CdTe prototype detector testing Anja Schubert The University of Melbourne 9 May 2011 Updates.

Slides:



Advertisements
Similar presentations
The Factors that Limit Time Resolution in Photodetectors, Workshop, University of Chicago, April 2011 What is known experimentally about timing determinants.
Advertisements

Performance of MPPC using laser system Photon sensor KEK Niigata university, ILC calorimeter group Sayaka IBA, Hiroaki ONO, Paul.
Study of Photon Sensors using the Laser System 05/7/12 Niigata University, Japan Sayaka Iba, Editha P. Jacosalem, Hiroaki Ono, Noriko.
1 Pixel Readout of GEMs Motivation The Setup Results Next Steps A. Bamberger, K. Desch, J. Ludwig, M. Titov, U. Renz, N. Vlasov, P. Wienemann, A. Zwerger.
Optical and Electrical Characterisation of Defects and Charge Transport in CdZnTe radiation detectors P.J. Sellin, S. Rath, M. Breese, A. Hossain, E.J.
Ionization. Measuring Ions A beam of charged particles will ionize gas. –Particle energy E –Chamber area A An applied field will cause ions and electrons.
Slide 1 Diamonds in Flash Steve Schnetzer Rd42 Collaboration Meeting May 14.
28 June 2002Santa Cruz LC Retreat M. Breidenbach1 SD – Silicon Detector EM Calorimetry.
Andrea Giammanco CMS Tracker Week April DS ROD Prototype: “final” optohybrids “final” CCUM integrated in the rod with new FEC_to_CCUM adapter (Guido.
FPIX0 Electronic Test Marina Artuso Paul Gelling Jianchun Wang  The system works fine with charge injection calibration  Gain curve, threshold, and noise.
Why silicon detectors? Main characteristics of silicon detectors: Small band gap (E g = 1.12 V)  good resolution in the deposited energy  3.6 eV of deposited.
Summary of CMS 3D pixel sensors R&D Enver Alagoz 1 On behalf of CMS 3D collaboration 1 Physics Department, Purdue University, West Lafayette, IN
First Results from Cherwell, a CMOS sensor for Particle Physics By James Mylroie-Smith
1 Alessandra Casale Università degli Studi di Genova INFN Sezione Genova FT-Cal Prototype Simulations.
FPIX_flip_chip_test module calibration tests November 14 th 2012.
Silicon Sensor with Readout ASICs for EXAFS Spectroscopy Gianluigi De Geronimo, Paul O’Connor Microelectronics Group, Instrumentation Division, Brookhaven.
Updates on GEMs characterization with APV electronics K. Gnanvo, N. Liyanage, K. Saenboonruang.
Calibration of the ZEUS calorimeter for electrons Alex Tapper Imperial College, London for the ZEUS Collaboration Workshop on Energy Calibration of the.
Paul Sellin, Radiation Imaging Group Time-Resolved Ion Beam Induced Charge Imaging at the Surrey Microbeam P.J. Sellin 1, A. Simon 2, A. Lohstroh 1, D.
M. McConnell1, J. R. Macri, J. M. Ryan, K. Larson, L. -A. Hamel, G
14 December nd CARAT Workshop, GSI, Darmstadt1 Radiation hardness studies with relativistic electrons Sergej Schuwalow, Uni-HH / DESY On behalf of.
Fully depleted MAPS: Pegasus and MIMOSA 33 Maciej Kachel, Wojciech Duliński PICSEL group, IPHC Strasbourg 1 For low energy X-ray applications.
Montpellier, November 15, 2003 J. Cvach, TileHCAL and APD readout1 TileHCAL- fibre readout by APD APDs and preamplifiers Energy scan with DESY beam –Energy.
Summary of CMS 3D pixel sensors R&D Enver Alagoz 1 On behalf of CMS 3D collaboration 1 Physics Department, Purdue University, West Lafayette, IN
Development of an ASIC for reading out CCDS at the vertex detector of the International Linear Collider Presenter: Peter Murray ASIC Design Group Science.
Performance limits of a 55  m pixel CdTe detector G.Pellegrini, M. Lozano, R. Martinez, M. Ullan Centro Nacional de Microelectronica, Barcelona, 08193,
TPC R&D status in Japan T. Isobe, H. Hamagaki, K. Ozawa, and M. Inuzuka Center for Nuclear Study, University of Tokyo Contents 1.Development of a prototype.
1 TRD-prototype test at KEK-FTBL 11/29/07~12/6 Univ. of Tsukuba Hiroki Yokoyama The TRD prototype is borrowed from GSI group (thanks Anton).
1 Report on analysis of PoGO Beam Test at Spring-8 Tsunefumi Mizuno July 15, 2003 July 21, 2003 revised August 1, 2003 updated.
Development of CCDs for the SXI We have developed 2 different types of CCDs for the SXI in parallel.. *Advantage =>They are successfully employed for current.
e- beam 2D sample e- beam 3D
NA62 Gigatracker Working Group 28 July 2009 Massimiliano Fiorini CERN.
Charge Collection and Trapping in Epitaxial Silicon Detectors after Neutron-Irradiation Thomas Pöhlsen, Julian Becker, Eckhart Fretwurst, Robert Klanner,
26 Apr 2009Paul Dauncey1 Digital ECAL: Lecture 2 Paul Dauncey Imperial College London.
Status of photon sensor study at Niigata University -- SiPM and MPPC -- Photon sensor mini workshop 05/9/16 University Niigata University.
JRA-1 ISRSI Consortium Meeting Perugia, INAF/IASF (BO,PA)
SIAM M. Despeisse / 29 th January Toward a Gigatracker Front-end - Performance of the NINO LCO and HCO Matthieu Despeisse F. Osmic, S. Tiuraniemi,
The development of the readout ASIC for the pair-monitor with SOI technology ~irradiation test~ Yutaro Sato Tohoku Univ. 29 th Mar  Introduction.
TCT measurements with SCP slim edge strip detectors Igor Mandić 1, Vladimir Cindro 1, Andrej Gorišek 1, Gregor Kramberger 1, Marko Milovanović 1, Marko.
DESY BT analysis - MPPC Saturation Correction - S. Uozumi Feb Sci-ECAL meeting 1.MPPC Gain Measurement (with LED data) 2.Inter-calibration of readout.
Pixel detector development: sensor
Timing in Thick Silicon Detectors Andrej Studen, University of Michigan, CIMA collaboration.
Construction and beam test analysis of GE1/1 prototype III gaseous electron multiplier (GEM) detector V. BHOPATKAR, E. HANSEN, M. HOHLMANN, M. PHIPPS,
Electron Transmission Measurement of GEM Gate Hirotoshi KUROIWA (Saga Univ.) Collaboration with KEK, TUAT, Kogakuin U, Kinki U, Saga U Introduction Motivation.
MuTr Chamber properties K.Shoji Kyoto Univ.. Measurement of MuTr raw signal Use oscilloscope & LabView Read 1 strip HV 1850V Gas mixture Ar:CO 2 :CF 4.
FBK 3D CMS pixel sensors preliminary lab measurements E. Alagoz 1, A. Krzywda 1, D. Bortoletto 1, I. Shipsey 1, G. Bolla 1, and G. F. Dalla Betta 2, M.
Development of the X-ray Detectors for XTP Cao Xue-lei Institute of High Energy Physics, CAS IHEP, Beijing.
Update on works with SiPMs at Pisa Matteo Morrocchi.
Edgeless semiconductor sensors for large-area pixel detectors Marten Bosma Annual meeting Nikhef December 12, 2011, Amsterdam.
Equalization of Medipix2 imaging detector energy thresholds using measurement of polychromatic X-ray beam attenuation Josef Uher a,b, Jan Jakubek c a CSIRO.
Philip Bambade, Pierre Barillon, Frédéric Bogard, Selma Conforti, Patrick Cornebise, Shan Liu, Illia Khvastunov Journée PHIL
Neutrinoless double beta decay (0  ) CdTe Semico nductor Band gap (eV) Electron mobility (cm 2 /V/s) Hole mobility (cm 2 /V/s) Density (g/cm 3.
PHOTOTUBE SCANNING SETUP AT THE UNIVERSITY OF MARYLAND Doug Roberts U of Maryland, College Park.
Silicon microstrip detector for imaging of fast processes at high intensity synchrotron radiation beam. Budker INP V.Aulchenko1,2, L.Shekhtman1,2, B.Tolochko3,2,
Fabio, Francesco, Francesco and Nicola INFN and University Bari
Jinfan Chang Experimental Physics Center , IHEP Feb 18 , 2011
Preliminary results from 3D CMS Pixel Detectors
Charge collection studies with irradiated CMOS detectors
Marcin Chrząszcz Cracow University of Technology Itamar Levy
Panagiotis Kokkas Univ. of Ioannina
Detection of muons at 150 GeV/c with a CMS Preshower Prototype
(proxy for Harris Kagan) for the RD42 Collaboration
The response function of semiconductor pixel detectors for highly ionizing particles Jan Jakubek, P. Soukup Institute of Experimental and Applied Physics,
BTeV Pixel Silicon Detector
BESIII EMC electronics
High Rate Photon Irradiation Test with an 8-Plane TRT Sector Prototype
LSO: Energy resolution
Beam Test Results for the CMS Forward Pixel Detector
Why silicon detectors? Main characteristics of silicon detectors:
Igor Mandić1, Vladimir Cindro1, Gregor Kramberger1 and Marko Mikuž1,2
Presentation transcript:

CdTe prototype detector testing Anja Schubert The University of Melbourne 9 May 2011 Updates

Abstract Tested the performance of MCSS CdTe prototype. Charge collection efficiency Energy resolution Energy calibration Gain variation Tests run at SAXS&WAXS beamline Updates – Anja Schubert2 9 May 2011

3 2mm CdTe sensor 19 pixels in hexagonal array Signal is passed through preamp then delay-line shaper 8 channels connected read-out at once 14-bit ADC Labview interface Updates – Anja Schubert Charge preamp Pulse Shaper Output Buffer 9 May 2011 det I/p output

CdTe versus Si CdTe Z 48/52 e mobility cm2V/s h mobility cm2V/s E for e/h 4.4 to 4.5eV Higher CCE at high E Need to collect electrons Low h mobility leads to trapped positive charge! Si cm2V/s cm2V/s 3.62 eV Low CCE at high E Can collect holes or electrons 9 May 2011

Pixel Layout Hexagonal pixels with 20um gap. Indium bump bonds to connect to chip are formed on circular pads. Uniform electrode on back of sensor (Pt) 9 May

General set-up 9 May 2011 det X-stage Collimated beam 20x20um SAXS|WAXS table Z stage SAXS|WAXS table Z stage sensor Vbias = - 600V Sensor thickness 2mm

Charge Collection Maps Scanned over a 400x400um region in 20um steps. Beam size 20x20um 1000 triggers per point Peak of each waveform is histogrammed. From this the mean energy is calculated via a Gaussian fit for each point. Gives map of collection efficiency 9 May 2011 Drift timeSensor thickness Charge spread Bias Voltage Hole mobility

CCE map Scanned pixels are scaled to their central value (ie max) and summed 9 May

Charge Sharing From the CCE map we can calculate the charge sharing due to diffusion and drift. Repeated for all possible cross sections: σ = 23±5µm 9 May 2011

Energy Resolution Moved to center of pixel (calculate from CCE map). Gaussian fit to photo peak gives energy resolution as sigma/mean Energies 20keV,18keV,16keV,12keV and 8keV Energy resolution was found to be around 12% at 20keV. 9 May 2011 Incident Energy (keV) Mean Pulse Height (DAC) Standard Dev of Pulse Height (DAC) Energy Resolution Stdv/mean (%) noiseNA E=20keV

Energy Calibration Calibration for Pixel 4. 9 May 2011

Gain Variation E is 20keV Calculated from energy distribution at the center of each pixel Gain variation of 500% 9 May 2011 ChannelPixelMeanStd Dev

Summary Energy resolution was found to be around 12% at 20keV, decreasing as energy was lowered. Charge sharing due to drift and diffusion is up to 28um. Gain variation between pixels 500%. 9 May 2011