Thursday Case of the Day The arrow indicates a region of high apparent FDG uptake posterior to the bladder. What is the cause of this artifact? a) attenuation.

Slides:



Advertisements
Similar presentations
What are PET basics?.
Advertisements

O Nearly all test-retest SUV measurements on the same scanner were within approximately 20% and 1.0 SUV units of each other. On different scanners SUV.
Cardiac Case Study Inferior wall attenuation and correction with CT-based Attenuation Correction Linda Campbell, BS GE Applications Scientist.
Completion of a Truncated Attenuation Image from the Attenuated PET Emission Data Johan Nuyts, Christian Michel, Matthias Fenchel, Girish Bal, Charles.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Introduction to PET/CT in Oncology: Practical Aspects Jeffrey T. Yap, PhD Department of Imaging Dana-Farber Cancer Institute.
Coincidence imaging today
International Atomic Energy Agency L 4 PROTECTION ISSUES IN CLINICAL METHODOLOGY.
Coincidence to Image: PET Imaging Jennifer White Marketing Manager SNS Workshop October 13, 2003.
Procedure Guidelines and Practical Applications For PET \CT Imaging by Dr. H. Hawesa RAD 466-Lecture 7.
Prognostic Value of Tumor Hypoxia,
WELCOME. PET/CT CASE REVIEW COURSE JANUARY 31 ST PASADENA CONVENTION CENTER LOS ANGELES RADIOLOGIC SOCIETY MID WINTER MEETING.
Ultrasound, Positron Emission Tomography, and Single Photon Emission Tomography Allen T. Newton, Ph.D. PAVE 2014.
Nuclear Medicine: Tomographic Imaging – SPECT, SPECT-CT and PET-CT Katrina Cockburn Nuclear Medicine Physicist.
Nuclear Medicine Physics
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part A.
Albert J. Chang, MD, PhD 1 Farrokh Dehdashti, MD 2 Perry W. Grigsby, MD, MS 1 Department of Radiation Oncology 1 Department of Radiology and Nuclear Medicine.
Improved Hybrid PET Imaging Using Variable Focal Length Fan-Slat Collimators Thomas C. Rust and Dan J. Kadrmas, Ph.D. Medical Imaging Research Laboratory.
Nuclear Medicine Physics Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Image Artifacts in Nuclear Medicine Part B.
Monday Case of the Day History: Multiple examinations from a CT scanner using a standard helical scan protocol show diffuse hypointense regions of artifact.
Monday Case of the Day A) The treatment was successful: The bremsstrahlung SPECT (Fig 2) indicates that 90 Y was deposited in the tumor. B) The treatment.
History: A 61-year-old man with suspected thoracic aneurysm underwent ECG-gated chest CT angiogram. He was instructed to hold his breath during the exam.
Introduction to Medical Imaging SPECT, Introduction to Medical Imaging SPECT, PET and Lesion Detection Guy Gilboa Course
Wednesday Case of the Day History: CT pulmonary angiography (image shown) was performed on a 24-year-old female patient with pleuritic chest pain and increased.
18 F-FDG Colonic Uptake by Oral Anti- Diabetic Drugs in PET/CT Scan Sang Gyu Kim, Hoon Hee Park, Jung Yul Kim, Young Kag Bahn Han Sang Lim, Jae Sam Kim,
PET Imaging Positron Emission Tomography
JESÚS SILVA-RODRÍGUEZ, PABLO AGUIAR, INÉS DOMÍNGUEZ-PRADO, MICHEL HERRANZ, ÁLVARO RUIBAL 18F-Choline: Is shine-through effect an issue for prostate SUV.
The Optimization of Reconstruction Method Reducing Partial Volume Effect in PET/CT 3D Image Acquisition Department of Nuclear Medicine, Samsung Medical.
Date of download: 7/8/2016 Copyright © 2016 SPIE. All rights reserved. Illustration of two types of artifacts: (a)–(d) type A and (e)–(h) type B, the first.
Fig FDG PET-CT in patient with cutaneous T-cell lymphoma with treatment response. Baseline (A) and post treatment (B) axial fused PET-CT images demonstrate.
Fig. 2. Arterial fluorodeoxyglucose (FDG) uptake related to atherosclerosis. On positron emission tomography (PET) coronal (A), PET/CT fusion coronal (B),
Tomography for Intraoperative Evaluation of Breast Tumor Margins:
Pulmonary artery stent infection
Correlation of 18F-FDG Avid Volumes on Pre–Radiation Therapy and Post–Radiation Therapy FDG PET Scans in Recurrent Lung Cancer  Nadya Shusharina, PhD,
Chronic pulmonary embolism
Impact of moving target on measurement accuracy in 3D and 4D PET imaging—a phantom study  Yunfeng Cui, PhD, James Bowsher, PhD, Jing Cai, PhD, Fang-Fang.
Thursday Case of the Day
Fig. 4. A 50-year-old woman with invasive ductal cancer involving left breast underwent breast MRI and PET-CT. (a, b) On axial breast MRI, irregular and.
Sunday Case of the Day Physics
Monday Physics Case of the Day Answers (1 of 3) Paul E
Sunday Case of the Day Physics
Monday Case of the Day Physics
Tuesday Case of the Day Physics History: A B MIP
Whole Body 18F-FDG PET/CT Imaging Staging of Malignant Melanoma:
18F-FDG Colonic Uptake by Oral Anti-Diabetic Drugs in PET/CT Scan
Transaxial PET without attenuation correction.
Figure 1 Carotid artery inflammation detected using 18F-FDG-PET
Trousseau Syndrome and the Unknown Cancer: Use of Positron Emission Tomographic Imaging in a Patient With a Paraneoplastic Syndrome  John A. Batsis, MBBCh,
Monday Case of the Day Physics
Wednesday Case of the Day
Wednesday Case of the Day
Future of Thoracic PET Scanning
Tumor Imaging Via Indium 111–Labeled DTPA– Adenosylcobalamin
MPHY8149: Image Registration
Nat. Rev. Urol. doi: /nrurol
Carmel G. Cronin, Michael Moore, Michael A. Blake 
PSMA-RADS Version 1.0: A Step Towards Standardizing the Interpretation and Reporting of PSMA–targeted PET Imaging Studies  Steven P. Rowe, Kenneth J.
Future of Thoracic PET Scanning
[68Ga]Pentixafor PET/CT and [18F]FDG PET/CT A–DMaximum intensity projections (MIP) of [68Ga]Pentixafor (A) and [18F]FDG PET/CT (B) of a 68‐year‐old male.
Volume 11, Issue 2, Pages (April 2016)
Thursday Case of the Day
MR-PET of the body: Early experience and insights
The role of positron emission tomography for non-small cell lung cancer  Albert J. Chang, MD, PhD, Farrokh Dehdashti, MD, Jeffrey D. Bradley, MD  Practical.
Ustekinumab treatment is associated with decreased systemic and vascular inflammation in patients with moderate-to-severe psoriasis: Feasibility study.
The influence of biological and technical factors on quantitative analysis of amyloid PET: Points to consider and recommendations for controlling variability.
Evaluation of Average CT to reduce the artifact in PET/CT
Fig. 5. A 62-year-old woman with mixed invasive ductal and mucinous cancer involving left breast underwent breast MRI and PET-CT. (a) On axial breast MRI,
A 54-year-old man with biopsy-proved hyperplastic hematopoietic bone marrow in the T11 vertebral body. A 54-year-old man with biopsy-proved hyperplastic.
Examples of imaging data in patients.
Figure 12a Pitfalls in N staging at CT and PET
Presentation transcript:

Thursday Case of the Day The arrow indicates a region of high apparent FDG uptake posterior to the bladder. What is the cause of this artifact? a) attenuation correction b) scatter correction c) reconstruction filter d) randoms correction History: A patient underwent a PET/CT study to investigate breast cancer Brad Kemp, Ph.D. Department of Radiology, Mayo Clinic, Rochester, MN Physics Figure 1: Transaxial IR CTAC image Scanning protocol: 15 mCi FDG, 65 min uptake 3D acquisition with 2 min/bed on a GE Discovery 690 PET/CT system Fully-3D iterative reconstruction (IR) with CT attenuation correction (CTAC), scatter correction and randoms from singles correction

Findings: The correct answer is a) This artifact is caused by the attenuation correction as there is a misregistration between PET and CT images. The motion of gas within the patient’s rectum between the CT and the PET images causes a mismatch in the attenuation distribution and a subsequent error in the attenuation correction. Note that this misregistration is due to internal organ motion and not gross patient motion. The motion of gas is apparent in the fused PET/CT image in Figure 2B. The PET images are overcorrected for attenuation in the region where the low-attenuation gas pocket was located for the PET acquisition. The IR NAC PET image shows symmetric perirectal activity (Figure 2C). Figure 2: (A) Transaxial IR CTAC PET image. (B) Transaxial fused PET/CT image. (C) Transaxial PET image with no attenuation correction (NAC). A B C

Lodge conducted an experiment to show that the artifact is caused by an attenuation correction error due to motion between CT and PET acquisitions [1]. See Figure 3 and 4. The results of the experimental suggest that the movement of gas within the rectum causes differences in the attenuation distribution between the CT and PET acquisitions. As such, the hot streak represents a region that is over-corrected for attenuation. Figure 4: Reconstructed PET images of the body phantom with increasing amounts of misregistration between PET and CT data. Hot and cold artifacts, as shown by the red arrows, increase as the misregistration increases. Discussion Figure 3: CT image showing body phantom with two cylindrical inserts. The upper insert contained radioactive water while the lower insert was a low attenuation medium. Red arrow shows direction of motion of PET distribution prior to attenuation correction.

Figure 5 shows images from a PET/CT exam in which the gas pocket within the rectum is stationary; the PET and CT images are properly registered and there is no error in the attenuation correction of the PET images. Discussion Figure 5: (A) Transaxial IR CTAC PET image. (B) Transaxial fused PET/CT image. (C) Transaxial IR NAC PET image. Reference Lodge MA, Muhammad AC, Udall DN, Wahl RL. Characterization of a perirectal artifact in 18 F-FDG PET/CT. J Nucl Med 2010;51: A B C