Brightfield Contrasting Techniques Kurt Thorn NIC.

Slides:



Advertisements
Similar presentations
Adjusting a Microscope 1Center components on optic axis 2Focus objective 3Focus condenser 4Adjust illumination lamp voltage (intensity) iris diaphragm.
Advertisements

Fundamentals of Photoelasticity
Microscopy Do you want a footer?.
Biology 177: Principles of Modern Microscopy Lecture 09: Polarization and DIC.
IPC Friedrich-Schiller-Universität Jena Contrast modes in light microscopy: Bright field.
Microscopy Outline 1.Resolution and Simple Optical Microscope 2.Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3.Newer.
Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display. 1 The Study of Microbial Structure: Microscopy and Specimen.
Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display. 1 The Dark-Field Microscope Image is formed by light reflected.
Copyright © The McGraw-Hill Companies, Inc. Permission required for reproduction or display. 1 Microscopy.
ERT107 MICROBIOLOGY FOR BIOPROCESS ENGINEERING Pn Syazni Zainul Kamal PPK Bioprocess.
POWERPOINT PRESENTATION ON POLARISED MICROSCOPE
Basic Illuminating Light Paths and Proper Microscope Alignment E. D. Salmon University of North Carolina at Chapel Hill.
VII Contrasting Techniques From Brightfield to Plas-DIC December 2008 Rudi Rottenfusser.
Blizard Advanced Light Microscopy club Making friends with your microscope.
Polarization of Light Waves
Microscopy Boot Camp /08/25 Nikitchenko Maxim Baktash Babadi.
II From “Beam Paths” to the “Microscope” October 2008 Rudi Rottenfusser – Carl Zeiss MicroImaging.
Properties of Multilayer Optics An Investigation of Methods of Polarization Analysis for the ICS Experiment at UCLA 8/4/04 Oliver Williams.
c = km/sec I F = I 0 x (cosθ) 2.
Simple Microscope Objective magnification Working Distance Eyepiece magnification
USE AND CARE OF THE MICROSCOPE LECTURE 1. MICROSCOPY u Light Microscopy: any microscope that uses visible light to observe specimens u Compound Light.
Theory & Appl. Light Microscopy Phase Contrast Optics.
Light Microscopy Sarah Heintz. Compound Microscope The microscope uses a lens that is close to the object and uses light to focus on the real image of.
Diffraction vs. Interference
3: Interference, Diffraction and Polarization
Microscope.
Chapter 5 Jones Calculus and Its Application to Birefringent Optical Systems Lecture 1 Wave plates Wave plates (retardation plates) are optical elements.
Review: Laws of Reflection and Refraction
Microscopy.
Nano-Electronics S. Mohajerzadeh University of Tehran.
Analytical Tools Microscopy Chapter 4 ©2010 Elsevier, Inc.
Identification of minerals with the petrographic microscope
TIRF Total Internal Reflection Fluorescence Microscopy specialized fluorescence microscopy technique specifically images a very thin optical section (50-250nm)
Microscopy. Scale Lenses and the Bending of Light light is refracted (bent) when passing from one medium to another refractive index –a measure of how.
Honors Microbiology: Chapter 3 Microscopy and Staining
FNI 2B OM 1 Optical Microscopes. FNI 2B OM2 Outline Justification History Components of the Optical Microscope Theory of operation  Basic Microscope.
Microscopy 1. UNITS OF MEASUREMENT 1 m = 1000 mm (millimeters) 1 m = 1000 mm (millimeters) 1000 mm = 1 µm (microns) 1000 mm = 1 µm (microns) Bacteria.
Agenda How to make the specimen visible – Definition of Contrast
Basic Microscopy – An Overview – October 2005 Protistology Course MBL, Woods Hole, MA.
Advanced Biology Visualizing Cells. The Human Eye  Resolution – The minimum distance two points can be apart and still be distinguished as two separate.
Interference in Thin Films, final
Basic Microscopy – An Overview – October 2005 Protistology Course MBL, Woods Hole, MA.
Light Microscopy By: Nicole Sullivan.
Unit 12: Part 1 Physical Optics: The Wave Nature of Light.
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Contrast Microscopy Koehler Illumination Dark Field Phase Contrast Differential Interference Contrast Hoffman Modulation.
Chapter 38 Diffraction Patterns and Polarization.
Announcements HW set 10 due this week; covers Ch (skip 24.8) and Office hours: Prof. Kumar’s Tea and Cookies 5-6 pm today My office hours.
Elliptical polarization. Linear polarization the two orthogonal components are in phase.
Principles and Practices of Light Microscopy II: Brightfield optics, resolution, and aberrations.
Brightfield Contrasting Techniques Kurt Thorn Nikon Imaging Center University of California, San Francisco USA.
Phase contrast and DIC Gerd A. Guenther 2011 Honorable Mention Nikon Small World photomicrograhy competition Image of a freshwater ciliate. Ryan McGorty,
Introduction to Light Microscopy
Chapter 24 Wave Optics. Young’s Double Slit Experiment Thomas Young first demonstrated interference in light waves from two sources in Light is.
(Image: T. Wittman, Scripps) Introduction to Light Microscopy.
(Image: T. Wittman, Scripps) Introduction to Light Microscopy.
 Bright-field  Dark-field  Phase Contrast  Fluorescence.
Microscopy.
Microscopy Group 2 Cabatit, Mendoza, Ramos, Rodriguez, Tan.
Chapter 5 Jones Calculus and Its Application to Birefringent Optical Systems Lecture 1 Wave plates Wave plates (retardation plates) are optical elements.
By c.Keerthana.  First described by Dutch physicist frits Zernike in  It is a type of light microscopy.  It is a contrast enhancing optical technique.
Light Microscope Terms and Practices.
Microscope.
Two-Beam Interference
METALLURGICAL MICROSCOPE
3.3 Other types of microscopy
ساختمان فیزیکی الیاف دکتر مصطفی یوسفی
LIGHT MICROSCOPY basic
LIGHT MICROSCOPY variations
Elliptical polarization
Presentation transcript:

Brightfield Contrasting Techniques Kurt Thorn NIC

Generating contrast in light microscopy Problem: Many biological specimens are thin and transparent and difficult to see. Solution: –Fluorescent staining –Brightfield contrasting techniques: DIC, Phase, others BrightfieldPhase ContrastDIC

Review: The Trans-illumination Microscope Sample Objective Tube lens Intermediate image plane Projection Eyepiece Object plane Back focal plane (pupil) Secondary pupil plane Camera Final image plane Imaging path Aperture iris Field iris (pupil plane) (image plane) (pupil plane)Light source Illumination path Collector Condenser lens Field lens The aperture iris controls the range of illumination angles The field iris controls the illuminated field of view

Reminder: Interference In phase Opposite phase + + = = constructive interference destructive interference

Amplitude and Phase Samples Initial beam, unpurturbed Amplitude sample Phase sample

Amplitude and Phase Samples Initial beam, unpurturbed Amplitude sample Phase sample

Optical path length (OPL) difference = t (n s -n m ) = 1 micron ( ) =.05 microns = 50 nm, which is about 1/10 the wavelength of green light (488 nm) t = sample thickness, ~ 1 micron n s = sample refractive index, ~ 1.38 n m = medium refractive index, ~ 1.33 Many biological samples are phase samples OPL nsns nmnm Higher refractive indices slow down light

Amplitude and Phase Samples Represented as Vectors Vector length: amplitude Rotation around circle: phase Only amplitude changes Only phase changes

How can we see a phase change? Problem: optical detectors are not phase sensitive; they only detect changes in amplitude Decompose phase shift into an unpurturbed wave and a perturbed wave

Converting a phase change to an amplitude change Unpurturbed wave (Surround wave) Perturbed wave (Diffracted wave) Resulting wave (Particle wave) Imagine shifting phase of surround wave by 90° Resulting wave is now reduced in amplitude, and this amplitude change is detectable

Phase contrast microscopy Need to selectively shift the phase of the surround wave. How to do this? The sample will scatter light in all directions, so if we illuminate with a small range of angles we can specifically alter the phase at those angles

The phase contrast microscope Sample Objective Tube lens Projection Eyepiece Camera Imaging path Aperture iris Field iris Intermediate image plane Object plane Back focal plane (pupil) Secondary pupil plane Final image plane (pupil plane) (image plane) (pupil plane) Light source Illumination path Collector Condenser lens Field lens

The phase contrast microscope Sample Objective Tube lens Projection Eyepiece Camera Imaging path Aperture iris Field iris Light source Illumination path Collector Condenser lens Field lens Illumination phase ring Defines range of illumination angles Phase contrast objective Selectively retards surround light

Phase Contrast in the Microscope Phase ring is typically built into objective

Phase Contrast – Further Refinements Typically, phase contrast ring in objective attenuates surround wave by ~75% Enhances visibility of small phase differences Small amplitude change Large amplitude change

Phase Contrast – Further Refinements Negative phase contrast –Retards surround wave, so objects which advance phase (low refractive index) are brighter Positive phase contrast –Advances surround wave, so objects which retard phase (high refractive index) are brighter

Back focal plane Front focal plane Alignment of phase rings

Limitations of Phase Contrast Halos result from diffracted light that is intercepted by phase ring Shade off is caused by greater diffraction at edges of objects than their centers

Limitations of Phase Contrast Poor optical sectioning due to limited illumination aperture. For sufficiently thick samples, can get more than 360°phase shift, meaning thin and thick regions will be identical in contrast.

Darkfield Microscopy Idea: throw away all non-diffracted light by illuminating with a condensor annulus at higher NA than the objective. No illumination light makes it through the objective, only diffracted light. Sample Objective Aperture iris Field iris Object plane Back focal plane (pupil) (pupil plane) (image plane) (pupil plane)Light source Collector Condenser lens Field lens

Darkfield Microscopy Ideal for small particles, unstained microorganisms, etc. Can easily see subdiffraction particles (e.g. 40 nm gold) Not so good for large objects

The idea: Use two beams and interferometry to measure the path length difference between adjacent points in the sample Differential Interference Contrast (DIC)  OPL nsns nmnm

What DIC accomplishes Converts relative differences in optical path length to differences in amplitude

Features of a DIC image 1.Contrast is directional 2.Contrast highlights edges 3.One end brighter, other is dimmer giving a pseudo – 3D image

Polarization Polarization: orientation of E-field. Most light sources produce unpolarized light – no preferred polarization angle

Polarizers Polarizers specifically transmit one polarization angle of light Crossed polarizers transmit no light X

Interference and polarization In phase + = linearly polarized = + circularly polarized Phase lag

Birefringence Birefringent materials have different indices of refraction for light polarized parallel or perpendicular to the optical axis. For unpolarized light two beams with orthogonal polarization are produced illuminated not along optical axis

Wollaston / Nomarski Prisms Two pieces of cemented calcite / quartz Produce orthogonally polarized beams propagating at different angles Placed at a back focal plane, this produces the two beams that will probe the OPL difference of our sample. 1f1f

The differential interference contrast (DIC) microscope Sample Objective Tube lens Projection Eyepiece Camera Imaging path Aperture iris Field iris Light source Illumination path Collector Condenser lens Field lens Wollaston Polarizer (analyzer) Polarizer

How DIC generates contrast Both beams see same OPL Emerge in phase Regenerate initial polarization No light makes it through analyzer X

How DIC generates contrast Beams see different OPL Right beams is phase retarded Generate elliptical polarization Light makes it through analyzer

How DIC generates contrast X Both beams see same OPL Emerge in phase Regenerate initial polarization No light makes it through analyzer

Role of Bias in DIC

Bias adjusment in de Sénarmont DIC

DIC is sensitive to specimen orientation

Can’t plate cells on or or cover cells with plastic. DIC doesn’t work on birefringent samples Phase DIC

Phase DIC is higher resolution than phase contrast

To provide cellular or organismal reference. Phase and DIC are more general (and less toxic) than fluorescence. Phase and DIC do degrade fluorescence performance slightly Combining Phase / DIC with fluorescence

Further reading micro.magnet.fsu.edu Douglas B. Murphy, “Fundamentals of Light Microscopy and Electronic Imaging” Hecht, “Optics” Acknowledgements Orion Weiner / Mats Gustafsson