SIZE EXCLUSION CHROMATOGRAPHY

Slides:



Advertisements
Similar presentations
Gas Chromatography Introduction 1.) Gas Chromatography
Advertisements

Foundation GPC Part 4 – Organic Gel Permeation Chromatography Columns and Applications.
Size Exclusion Chromatography
Foundation GPC Training Course Theory. Nomenclature Gel Permeation ChromatographyGPC Size Exclusion ChromatographySEC Gel Filtration ChromatographyGFC.
1 HPLC Lecture Mobile Phase Selection in Partition Chromatography Optimization of the mobile phase composition and polarity is vital for obtaining.
Separation of molecules and determination of there molecular weight by gel filtration chromatography. Experiment 7 BCH 333.
Chromatography for Protein purification 1
Chapter 32 HIGH-PERRORMANCE LIQUID CHROMATOGRAPHY High-performance liquid chromatography (HPLC) is the most versatile and widely used type of elution chromatography.
Size-exclusion chromatography (SEC) Gel permeation chromatography (GPC) Gel Filtration Chromatography (GFC)
Four basic types of column chromatography where mobile phase is a liquid Partition Chromatography Bonded-Phase Liquid-Liquid Adsorption Chromatography.
High Performance Liquid Chromatography. HPLC originally refered to: High Pressure Liquid Chromatography currently refers to: High Precision Liquid Chromatography.
Protein Purification and Analysis Solubility of proteins important for purification: 60-80% soluble, 20-40% membrane Size of proteins varies Some proteins.
Gel Filtration Chromatography.
In the name of god.
An Extended Range of Columns for Aqueous Size Exclusion Chromatography Greg Saunders, Graham Cleaver and Ben MacCreath, Varian, Inc., Church Stretton,
Basics of Chromatography Jony Mallik. Introductory Principles Chromatography is a combination of two words; * Chromo – Meaning color * Graphy – representation.
Chromatography Desalting and Affinity. Chromatography Technique to separate components of a mixture by passing them through a matrix. –A solvent is used.
Chemistry 4010 Lab It’s all about PROTEINS… It’s all about PROTEINS…
High Performance Liquid Chromatography High Performance Liquid Chromatography Chem. 331.
Gel filtration chromatography y/college/voet/ /animat ed_figures/ch05/f5-6.html.
Chapter 4-1 Chromatography
1 HPLC HIGH PERFORMANCE LIQUID CHROMATOGRAPHY High Performance Liquid Chromatography (HPLC) is one of the most widely used techniques for identification,
High Performance Liquid Chromatography
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Magnet Analytical Chemistry Unit 4
Protein Purification and Analysis Solubility of proteins important for purification: 60-80% soluble, 20-40% membrane Size of proteins varies Some proteins.
Electrophoresis PAGE Dr Gihan Gawish.
1 High Performance Liquid Chromatography (HPLC) Lecture 40.
High Performance Liquid Chromatography Instrumentation.
1. Fast Protein Liquid Chromatography 2 FPLC A semi-automatic microprocessor controlled machine used primarily for the separation of macromolecules A.
Factors affecting electrophoresis The electric field ·       Voltage. If the separation of the electrodes is d (cm) and the potential difference.
Gihan Gawish.Dr High Performance Liquid Chromatography.
Agarose Gel Electrophoresis
Proteins separation and analysis using Fast Protein Liquid Chromatography Ayelet David, Ph.D Dept. of Clinical Pharmacology.
High Performance Liquid Chromatography. The chromatogram is a record of detector output Vs time as the analyte passes through the chromatography.
Introduction  High-performance liquid chromatography (HPLC) is a form of liquid chromatography.liquid chromatography  The main purpose is to separate.
Best Broken into four categories
Chromatography High Performance Liquid Chromatography HPLC Chapter Dr Gihan Gawish.
HPLC.
High Performance Liquid Chromatography
Ch 21 – Principles of Chromatography and Mass Spectrometry Ch 22 – Gas and Liquid Chromatography.
HPLC.
Size-exclusion Chromatography prof. aza Department of Pharmacy, Andalas University STIFI Perintis, Padang STIFAR, Pekan Baru STIFI Bhakti Pertiwi, Palembang.
1 High Performance Liquid Chromatography (HPLC) Lecture 40.
T.A Nouf Alshareef KAU-Faculty of Science- Biochemistry department Analytical biochemistry lab (Bioc 343) 2012 False colour scanning.
Ashraf M. Mahmoud, Associate professor CHROMATOGRAPHYCHROMATOGRAPHY.
GEL FILTRATION CHROMATOGRAPHY Size Exclusion Chromatography
Food Analysis Lecture 18 (03/27/2012) Basic Principles of Chromatography (3) Qingrong Huang Department of Food Science Read Material: Chapter 27, page.
BY G.SHRAVANI ION EXCHANGE AND GEL PERMEATI ON CHROMATOGRAPHY 1G.SHRAVANI.
Gel Electrophoresis + restriction enzymes Lecturer: Bahiya Osrah.
Tutorial 09 Overview of other chromatographic methods 1.
Principles of chromatography
Size exclusion Chromatography
 Laboratory technique for the Separation of mixtures  Chroma -"color" and graphein - "to write”.  Colour bands - separation of individual compounds.
Bioseparation II Chromatography Techniques. Chromatography Most widely used purification technique used for biomolecules. Most widely used purification.
High Performance Liquid Chromatography Presented by Dr. Kamal Modi 2 nd Year Resident.
Gel Permeation Chromatography Size Exclusion Chromatography
BCH BIOCHEMICAL METHODS (2 UNITS) DR AKINLOYE’S ASPECT
Lecture 2: Introduction to Chromatography and Size- Exclusion Chromatography Yaj.
Chapter 4-3 Chromatography Gel Chromatography Dr Gihan Gawish.
Gel Filtration Chromatography.
Size Exclusion Chromatography (SEC)
High Performance Liquid Chromatography
Gel Filtration Chromatography.
Four basic types of column chromatography
High Performance Liquid Chromatography
High-Performance Liquid Chromatography
M.S COLLEGE OF ARTS, SCIENCE, COMMERCE AND B.M.S
High performance liquid chromatography (HPLC)
High-Performance Liquid Chromatography
Presentation transcript:

SIZE EXCLUSION CHROMATOGRAPHY Presented by MANDAWI TRIPATHI

CONTENT Introduction Principle Material Instrumentation Advantage Disadvantage Application Gel Electrophoresis Size Exclusion Chromatography 2

INTRODUCTION Size-exculsion chromatography (SEC), also called gel-filtration or gel-permeation chromatography (GPC), uses porous particles to separate molecules of different sizes. It is generally used to separate biological molecules, and to determine molecular weights and molecular weight distributions of polymers It is usually applied to large molecules or macromolecular complexes such as proteins and industrial polymers.

When an aqueous solution is used to transport the sample through the column, the technique is known as Gel-filtration chromatography . When an organic solvent is used as a mobile phase,the technique is known as Gel-permeation chromatography. The separation of molecules is called fractionation. Size of pores in beads determines the exclusion limit (what goes through the beads and what goes around the beads)

PRINCIPLE A mixture of molecules dissolved in liquid (the mobile phase) is applied to a chromatography column which contains a solid support in the form of microscopic spheres, or “beads” (the stationary phase). The mass of beads within the column is often referred to as the column bed. The beads act as “traps” or “sieves” and function to filter small molecules which become temporarily trapped within the pores.

Larger molecules pass around or are “excluded” from the beads . Large sample molecules cannot or can only partially penetrate the pores, whereas smaller molecules can access most or all pores. Thus, large molecules elute first, smaller molecules elute later, while molecules that can access all the pores elute last from the column. Particles of different sizes will elute (filter) through a stationary phase at different rates.

Total column volume (Vt) Vt = Vg + Vi + V0 where Vg --is the volume occupied by the packing Vi --is the volume of solvent in the pores V0 --is the free solvent volume (similar to injection volume)

COMPONENTS OF A SEC Stationary Phase The Mobile Phase The Columns The Pump Detectors

STATIONARY PHASE: Stationary Phase Semi-permeable, porous beads with well-defined range of pore sizes . Beads are crosslinked polymers Degree of crosslinking is controlled carefully to yield different pore sizes. Smaller pore sizes are used for rapid desalting of proteins or for protein purification. Intermediate pore sizes are used to separate relatively small proteins.

Very large pore sizes are used for purification of biological complexes. Stationary phase used for gel exclusion chromatography include dextran (Sephadex™), polyacrylamide and dextran-polyacrylamide (Sephacryl™). Each is available with a variety of different ranges of pore size in the beads, permitting separation of macromolecules of different size

A good stationary phase should have following properties: It should be chemically inert. It should be inexpensive. It should not react with component to be separated. It should not react with eluent. It should be colorless, uniform in size and shape. It should be mechanically stable.

Soft gel e.g.- dextran(Sephadex), Polyacrylamide gels Separation of proteins. Semi-rigid gel e.g.- bio beads Separation of non-polar polymers in non-polar solvents. Highly rigid gels and glasses Separation of polar systems.

Dextran A homopolysaccharide of glucose residues. It’s prepared with various degrees of cross-linking to control pore size. It’s bought as dry beads, the beads swell when water is added. The trade name is sephadex. It’s mainly used for separation of small peptides and globular proteins with small to average molecular mass.

TYPICAL SEPARATION RANGES THAT CAN BE ACHIEVED USING SEPHADEX ARE GIVEN BELOW. MOLECULES RANGING FROM 100 TO 600,000 DA CAN BE SEPARATED DEPENDING ON THE TYPE OF SEPHADEX CHOSEN SEPHADEX RANGE G10 100-800 Da G15 (500-1500 Da), G25 (1000-5,000 Da), G50 (1,500-30,000 Da) G75 (3,000-80,000 Da ) G100 (4,000-150,000 Da ) G150 (5,000-300,000 Da) G200 (5,000-600,000 Da).

Polyacrylamide these gels are prepared by cross linking acrylamide with N,N-methylene bis acrylamide. The pore size is determined by the degree of cross-linking. The separation properties of polyacrylamide gels are mainly the same as those of dextrans. They are sold as bio-gel P. They are available in wide range of pore sizes.

Agarose Linear polymers of D-galactose and 3,6 anhydro-1-galactose. It forms a gel that’s held together with H bonds. It’s dissolved in boiling water and forms a gel when it’s cold. The concentration of the material in the gel determines the pore size. The pores of agarose gel are much larger than those of sephadex or bio-gel p. It’s useful for analysis or separation of large globular proteins or long linear molecules such as DNA

Mobile phase The liquid used to dissolve the biomolecules to make the mobile phase is usually called a buffer. The mixture of biomolecules dissolved in the buffer is called the sample. The choice of mobile phase to be used in any separation will depend on the type of separation to be achieved and component to be separated. The most common eluents in for polymers that dissolve at room temperature.e.g.-Tetrahydrofuran,Chloroform, Dimethyl formamide.

MOBILE PHASE material solvent Synthetic elastomers ( polybutadiene , polyisoprene ) Toluene PS, PVC, Styrene-Butadiene Rubber , Epoxy resins Tetrahydrofuran Polyolefins Tri- chloro -benzene Polyurethane Di- methylformamide Proteins, polysaccharides Water / Buffers

SOLVENT SELECTION The solvents used for mobile phase of SEC are limited to those follows following criteria: The solvent must dissolve the sample completely. The solvent has different properties with solute in the eluent: typically with solvent refractive index (RI) . solvent must not degrade the sample during use Otherwise, the viscosity of eluent will gradually increase over times. The solvent is not corrosive to any components of the equipment

Mobile Phase Preparation high purity of solvent is recommended. Filter mobile phase solvents using 0.5 micron filter to remove any particular impurities such as dusts, insoluble salts. Antioxidant is added to trichlorobenzene to keep solvent stable in high temperature. Other additives eliminate adsorption or interaction of solutes with column packing materials Size Exclusion Chromatography

Sample preparation The sample solutions are supposed to be prepared in dilute concentration (less than 2 mg/mL) A good solvent can dissolve a sample in any proportion in a range of temperatures. Samples with broad molecular weight distribution may require higher concentrations. It is recommended to filter the sample solutions before injecting into columns in order to get rid of clogging and excessively high pressure problems.

Agitation and filtration Generally filtration is required to remove insoluble impurities. Do not agitate and filter samples that contain very high MW (>1 million).

COLUMNS Commercially Available Columns analytical column- 7.5–8mm diameters. Preparative columns-22–25mm for. Usual column lengths-25, 30, 50, and 60 cm. Recently, narrow bore columns- 2–3mm diameter have been introduced, which save time and solve

Selecting SEC column Shorter columns save time and solvent. Small particles (typically 5 mm) provide a better resolution. On the other hand, 5 mm (or even 3 mm) packings are more sensitive towards contamination by samples containing impurities. Particles as large as 20 mm have been recommended for very high-molecular-weight polymers. ion.

Small particle size packings can sometimes result in shear degradation of large polymer molecules because the space between particles is very narrow. Columns with different porosity or mixed-bed columns, provide a better separation.

Handling SEC Columns A column set in SEC should be always run in the same mobile phase.(isocratic) SEC columns should never be operated in a backward direction. Care should also be taken in connecting columns or in sample injection. Replacing a clogged inlet frit is a dangerous operation which can reduce column performance. A damaged or dirty check valve of pump, can also reduce column life.

PUMP A highly constant flow rate has to be maintained during the entire chromatogram. This is very important in SEC. A change of the flow rate of only 0.1% can cause an error in molar mass of up to 10%. Most pumps can only reproduce the flow rate to 0.2–0.3%. In-line filters in the solvent reservoir may prevent particles from coming into the pump heads, which might damage the check valves or the pump seals. Types-)Syringe pumps, Reciprocating pumps

DETECTOR Concentration sensitive detectors Bulk Property Detectors- Refractive Index (RI) Detector Solute Property Detectors- Ultraviolet (UV) Absorption Detector Evaporative Detectors- Evaporative Light Scattering Detector (ELSD) Molar mass sensitive detectors Light Scattering Detectors Low Angle Light Scattering (LALS) Detectors Multiangle Light Scattering (MALS) detectors

Viscosity Detectors- Differential Vscometers Other :- Flame Ionization Detector (FID), A Mass Spectrometer or A Fourier Transform Infrared (FTIR) Spectromet

ADVANTAGES Short analysis time. Well defined separation. Narrow bands and good sensitivity. There is no sample loss. Small amount of mobile phase required. The flow rate can be set.

DISADVANTAGES Limited number of peaks that can be resolved within the short time scale of the GPC run. Filtrations must be performed before using the instrument to prevent dust and other particulates from ruining the columns and interfering with the detectors. The molecular masses of most of the chains will be too close for the GPC separation to show anything more than broad peaks .

APPLICATION Proteins fractionation Purification Molecular weight determination. Separation of sugar, proteins, peptides, rubbers and others on the basis of their size. This technique can be determine the quaternary structure of purified proteins.

SEC is a widely used technique for the purification and analysis of synthetic and biological polymers, such as protein, polysaccharides and nucleic acid. Various species of RNA and viruses have been purified using agarose gels. For Desalting For copolymerisation studies