Patent Ductus Arteriosus Occlusion Device Christina Mathieson Trung Nguyen Advisor: Dr. Tom Doyle BME 273 – Senior Design Project.

Slides:



Advertisements
Similar presentations
CONGENITAL HEART DISEASE.
Advertisements

Acyanotic Heart Disease PRECIOUS PEDERSEN INTRODUCTION Left to right shunting lesions, increased pulmonary blood flow The blood is shunted through.
Sheep Heart Dissection. Before:  Draw what you think a heart would look like if it was sliced in half from top to bottom.
Congenital Heart Disease Cheston M. Berlin, Jr., M.D. Department of Pediatrics.
Congenital Heart Defects Fred Hill, MA, RRT. Categories of Heart Defects Left-to-right shunt Cyanotic heart defects Obstructive heart defects.
TOPIC 2 Group A.
Acyanotic Congenital Heart Disease
Congenital Cardiac Defects
Website: – Munro – Cadaver Dissection – Ligation 1940 – 50 years later surgical Rx. PDA closure 1971 – Cather based closure Rx. Options.
Patent Ductus Ateriosis PDA Muhammad Syed MD Heart.
Paediatrics Revision Session Cardiology
PATENT DUCTUS ARTERIOSUS By: Nicole Stevens. Patent Ductus Arteriosus is a functional connection between the pulmonary artery and the descending aorta.
Introduction to cardiac conditions
2 Heartworms and Heart Murmurs Partnership for Environmental Education and Rural Health Texas A&M University An introduction to problems.
Congenital Heart Disease
Congenital Heart Defects. Eight out of every 1,000 infants have some type of structural heart abnormality at birth. Such abnormalities, known as congenital.
Congenital Heart Defects
Developmental Defects of Cardiovascular System
The Artificial Heart: A Design Example BIOE 1000 October 18, 2001.
Copyright © 2013 Wolters Kluwer Health | Lippincott Williams & Wilkins Chapter 40 Nursing Care of the Child with a Cardiovascular Disorder.
Principal Groups of CHD
Islamic University of Gaza Faculty of Nursing Pediatric Nursing
CONGENITAL DISEASES Dr. Meg-angela Christi Amores.
Introduction to Ventricular Assist Device (VAD)
Heart Disease An introduction to problems that can occur in the heart Partnership for Environmental Education and Rural Health Texas A&M University
INTRODUCTION A 35 year old woman with transposition of the great arteries repaired with a Mustard procedure attends your clinic for annual follow-up. Her.
Heart Pathology Chapter 11. Arrhythmia Any abnormal rhythm of the heart Atrioventricular-impulses are not through the Av node to the bundle of His –Causes.
CASE 2 Group B. Baby Fiona Goodchild was born on May 3, It was noted that she had a low birth weight of 2600g. Normally, the average weight of a.
Congenital heart disease
3/10/2003 Patent Ductus Arteriosus Occlusion Device Oral Presentation #3 Group 6 David Brogan, Darci Phillips & Daniel Schultz Advisor: Dr. Thomas Doyle.
- Describe the clinical features that point to the presence of a congenital heart malformation. - Describe the general classification of heart diseases.
Vanessa Beretta & Dan Fleming. About CHD A congenital heart defect also known as CHD is a defect in the structure of the heart and great vessels. Most.
Congenital Heart Disease Dr. Raid Jastania. Congenital Heart Disease 8 per 1000 live birth Could be minor defect or major defect Cause – unknown –Genetic:
Patent Ductus Arteriosus Occlusion Device David Brogan *, Darci Phillips *, Daniel Schultz *, Dr. Thomas Doyle ‡ * Vanderbilt University Biomedical Engineering.
Cardiovascular surgery, Congenital heart disease Dr. Robin Man Karmacharya, Lecturer, Department of Surgery, Dhulikhel Hospital.
Congenital Cardiac Lesions. Overview Three Shunts of Fetal Circulation Ductus Arteriosus Ductus Arteriosus Protects lungs against circulatory overload.
Patent Ductus Arteriosus
Pulmonary Flow Resistive Device Taya Furmanski Albert Attia Advisor: Thomas Doyle, M.D. April 9, 2003.
Patent Ductus Arteriosis (PDA)
Cardiothoracic Surgery. Topics Valvotomy and valve replacement Open heart surgery and cardiac bypass surgery Correction of congenital heart diseases Heart.
Fetal Circulation Mike Clark, M.D.. Figure (a) Day 20: Endothelial tubes begin to fuse. (b) Day 22: Heart starts pumping. (c) Day 24: Heart continues.
Fetal Circulation. Salient Features of Fetal Growth Placenta is very active.Placenta is very active. Liver and lungs are passive.Liver and lungs are.
DR AFTAB YUNUS FRCSEd. CHAIRMAN CARDIAC SURGERY
(Lub dub, lub dub).  You will need to be able to label the following parts of the circulatory system AND know what their “job” is:  Left and right atrium.
Patent Ductus and Vascular Occlusion Devices Ornob Roy Elaine Isom Michelle Carlsen Brad Klosterman Advisor: Dr. Thomas Doyle.
The Child with a Cardiovascular Disorder
Patent Ductus and Vascular Occlusion Devices Michelle Carlsen Elaine Isom Brad Klosterman Ornob Roy Advisor: Dr. Thomas Doyle.
Lecture II Congenital Heart Diseases Dr. Aya M. Serry 2015/2016.
Transposition of the Great Arteries. What is TGA 1.Congenital heart defect 2.Large blood vessels that carry blood from the heart to the lungs are improperly.
Pulmonary Flow Resistive Device Taya Furmanski Albert Attia Advisor: Thomas Doyle, M.D. March 17, 2003.
Hypoplastic Left Heart Syndrome By: Tyler Nickels, Amanda McKellar, Kassie Herp, Zachary Zwiernikowski, & Amanda Mathy.
Patent Ductus Arteriosus (PDA) March of Dimes Fall 2008 November 11, 2008.
Patent Ductus and Vascular Occlusion Devices Michelle Carlsen Elaine Isom Brad Klosterman Ornob Roy Advisor: Dr. Thomas Doyle.
H. Amoozgar, MD Professor of pediatric cardiology Shiraz University of Medical Sciences, Shiraz, Iran TRANSCATHETER CLOSURE OF LARGE CORONARY-CAMERAL FISTULAE.
Chris Burke, MD. What is the Ductus Arteriosus? Ductus Arteriosus  Allows blood from RV to bypass fetal lungs  Between the main PA (or proximal left.
Congenital Heart Disease
Occlusion Device for Patent Ductus Arteriosus Christina Mathieson & Trung Nguyen Advisor: Dr. Tom Doyle BME 273 – Senior Design Project.
2/14/2003 Patent Ductus Arteriosus Occlusion Device Oral Presentation #2 Group 6 David Brogan, Darci Phillips & Daniel Schultz Advisor: Dr. Thomas Doyle.
4/11/2003 Patent Ductus Arteriosus Occlusion Device Oral Presentation #4 Group 6 David Brogan, Darci Phillips & Daniel Schultz Advisor: Dr. Thomas Doyle.
Heart.
Issues and Current Situations in the Development of Endovascular Treatment Devices for Pediatric Cardiology in the US – US Industry Dan Gutfinger, MD,
Cardiac Manifestation of DiGeorge Syndrome
CONGENITAL HEART DISEASES II
Flow of the Blood Through the Heart
Trans catheter closure of Preterm Ductus Arteriosus
Patent Ductus Arteriosus Occlusion Device
Medical Therapeutics: November 3, 2017
Patent Ductus Arteriosus Occlusion Device Oral Presentation #1
CHAMPION Trial design: Patients with recent hospitalization for heart failure were implanted with a pulmonary artery pressure monitor and randomized so.
Circulatory system Quiz
Presentation transcript:

Patent Ductus Arteriosus Occlusion Device Christina Mathieson Trung Nguyen Advisor: Dr. Tom Doyle BME 273 – Senior Design Project

Patent Ductus Arteriosus: General Information Connection of aorta and pulmonary artery Connection of aorta and pulmonary artery Functionally beneficial prior to birth Functionally beneficial prior to birth Maintained by low oxygen and high prostaglandins Maintained by low oxygen and high prostaglandins Usually closes within hours of birth Usually closes within hours of birth Abnormal if patent after 3 months Abnormal if patent after 3 months 1 in in 2500 Female:male 2:1 Female:male 2:1

Patent Ductus Arteriosus: Causes All factors unknown All factors unknown Prematurity of infant Prematurity of infant Rubella infection Rubella infection Random Random

Patent Ductus Arteriosus: Effects Increased volume load on left atrium and ventricle Increased volume load on left atrium and ventricle Distention and dilation Distention and dilation Delay in postnatal pulmonary vascular changes Delay in postnatal pulmonary vascular changes Growth retardation, breathlessness, poor feeding Growth retardation, breathlessness, poor feeding Congestive heart failure Congestive heart failure Bacterial endocarditis, aneurysm of ductus Bacterial endocarditis, aneurysm of ductus

Patent Ductus Arteriosus: Shapes Type A: Cone Type D: Aneurismal Type B: WindowType C: Tube Type D: Irregular Cone

Current Treatments: Surgery Left posterolateral thoracotomy Left posterolateral thoracotomy Ligation or dissection Ligation or dissection Chest tube 24hrs Chest tube 24hrs Hospital 6-7 days Hospital 6-7 days 6-8 month recovery 6-8 month recovery

Current Treatments: Coils Most common method Most common method Use multiple coils Use multiple coils Materials: Materials: Surgical grade stainless steel Surgical grade stainless steel Dacron fibers Dacron fibers Problem: Problem: Cannot occlude smaller PDAs Cannot occlude smaller PDAs Easily dislodge when too many coils Easily dislodge when too many coils

Current Treatments: Rashkind Double Umbrella Double Umbrella Polyurethane fabric on stainless steel frame Polyurethane fabric on stainless steel frame Wire frame limits sizes Wire frame limits sizes Must pre-order based on approx. size of PDA Must pre-order based on approx. size of PDA

Our Goals PDA occlusion device PDA occlusion device Delivered via 6 – 7F catheter Delivered via 6 – 7F catheter Effective for all ages Effective for all ages Conforms to various shapes of PDA Conforms to various shapes of PDA Able to retract and reposition Able to retract and reposition Biocompatible material (foam) Biocompatible material (foam) Low cost Low cost

Project Timeline DecJanFebMarApr Research****** Advisor Meeting ***** Material Decision ** ** Design****** Prototype****** Testing*** Currently, have basic design idea. Goal for this month: develop testing prototype, determine material, create delivery system

Work Completed Research Research PDA background: current treatments, material ideas, catheter sizes PDA background: current treatments, material ideas, catheter sizes Device shapes: Rashkind, coils, etc. Device shapes: Rashkind, coils, etc. Determined Shape: Mushroom Determined Shape: Mushroom Ideas: Ideas: Delivery system Delivery system Materials Materials

Mushroom Shape Flat top in Aorta Flat top in Aorta Conical / Tubular base in PDA Conical / Tubular base in PDA Top prevents dislodging Top prevents dislodging Base provides surface for coagulation Base provides surface for coagulation

Delivery System 6-7 F catheter 6-7 F catheter Similar to Rashkind implantation procedure Similar to Rashkind implantation procedure Enter PDA via Pulmonary Artery Enter PDA via Pulmonary Artery

Future Work Modify mushroom shape to prevent puckering Modify mushroom shape to prevent puckering Improve delivery system idea Improve delivery system idea Finalize material selection Finalize material selection Build prototype Build prototype Look into methods of testing device Look into methods of testing device