Misuse of Combustible Gas Meters Supervisor Training 1/11/08.

Slides:



Advertisements
Similar presentations
Monitoring Equipment and Detection Devices Chapter 14.
Advertisements

Mine Gases Atmospheric Testing C.4. Contributed by: Richard Black Industrial Scientific Corp.
MATERIAL SAFETY DATA SHEET
INDUSTRIAL HYGIENE - GASES, VAPORS, AND SOLVENTS UNIVERSITY OF HOUSTON – DOWNTOWN FALL 2013.
Flash Point-Ignition Point
SOLARIS 3-GAS METER Complete operation of Solaris multi-gas meter can be found on the LAFD Ethernet.
Sensor Operating Theory
Hazmat Industry Technician. Types of Equipment Monitoring equipment Detection devices Air sampling systems.
Arson. The Common Accelerants  The most commonly used accelerants are petrol, kerosene, mineral turpentine and diesel.  These accelerants are generally.
Chemistry of Fire.
PROPERTIES OF FLAMMABLE MATERIALS. Flammability Flammable Flammable –Capable of being ignited and of burning –Synonymous with combustible.
© BLR ® —Business & Legal Resources What’s the Big Deal? Flammable and combustible liquids: Are easily ignited and ignite with explosive force Burn.
Mine Rescue Gas Detection
Basic Air Monitoring Concepts And Use Of The Altair 4 Gas Meter
Incident Date:1/11/2012 Time:07:30 AM Fatality: 25 injuries: 131 Total losses:
Fires and Explosions.
Air Monitoring Presented By: Etech Environmental & Safety Solutions, Inc.
What is methane (CH4)? a. A colorless, odorless and tasteless combustible gas. b. A colorless, odorless and tasteless poisonous gas. c. A colorless, odorless.
Safe Handling of Flammable and Combustible Liquids
Part 2 - Dilution Ventilation (General Ventilation)
Chapter 14 Chemical reactions
Chapter 15 Chemical reactions.  Any material that can be burned to release thermal energy is called a fuel.  Most familiar fuels consist primarily of.
1 When entering a confined space Before and during normal working in a Process or Hydrocarbon Area When breaking containment on a process system When hot.
Fire Triangle In order to understand how fire extinguishers work, you first need to know a little bit about fire. Four things must be present at the same.
Vincent J. Giblin, General President 1293 Airport Road Beaver, WV Phone: (304) Fax: (304) Air Monitoring.
Oil Heat Properties of Fuel Oil.
MULTI-GAS MONITOR FOR CONFINED SPACE APPLICATIONS
Properties of Fuels.
Combustible Gas & Oxygen Meter. Combustible gas indicators (CGIs) measure the concentration of a flammable vapor or gas in air, indicating the results.
Copyright © Flammable & Combustible Liquids.
HazMat Technician Mod 1 Unit 2 Slide 1 Ohio Hazmat & WMD Technician Module 1 Unit 2 Field Survey Instruments.
1 BUILDING TRADE SAFETY CONFINED SPACES. 2 BUILDING TRADE SAFETY Confined spaces.
WELCOME TO Gas Tester Training O2, LEL Monitoring
MODULE 5 Complete and Incomplete Combustion Characteristics.
1 Combustion Oil Fired Equipment. 2 OIL Combustion of oil fired equipment has the same basics of gas. Remember, Oil is a liquid and this must be converted.
HAZCOM Hazard Communication Standard “The Right-to-Know”
4.3.2 Characterization of Flammable Vapor Hazards
Safety Determine conditions where fires and /or explosions can occur.
FIRES AND EXPLOSION.
Chapter 2: Chemical Properties and Information Resources on Hazardous Chemicals.
FIRES AND EXPLOSION LECTURE 10.
PLANT DESIGN.
Hazardous Materials Subpart H. Subpart H Standards Compressed Gases Acetylene Hydrogen Oxygen Nitrous Oxide Flammable.
SECTION 2 TERMINOLOGY. HAZARDOUS MATERIALS PROPERTIES.
FLAMMABLES. 1. PROPERTIES OF FLAMMABLES Combustion (redox) reactions Solids, gases, most commonly liquids Fire Tetrahedron Ignition sources External-
Chapter 13 - Analytical Instrumentation 1 Chapter 14 Analytical Instrumentation.
General Monitoring Practices Sugar Grove Fire Department Training and Education.
Prof. Jiakuan Yang Huazhong University of Science and Technology Air Pollution Control Engineering.
Power Plant Engineering
2015 Hazmat Operation Refresher. Course Objectives We will be refreshing on: Detectors Detectors Hazmat Responses:Hazmat Responses:  Gas Leak – Natural.
Gas sensing Panca Mudji Rahardjo, ST.MT Electrical Engineering - UB.
1 LANDFILL GAS TRAINING. 2 INTRODUCTION  TYPE OF MONITORING -Screening -Specialized  GOALS.
Chemical Characteristics Assessing Hazards Presented by: Eureka Educators.
Hazards of Confined Spaces. INSTRUCTIONAL GOAL The participant till understand the various hazards associated with confined space entries and the need.
Bi-Weekly Safety Meeting November 29, The days of using canaries in mines as primitive gas detectors are long gone..they've been replaced by fixed.
1 Emergency and Disaster Response to Chemical Releases Monitoring Module 4.
Atmospheric Monitoring. ATMOSPHERIC MONITORING INSTRUCTIONAL GOAL The participant will understand the need for atmospheric monitoring and the types, uses,
CHAPTER 6 Gases and Vapors. Learning Objectives Describe the various properties of the chemicals that are most important to the practice of industrial.
Safe Handling of Flammable and Combustible Liquids.
ArsonAnalysis of Fire Arson is defined as purposely setting fire to a house, building or other property. Arson is the second leading cause of death by.
Fire Loss Control - Basic Elements
Fire Behavior & Extinguisher Use.
Safe Handling of Flammable and Combustible Liquids
Confined Space Hazards
Basic Hazardous Materials Terms
Flammable & Combustible Liquids
Safe Handling of Flammable and Combustible Liquids
Basic Hazardous Materials Terms
Gas Monitoring Basics Presented by Dan Klimek Sisk & Company.
Presentation transcript:

Misuse of Combustible Gas Meters Supervisor Training 1/11/08

Case History  An incident investigation raised concerns that users of combustible gas meters may be greatly underestimating fire and explosion hazards.  Errors result when meters calibrated on a specific gas are used to measure other explosive gases or vapors.  In one case a methane calibrated meter was used to measure gasoline vapors; the meter gave a reading of 14% LEL (Lower Explosive Limit) while the actual concentration in the tank was 73% of the LEL.

How Combustible Gas Meters Work  Most combustible gas detectors these days literally measure the contaminant by combustion at a catalytic detector. –Older style used the wheat-stone bridge method.  The heat produced is used as a measure of the “explosivity” of the contaminant in air.  Different compounds produce different amounts of heat when they are burned. So the meters respond differently to different chemical mixtures in air.

Meter Variation  Combustible gas meters can only be expected to respond accurately to the gas for which they were calibrated.  To measure other gases with the same meter, consideration must be given to the specific properties of the gas and of the detector.  Most manufacturers have responded to the need to estimate more accurately the concentration of other gases by providing correction factors which allow calculation of percent LEL from the measured level.

What is LEL & UEL? When vapors of a flammable or combustible liquid are mixed with air in the proper proportions in the presence of a source of ignition, rapid combustion or an explosion can occur. The proper proportion is called the flammable range and is also often referred to as the explosive range. The flammable range includes all concentrations of flammable vapor or gas in air, in which a flash will occur or a flame will travel if the mixture is ignited.

What is LEL & UEL? There is a minimum concentration of vapor or gas in air below which propagation of flame does not occur on contact with a source of ignition. There is also a maximum proportion of vapor in air above which propagation of flame does not occur. These boundary-line mixtures of vapor with air are known as the lower and upper flammable limits (LFL and UFL) respectively, and they are usually expressed in terms of percentage by volume of vapor in air. The LFL is also known as the lower explosive limit (LEL). The UFL is also known as the upper explosive limit (UEL).

Flammable Limits In popular jargon, a vapor/air mixture below the lower flammable limit is too "lean" to burn or explode, and a mixture above the upper flammable limit is too "rich" to burn or explode. No attempt is made to differentiate between the terms flammable and explosive as applied to the lower and upper limits of flammability.

Limitations of Meters  These instruments can not be relied upon to: –measure highly toxic gases such as hydrogen sulphide. –respond accurately in atmospheres which do not contain 20.9% oxygen in air. –provide a reliable indication of the degree of explosive hazard when meters are not calibrated before and after each use. –compensate for poor field sampling technique or an operators failure to consider the work environment.

Limitations of Meters  ALWAYS remember… 1% = 10,000 ppm  QUESTION… OSHA’s acceptable limit for O2 is 19.5%-23.5%. Your meter gives you an O2 reading of 19.9%. Would you enter this space based on this information?

Limitations of Meters  Oxygen concentrations in air, other than those normally occurring in the atmosphere (i.e. 20.9%), may result in underestimating the explosive hazard.  Meter response depends on its ability to burn the combustible gas.  If there is not enough oxygen to support combustion, the meter then would give a lesser LEL, even if high levels of combustible gas were present.

Limitations of Meters  Since it is quite common for the meter sensors to fail, bump-test before use is required.  Exposure of the sensor to moisture, lead compounds, silicon compounds, or chlorinated hydrocarbons may result in instrument failure during use.

How to use Correction Factors K-Factors & Correction Factors  Various means of correcting for different meter response have been employed by a number of manufacturers.  In each case, however, the correction factors given are specific to the model of meter and the gas/vapor being measured.  Individual suppliers must be contacted to obtain the appropriate correction factors for your meter.

How to use Correction Factors K-Factor Formula K-factor (calibration gas) x observed LEL K-factor (measured gas/vapor)

How to use Correction Factors Trick Question #1…  In a tank used to store anhydrous ammonia an operator records a reading of 18% LEL using a meter calibrated for methane gas (what you use).  What is the actual % LEL in the tank?  Given: –Observed reading: 18% LEL –ammonia K-factor = –Calibration gas was methane, K- factor = 112.0

How to use Correction Factors Answer..  Then: K-factor (calibration gas) x observed LEL K-factor (measured gas/vapor) x 18 = 14% 141.7

How to use Correction Factors Trick Question #2  The area around a leaking gas line gives a reading of 56% LEL on a meter calibrated with methane.  What is the actual % LEL in the area? Given: Observed reading = 56% LEL Correction factor (propane) = 1.1

How to use Correction Factors Answer: Corrected % LEL = correction factor x observed % LEL = 1.1 x 56 = 62%

Correction Factors for some of your chemicals Some flammables have a correction factor as high as 2.5!!! (xylene & Hexane)

Four Gas Meters  Your meters are 4-gas meters, meaning they measure ONLY four gases! –Oxygen –LEL –H2S –CO  A few meters have a NH3 sensor instead of a CO or H2S sensor

Four Gas Meters  What do you do when you want to measure a potential toxic atmosphere for Sulphuric Acid? –Will your four gas meter work?  Colormetric/Diffusion Detector Tubes are your answer.

Response Times  When sampling an area, one sample is NOT enough  Sampling a PRCS –Top - Middle - Bottom –Vapor Density  Last Trick Question of the day…

Response Times  You are sampling a tank for entry and the tank is 50 ft tall. You are on top of the tank and lower your 50 ft sample hose to the bottom of the tank. It takes ~5 minutes for the air sample to reach the instrument. You then sample the middle of the tank (25 ft down)…how long should you leave the sample tube at the 25 ft mark, since it is half the distance of your first sample?  A. 5 minutes  B. 2.5 minutes  C. I do not have to sample the middle?

Industrial Scientific M40  FREE On-Line Video for Operation of the M40 Meters can be viewed at: Click on “Services” tab and select “Training” from drop down On the “Training Page” Click on “FREE Online Video Training” link just below the image of your M40 meter. Once on that page you will see an image of your meter, CLICK on it and begin with the first 2 minute video.