Liquid based cytology Liquid based Cytology.

Slides:



Advertisements
Similar presentations
Gross Techniques In Surgical Pathology. Introduction The routine work associated with a surgical pathology specimen includes gross & microscopic examinations.
Advertisements

Cell Culture Facilities and Equipment Presenter: Dr. R. Rahbarghazi.
Slide 1, 4/12/2015 of DNA.ppt  Purdue University Cytometry Laboratories DNA1.ppt.
ViahanceTM: Dead Cell, Stripped Nuclei and Free Oligonucleotide Removal Kit Instructions ViahanceTM dead cell removal kit enhances the viability ratio.
What is Flow Cytometry? Introduction to Flow Cytometry
Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display. 1 The Dark-Field Microscope Image is formed by light reflected.
New Research Directions in the Rowlen Group Detection of cancer cells in blood Isolation of cancer cells using dielectrophoresis Multispectral analysis.
How to do cervical pap smear
Introduction to Pathology. What is pathology Pathology is the scientific study of disease. In its broadest sense, it is the study of how the organs and.
CHAPTER 37 MLAB 1415 HEMATOLOGY JOANNA ELLIS, MLS(ASCP) Optical Light Scatter and Flow Cytometry.
BACTERIAL STAINING.
RBC Manual Count.
Assays Molecular Diagnostics CSULA. What’s a molecular diagnostic assay? n A laboratory test for the presence or absence of a particular type of molecule.
King Saud University Riyadh Saudi Arabia
1 Module AFB MICROSCOPY. 2 Content Overview Collection of sputum specimens Quality of sputum specimens Processing of sputum specimens Registration of.
What is Flow Cytometry? Flow Cytometry uic Introduction to Flow Cytometry IGC Workshop IGC – Nov 12, 2009 Cell Sorting.
Microscope.
The Role of the Cytology Laboratory
Cytopathology-1 DR. MAHA AL-SEDIK.
Antigen antibody reactions
TIRF Total Internal Reflection Fluorescence Microscopy specialized fluorescence microscopy technique specifically images a very thin optical section (50-250nm)
Cytology and Cytological Techniques
Measurement & Microscopes
Microscopy Compound light microscope is composed of: 1- stand 2- stage 3- substage ( condenser, diaphragm) 4- body tube (carrying lens system) 5- light.
Amal Almuhana  A concentration technique is performed mainly to separate the parasites from fecal debris.  The concentration procedure not only.
2A: Microscopy. Post Lab 2 is assigned today and due by the time your lab meets next. Pre Lab 3 will be available on Wednesday at 5 PM and is also due.
Microarray Technology
Flow Cytometry at Boston University Medical Campus Introduction to some methods that we offer Yan Deng (X4-5225), Gerald Denis (X4-1371),
CHAPTER 3 OBSERVING MICROORGANISMS THROUGH A MICROSCOPE
Introduction To Flow Cytometry By Noha Kamel. Flow cytometry is a method of measuring multiple physical and chemical characteristics of particles by optical.
Dr Gihan Gawish Hydrodynamic focusing is a technique used to provide more accurate results from flow cytometers or Coulter counters for determining the.
Basic Principles in Flow Cytometry
 Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid.
1 Atomic Absorption Spectroscopy Lecture Emission in Flames There can be significant amounts of emission produced in flames due to presence of flame.
Microarray - Leukemia vs. normal GeneChip System.
The ThinPrep ® PapTest ™ PAUL SULTANA MLS Thursday, 22 October 2015.
NucleoCounter YC-100 Presentation
Blood Collection and Handling of Blood Samples. Collecting your Sample Determine which ________________ are needed. Determine the __________________ you.
Mitosis  Is a process of cell division which results in the production of two daughter cells from a single parent cell. The daughter cells are identical.
Flow Cytometry Basic Training. What Is Flow Cytometry? Flow ~ cells in motion Cyto ~ cell Metry ~ measure Measuring properties of cells while in a fluid.
Grading And Staging Grading is based on the microscopic features of the cells which compose a tumor and is specific for the tumor type. Staging is based.
Starter: Microscopes Which image is from the light microsope? How do you know?
What is a Stain A stain is a substance that adheres to a cell, giving the cell color. The presence of color gives the cells significant contrast so are.
Dalia Kamal Eldien Mohammed. introduction  Types of blood specimen include  Venous blood  Capillary blood.
Red Blood Cell (RBCs) Count
CHAPTER 3 OBSERVING MICROORGANISMS THROUGH A MICROSCOPE Dr. Reitano SUFFOLK COUNTY COMMUNITY COLLEGE.
Antigen-Antibody Interactions: Selected Tests (Contd.)
Lab 3 Goals and Objectives :
New Pap Technologies Pap smear screening under scrutiny since late’80s
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
Tools of the Biologist Simple Microscope- Magnifying glass Light Microscope- Using light to produce an enlarged view of the object Magnification- The ratio.
Chapter 1: The Nature of Analytical Chemistry
Identification of Cancer Stem Cells using Flow Cytometry Analysis تعيين الخلايا الجذعية السرطانية باستخدام تحليل التدفق الخلوي Dr. Ayat Al-Ghafari Biochemistry.
Lab 1 – Part 1: Media Types and Uses
Diagnosis of Endoparasitism
Preparing Slides There are 4 different ways of preparing slides – the one you pick is dependent on the specimen and resolution required. The four types.
Preparation of blood film Dalia Kamal Eldien MSc in microbiology Practical NO-4-
1 SOFTWARE FOR VERIFICATION OF HOMOGENEITY RESULTS FROM AUTOMATED URINE ANALYZERS IN REAL TIME Waldemar Volanski Ademir Luiz do Prado Geraldo Picheth et.
CHAPTER 3 OBSERVING MICROORGANISMS THROUGH A MICROSCOPE
Microscope.
Flowcytometry.
Principles of Laboratory Diagnosis of Infectious Diseases
SPECIAL STAINS FOR NUCLEIC ACID.
Bright-Field Microscopy
Practical clinical chemistry
Department of Pathology
Flow Cytometry and Sorting Part 3
Methods of Sample Preparation
The Art of Flow Cytometry
Introduction To Medical Technology
Presentation transcript:

Liquid based cytology Liquid based Cytology

Background The conventional pap smear has been the most successful screening test Screening every 3-5years has resulted in a 70% reduction in incidence

Why LBC (Liquid based cytology) has been introduced ? Continuing improvement to the Cervical screening Programme Limitations of conventional cytology Modernisation of the technique Future benefits Extra tests – HPV, chlamydia, Neiseria Gonorrhoea.

Limitations Of Conventional Smear (From UK studies) False Negative Rate of up to 55%1 Sampling and interpretive errors Ambiguous reports of 6.4%2 70% are truly negative 30% represent more severe abnormality Inadequate specimens of 9.7% 2 1. Hutchinson et al., AJCP, Vol 101-2; 215-219, 2. DOH Statistical Bulletin 2000/2001

Sources Of False Negatives Sampling issues (70%) cells not collected on the sampling device cells collected, but not transferred to the slide Thicker and thinner areas Nuclear feathering artifact Interpretive issues (30%) abnormal cells present on slide but either not seen or misinterpreted Blood/mucus Air drying artifact Blood mucus overlap not preserved well

What does 'Liquid Based Cytology' mean? Literally it means “cytology (the study of cells) through a liquid medium.”

Liquid-based cytology techniques he TransCyt® filter has been plunged into the sample, it rotates at a high speed and facilitates cell and mucus dispersion. A vacuum is then applied to the filter, which collects cells on a 5 μm porosity membrane. A software program allows a homogeneous deposition of cells until saturation. The TransCyt filter is then inverted and a positive pressure allows cells to adhere to an electronegative slide. After insertion of another TransCyt filter and of another slide, the whole procedure may be repeated until the entire sample has been treated.

Technique of LBC Step : I Collection of sample

Specimen Collection PreservCyt Solution. Capped, labeled, and sent to the laboratory equipped with a ThinPrep 2000 Processor . Composition Buffered methanol No active ingredient Storage 15 to 30 C for 6 weeks

Thinprep Processor Cell dispersion Cell Collection Cell Transfer

Thin Prep processor (1)Cell dispersion Swirling the sampling device in the preservation solution Strong enough to separate debris and disperse mucus, but gentle enough to have no adverse effect on cell appearance.

Thin Prep processor (2) Cell Collection A gentle vacuum is created within the ThinPrep Pap Test Filter, which collects cells on the exterior surface of the membrane. Cell collection is controlled by the ThinPrep 2000 Processor’s software that monitors the rate of flow through the ThinPrep Pap Test Filter.

Thin Prep processor (3) Cell Transfer After the cells are collected on the membrane, the ThinPrep Pap Test Filter is inverted and gently pressed against the ThinPrep Microscope Slide. Natural attraction and slight positive air pressure cause the cells to adhere to the ThinPrep Microscope Slide resulting in an even distribution of cells in a defined circular area.

Summary

Advantages of LBC

Conventional smear Conventional Pap Smear (Macroscopic) • The manual smearing method is readily noted macroscopically. Conventional slide experience is transferable to the ThinPrep process. Cytology does not have to be relearned, base knowledge is merely refined.

The prepared slide with the new ThinPrep TP (Macroscopic) • The key difference in presentation is no longer seeing the smear pattern. ThinPrep takes the same material which is concentrated onto the center of the slide in a thin, uniform layer. Specimen preparation is standardized, eliminating the inconsistency associated with manual preparations. The prepared slide with the new ThinPrep

What we see under the microscope conventional smear CP • Microscopically, the uneven distribution of cellular material associated with the CP smear pattern is evident.

TP Same Patient • Tissue architecture is maintained. TP rearranges the relationship of cell groups on the glass slide. A group/sheet of endocervical cells present represents this. What we see under the microscope. Notice the clean back ground and how well the cells are dispersed rendering easier to interpretation.

Disadvantages of LBC

Disadvantages of LBC

Manual membrane based method of LBC Relatively inexpensive method Equipments required- vortexer and laboratory centrifuge

Solution required are; Fixative solution Surepath preservation fluid Lab prepared- water, NaCl, Na Citrate, 10% formalin,alcohol Polymer solution Agarose, polyethylene glycol Alcohol, poly-L-lysene

Procedure Collection of sample Specimen is vortex mixed Centrifuge at 800 g for 10 min Vortex mix Add 1-2 ml of polymer solution to tube Decant fixative and blot excessive fixative Allow 3-6 drops of suspension to glass slide Allow to dry Stain with conv. Pap

Laser Scanning Cytometry

What is the need? Our limited ability to undertake accurate, quantitative measurement of cellular and subcellular factors Established technologies in clinical pathology, including conventional microscope-based histopathology and histochemistry, fluorescence microscopy, flow cytometry and computer-based image cytometry, all have limitations.

Introduction Imaging + cytometric analysis Not random, but event-based. It is closely related to conventional flow cytometry, which also analyzes individual cells that meet certain characteristics (and is also event-based). Both are therefore cytometric techniques Image or scanning cytometry (IC) combines imaging and cytometric analysis in a single technology platform. Rather than randomly imaging an entire field (like a microscope does), it selects, images and measures cells that meet certain user-adjustable criterion (such as size or fluorescence). It is therefore not random, but event-based. It is closely related to conventional flow cytometry, which also analyzes individual cells that meet certain characteristics (and is also event-based). Both are therefore cytometric techniques Unlike conventional flow cytometry, IC usually analyzes cells fixed to a horizontal surface. However, the technology (light sources, detectors, etc.) is very analogous to traditional flow cytometry. With scanning cytometry, imagery becomes a parameter, and relates to the other parameters (scatter and fluorescence).

Limitations of Flow cytometry time-resolved events such as enzyme kinetics cannot be analyzed. Simultaneous study of Morphology of the measured cell is not possible. Cell analysis is at zero spatial resolution. The cell once measured cannot be re-analyzed with another probe(s)

5. Analysis of solid tissue requires cell or nucleus isolation, a procedure that may produce artifacts and loss of the information on tissue architecture. 6.size samples such as fine needle aspirates, spinal fluid, thus, are seldom analyzed by FC. 7. The measured sample cannot be stored for archival preservation.

Introduction 2 manufacturers: CompuCyte Corp. (Cambridge, MA) Olympus Optical Co. (Tokyo), offers many advantages of flow cytometry but has no limitations listed above. The analytical capabilities of LSC, therefore, complement these of FC, and extend the use of cytometry in many applications

Principle

Lasers Lasers Photomultipliers The microscope (Olympus Optical Co.) is the key part of the instrument, and it provides essential structural and optical components (Fig. 1). The specimen deposited on a micro-scope slide on the stage of the microscope is illuminated by laser beams that rapidly scan the slide. The beams from two lasers (argon ion and helium-neon) spatially merged by dichroic mirrors are directed onto the computer controlled oscillating (350 Hz) mirror, which reflects them through the epi-illumination port of the microscope and images through the objective lens onto the slide. The mirror oscillations cause the laser beams to sweep the area of micro-scope slide under the lens. The beam spot size varies depending on the lens magnification, from 2.5 (at 40x) to 10.0 m (at 10x). The slide, with its xy position monitored by sensors, is placed on the computer-controlled motorized microscope stage, which moves at 0.5 m steps per each laser scan, perpendicularly to the scan. Laser light scattered by the cells is imaged by the condenser lens and its intensity recorded by sensors. The specimen-emitted fluorescence is collected by the objective lens and part of it is directed to a charge-coupling device(CCD) camera for imaging. Another part is directed to the scanning mirror. Upon reflection,it passes through a series of dichroic mirrors and optical emission filters to reach one of the four photomultipliers. Each photomultipler records fluorescence at a specific wavelength range, defined by the combination of filters and dichroic mirrors. A light source, additional to the lasers, provides transmitted illumination to visualize the objects through an eyepiece or the CCD camera. Photomultipliers

Interpretation

Thresholding on flow cytometer Setting the threshold (or discrimination) on the flow cytometer allows us to eliminate unwanted cells (like erythrocytes) from the saved analysis

Thresholding (or contouring) on the laser scanning cytometer On the laser scanning cytometer, any event that is above the threshold (usually DNA fluorescence or sometimes forward scatter) is also considered a “cell” and is also displayed for all parameters. The instrument marks such events with a red “contour”, which contains the amount of fluorescent signal above the threshold. Thresholding on the LSC is therefore referred to as “contouring”

Event-based contouring on the iCys The red region is the event or thresholding contour. It is analogous to “thresholding” or “discrimination” on the flow cytometer. It defines the minimum signal intensity that defines a cell. Everything else below it is ignored. Like in cytometry, you need a universal parameter (like scatter or DNA luorescence) as the trigger for threshold contouring The green region is the inte grating contour. You can set this any number of pixels out from the thresholding contour and measure the brightest pixel (Max Pixel), or the total fluorescence (Integral) within the green region. The blue regions are the background contours. You can also set these any number of pixels out from the integrating contour, away from the cell. The signal between them is interpreted as the autofluorescence background, which is subtracted from the other signals.

Parameters studied by LSC Integrated fluorescence intensity The maximal intensity The Integration area The perimeter of the integration contour (in micrometers). The fluorescence intensity integrated over the area of a torus of desired width defined by the peripheral contour located around (outside) of the primary integration contour. 1. Integrated fluorescence intensity representing the sum of intensities of all pixels (pic-ture elements) within the integration contour area. The latter may be adjusted to a de-sired width with respect to the threshold contour (Fig. 2). 2. The maximal intensity of an individual pixel within this area (maximal pixel). 3. The integration area, representing the number of pixels within the integration contour. 4. The perimeter of the integration contour (in micrometers). 5. The fluorescence intensity integrated over the area of a torus of desired width defined by the peripheral contour located around (outside) of the primary integration contour. For example, if the integration contour is set for the nucleus, based on red fluorescence (DNA stained by propidium iodide, PI), then the integrated (or maximal pixel) green fluorescence of FITC (fluorescein isothiocyanate)-stained cytoplasm can be measured separately, within the integration contour (i.e., over the nucleus) and within the periph- eral contour (i.e., over the rim of cytoplasm of desired width outside the nucleus). All above values of fluorescence (1,2,4) are automatically corrected for background, which is measured outside the cell, within the background contour (Fig. 2). 6. The Xy coordinates of maximal pixel locating the measured object on of the microscope stage. 7. The computer clock time at the moment of measurement. The measurements by LSC are relatively rapid; having optimal cell density on the slide up to 5000 cells can be measured per minute. The accuracy and sensitivity of cell fluorescence measurements by LSC are comparable to the advanced flow cytometers

6. The X-Y coordinates of maximal pixel locating the measured object on of the microscope stage. 7. The computer clock time at the moment of measurement.

Applications in Cytology and Pathology

In Cytology It negates disadvantages of Flow cytmetry Image analysis Requires sufficient amount Verification /reproducibility not possible Not for adherent cells Flow cytmetry Experience of cytologist needed Image analysis

Circulating tumor cells Quantification monitoring for potential metastasis/recurrence Molecular studies therapeutic purposes

Apoptosis studies Characterized by certain morphological features e.g. nuclear fragmentation, nuclear condensation Methods e.g. annexin V, loss of transmembrane potential in mitochondria, analysis of nuclear fragmentation, alteration of DNA condensation. Intensity of maximal pixel of fluorescence is a sensitive marker of local hypo- or hyper-chromicity within the cell, reflecting low or high concentration (density) of the fluorescent probe. One of the early applications of LSC along this line was to detect chromatin conden-sation. Namely, DNA in condensed chromatin, such as mitotic or apoptotic cells, shows in-creased stainability (per unit area of chromatin image) with most fluorochromes. Thus, even when integrated fluorescence of the analyzed cells (representing their DNA content) is the same, the intensity of maximal pixel of the cell with condensed chromatin is higher than that of the cell with more diffuse chromatin structure. Maximal pixel of the DNA-associated fluorescence was used as a marker to distinguish mitotic and immediately post-mitotic G 1 cells from interphase cells [5,6]. Although mitotic cells can be recognized by FC using a variety of markers [reviewed in 7], the advantage of this approach by LSC is that a single fluorochrome is used to discriminate between G 1 vs S vs G 2 vs M phase cells.

Immunophenotyping of leucocyte Especially useful when amount of obtained sample is minimum e.g. neonate, critically ill patients Upto 5 flurochromes in < 15 microlit of peripheral blood. Analysis by 2 methods Specific Immunostaining e.g. CD45 DNA staining with 7-AAD- difference in intensity with different leucocytes

Ploidy and DNA index Anueploid number characteristic of a malignant cell e.g. Gastric, colon, kidney, head and neck etc. Prognostic marker in several tumors LSC measures amount of DNA in cell

Tumor cells are identified and gated based upon cytokeratin staining. To verify the morphology of the two populations, cells can be restained with Wright-Giemsa or H & E. Cells from each region can be relocated and visualized by the CompuSort™ process.

Bacterial detection detect live and dead bacteria, estimate cell numbers, and calculate live/dead ratios. The methods are easier and more accurate than traditional, manual counts. propidium iodide (PI), unable to permeate the intact cell membrane of a living cell, but does label dead cells with red fluorescence. SYTO® 16(Molecular Probes) will label the nucleic acids of living cells with green fluorescence. Shown here are E. coli bacteria captured on a membrane and stained with PI and SYTO 16

Disadvantages Expensive instrumentation Exact DNA content of paraffin block is difficult to determine.