Force-Sensing. Laparoscopic Grasper Client: Charles P. Heise, MD, FAS, CRS Advisor: William L. Murphy, Ph.D Team Leader: Adam Dahlen Communication: Darshan.

Slides:



Advertisements
Similar presentations
Auto-suture device for nasal surgery
Advertisements

Instrumentation (AMME2700) 1 Instrumentation Dr. Xiaofeng Wu.
BASIC SENSORS AND PRINCIPLES. ① Strain gages Measurement of extremely small displacement ① Potentiometers Translational and Rotational displacement ②.
ENTC 3320 Absolute Value.
Microprocessor Motor Control Spring Introduction  Stamp projects Robots  Sensors  Motor control  Logic Rocketry  Reading acceleration (“g”
Trey DeLong Lacey Gorochow Brian Rappa Adam Vandergriff Sandra Wadeer Advisor: Dave Martinez VP of Sales at Zimmer, Inc. Surgical Consult: Dr. Michael.
PRESENTATION ON LOAD CELL AND ITS APPLICATION
Measurement of force torque and pressure
Instrumentation (AMME2700) 1 Instrumentation Dr. Xiaofeng Wu.
Senior Design: Tachometer Calibration Device Team 4: Jennifer Egolf, Matthew Hagon, Michael Lee, Christopher Pawson Sponsor: DuPont Advisor: Dr. Glancey.
Multidisciplinary Engineering Senior Design Project 6218 Soft Tissue “Tensile” Tester Preliminary Design Review 2/24/2006 Project Sponsor: Bausch & Lomb.
© 2012 Pearson Education. Upper Saddle River, NJ, All rights reserved. Electronic Devices, 9th edition Thomas L. Floyd Electronic Devices Ninth.
REINFORCED CONCRETE Reinforced concrete is a composite material which utilizes the concrete in resisting compression forces, and steel bars and/or.
GROUP 2: Martha Ingram, Megan Johnston, Chelsea Samson MENTORS: Dr. Michael Holzman, Dr. William Beck, Dr. Benjamin Poulose AN OPENING IN THE ABDOMEN,
Topic 1 different attributes that characterize sensors ETEC 6405.
Esophageal Strictures William Stanford – Leader & BSAC Daniel Frost – BWIG Thomas Fleming – Communicator Client: Dr. Mark Reichelderfer, MD Advisor: Professor.
Mechanical Properties
Loading Apparatus for High Velocity Tissue Rupture Mechanical Engineering Dalhousie University Senior Design Project Winter 2010.
Tower Light Show Power Supply Team Members: Jesse Walson Jake Wagner Dave Miller Sponsor: Robert Rinker Team Advisor: Touraj Assefi.
T.M.F.T: Thermal Mechanical Fatigue Testing
An Opening in the Abdomen An Opening in the Market: Hernia Tensiometer GROUP 2: Martha Ingram Megan Johnston Chelsea Samson.
Group 1: Material Testing Device Jobe Dyson Brie Witherspoon John Chandler Brett Newstead.
Introduction to Engineering Lab 4 – 1 Basic Data Systems & Circuit Prototyping Agenda REVIEW OF LAB 3 RESULTS DEFINE a DATA SYSTEM  MAJOR TYPES  ELEMENTS.
Interfacing Pressure Sensor to logic device input pin Click within the blue frame to access the control panel. Only use the control panel (at the bottom.
CHAPTER 11 Op-Amp Applications. Objectives Describe and Analyze: Audio mixers Integrators Differentiators Peak detectors Comparators Other applications.
With strain gauge sensors, a Wheatstone bridge may not bridge the gap This presentation is partially animated. Only use the control panel at the bottom.
Robotic Arm for Minimally Invasive Surgery Team: Brenton Nelson, Ashley Huth, Max Michalski, Sujan Bhaheetharan BME 200/300 October 14, 2005.
Team: Charlie Haggart, Gabriel J. Martínez-Díaz,
Applications of OP-AMP. Introduction Operational amplifier using IC's is inexpensive, versatile and easy to use. For this reason they are used not only.
INTRODUCTION Strain gauge measurement is a point strain measurement method developed in the USA c.a independently by E. Simmons and A. Ruge. It.
Laparoscopic Banding Device Client: Dr. Thomas M. Julian, M.D. Advisor: Mitch Tyler Leader/BSAC: Gina Stuessy BWIG: Kailey Feyereisen Communications: Anna.
Development of Depth Gage Instrument Brian Cost Justin Johnson Tyler Kibbee March 20, 2008.
Bioactive Interference Screws for ACL Reconstruction
Abstract There is a high amount of necessity in third world countries for many concepts in mechanical engineering to inspire learning and choosing engineering.
18.1 Introduction Powder metallurgy is a process by which fine powdered materials are blended, pressed into a desired shape, and then heated to bond.
Biomechanical Tissue Stimulator Group 1: Matt Brady (BME/EE) Ankeet Choxi (BME) Misha Kotov (CS) Steven Manuel (ME) Adviser: Dr. V. Prasad Shastri.
Mid-Term Presentation February 28, Team Members Charlie Mraz EE Team Leader Analog Design PCB Layout Allen Joiner EE Power Supply Purchasing/Finance.
BME Design 200/300 Glaucoma Medication Reminder. Intellectual Property  All information provided by individuals or Design Project Groups during this.
Dr. Thomas M. Julian – Client Professor Mitch Tyler – Advisor Anna Moeller– Team leader Tyler Witt – BWIG Gina Stuessy – Communications Kailey Feyereisen–
Ring Removal Device Team: Evan Rogers, Sujan Bhaheetharan, Tyler Allee, Steve Noel BME 301 February 24, 2006.
PEAK INVERSE VOLTAGE Using the ideal diode model, the PIV of each diode in the bridge rectifier is equal to V2.This is the same voltage that was applied.
Review of the Ergo Measurement Tools Jason Beardsley Bob Fox Jason Beardsley Bob Fox.
Gas Pressure Meter (Engineering World Health) Advisor: Prof. Ken Gentry Client: Prof. John Webster, Engineering World Health Engineering World Health Team.
Biofeedback and Stress Management John Harrison – Leader Albert Kwansa – Communicator Eric Lee – BSAC Brenton Nelson – BWIG Dr. Daniel Muller – Client.
Auto-Suture Device for Nasal Surgery Client: Dr. Benjamin Marcus Department of Otolaryngology Advisor: Professor Tyler Jennifer Wager ~ Leader Mark Yarmarkovich~
Ophthalmic Dose Compliace Monitor. Our Team Benjamin Roedl – Team Leader Benjamin Roedl – Team Leader Nina Lewis – Team Leader Nina Lewis – Team Leader.
Gas Pressure Meter (Engineering World Health) Advisor: Professor Naomi Chesler Client: EWH and Professor John Webster Team Members: Ksenija Bujanovic-
Hemostatic Compression Device Jill Doyle Reed Gioe Christopher Miller Aut Singhchinsuk.
Laparoscopic Grasping Instrument by Lynn Murray, Becky Jones, Ann Sagstetter and Richard Bamberg.
Scale to Measure Nighttime Weight Change Joe Ferris – BWIG Kyle Herzog – Team Leader Ben Engel – Communicator Eric Printz – BSAC Client: Timothy Juergens,
Self-Contained Endoscopic Video Camera
DEVICE FOR THE IMPLANTATION OF NEURAL ELECTRODE ARRAYS
Absolute Value.
MECH 373 Instrumentation and Measurements
MECH 373 Instrumentation and Measurements
Intracranial Pressure Monitor
MECH 373 Instrumentation and Measurements
Laparoscopic Banding Instrument
Peak Inspiratory and Expiratory Flow Meter
POWER AMPLIFIERS.
Retractor for Breast Surgery
MSD P15280 RIT HOT WHEELZ TEST BENCH
Prepared by Dr. Mohamed Ahmed Awad
Mechanical Measurements and Metrology
Force-Sensing Laparoscopic Grasper
Motorized Wheelchair Mount
Copier Jam Detector Design Problem
Applied Technology High School (ATHS)
TOPIC- Load cell A load cell is a transducer that is used to create an electrical signal whose magnitude is directly proportional to the force being.
Presentation transcript:

Force-Sensing

Laparoscopic Grasper

Client: Charles P. Heise, MD, FAS, CRS Advisor: William L. Murphy, Ph.D Team Leader: Adam Dahlen Communication: Darshan Patel BWIG: Andrew Eley BSAC: Clara Zhang

Problem Statement Goal: design and build a prototype laparoscopic grasping tool –Provides feedback auditory or tactile Tissue-damaging force –Redesigned jaw mechanism Eliminate pinching at the pivot point of the jaws –Educational tool for surgical teaching

Background Laproscopy: minimally invasive surgery viewed by video screen (see figure on right) –Minimal incisions (5-10mm) Current Problems –Unknown force on tissue –Grasper pinching at pivot point –Causes damage to tissue

Wikipedia [Online]. March 3, A surgeon's view using a laparoscope. The top photo is of a gallbladder. The bottom photo is a cystic duct.

Design Constraints Foremost, re-design grasper to reduce pinching to a floating-point jaw mechanism Increase awareness by providing feedback Dimensions: –Fit to 5mm port –About 30cm long Varying closure setting Autoclavable

Design Specifications

“floatable” jaw to reduce pinching of bowel Provide feedback of pinching pressure must fit through 5mm port must be about 30 cm long optional ratcheting for locking autoclavable or disposable grasp a large portion of bowel without causing damage design must not be hazardous to surgeons or the patient No separate or loose parts that could possibly get lost in the patient must be precise enough to ensure that no damage is done to the tissue grasping mechanism must be solid with little slack durable material such as stainless steel The circuit should need little maintenance This device should last several years in a hospital environment. The handle should be easy to use and grasp by a surgeon. One large model will be prototyped for show; if successful, further manufacturing of the recommended design can be done and utilized for future testing or redesign. Cost should be under a few hundred dollars FDA approval is not required at this stage

Effect of jaw design on grip security, measured as peak load. Conducted independently of squeeze pressure. Damian D. Marucci, John A. Cartmill, William R. Walsh, Christopher J. Martin. Patterns of Failure at the instrument-Tissue Interface. Journal of Surgical Research, 2000 Jaw Design Research

Grey boxes = tissue samples that tore with respective jaw patterns No tearing occurred at any squeeze pressure with plane and wave pattern jaws. Damian D. Marucci, John A. Cartmill, William R. Walsh, Christopher J. Martin. Patterns of Failure at the instrument-Tissue Interface. Journal of Surgical Research, 2000

With increasing the jaw size and contact area (from 32mm 2 to 64mm 2 to 128mm 2 ), the allowable force before tissue damage was increased from 15 to 37 N. the maximum pressure applicable to the tissue without damage was computed to be 4.7 x10 5 Pa. E. A. M. Heijnsdijk, H. deVisser, J. Dankelman, D. J. Gouma. Slip and damage properties of jaws of laparoscopic graspers. Surg Endosc, 2004.

Research Conclusions 2 mm wave pattern is the optimal design option in terms of both grip security and minimal tissue damage. 4.7x105 Pa is the maximal pressure applicable to the tissue. With a safety factor of 1.5, the threshold pressure is: (1/1.5)x (4.7x105 Pa) =3.12x105Pa

Design Alternatives Strain Gages: –Widely accepted method of measuring strain on bendable materials –Small (reduces obstruction) –Inexpensive (under 10 dollars) –High-temperature resistant models (immovable after fixation and reduces limitations on placement location) –Functions by producing a change in resistance when strain is applied

Strain Gages on the Shaft Measure the compression and tension stresses on the shaft

Less clutter on the instrument, which will create a minimal nuisance to the user. Custom molds necessary for each gage during fixation procedure Manufacturing difficulty and practicality of putting the strain gages in the 5mm diameter of the shaft Increased cost for high-temperature resistant materials Less accurate, as compression and tension forces are small in comparison to those in the handle Signal conditioner necessary to supply voltage to and monitor output from the sensors

Strain Gages on the Handle Measure the compression and tension stresses on the handle

Larger fixation area on handle than on the shaft Larger compression and tension forces in the handle than on the shaft Extraneous deformation due to variations in handling Disadvantages of first design

Final Design Jaw Structure Design

The parallel jaw mechanism extends when the actuator is pulled.

Circuit Design

Buzzer circuit powered by a 6.0V battery which directly drives the supply voltage of an operational amplifier and a 3.3 V voltage regulator Potentiometer used to divide the 3.3 V and inputs to the inverting side of an opamp Opamp used as a voltage comparator which drives a buzzer if the non-inverting input voltage is higher then the inverting input, or the reference voltage. The variable voltage is dependent on the force sensor. As the force sensor is compressed the voltage goes up. If the sensor is pressed too much, the voltage will rise above the reference voltage causing the comparator to output 6.0V and turning on the buzzer.

Prototype Design: –Made to show feasibility of design idea and provide a physical reference –Not constructed to proposed scale of design due manufacturing constraints of available machining equipment –Tests were conducted to demonstrate that an equal force of N was experienced at various locations across the jaws when a tensile force of N was applied to the actuator Circuit: –Can be calibrated to specific value chosen by user to demonstrate a threshold force applied to the handle, which is proportional to the force experienced at the jaws

An example run was conducted with a compressive force of 10 N between the jaws of the Stryker laparoscopic grasper model. –Voltage was increased until 3.8 V, sounding the buzzer. –The threshold voltage would vary when connected to the proposed scale model design, as the ratio of input to output force would differ from the prototype model.

Future Work Construct an actual model to the design specifications and dimensions provided using stainless steel Conduct test of ratio of input to output force of scaled design Conduct testing of the circuit connected to the scaled design to set the threshold voltage to sound the buzzer when an output force of 10 N was observed.