Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun

Slides:



Advertisements
Similar presentations
Analog Imaging II Intensifying Screens
Advertisements

Technical Aspects: the machine the image
Advanced Biomedical Imaging Lecture 3
Radiographic Film & Intensifying Screens
The Bases x-ray related physics
Formation Characteristics
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Primary Exposure Factors IV
Chapter 18 Radiographic Exposure
IMAGE FORMATION  Introduction  The Invisible and Visible Image  Image Characteristics.
Components of Radiographic Image Quality
Radiographic Quality Chapter 5.
Digital Radiography.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
Resident Physics Lectures
MAMMO QC – covered in week 8
X-ray radiography Seminar The "geometrical" unsharpness of X-ray images is limited by finite dimension of the source, i.e. the window on.
Dr. Mustafa Zuhair Mahmoud Mr.Ali B Alhailiy بسم الله الرحمن الرحيم 1.
Chapter 15 Radiographic Intensifying Screens And Cassettes
Chapter 9 Film & Film Holders.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Image Formation Differential Absorption of X-rays by the Patient  Absorption of non-attenuated x-rays by screen  Emission of light  Formation of Film.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Intensifying Screens Kyle Thornton DMI 50B.
RAD 354 Chapt. 13 Intensifying Screens
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
ADC PRE-PATIENT COLLIMATION POST-PATIENT COLLIMATION.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Chapter 7 Dental X-Ray Film.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Quality Control.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
IMAGE QUALITY REVIEW RT
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
Image Receptor Systems
Image Quality Radiographic Resolution.
Mammography Questions and Answers
Exposure Factors or Prime Factors
Chapter 6 Image Receptors Radiology. Introduction To further understand radiography, one must know how a permanent record is produced using x-rays. To.
Radiographic Quality Visibility and Sharpness
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
RAD 254 Chapter 13 Intensifying Screens Physical purpose: to convert x-ray photons into light photons (done at the phosphor layer)
Quality Assurance in the Dental Office
Analog Imaging III By Professor Stelmark.
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
Grids George David Associate Professor Department of Radiology
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Image Receptors Film Construction, Film Handling, Cassette Construction, DarkRoom, Digital Imaging 1.
Fluoroscopy. Real-time imaging Most general-purpose fluoroscopy systems use TV technology, operating at 30 frames/sec May be recorded (barium swallow.
Film Imaging Lavin: Chapter 5 CTVT: Ronnie’s snood
Fluroscopy and II’s. Fluroscopy Taking real time x-ray images Requires very sensitive detector to limit the radiation needed Image Intensifier (II) is.
Image Receptor Unsharpness By Professor Stelmark.
DIGITAL RADIOGRAPHY.
Sergeo Guilbaud School of Radiologic Sciences
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
B ASIC P HYSICS OF D IGITAL R ADIOGRAPHY By : Maisa Alhassoun Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS &
Radiology. The Cathode… Provides a source of electrons and directs the electrons towards to anode. The cathode has a coiled wire filament that emits electrons.
INTENSIFYING SCREENS.
Resident Physics Lectures
Optimisation of Patient Protection for Radiography
Imaging on Film RVT: Chapter 5
Quality Control Testing of Screen Speed
Resident Physics Lectures
Fluoroscopic Image Intensifier Image production
Fluoroscopic Image Intensifier Image Production
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun

Intensifying Screens A. Screens -Intensifying screens contain phosphor crystals that absorb x-ray photons and emit many more visible light photons, which expose the film as shown in Fig The screen therefore converts the x-ray pattern to a light pattern, which is recorded on film.

-Intensifying screens improve the efficiency of radiographic imaging over use of film alone. -The use of intensifying screens decreases the mAs required for a given film density, resulting in a lower patient dose. -Shorter exposure times also decrease x-ray tube loading and image blur caused by patient motion.

-Intensifying screens are typically made of a polyester base for support with a 40 to 200 μm thick phosphor layer. -About 50% of the radiation incident on a cassette containing a pair of screens will be absorbed. -Absorption efficiency refers to the percentage of x-ray photons absorbed in a screen and is typically 25%.

-Conversion efficiency refers to the percentage of the absorbed x-ray energy that is converted into light energy and is typically 10%. -The intensification factor is the ratio of exposures, without and with intensifying screens, required to obtain a given film density. -Intensification factor depends on the absorption and conversion efficiency of the screen. -Typical intensification factors are 30 to 50.

B. Screen materials -Screens contain high atomic number materials to maximize the absorption of x-rays. -Calcium tungstate (CaWO4) was used in intensifying screens until about 1970.

-Tungsten has a high K-shell binding energy (69.5 keV) compared with the mean photon energy levels normally used in diagnostic radiology; therefore, absorption was less than optimal. -For example, an examination performed at 100 kVp corresponds to a mean photon energy level of about 40 keV, which is well below the K-shell binding energy in the screen.

-Rare earth screens are "faster" than calcium tungstate because they have a higher absorption efficiency at the mean x-ray energies normally used in radiology. -Rare earth screens also have a higher conversion efficiency and produce more light for a given amount of deposited x- ray energy. -Table 4.2 summarizes the K-shell binding energy levels of the rare earth elements used in screens.

-Gadolinium oxysulfide (Gd2O2S) emits mainly green light, and lanthanum oxybromide (LaOBr) and CaWO4 emit mainly blue light. -The light output—that is, wavelength (λ) or color—of a screen and the light sensitivity of the film must be matched (spectral matching). -Conventional film is sensitive to ultraviolet and blue light. -Orthochromatic film is also sensitive to green light.

C. Cassettes

-The film and screens are held in a light-tight cassette. -Screens are usually permanently mounted inside the cassette. -A thin layer of resilient foam holds the screen tightly against the film when the cassette is closed.

-The front of the cassette is made of a minimally attenuating material, such as aluminum or carbon fiber. -Dual-screen, dual-emulsion systems are frequently used to improve x-ray absorption. -Intensifying screens can be significant sources of image artifact. -Scratches, stains, hair, dust, cigarette ash, and talcum powder are all potential sources of image artifacts.

-As part of a quality control program, all screens should be regularly cleaned. -Cassettes should also be evaluated for good screen/film contact. -Screen/film contact is evaluated by taking an image of a wire mesh and ensuring that the resultant image permits visualization of the mesh.

D. Screen/film speed -The speed of a screen/film combination is inversely related to the exposure required to produce a given density. -As the speed increases, the exposure required decreases.

-The speed of a screen/film combination used in radiology ranges from approximately 50 to Screen/film speed is normally expressed relative to a CaWO4 standard assigned a speed of Screen speed increases with increasing screen thickness, absorption, and conversion efficiency. -Both screen and film must be specified when assigning speed to any screen/film combination.

-High-speed screens are generally thicker and have decreased spatial resolution. -Thick screens increase image blur because of increased diffusion of light in the screen before striking the film. -Detail screens are thinner and therefore slower, but they have better spatial resolution.

-Fast screen/film combinations are used for abdominal studies: slow films are used for extremity examinations. -Single-emulsion single-screen systems are used for bone detail and mammography.

-To ensure the correct film density, phototiming systems are generally used. -A phototimer measures the actual amount of radiation incident on the screen/film and terminates the exposure when the correct amount has been received. -Table 4.3 lists a range of screen types used in radiology, as well as their clinical applications.