Second run with the Copper filter WA104 technical meeting – 23/03/2016.

Slides:



Advertisements
Similar presentations
Inert Gas Purification Systems
Advertisements

A. Alton 1, H. O. Back 2, C. Galbiati 2, A. Goretti 2, C. Kendziora 3, B. Loer 2, D. Montanari 2, P. Mosteiro 2, S. Pordes 3 1)Augustana College 2)Princeton.
Inert Gas Purification Systems Why do we need them??? Many of the materials used in research and development today are extremely air sensitive, and can.
LAPD Run 2 Results and Notes Terry Tope
LBNE 35 ton prototype Phase 1 summary Terry Tope Fermi National Accelerator Laboratory All Experimenters’ Meeting – Fermilab – May 19, 2014.
Cryogenic discussion and update 30 October 2014Neutrino platform.
Current. Current Current is defined as the flow of positive charge. Current is defined as the flow of positive charge. I = Q/t I = Q/t I: current in Amperes.
Materials Qualification for long drift Liquid Argon TPC Effect of water on electron drift lifetime Plans Material Tests, purging and H20 studies, Purity.
RPC Purifiers’ studies /status/ IR, RG
© Copyright 2002 ABB. All rights reserved. - Microelectromechanical Systems for Process Analytics IFPAC 2003 Dr. Berthold Andres ABB Automation Products.
AIR CONDITIONING (COOLING) UNIT 40 TYPICAL OPERATING CONDITIONS
Fundamental Techniques and Measurements  Mass Measurements  Volume Measurements  Preparation of a solution of known concentration  UV-Visible Spectrophotometer.
This tutorial is a PowerPoint ® file which includes recorded narration. Automatic slide sequencing is coupled to the sound track. For the most effective.
Depleted Argon from Underground Sources Henning O. Back – Princeton University For the DarkSide experiment, E-1000 Augustana College – D. Alton FNAL –
Gas Properties. Gasses A state of matter characterized by: A state of matter characterized by: Filling the container it occupies Filling the container.
LBNE R&D Briefing May 12, 2014 LBNE R&D Briefing May 12, 2014 LArIAT and LBNE Jim Stewart LArIAT EPAG Chair BNL LBNE LARIAT-EPAG J. Stewart BNL T. Junk.
Observations on possible factors affecting LAr purity on the 50kton TPC from the viewpoint of the internal walls Alejandro S. Díaz Chemical Engineering.
Copyright © 2009 Pearson Education, Inc. Chapter 25 Electric Currents and Resistance.
Techniques for determination of deep level trap parameters in irradiated silicon detectors AUTHOR: Irena Dolenc ADVISOR: prof. dr. Vladimir Cindro.
Grade 9 Science – Unit 4 – Electricity #2 Series and Parallel Ohm’s LawResistance, Current and Voltage More R, V and I plus Power and Efficiency Producing.
RICH 2007Dirk Wiedner / LHCb1 The Usage of N-Perflourcarbons as RICH Radiators.
Unit One Review Accuracy and Precision The important things to remember about accuracy and precision: You want measurements that are both accurate and.
4/5/2005FLARE Cryogenics, RLSchmitt1 Argon Refrigeration Supply Purity.
Liquid Argon Time Projection Chamber: Purity and Purity Monitoring DAVID GERSTLE – LArTPC – YALE UNIVERSITY/FNAL 31 May 2006 – FNAL Users’ Meeting Materials.
Materials Qualification for long drift Liquid Argon TPC Effect of water on electron drift lifetime Plans S. Pordes Argon Purity1Integrated R & D Plan Review.
Hans Jostlein, May 24, Large Liquid Argon Purity Demonstration— Phase I: Gas Flushing -What -Why ? -How ? -When ? -Define “Success”
WA104 Cryogenics Design Status David Montanari / Johan Bremer May 21, 2015 Rev. 1.
Poster Design & Printing by Genigraphics ® Neutrino Interactions Studying the properties of neutrinos will shed light on the origin of the.
V B EE ???? + -  V>0 Hall Effect h. v B EE >0 + -  V>0 v B EE ???? + -  V
The RPC Closed loop Gas System Slides from Bianco and Guida.
Materials Qualification for long drift Liquid Argon TPC Effect of water on electron drift lifetime Plans Material Tests, purging and H20 studies, Purity.
S. Pordes - FNAL May 13th Purity Monitor - Description and Experience at Fermilab Production of Clean Argon - Description of Filtering System and.
Preliminary tests of resistive materials for ProtoDUNE CPAs. 1.
Neutrino Cryogenics Requirements Meeting 25/09/2014 Day 1 review.
Preliminary results on Copper filters, with FLIC test chamber F. Pietropaolo, A. Zani, M. Meli CENF General Meeting – 22/10/2015.
UM PPS Lab Activities Mid-size Panel Tests PPS meeting January 15, 2012 Claudio, Curtis, Dan, Ethan, Riley.
Atomic Absorption Spectroscopy
Halliday/Resnick/Walker Fundamentals of Physics
Concept Design for LBNF Far detector (LAr single phase)
1/26/2009 S. Pordes LAr R & D Briefing 1 Liquid Argon presents prospect of continuously live Imaging Calorimeter C. Rubbia.
Ions in LAPD Can we Observe Them in Long Bo ? 3/6/2016 Ions in LAPD Sarah Lockwitz and Hans Jostlein, Fermilab 1 Sarah Lockwitz and Hans Jostlein Fermilab.
Cryogenics Fabrication Plans & Constraints David Montanari (on behalf of the CERN Neutrino Cryogenic Group) ProtoDUNEs Meeting Feb 23, 2016.
Materials Qualification for long drift Liquid Argon TPC Effect of water on electron drift lifetime Plans Material Tests, purging and H20 studies, Purity.
S. Colafranceschi - Ischia CMS RPC Italia1 CMS RPC studies at ISR Stefano Colafranceschi Laboratori Nazionali Frascati Università di Roma – La Sapienza.
S.Colafranceschi 20 June Stefano Colafranceschi LNF Frascati CERN Universita’ di Roma – La sapienza An overview on gas studies, purifier materials.
35ton and LAPD cryostats and future DUNE detector development.
TRD Gas System Layouts Production Readiness Review 03 – Nov F. Hahn.
35 Ton LAr Impurity Distribution Measurements and CFD simulation Erik Voirin – Fermilab – Thermal and Fluids Engineering Group / Engineering.
PSA PLANT OXYGEN GENERATION SYSTEMS. 1. PSA Description 2. Molecular Sieves 3. Process ( Adsorption and Desorption) 4. Application for PSA process 5.
CHROMATOGRAPHY Chromatography is used to separate and analyse small amounts of mixtures Methods involve a stationary phase and a mobile phase. There are.
Methane Update and Run Plan Ben Jones. Configuration 8 inch PMT with TPB plate - 8” from alpha source - Now includes magnetic shield - 2 inch PMT without.
1 2A-08 Buoyancy of Air IF SENSITIVE WEIGHING OF AN OBJECT IS REQUIRED, UNEQUAL BUOYANT FORCES COULD AFFECT THE RESULTS. mbgmbg magmagρ air gV b ρ air.
Low radioactivity argon from underground sources Henning O. Back – Princeton University LIDINE 2013.
Cryogenic Test Stand Facilities at PAB: Stephen Pordes Neutrino R&D Workshop January 20 –
Study of the cryogenic THGEM-GPM for the readout of scintillation light from liquid argon Xie Wenqing( 谢文庆 ), Fu Yidong( 付逸冬 ), Li Yulan( 李玉兰 ) Department.
1 Cary Kendziora Fermi National Laboratory Cary Kendziora Fermi National Laboratory CEC-ICMC 2013 Anchorage Alaska – June, 2012.
In-Vessel Electronics Infrastructure Craig Thorn LAr40 Cold Electronics Far Site Review, December 6-9, 2011.
Neutrino Platform Proximity Cryogenics Functional Design M.Chalifour 14/04/2016 M. Chalifour - TE/CRG-CI 2.
ProtoDUNE Single Phase (NP-04) P&ID David Montanari (on behalf of Mark A, Johan B, Michel C, Joaquim C, Kevin H) CERN Neutrino Platform General Meeting.
Ed Connor, GC Product Specialist
Installation of the T600 at Fermilab
Proximity Cryogenics P&IDs meeting
David Montanari / Johan Bremer Jun 11, 2015 Rev. 1
Follow-up of the tests with the copper filter
Recirculating CO2 System
Circuits Current and Resistance
Series 5300 Lithium Cell Formation System
CHROMATOGRAPHY.
Presentation transcript:

Second run with the Copper filter WA104 technical meeting – 23/03/2016

Summary first run WA104 Technical Meeting2 With the copper-based filter operated in liquid phase during filling and in gas phase during Argon recirculation (1 full volume/4-5 days), a purity higher than 10 ms (above sensitivity of our detector) was reached in a short time. The 50 l LAr-TPC was operated for several days at max purity, then O 2 was injected in steps for a total of ~16 g of O 2, over the whole run. Purity started getting worse after injection of ~1.5 g of O 2 but with a very slow rate of change: still 10 us electron lifetime after injection of 16 g of O 2, suggesting that the filter trapping/saturation mechanism is not completely understood. It is also possible that other impurities are present in the detector that are not trapped by the filter. Fitting the slow component  s of the LAr scintillation light, yielded a value around 0.8  s ( to be compared with  s > 1.2  s prior to oxygen injection). Before the end of test the amount of N 2 and H 2 O in the recirculating gas was measured with argon gas analyzer: 16.6 ppm and < 0.5 ppm, respectively. These are partially compatible with the measured slow light component and electron lifetime and tend to confirm the presence of additional impurities in the LAr-TPC not trapped by the molecular sieve and the copper.

Filter composition: - Molecular Sieve (Type 4A mm) —> 2.46 kg - Copper (BASF Cu-0226 S 14x28 mesh or Q-5)—> 2.96 kg (  = 0.85 kg/dm 3 ) Copper fills 45 cm of the filter Adsorption capacity, as estimated at FNAL : 0.5 g O2 /kg in LAr, 5 g O2 /kg in GAr 3WA104 Technical Meeting

4 FLIC Cryo-set up - reminder New filter with copper bed installed BEFORE recondenser

Overall purity plot WA104 Technical Meeting5

Further measurements WA104 Technical Meeting6 A new run was performed in late December, taking data for few days with the filter left as in previous test: no tracks could be collected during this run, which seems to confirm that the Copper filter was saturated; the measurement of the slow LAr light component was compatible with no contamination by other impurities, as it had been in the previous run, before starting Oxygen injection. In January 2016 the filter was sent back to CERN cryolab and is was regenerated with the following prescriptions: Argon Flow at high Temperature (> 200°) to activate molecular sieve, followed by Ar(98%)+H 2 (  2%) flow for Copper activation. Water content was continuously recorded to get an estimate of the trapped impurities level both in Mol. Sieve and Copper. Far less than 15 g of water was actually extracted during the purification phase. A new run with the LAr-TPC followed. A new bottle of Ar60+O 2 55 (1% molar), with much lower content of impurities other than Oxygen, was used for filter saturation (O > H 2 0 ≤ 0.5 ppm-mol; CO,CO 2,H 2 ≤ 0.1 ppm-mol).

Second run A new run was performed in the past February, after the filter was regenerated by the Cryolab group at CERN. The whole system was kept identical to previous run (cryo plant, chamber), as well as the procedure: After filling, initial life-time  30  s; we recirculated until overcoming 3 ms (about one week), then started Oxygen injection. The only difference with respect to the past run is related to the much purer gas bottle used. We forced fast recirculation, one volume in 2 days, by increasing heat input to the chamber, to fast achieve high purity. Ar-O 2 mixture was then injected at g O2 /h, for a total of 17.5 g O2. WA104 Technical Meeting7

8 Purity during O 2 injection phase

Comments WA104 Technical Meeting9 This time we observed purity increase always compatible with the recirculation time, for O 2 injection up to 13.5 g. Then, an abrupt decrease of the purity follows, with “vanishing tracks”. This happens in a few hours, and it is confirmed by the change in the slow component of the LAr scintillation light, cut in half after the purity drop. This behaviour is well compatible with Oxysorb filters and what we expected from a correctly activated cartridge. As well, the amount of Oxygen needed to saturate the Copper filter is around 15 g, compatible with the previous run and data shared by American colleagues. Such results strengthen the hypothesis that the odd outcome of the past run may be due to low grade of the Oxygen used, implying a large amount of unknown impurities injected along with O 2.

Saturation Region WA104 Technical Meeting10 Purity almost goes to “zero” in around 6 hours. At g/h, this means (roughly) more than 1.5 g of O 2 -> 10% of the full capacity. This translates in 10% length over which the filter looses 100% efficiency (interface) -> over the length of the filter this means (rough estimate) 4.5 cm. Oxysorb interface is smaller (purity decrease sharper), probably due to larger size of allumina pellets in the copper filter. d d = f(mean free path) = f’(pellet size) Full portion

Vanishing track example WA104 Technical Meeting11 COLLECTION INDUCTION WIRES * * = 280 t-sample, or 112  s Run 3024, ev Around 2-3 hours after purity starts dropping.

Summary and next steps WA104 Technical Meeting12 A new run has been performed with the copper filter and the ICARUS 50l chamber: we verified saturation in gas is achieved with  15 g of O 2. High-purity oxygen was injected and the filter behaviour reflected common knowledge: 100% efficiency up to saturation, followed by an abrupt decrease in purity (6 h at g O2 /h). A new test will be put in place, to characterise purification in liquid phase. Simpler circuit, no use of the 50l TPC. Immersed LAr pump and filter, connected to the Ar-O 2 bottle and to long exhaust at the output (to gasify the liquid and avoid air back-diffusion). In series to the copper filter, we will place a small, glass-transparent cartridge of Oxysorb, which will change colour as soon as it start operating: this will tell us exactly when the copper filter is full. A scheme of the full system is in development.