CEQ Sequencing and Fragment Analysis Troubleshooting CEQ User Group Meeting Noreen Galvin Ph.D.

Slides:



Advertisements
Similar presentations
Maximizing the Efficacy of Melt Profiling through Stringent PCR Optimization Steven F. Dobrowolski, PhD.
Advertisements

Subcloning Techniques
PCR, Gel Electrophoresis, and Southern Blotting
Troubleshooting DNA Sequences: Guidelines and Suggestions.
Manipulating DNA: tools and techniques
PCR way of copying specific DNA fragments from small sample DNA material "molecular photocopying" It’s fast, inexpensive and simple Polymerase Chain Reaction.
Technology: An Introduction to GA 310 Instrument and troubleshooting.
SEPARATION OF DNA FRAGMENTS BY LENGTH. Organic molecules such as DNA are charged. DNA is negatively charged because the phosphates (red circles) that.
Chemical Synthesis, Sequencing, and Amplification of DNA
FISH 543 / OCEAN 575 Molecular Techniques
Introduction to DNA.
General Genetics. PCR 1.Introduce the students to the preparation of the PCR reaction. PCR 2.Examine the PCR products on agarose gel electrophoresis.
Genomic DNA purification
HIV GENOTYPE ASSAY Anabelia Perez, MLT (ASCP) Molecular Technologist August 6, 2008.
UALR Biosciences Core Facility High Throughput DNA Sequencing Fragment Analyses Real Time PCR In situ Hybridization Proteomics.
Bioinformatics/PCR Lab How does having a certain genetic marker affect chances of getting brain cancer?
Polymerase Chain Reaction
Laboratory: Unit 3: PCR (pages 54-55) Lecture: PCR & primer stock preparation In-Class Writing: abstract for AEM 63: , 1997 (page 68) Hand In: abstract.
CULTURE INDEPENDENT ANALYSIS OF MICROBIAL COMMUNITIES IN SOIL
Polymerase Chain Reaction (PCR)
EDVOKIT#300: Blue/White Cloning of a DNA Fragment
Cat # SL Store at 4~23 0 C DiatoCLEAN™ DNA Purification Kit Quick Protocol Small 300 Preps Large 600 Preps Gaither Drive Gaithersburg,
Recombinant DNA Technology………..
Journal Meeting Jung-Yun Ko DNA Sequencing & ABI DNA Sequencer.
PCR Troubleshooting Virginia Balke
Tina Doss Applied Biosystems
A technique to make a lot of DNA from only a little!
PCR Forensics. Today’s Lab There has been an outbreak of Salmonella poisoning in the Student Union cafeteria at Stanford University cafeteria. You have.
Polymerase Chain Reaction (PCR) Developed in 1983 by Kary Mullis Major breakthrough in Molecular Biology Allows for the amplification of specific DNA fragments.
Molecular Testing and Clinical Diagnosis
Dye-binding/High-resolution Thermal Denaturation PCR-what makes or breaks fragment analysis Steven F. Dobrowolski, PhD.
From the Seed Sample to DNA II: DNA Isolation, Quantification, & Normalization Beni Kaufman.
PCR is used in; Cloning into plasmid vectors DNA sequencing Genetic screening DNA based phylogeny Functional analysis of genes Identification of DNA fingerprints.
The polymerase chain reaction
PPT-1. Experiment Objective: The objective of this experiment is to amplify a DNA fragment by Polymerase Chain Reaction (PCR) and to clone the amplified.
The polymerase chain reaction
Polymerase Chain Reaction (PCR) Nahla Bakhamis. Multiple copies of specific DNA sequences; ‘Molecular Photocopying’
Amplification of a DNA fragment by Polymerase Chain Reaction (PCR) Ms. Nadia Amara.
Molecular Genetic Technologies Gel Electrophoresis PCR Restriction & ligation Enzymes Recombinant plasmids and transformation DNA microarrays DNA profiling.
PCR With PCR it is possible to amplify a single piece of DNA, or a very small number of pieces of DNA, over many cycles, generating millions of copies.
Polymerase Chain Reaction Chromosome 16: PV92Alu PCR TM.
Introduction to PCR Polymerase Chain Reaction
Lab 22 Goals and Objectives: EDVOKIT#300: Blue/White Cloning of a DNA Fragment Calculate transformation efficiencies Compare control efficiency to cloned.
Restriction Enzyme Store in -20 C Digestion. Restriction Enzyme Buffer NaCl or KCl, TrisHCl, MgCl 2, DTT Different salt: varied activity Buffer supplied.
PCR mediated mutagenesis 2013 년도 2 학기 생화학 실험 (2) 5 주차 조교 : 안성원.
Kevin Chen.  A method of amplifying or copying DNA fragments.
Lecturer: Bahiya Osrah Background PCR (Polymerase Chain Reaction) is a molecular biological technique that is used to amplify specific.
Rajan sharma.  Polymerase chain reaction Is a in vitro method of enzymatic synthesis of specific DNA sequences.  This method was first time developed.
HRM Assay and Optimization 1. DNA Quality 2. Amplicon LengthAmplicon  lengths of 100–300 bp 3. Primer Selection 4. Dye Selection 5. MgCl2 Concentration.
PCR The Polymerase Chain Reaction PCR The Polymerase Chain Reaction.
Presented by: Khadija Balubaid.  PCR (Polymerase Chain Reaction) is a molecular biological technique  used to amplify specific fragment of DNA in vitro.
Introduction to PCR Polymerase Chain Reaction
Polymerase Chain Reaction (PCR)
EDVOKIT#300: Blue/White Cloning of a DNA Fragment
DNA Sequencing Techniques
PCR TECHNIQUE
POLYMERASE CHAIN REACTION TECHNIQUES
Primer-Dimer Formation: The Problem and the Solution
Polymerase Chain Reaction
DNA EXTRACTION Protocol and notes 9/17/2018.
BIOTECHNOLOGY BIOTECHNOLOGY: Use of living systems and organisms to develop or make useful products GENETIC ENGINEERING: Process of manipulating genes.
Polymerase Chain Reaction
Lab 8: PTC Polymerase Chain Reaction Lab
Polymerase Chain Reaction (PCR) technique
mRNA Sequencing Sample Preparation
PCR types and Trouble shooting
ChIP DNA Sample Preparation
The polymerase chain reaction
Genomic DNA Sample Preparation
Setting up your Sequencing reaction
Presentation transcript:

CEQ Sequencing and Fragment Analysis Troubleshooting CEQ User Group Meeting Noreen Galvin Ph.D.

Introduction Run Method Summary Troubleshooting Tools Normal traces Current problems Raw data problems Mixed signals Difficult templates

Run Methods Sequencing –LFR-1 = 700bp –LFR-a = >800bp –LFR-b = 600bp –LFR-c = 500bp –Seq-Test = Sequencing Test Sample/pUC only Fragment Analysis –Frag1 –Frag2 –Frag3 = 400bp frags –Frag4 = 600bp frags –SNP-1 = SNP anaysis –Frag-Test = Fragment Test Sample only

Troubleshooting Tools Sequencing Test Mix (Part ) –Instrument –Gel –Capillary –Method pUC control and -47M13 Primer –Reagents –Technique Samples –DNA & Primer These three tests rely on the analysis of Raw Data signal, Current profiles and Analysed Data integrity

Troubleshooting Tools Sequencing Test Mix & pUC18dG/-47 –Method Seq-Test >98% accuracy (2%N) Align with pUC18dG

Sequence Trace

Abnormal Sequence Trace!

Normal Raw Data and Current LFR-1 Method

Current Problems Current problems are associated with the electrokinetic injection of sample and the subsequent separation of the DNA sequencing fragments in the Linear Polyacrylamide (LPA) matrix Electrokinetic Injection –All negative charged molecules loaded DNA sequencing products DNA template RNA Protein Salts

Current Problems Gradually decreasing current => short reads

Current Problems Gradually decreasing current => short reads –Some excess of supercoiled DNA template Re-inject sample Reduce injection Reduce [template] Pre-heat treatment –96°C 1 minute

Current Problems “Crashed” current –Short/failed read –Inaccurate base calling

Current Problems Crashed current –Reduce [template] –Pre-heat treatment 96°C 1 minute Cleave template –Restriction Remove template after reaction –Magnetic beads

Current Problems –Optimizing 96°C –1 minute –2 minutes –3 minutes 85°C –5 minutes

Current Problems Identical current crash in all eight capillaries –Air bubble in manifold Manifold purge and re-run samples

Current Problems Current too high –>10µA on LFR-1 Data starts at 20 minutes –Poor resolution –Gel degraded Excess hours on CEQ

Raw Data Signal Problems Full available range –Min Signal > 3x baseline –Max signal < 130,000 units High Background –Reduced sensitivity with lower signals Signal too low –Short/no read –Minimum signal:noise > 3:1 Signal too high –Base calling errors Too many peaks –Multiple sequencing products in one tube

Raw Data Background If >6000 rfu (relative fluorescence units) –Clean capillary window MQ water 70% Ethanol Blow dry

Normal Raw Data and Current LFR-1 Method

Low Raw Data Signal No Signal & no dye peaks No signal but dye peaks present

Low Raw Data Signal No Signal & no dye peaks –Lost pellet –(Dye degradation) Instrument Problem Centrifuge CEQ No signal but dye peaks present –Chemistry problem [Template DNA] Salts in sample [primer] Poor primer Polymerase inhibitor –Instrument Problem Thermal cycler

Low Raw Data Signal [DNA Template] Too little DNA Template –Not enough reaction products Add more DNA Correct [DNA Template] femtomoles for dsDNA femtomoles for ssDNA femtomoles for PCR products

DNA Quantity Table

Quantitation of [Template DNA] –Agarose Gel –Fluorescence Picogreen –Spectrophotometer

Molecular Weight Markers

Low Raw Data Signal Salt in template or primer –Competition for injection Add less template Purify DNA template Re-synthesize primer

Low Raw Data Signal Primer Problems Low [primer] Inefficient primer binding –Primer design –Cycling conditions Cycle number Annealing temperature

Primer Design Ideal Primer length 22-25bp 50% G/C –Tm >55°C Annealing temp + 5°C Fwd (Tm 43.5) Rev (Tm 79.8)

Primer Design Primer Interactions –Hairpins –Primer dimers

Primer Quality Poor quality primer –Degraded Freeze-thawing Store in MQ water or 10mM Tris –Poor synthesis Check primer integrity on CE/HPLC –Poor purification Minimum = desalting

Low Raw Data Signal Polymerase Inhibitors Polymerase inhibitors –DEPC water –Phenol –EDTA –DNA binding proteins –Other…. Purify Template –Spin columns Store primer –Water or Tris

Low Raw Data Signal Dye Degradation Typically red/black dyes degrade first –Dyes stable in: Reaction buffer SLS under mineral oil –Bad formamide Use CEQ kit SLS (part ) Do NOT freeze-thaw –Heating pellets to dry them Dry in vacuum or air –No mineral oil overlay in CEQ plate Always overlay samples in SLS

Low Raw Data Signal Dye Degradation Typically red/black dyes degrade first

Low Raw Data Signal Ethanol Contamination Typically red dye affected the most Massive noise in the D1 channel Low sample signal

Instrument Issues Thermal Cycler –Heated lid –Sample evaporation with plates/caps –Block problem (PCR may still work) Centrifuge –Refrigerated –Forces > 1000 x g for plates, >10,000 x g for tubes CEQ –Check with sequencing test sample Resolution and accuracy Capillary, gel, instrument

Raw Data Signal Too High When the signal is too high this causes the data to be “truncated” (off-scale) This can result in the addition of erroneous peaks in the base calling Data Truncated (off- Scale)

Raw Data Signal Too High Reduce injection of samples Reduce [DNA template] Reduce cycle number Check [DNA template] before reaction

Correct Raw Data Signal Analysis Poor or Fails Insertions Caused by: –Mixed Templates –PCR Primer Carryover –Primer Mis-priming

Correct Raw Data Signal Analysis Poor or Fails Template contamination

Correct Raw Data Signal Analysis Poor or Fails Template contamination –PCR products: Multiple products –Check on agarose –Purify from agarose –Plasmids More than one colony –Pick another colony –Re-plate

Low Level Contamination of PCR Template

Reduce injection Reduce amount of template Increase annealing temp in PCR reaction Re design PCR primers

Correct Raw Data Signal Analysis Poor or Fails Primer contamination –PCR templates – primer carry-over Purify by spin column or ExoSap Primer degredation –N-1 peaks Fresh primer

Primer Degradation N-1 peaks Pre-peak reduction in S/W Resynthesize primer

Pre-peak Reduction

Difficult Templates Short PCR Products –Difficult to purify G/C rich samples –Form hairpins PolyA or polyT –Enzyme slippage

Short PCR Products (<200bp) Low recovery on Spin columns

Short PCR Products (<200bp) Low recovery on Spin columns –Exo-SAP purification ExonuleaseI + Shrimp Alkaline Phosphatase –0.05u SAP + 0.1u Exo per µl PCR reaction –37°C for 60 minutes –Inactivate at >75°C for 15 minutes –Hold at 4°C

G/C rich templates

Range of Heat Treatments –From 1 to 5 minutes Adjust thermal cycling conditions: –Hot start 96 ° C for two minutes –Thirty cycles of 96 ° C 30seconds 60 ° C 4 minutes –Hold 4 ° C

G/C rich templates Chemistry Modifications –DMSO DO NOT use –Single Strand Binding Protein (Stratagene) 0.5 µg per 20 µl DTCS reaction Better with DTCS than Quickstart –Betaine Final concentration 0.7 to 2M per sequencing reaction Good for both DTCS and Quickstart –DeazaGTP Replace dNTP mix in DTCS kit –Separate components from Sigma, USB etc Add dNTPs to a final concentration of 0.6mM –Use deazaGTP in place of dITP Use 72 ° C extension temperature

PolyA and PolyT Problem is enzyme slippage

PolyA and PolyT Problem is enzyme slippage –Pre-peak reduction may help Sequence specific –Redesign primer position –Try both Quickstart and DTCS chemistry

SEQUENCING Conclusion Troubleshooting Logic –“Big picture” patterns Look for problems common to: –One primer –One DNA template –One purification method –One user! –Etc….. Please ask us for help!