Journal Club Postmarket Surveillance of Point-of- Care Glucose Meters through Analysis of Electronic Medical Records L.F. Schroeder, D. Giacherio, R. Gianchandani,

Slides:



Advertisements
Similar presentations
May F. Mo FDA/Industry Statistics Workshop
Advertisements

Independent Project HCAI 5221 Evaluation of Point of Care Bedside Glucose Monitors for Use in a Specialty & Transplant Hospital Mary A. Weidner, BS, MT(ASCP)
Regulatory Clinical Trials Clinical Trials. Clinical Trials Definition: research studies to find ways to improve health Definition: research studies to.
University of Minnesota – School of Nursing Spring Research Day Glycemic Control of Critically Ill Patients Lynn Jensen, RN; Jessica Swearingen, BCPS,
Errors in Chemical Analyses: Assessing the Quality of Results
Alternative Calibration Strategies for the Clinical Laboratory: Application to Nortriptyline Therapeutic Drug Monitoring M.T. Olson, A. Breaud, R. Harlan,
Harmonization of Measurement Results of the Alcohol Biomarker Carbohydrate- Deficient Transferrin by Use of the Toolbox of Technical Procedures of the.
MODELING METHOD Glucose-Insulin System Model CLINICAL DATA DATA: Are taken from the SPRINT [3] TGC cohort totalling 393 patients and ~40,000 patient hours.
The Leader in Point-of-Care Informatics MEDICAL AUTOMATION SYSTEMS © 2006 Medical Automation Systems RALS-TGCM Web-based Graphical Reports for Glycemic.
Glycemic Control in Acutely Ill Patients Martin J. Abrahamson, MD FACP Associate Professor of Medicine, Harvard Medical School Senior Vice President for.
Gestational diabetes mellitus (GDM), a common medical complication of pregnancy, is defined as “any degree of glucose intolerance with onset or first.
Rattan Juneja MD¹; Michael E. Stuart, MD 2,3 ; Sheri A. Strite 3 Indiana University School of Medicine, Indianapolis, Indiana¹ University of Washington,
Long-Term Biological Variation of Serum Protein Electrophoresis M-Spike, Urine M-Spike, and Monoclonal Serum Free Light Chain Quantification: Implications.
Laboratory Quality Control
Utilization of Assay Performance Characteristics to Estimate Hemoglobin A 1c Result Reliability A. Woodworth, N. Korpi-Steiner, J.J. Miller, L.V. Rao,
Association of 1,5-Anhydroglucitol with Diabetes and Microvascular
Microsatellite Instability Detection by Next Generation Sequencing S.J. Salipante, S.M. Scroggins, H.L. Hampel, E.H. Turner, and C.C. Pritchard September.
Unit 4: Monitoring Data Quality For HIV Case Surveillance Systems #6-0-1.
Journal Club Electronic Medical Record–Based Performance Improvement Project to Document and Reduce Excessive Cardiac Troponin Testing S.A. Love, Z.J.
Quality Assurance in the clinical laboratory
Low 25-Hydroxyvitamin D and Risk of Type 2 Diabetes: A Prospective Cohort Study and Meta-analysis S. Afzal, S.E. Bojesen, and B.G. Nordestgaard February.
Chemometrics Method comparison
Unit #7 - Basic Quality Control for the Clinical Laboratory
Validity and Reliability Dr. Voranuch Wangsuphachart Dept. of Social & Environmental Medicine Faculty of Tropical Medicine Mahodil University 420/6 Rajvithi.
Quality Assessment 2 Quality Control.
Journal Club Increased Rheumatoid Factor and Deep Venous Thrombosis: 2 Cohort Studies of 54,628 Individuals from the General Population C.L. Meyer-Olesen,
Measurements for 8 Common Analytes in Native Sera Identify Inadequate Standardization among 6 Routine Laboratory Assays H.C.M. Stepman, U. Tiikkainen,
Multiple Choice Questions for discussion
Determining Sample Size
Inference in practice BPS chapter 16 © 2006 W.H. Freeman and Company.
Sugar control in Critical care unit Senior clinical pharmacist : Lihua Fang Koo Foundation Cancer Center.
Journal Club Influence of Glycosylation on Diagnostic and Prognostic Accuracy of N-Terminal Pro–B-type Natriuretic Peptide in Acute Dyspnea: Data from.
Epidemiology The Basics Only… Adapted with permission from a class presentation developed by Dr. Charles Lynch – University of Iowa, Iowa City.
AJ Le Compte, CG Pretty, GM Shaw, JG Chase Introduction Elucidating links between glycemic control and clinical outcome requires reliable discrimination.
© Copyright 2009 by the American Association for Clinical Chemistry Glucose Meter Performance Criteria for Tight Glycemic Control Estimated by Simulation.
Lecture 12 Statistical Inference (Estimation) Point and Interval estimation By Aziza Munir.
Biostatistics: Measures of Central Tendency and Variance in Medical Laboratory Settings Module 5 1.
Improving Patient Outcomes GLYCEMIC CONTROL IN PERI-OPERATIVE PATIENTS UTILIZING INSULIN INFUSION PROTOCOLS.
T.S. Tran, J.R. Center, M.J. Seibel, J.A. Eisman,
Statistical Analysis. Statistics u Description –Describes the data –Mean –Median –Mode u Inferential –Allows prediction from the sample to the population.
Prof. of Clinical Chemistry, Mansoura University.
Chapter 5 Errors In Chemical Analyses Mean, arithmetic mean, and average (x) are synonyms for the quantity obtained by dividing the sum of replicate measurements.
Journal Club Recalibration of Blood Analytes over 25 Years in the Atherosclerosis Risk in Communities Study: Impact of Recalibration on Chronic Kidney.
EVALUATION OF CONVENTIONAL V. INTENSIVE BLOOD GLUCOSE CONTROL Glycemic Control in Critically Ill Patients DANELLE BLUME UNIVERSITY OF GEORGIA COLLEGE OF.
Prospective Evaluation of B-type Natriuretic Peptide Concentrations and the Risk of Type 2 Diabetes in Women B.M. Everett, N. Cook, D.I. Chasman, M.C.
PCB 3043L - General Ecology Data Analysis. OUTLINE Organizing an ecological study Basic sampling terminology Statistical analysis of data –Why use statistics?
PowerHour Information 03/09/2011.  Background  Description  Vision  Mission  Measurements  Participation Requirements.
Achieving Glycemic Control in the Hospital Setting Part 1 of 3
Medical Statistics as a science
ABSTRACT Hyperglycaemia is prevalent in critical care, and tight control reduces mortality. Targeted glycaemic control can be achieved by frequent fitting.
Quality control & Statistics. Definition: it is the science of gathering, analyzing, interpreting and representing data. Example: introduction a new test.
Dialysis Clinic Process Improvement Analysis Adam Roy Dorothy Fisher Janelle Bryant Joel Mikuta Sherri Mesquita Michelle Himes.
Comparability of methods and analysers Nora Nikolac
IMPORTANCE OF STATISTICS MR.CHITHRAVEL.V ASST.PROFESSOR ACN.
Achieving Glycemic Control in the Hospital Setting (Part 2 of 4)
Applying New Science to Drug Safety Janet Woodcock, M.D. Acting Deputy Commissioner for Operations April 15, 2005.
BC Cancer Agency CARE & RESEARCH Breast Cancer Mortality After Screening Mammography in British Columbia Women Andrew J. Coldman, Ph.D. Norm Phillips,
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
 Routine viral diagnostics: indirect and direct detection of viruses. ◦ Indirect detection: serological tests; ◦ Direct detection:  Viral antigens;
Question Are Medical Emergency Team calls effective in reducing cardiopulmonary arrest rates in the general medical surgical setting? Problem The degree.
Table 1. Methodological Evaluation of Observational Research (MORE) – observational studies of incidence or prevalence of chronic diseases Tatyana Shamliyan.
Diagnostic clinical chemistry
Management of Morning Hyperglycemia Following Cardiac Surgery LUMC 2ICU CV-Surgical Team CV AnesthesiaNursing Staff Pharmacy StaffAnesthesia Residents.
Quality Assessment.
Impact of State Reporting Laws on Central Line– Associated Bloodstream Infection Rates in U.S. Adult Intensive Care Units Hangsheng Liu, Carolyn T. A.
PCB 3043L - General Ecology Data Analysis.
Practical clinical chemistry
مدیریت داده ها و اطلاعات آزمایشگاه پزشکی
Choice of Methods and Instruments
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

Journal Club Postmarket Surveillance of Point-of- Care Glucose Meters through Analysis of Electronic Medical Records L.F. Schroeder, D. Giacherio, R. Gianchandani, M. Engoren, and N.H. Shah May © Copyright 2016 by the American Association for Clinical Chemistry

Introduction Strict glycemic control protocols in critically ill patients (see editorial) Critically ill patients have a high incidence of hyperglycemia Several observational studies and RCTs found maintenance of normoglycemic states (strict glycemic control, SGC) reduces mortality significantly After widespread adoption of SGC, many larger studies found no benefit and increased frequency of hypoglycemia under SGC SGC study designs Several differences in design of the early vs later SGC trials may account for different findings, e.g., differences in control blood glucose targets Another difference was that early studies utilized blood gas analyzers to measure glucose while later studies (and most clinical implementations) utilized glucose meters FDA has not approved most glucose meters for use in the critically ill 2 Editorial. van Hooijdonk RTM, Krinsley JS, and Schultz MJ. DETECT the Extremes That Usually Remain Undetected in Conventional Observational Studies. Clin Chem 62; 5: 668.

Introduction Maltose interference with GDH-PQQ glucose meters Between , FDA received 13 reports of death associated with meters using GDH-PQQ methodology Maltose, or other non-glucose sugars that metabolize to maltose, used in some medications including peritoneal dialysis fluid and immunoglobulin therapies, reacted with these meters Measurements were falsely increased up to 15 times the actual glucose value, triggering unnecessary and dangerous insulin administration FDA and CMS regulation 01/2014: FDA released draft guidance to the test manufacturing industry on increased accuracy requirements for approval in critically ill patients 09/2014: FDA approved first (and so far only) glucose meter for use in critically ill 11/2014: CMS memorandum (since reverted to draft status) that off-label use of glucose meters in critically ill patients would be considered laboratory developed testing and regulated as high complexity requiring institutional validation studies 3

Introduction Current US regulatory mechanisms to ensure safety of medical devices FDA premarket approval (studies on the order of thousands of samples; likely not large enough to cover the heterogeneous populations and polypharmacy to be encountered once the meter is offered to the public) FDA postmarket surveillance: passive adverse event reporting into the Maude database Proficiency testing programs: however, proficiency testing material for point of care glucose is non-commutable (may introduce matrix-effects) and thus, cannot be used to compare bias between different meters Data mining EMRs To Evaluate Coincident Testing (DETECT) Routine hospital draws for basic metabolic panels often occur simultaneously with POC glucose meter measurements in ICU patients In these patients, arterial or venous line blood is used typically for both measurements (i.e., no fingersticks) These events represent an opportunity to assess accuracy of POC glucose meters; this coincident testing analysis is occasionally employed by laboratories, but has never been validated 4

Objectives Conduct a gold standard study to validate DETECT Refine DETECT through successive use of filters to obtain the cleanest set of coincident events Compare the ability of DETECT to match the gold standard estimates of bias, random error, and percent of events outside quality goals 5

Method Gold standard study to validate DETECT (bedside ICU study) Nurses asked to flag events where an arterial draw was obtained for central laboratory glucose testing near in time to a POC glucose measurement on arterial blood from the same patient (+/- 5 minutes), without contemporaneous patient management changes (e.g., insulin or glucose infusion changes) We also compared DETECT estimates of accuracy to a laboratory validation study in critically ill patient samples 6

7 DETECT method Figure 1. Routine orders for POC and central laboratory glucose in hospital settings occasionally occur in the same patient at nearly coincident times. DETECT queries the EMR for these events. Additional filtering steps are performed on these coincident events, on the basis of laboratory turnaround times, location, time period, presence of central laboratory repeat testing, and presence of peripheral lines.

8 DETECT method: filters Outliers: events with large differences between POC and central laboratory glucose, calculated using median absolute deviation Table 1. Study size and outlier analysis.

Bias and random error Bias: average difference between the POC glucose measurement and central laboratory method Random error (composed of the two sources below) Imprecision: dispersion of results due to repeated testing of the same sample Random bias: this is actually a patient-specific bias (due to, e.g., hematocrit interferences) that appears as random noise in a cohort study Only random error is estimable by DETECT (as dispersion of POC and central laboratory differences); repeated measurements of the same sample do not occur in routine clinical practice. 9 Note: outliers were excluded when calculating bias and random error

10 Results Bias estimates between DETECT, bedside ICU study, and laboratory validation study were indistinguishable even with minimal filtering Random error estimates from DETECT continued to converge on those from the bedside ICU study with increased filtering (up to a point) Figure 3. Effect of filtering steps on overall bias and random error. Bias and random error calculated over all study glucose values (51–208 mg/dL) and for each sequential filtering step in DETECT. Also plotted are the bias and random error estimated by the laboratory validation study and the bedside ICU study. CIs are shaded and calculated as described in Methods.

11 Throughout glucose range Bias estimates between fully filtered DETECT, bedside ICU study, and laboratory validation study were largely overlapping throughout the range of glucose studied (range was limited by availability in the bedside ICU study) Random error estimates from fully filtered DETECT overlapped with bedside ICU study estimates, although at low glucose, laboratory validation estimates were lower Figure 2. Moving average estimates of bias and random error. Bias and random error are plotted with a ±20 mg/dL glucose window. Bias is calculated as the mean difference between POC and central laboratory testing results for all coincident events in a window. Random error is calculated as the SD of the percent POC and central laboratory differences in each window. CIs are shaded. Results

12 Glycolysis due to specimen transport and processing Central laboratory testing includes a period of time in which blood cells are in contact with plasma The longer blood cells are in contact with plasma, the more glucose they metabolize This can cause a positive bias in accuracy studies comparing POC glucose to the central laboratory. Results Supplemental Fig 2. Associations of central testing time delay with POC bias. Linear regression of POC & central laboratory testing result differences vs central laboratory testing bedside collection to result time delay, to assess for continued glycolysis in the collection tube.

13 Note: outliers included in this analysis Results Performance of meters compared to quality goals: DETECT vs bedside ICU (unadjusted and adjusted for central laboratory delay) Figure 4. Effect of central laboratory testing delay on POC quality measures. Bland–Altman plot showing differences between POC and central laboratory results, plotted against central laboratory results. Dashed lines represent CLSI quality goals. (A), Based on unadjusted central laboratory values. (B), Based on central laboratory values adjusted for time delay between blood collection and result verification, showing a reduction in the number of events outside quality goals.

14 Results Physiological interference In multivariable regression against analytes from the basic metabolic panel and CBC, only hematocrit displayed clinically significant interference Supplemental Fig 2. Associations of hematocrit with POC bias. Linear regression of POC bias (%) vs hematocrit to characterize interference.

Questions 1)How likely is it that inaccuracies in glucose meter performance hindered latter RCTs of strict glycemic control? 2)Which filters are most important to implement when using DETECT? How does this vary by application? Which filters are the easiest to implement? 3)With the large data sets provided using DETECT, what novel analyses could be performed? 15

Thank you for participating in this month’s Clinical Chemistry Journal Club. Additional Journal Clubs are available at Download the free Clinical Chemistry app on iTunes for additional content! Follow us 16