’ Image slicer Long slit Increase of efficiency at give R, or vice versa

UVES params

UVES: Efficiency

Same as UVES, but: For IR ( = 1 – 5.4 µm) Can use adaptive optics Cryogenics needed No cross-disperser. Only 1 order at a time! -> Pre-disperser

AO

CRIRES: Characteristics

CRIRES: Spectral format

CRIRES: Long slit

Peculiarities of the IR domain: Detector technology (no Si-CCDs, behavior less ‘ideal’ in terms of noise, flat-field stability, etc.) Thermal and sky back-ground -> remove e.g. by ‘nodding’ over long slit and put instrument ‘cold’ Atmospheric absorption and emission features Advantage: Use of adaptive optics to reduce slit width.