Using a Digital Data Logger North Carolina Immunization Program (NCIP)

Slides:



Advertisements
Similar presentations
Product Presentation. WHAT IS IT? Product Presentation.
Advertisements

Laptop Usage & Care Asad M. Nafees. Inserting a Battery Pack To insert a battery pack into the battery bay: 1. Turn the notebook upside down on a flat.
MHSL - Dolphin Training Class
Using the camera and the wireless microphone Howard Goldbaum 9/17/2012.
Georgia Vaccines for Children Program
Olympus DM-20 Digital Voice Recorder We’re high tech now, baby!
Working with Data Loggers. Data Logger Loaner Program In order to loan loggers from Sensor Switch, please visit then.
LogTag Training January 2015
Vaccine Storage and Handling Webinar Is Your Vaccine Safe? Presentation to: Georgia Chapter of American Academy of Pediatrics Presented by: Sherri Grady,
TLA. Replacing The Battery On the Back of the pager press in this button while moving the door away from the pager, then lift up. On the Back of the pager.
 Description of PressurePro System  Identify system components  Display and Operational Modes  Sensor alerts  Checking of tire pressures  Manual.
DR-2000 Customer Benefits Security; Tamper resistant programming protects your customer and you! Low maintenance, long battery life & long squeeze tube.
How to Perform Preventive Maintenance of the Personal Computer Properly Shut Down of the Computer Properly Shut Down of the Computer Unplug all Connection.
PDF Logger. Smaller than a credit card Begins recording with a button press Stops with a button press Accurate temperature recording Use inside a container.
If you use your computer more than a few hours per day, we suggest you not to turn off your computer at the end of the day. Instead turn off only the.
HOW WELL DO YOU KNOW THE BASICS OF USING YOUR COMPUTER?
Kaleida Health Point of Care Glucometer Testing 2011
70-290: MCSE Guide to Managing a Microsoft Windows Server 2003 Environment Chapter 2: Managing Hardware Devices.
DataLogger Operation - 1 DataLogger ® Operation Procedures.
Installing Fridge-tag® 2
EPOCH 1000 File Management Data Logging and Reporting
(Smart Loop System) Installation Guide. Rev B Shipping Box Contains Everything that is included in the box being shipped to a customer should contain.
Water Quality Supplies Waste bottle 1413 solution pH 10 solution pH 4 solution Solution cups Water sample cup Hanna Meter.
AccuMax Multi-Point Injection Mechanics
Computer Storage Devices Principles of Information Technology Lytle High School Click to continue.
Heart Rate The Heart Rate program allows the user to set a target heart rate. The system will alter it’s intensity (RPM’s) to maintain the targeted rate.
Welcome to the Milagrow Universe
DEMONSTRATION FOR SIGMA DATA ACQUISITION MODULES Tempatron Ltd Data Measurements Division Darwin Close Reading RG2 0TB UK T : +44 (0) F :
What is the It is the Next Generation, Calibration Station for the GasBadge ® Plus Personal Monitor
COBAE - Smart Classroom Technology Vic Catalasan For immediate in-class assistant please call X1500 or
9000U Quick Guide Marketing & RMA Department XELTEK 8/28/2015.
TRANSMILLE PATs The New Alternative To Fast, Easy & Efficient Portable Appliance Testing.
Submitted by Ravi Prakash PGT(Computer Sc.) KV39 GTC Varanasi
TrekView DMS Instructions. 1. Log in to TrekView DMS.
The Mavica for the Maverick: A How-to Guide Operating instructions for the Sony Mavica CD350 Digital Camera An Easy-to-Follow Guide to Everything You Need.
Student Support Services Laptop Workshop. What will you Find The actual Computer The Power Cord (Comes in two parts.)
Integrated Probe Storage Easy to Read LCD Display Bright LED Indicator Light Simple 4 Arrow Navigation with Enter Key (Circle) By the Numbers: Food.
Model 1621M Trainer’s Guide. General Overview Model 1621M is a Geiger Meuller based Personal Radiation Detector (PRD) Complies with ANSI Used for.
Digital Camera Basics Misty Freeman Georgia CTAE Resource Network 2010.
Gauge Operation and Software by Scott A. Ager. Computer Recommendations 750 MHz Pentium III 64 Meg SRAM 40 Gig Hard Drive 1024 x 768 graphics CD Writer.
1 Sport-Ident 101 Barry McCashin Edmonton Overlanders Orienteering Club January 22, 2006 Edmonton AB © BGMcC 2006 Part II B Programming with SI Manager.
WS50(Stratacache) Software updating guide (By Using USB)
Digital Literacy: Computer Basics
Puissant Panel User Training Esc.. 09:11:35 ACTIVE # * 3 6 RESET SYSTEM CLEAR DISPLAY TEST DISPLAY SILENCE BUZZER ESCAPE ENTER SILENCE.
AAPT workshop W03 July 26, 2014 Saint Cloud State University, MN, USA
PRO OX ® -100 Kit PROGRAMMABLE DIGITAL OXYGEN MONITOR US & FOREIGN PATENTS ISSUED & PENDING STATE-OF-THE-ART TECHNOLOGY FOR PRECISE OXYGEN MONITORING MADE.
42nd NIC Conference March Automated Temperature Monitoring for Vaccine Storage By Debbie Jo Trinidade, RN, MPH San Luis Obispo County Public Health,
Photo recovery from water damaged XD memory card recovery-from-water-damaged-xd-memory-card.
All Freezer Advance. When the Freezer is first supplied with power or if a power outage occurs, the display will indicate this by flashing the temperature.
Planned Preventive Maintenance for Fridges To improve your fridge’s function do the following tasks all the time: 2) Defrost your fridge once a month OR.
Everything you need to know about your Digital Data Logger - Daily Use
CASD Mobile Labs.
How to SWIVl: Using a Gen 2 SWIVL RObot
How to power down your NAS safely
AUTOMATED security system
CA05FS Owner’s Manual Fuel Cell Technology Breath Alcohol Tester.
Return/Wasted Vaccine Reporting
HOW WELL DO YOU KNOW THE BASICS OF USING YOUR COMPUTER?
Temperature Monitoring and Excursions
VFC 400 (Data Logger) Updated 6/22/17
PUMA-5 for Law Enforcement
Vaccine storage Vaccine Storage
Digital Chart Recorder Operation
Temperature Review and Reporting
Vaccine Storage and Handling
How to Perform Preventive Maintenance of the Personal Computer
Smoke Alarm Installation Program
Proper vaccine storage in ltc
Proper vaccine storage in ltc
How To Use The FD92 Digital Camera
Presentation transcript:

Using a Digital Data Logger North Carolina Immunization Program (NCIP)

WHY A DIGITAL DATA LOGGER? Provides accurate temperature history Reflects actual vaccine temperatures Provides an indication of length of time a unit has been operating outside recommended storage temperature Less expensive than replacing vaccines wasted due to inaccurate temperature readings - There are a variety of devices available 2

CDC RECOMMENDATIONS (REQUIRED OF ALL NEW PURCHASES AND FOR ALL PROVIDERS AS OF JANUARY 1, 2018) Digital display on outside of storage unit Detachable probe in a bottle filled with a thermal buffer, like glycol Continuous monitoring and recording capabilities Display of current, as well as minimum and maximum temperatures. 3

BEFORE YOU GET STARTED 4 Models are available in a single or double probe arrangement -How many units do you have and how close together are they? All models require a dedicated outlet without a ground-fault-interrupter and without the use of a surge protector With double units, determine which unit is “Probe 1” and which unit is “Probe 2”

ITEMS SUPPLIED 5 Traceable Excursion-Trac™ Thermometer (“Data Logger”) A/C Adapter (primary source of power)

ITEMS SUPPLIED 6 AAA Batteries installed, pull tab to activate batteries NOT YET! Glycol bottle with probe cable P1 (single bottle) – Two bottle probe includes P1 and P2 cable/bottle

ITEMS SUPPLIED 7 Instructions Certificate of Calibration DO NOT LOSE! Static Suppressor for probes Velcro and Magnets DO NOT USE WITH PROBE

STEP #1: PLACE THE PROBE 8

STEP #2: POWER SUPPLY 9 Thermometer must be powered at all times with supplied AC adapter Insert AC adapter plug into the unit, and plug the AC adapter directly into a wall socket (do not plug into a surge protector) Backup AAA batteries will power thermometer for one week in the event of a power failure - Pull tab to activate batteries on back of unit

STEP #3: STATIC SUPPRESSOR 10 The static suppressor is installed to eliminate interference due to radio frequency. Static within the cable can cause the unit to malfunction.

STEP #4: SET ALARM LIMITS 11

SETTING REFRIGERATOR ALARM LIMITS 12 Refrigerator settings in Celsius Low alarm 2°C High alarm 8°C Refrigerator settings in Fahrenheit Low alarm 35°F High alarm 46°F Current room temperature while setting alarm

SETTING FREEZER ALARM LIMITS 13 Freezer settings in Celsius Low alarm -50°C High alarm -15°C Freezer settings in Fahrenheit Low alarm -58°F High alarm 5°F Current room temperature while setting alarm

STEP #5: SET THE DATE & TIME 14 Setting the year, day, month and time is an important step to ensure the Excursion-Trac™ properly records the actual time and date of temperature readings Slide the DISPLAY switch to the DATE/TIME position Press the SELECT button to enter the setting mode Press the ADVANCE button to increase the selected date or time

STEP #6: SET THE LOGGING INTERVAL 15 Set the unit to record at intervals of 30 minutes To increase the logging interval to read at 30 minute intervals, press the ADVANCE button until 0:30 is displayed Slide the MEM VIEW switch to the OFF position to complete/close the setting Use the CLEAR button to erase the memory of the data logger the first time it is put in use.

STEP #7: ENABLE ALARMS 16 ENABLING/DISABLING ALARMS Wait for probe to cool down to an in-range temperature Slide the ALARM switch to the ON or OFF position to enable or disable the alarms Do not disable alarms

STEP #8: PLACE THE DATA LOGGER 17 Guide wires to ensure that staff do not accidentally pull the probe out of place or unplug the wires Packing tape may be used to hold wires in place

REGULAR USE OF YOUR DATA LOGGER 18

VIEWING MINIMUM/MAXIMUM MEMORY 19 The minimum or maximum temperature stored since the last clear of min/max memory Slide PROBE SWITCH to the desired P1 or P2 Slide the DISPLAY switch to the MIN/MAX position Thermometer will display the current, minimum, and maximum temperatures for the probe

RECORDING TEMPERATURE 20 PROVIDERS MUST MANUALLY RECORD ALL OF THE DATA BELOW EVERY DAY THAT YOUR CLINIC IS OPEN! Opening (or morning) temperature with date, time, and recorder’s initials Minimum and maximum temperatures (morning only). Press the CLEAR button once recording min/max temperatures Closing (or afternoon) temperatures with date, time, and recorder’s initials PLEASE VISIT IMMUNIZE.NC.GOV TO PRINT NEW LOGS

DOWNLOADING STORED DATA 21 Insert a formatted USB flash drive into the logger P1 and P2 light will come on Leave the USB flash drive in the logger until the light goes off File automatically creates a.csv file on the USB drive Once the light is off, remove the USB flash drive

REVIEWING STORED DATA 22 Plug your USB drive into a computer Open the CSV file created by the data logger Review all temperatures to ensure that all readings are in-range Data review should be completed by a different staff member than the staff who manually records temperatures Save file for 3 years

WHAT IF I FIND AN OUT-OF- RANGE TEMPERATURE? 23 Thermometer buzzer will sound when an out- of-range temperature occurs LED for the alarming temperature channel will flash (P1 or P2) Active alarm may be cleared by pressing the CLEAR SILENCE ALM button or disabling the alarm by sliding the ALARM switch to OFF position CALL THE NCIP STORAGE AND HANDLING PHONE LINE IMMEDIATELY UPON DISCOVERY OF AN OUT-OF-RANGE TEMPERATURE REGARDLESS OF HOW FAR OUT OF RANGE OR HOW LONG IT WAS OUT OF RANGE!

Instructional videos on how to set up the unit may be found at CyU8s-4px3DGg3AlmqpghQQ CyU8s-4px3DGg3AlmqpghQQ NCIP Storage and Handling phone line at NCIP Regional Staff ADDITIONAL RESOURCES