Process Measurements Division The New NIST Gas Pressure Scale: Reduced Uncertainties Based on Large Diameter Piston Gauges Douglas A. Olson National Institute.

Slides:



Advertisements
Similar presentations
Ground-based FTIR Update CAVIAR Meeting NPL, 29 th September 2010 Tom Gardiner, Marc Coleman.
Advertisements

Gases.
Report of the ISO/TC108 Representative to the 7 th Meeting of the CCAUV BIPM, Sèvres, October Hans-Jürgen von Martens Representative of ISO/TC 108.
A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders Michael Bair Director of Metrology-Pressure Fluke Calibration.
GASES! AP Chapter 10. Characteristics of Gases Substances that are gases at room temperature tend to be molecular substances with low molecular masses.
Liquids and Gasses Matter that “Flows”
11/8/20074th Challenge IFPSC1 Establishing Benchmarks for the Fourth Industrial Fluid Properties Simulation Challenge James D. Olson The Dow Chemical Company.
Copyright Alan Rowley Associates Steps to an Accurate Result Select a method and validate it as suitable for the purpose envisaged. Establish that.
NCSLI 2007 A PRIMARY FLOW CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard.
CMCs and the BIPM Key Comparison Database Raul Fernando Solís Betancur 2012.
Twinning Project RO2006/IB/EN/09 1 Saxony-Anhalt State Environmental Protection Agency Wolfgang GarcheBukarest Wolfgang Garche Saxony-Anhalt.
1 Introduction to the GUM ( International Guidelines for calculating and expressing uncertainty in measurement) Prepared by Les Kirkup and Bob Frenkel.
CMCs and the BIPM Key Comparison Database Raul Fernando Solís Betancur 2015.
Selection of measuring instruments
1 Practical Interpretation of Unbroken Chain in Metrological Traceability as to VIM 3 Center for Measurement Standards (CMS) Industrial Technology Research.
Miruna DOBRE – SMD (Belgium) Stephanie BELL – NPL (United Kingdom) Dolores del CAMPO – CEM (Spain) Martti HEINONEN - MIKES (Finland)
© NMISA 2010 INTERNATIONAL ACTIVITIES AT THE NMISA HUMIDITY LABORATORY DURING 2009/2010 Deona Jonker.
Physical Properties: Melting Temperature Boiling Temperature Color
Chapter 10: Fluids Three (common) phases of matter: 1. Solid: Maintains shape & size (approx.), even under large forces. 2. Liquid: No fixed shape. Takes.
Introduction and Chapter 1
1 Chapter 5: GASES. 2  In this chapter we will:  Define units of pressure and volume  Explore the properties of gases  Relate how the pressure, volume,
1 Gases Chapter Properties of Gases Expand to completely fill their container Take the Shape of their container Low Density –much less than solid.
Execution Plan March 2007-March 2008 SIM General Assembly Ottawa September 2007.
Introduction and Vectors
Fluid Properties: Liquid or Gas
CHAPTER 5: PRESSURE 5.1 Pressure and Its Units
Dr. Ilya Kuselman The National Physical Laboratory of Israel (INPL) Givat Ram, Jerusalem,
Time and Frequency activities at CENAMEP Panama, 2015.
Fakulti Kejuruteraan Mekanikal, Universiti Teknologi Malaysia
TECO2005WRC/PMOD Quality Management1 The World Radiometric Reference and its Quality System ISO/IEC Wolfgang Finsterle, Isabelle Rüedi, and Silvio.
The Nature of Gases. I.The Kinetic Theory and a Model for Gases A.Assumptions of the Kinetic Theory 1.Gases consist of large numbers of tiny particles.
Introduction to Basic Concepts of Thermodynamics
Measurement & Calculations Honors Chemistry Chapter 2.
Plasma Density Structures in the Inner Magnetosphere Derived From RPI Measurements B. Reinisch 1, X. Huang 1, P. Song 1, J. Green 2, S. Fung 2 V. Vasyliunas.
ME 322: Instrumentation Lecture 4 January 26, 2015 Professor Miles Greiner Had extra time, could add a few more slides next year Lab Guidelines and grading,
Experimental and Numerical studies on Bulk Micromegas SINP group in RD51 Applied Nuclear Physics Division Saha Institute of Nuclear Physics Kolkata, West.
Aerospace Engineering Laboratory I  Basics for Physical Quantities and Measurement  Physical Quantity  Measured Quantity VS Derived Quantity.
Effects of Breathing on an Interferometer Susan Gosse Daniel Freno Junior Lab II.
Chapter 10 Physical Characteristics of Gases The Kinetic-Molecular Theory of Matter In the late 19 th century the Kinetic-Molecular Theory was developed.
ChemE 260 Foundation of Thermodynamics March 29, 2005 Dr. William Baratuci Senior Lecturer Chemical Engineering Department University of Washington TCD.
NCSL INTERNATIONAL SERVING THE WORLD OF MEASUREMENT SINCE 1961 Real World Measurement in a Nuclear Plant.
2006 NCSL International Workshop and Symposium August 07 Management of the AC-DC voltage transfer difference comparison in the SIM region Sara Campos CENAM.
1 THE PARTICIPATION IN THE CIPM MRA OF VMI Country Report of Vietnam at 14th meeting of APMP developing economies committee (DEC) Xi'an-China. 10th June.
Requirements of EU Directives for Gas Analysis through ISO & Traceability to National Metrology Institutes Martin Milton, Peter Woods and David Nettleton*
IMEKO TC 16 (Pressure) International Conference Cultivating metrological knowledge November, 27 – 30, 2007, Merida, Mexico A 36 MPa PRESSURE BALANCE IN.
Barometry  Evangelista Torricelli credited with invention of barometer.  Gasparo Berti may have been first.
A WEIGHT EXCHANGE SYSTEM FOR AN ABSOLUTE PRESSURE BALANCE Authors: H. Kajastie, J. Manninen, M. Rantanen, K. Riski and S. Semenoja, Centre for Metrology.
CCM Working Group on CMCs 1 Report on 11 May 2011 meeting of CCM-WGCMC & on Related Issues Chris Sutton, Chair Measurement Standards Laboratory of New.
The Traceability of Conductivity Meter and Cell Calibration in the Low Conductivity Ranges Jun Bautista, Director, Genzyme Metrology Genzyme Corporation,
Automation of the 27.1 kN Dead Weight Machine at NIST Kevin Chesnutwood, Samuel Ho NIST Mass and Force Group Stop 8222, 100 Bureau Drive Gaithersburg,
Unit 10 part 1: Kinetic Molecular Theory: GASES and Gas Laws.
Key Comparisons in Electrochemistry Significance to Metrology and Global Trade Kenneth W. Pratt* and David L. Duewer National Institute of Standards and.
APLMF 12 VOLUME AND FLOW LABORATORY OF VMI APPLYING PROCESS FOR CMC Dr Nguyen Hong Thai Volume and Flow Lab- VMI 6/25/
Putting Confidence Into Your Lab’s Results Alan Steele, Barry Wood & Rob Douglas National Research Council Ottawa, CANADA National.
APMP TC Fluid Flow Initiative Project Traceability Scheme Study for LPG Metering Final Report at APMP 2012 TCFF Meeting Chun-Min Su CMS/ITRI, Chinese Taipei.
Process Measurements Division Calibration of Hydraulic Piston Gauges Operating in Differing Fluids Dana R. Defibaugh and Douglas A. Olson National Institute.
1 Activities of TCM to the DEC Economies A.K. Bandyopadhyay Chair, TCM National Physical Laboratory New Delhi – , India.
Additive Manufacturing Measurement Uncertainty Emerzon Cruz Guandique
Ranges of Magnitudes & Quantities
Lecture 66 – Lecture 67 – Lecture 68 Gases Ozgur Unal
Product Presentation Calibration Services.
Key Comparisons and the MRA Emma Woolliams
News from the Working Group on Fluid Flow
CMC Submission and Reviewing Process
1- ما تعليقك على هذه الصورة ؟؟
By : M. Dlamini D. Jonker E. P. Tarnow
The Empirical Gas Laws Lesson 2.
Progress Report of APMP.M.FF-K2b
The Empirical Gas Laws.
Metrology Calibration and Measurement Capabilities
Presentation transcript:

Process Measurements Division The New NIST Gas Pressure Scale: Reduced Uncertainties Based on Large Diameter Piston Gauges Douglas A. Olson National Institute of Standards and Technology James Schmidt, Greg Driver, Walt Bowers, and Dana Defibaugh NCSL International Workshop and Symposium August 5, 2008

Process Measurements Division Outline of Talk Existing gas pressure scale Primary standard piston gauges (PGs) and their characterization Revised pressure scale based on PGs Characterization of PGs traceable to primary Validation of uncertainties via international comparisons Status of Calibration Measurement Capabilities (CMCs)

Process Measurements Division Pressure Realization Using Piston Gauges masses, m,  m gravity, g cylinder piston fluid pressure, p a ambient density,  a Artifact-based SI unit derived from fundamental quantities of mass and length Pressure results from force generated in gravitational field Artifact performance characterized by “effective area”, A e If A e determined by dimensional metrology, PG is primary standard

Process Measurements Division Present Gas Pressure Scale: Circa 1990 Pressure realization through gas piston gauge (PG) artifacts, 10 kPa to 110 MPa –Overlaps with Hg manometer (UIM) up to 360 kPa Multiple overlapping ranges of decreasing piston diameters (336 mm 2 to 8.4 mm 2 ) Working standards are GEIS (Ruska) 2465 and 2470 piston gauges Traceability to Gas Thermometer Manometer (GTM) at 27 kPa and 95 kPa in 1989 Standard uncertainties: 6.5 ppm to 20 ppm Up to 7 MPa, distortion is neglected, incorporated as higher uncertainty

Process Measurements Division Greg Driver performing crossfloat calibration

Process Measurements Division Why Change? GTM dismantled early 1990s, original traceability no longer possible Discrepancy exists for PGs traceable to GTM and UIM Direct traceability was limited to 95 kPa Customers want lower uncertainties –other NMIs can do better. Present NIST values conservative –High quality commercial PGs have much better repeatability than NIST uncertainty Large diameter (35 mm & up) PGs can now be characterized as primary standards Easier calibration of secondary standard PGs to primary PG than to manometers

Process Measurements Division PG38 and PG39: Acquired by NIST 1989 from Ruska Instrument Co. Used as check standards for UIM J. Schmidt performed extensive modeling of dimensional data and distortion Metrologia 43 (2006) establishes PG38 and PG39 as primary standards Upright Inverted

Process Measurements Division Summary of Characterization and Validation Dimensional measurements and force modeling Comparisons to NIST 3 m UIM Crossfloat comparisons of PG38 vs PG39 Characterization equation: Uncertainties: u(A 0 ), u(b)

Process Measurements Division Dimensional Measurements and Force Modeling PG38: NIST Dim 1989, PTB dim 2003 PG39: NIST Dim 1989, PTB dim 1999, 2003 PTB dim: diameters, roundness, and straightness of piston and cylinder –12.5 nm and 25 nm standard uncertainty on piston and cylinder diameters translates to 1 ppm std. unc. on effective area. –However, forces on piston flanks must be modeled and included. Change in diameters between 1999 and 2003 was 0.1 to 0.8 ppm A 0 (Schmidt) assembled dimensional data into “bird-cage”. Used various flow models in gap (viscous, viscous/molecular) b from elasticity theory, FEA, and flow models A 0, b chosen from distribution of modeling results

Process Measurements Division Dimensional Measurements and Force Modeling u(A 0 )/A 0 = 3 ppm b = 8.97x Pa -1, u(b) = 1.12x Pa -1 At 1 MPa: u(A e )/A e = 3.2 ppm

Process Measurements Division Comparisons to NIST 3 m UIM to 360 kPa Range of b / Pa via UIM PG38PG to to 8.8 A 0 agrees within 1-2 ppm. Stable over time b via UIM slightly lower, but limited to 360 kPa

Process Measurements Division Crossfloat Comparisons of PG38 and PG39 Operating both inverted/upright allows testing 4 distortion ratios Agreement between experiment and theory confirms b, u(b) Area ratio further confirms dimensional data, UIM

Process Measurements Division Revised Pressure Scale to 17 MPa Primary Standard UIM 0.3 MPa Dimensions and Model PG38 1 MPa PG39 1 MPa PG MPa PG MPa PG13 7 MPa PG35 7 MPa PG32 17 MPa PG23 17 MPa PG MPa PG MPa PG MPa PG MPa DHI380 7 MPa b Gauge Max P (MPa) 2465 Series Area (mm 2 ) PG22, PG36 0.2TL336 PG28, PG TTL336 PG34, PG37 1.4C84 PG13, PG35 7V8.4 PG23, PG32 17D8.4 DHI380: DHI PG7601, 49 mm 2

Process Measurements Division Propogation of Scale TL, TTL, and C series: direct crossfloat vs PG38 and PG39 in gauge mode –Elimination of intermediate step, less extrapolation Crossfloat vs “twin” of series as check V series (7 MPa): crossfloat vs C series –elasticity theory on PG7601 –crossfloat to PG13 to DHI 380 to set b D series: crossfloat vs V series, extrapolation of distortion

Process Measurements Division PG28 (TTL472) Type B uncertainty from measurement equation using PG38/PG39 uncertainty Type A from crossfloat and data fit -4.8 ppm shift from 1994 traceability Uncertainty bars: k=1 4.2 ppm (min)

Process Measurements Division Area Ratio, PG29/PG28 Direct crossfloat confirms traceability to PG38, PG39 Area ratio consistent with existing traceability

Process Measurements Division TL and C Series PG22 (TL460) PG34 (C74) C series requires modest extrapolation to 1.4 MPa u/A: 5.2 ppm (min) u/A: 4.3 ppm (min)

Process Measurements Division V Series (PG13: V373) b via calibration to DHI380 (PG7601) –PG7601: cylinder diameters uniform –b DHI380 from elasticity theory –b DHI380 = -2.36x Pa -1 b PG13 = 2.66x Pa -1 Primary MethodIntermediate Stepsb (Pa -1 ) / Elasticity theory, DHI380PG13 v DHI380 Xfloat2.66 FEA model, PG13none3.30 FEA model, PG23 (2465 D)PG13 v PG23 Xfloat2.64 Capacitive meas., PG13none1.95 Capacitive meas., PG23PG13 v PG23 Xfloat1.16 b of PG13 determined from independent methods 7.9 mm D32 mm D DHI380

Process Measurements Division PG13 (V373) A 0 from crossfloat to C series with fixed b 10:1 area ratio in crossfloat Uncertainty u(b) from PG38, PG39 u(A e ) ranges from 5.8 to 9 ppm (was 20 ppm)

Process Measurements Division D series: PG23 (D49) Crossfloat to PG13, PG35 (V series) to 7 MPa b fixed at common average for PG23 and twin PG32 –PG23 v PG32 crossfloat shows same b in both Uncertainty due to b becomes dominant above 7 MPa u/A: 7 ppm (min)

Process Measurements Division New vs old uncertainties of standards TL, TTL, C Series V and D Series Improvement through direct traceability to PG38/39 Improvement through better distortion estimation

Process Measurements Division Summary of Revised Uncertainties of Standards StandardPressure Range Existing Uncertainties (k=2) Revised Exp. Uncertainties (k=2) UIM1 Pa to 360 kPa5.2 ppm PG38/PG3920 kPa to 1 MPaN/A6.0 to 6.4 ppm PG22/PG3610 kPa to 150 kPa23 ppm10.3 to 23.6 ppm PG28/PG2920 kPa to 350 kPa13 ppm8.4 to 11.1 ppm PG34/PG3740 kPa to 1.4 MPa18 ppm8.5 to 11.6 ppm PG13/PG MPa to 7 MPa40 ppm11.6 to 18 ppm PG23/PG320.7 MPa to 17 MPa33 ppm14 to 40 ppm

Process Measurements Division Validation of uncertainties via international comparisons CCM.P.M-K1c Key Comparison 5 participants, NIST possesses transfer standards (C-415, V-762) 2 ranges: 79 kPa to 896 kPa, 621 kPa to 6.79 MPa Recent data can be compared to Key Comparison Reference Values –C-415 calibrated against PG38 and PG39 –V-762 calibrated against PG13

Process Measurements Division C-415 agrees to KCRV NIST area agrees to KCRV to within 3.5 ppm or less Agreement to KCRV always better than standard uncertainty Agreement to other NMIs always better than standard uncertainty NIST vs KCRV k=1 unc. bars

Process Measurements Division V-762 agrees to KCRV NIST agreement to KCRV always better than standard uncertainty Agreement to other IMGC, LNE, NRLM always better than standard uncertainty. Agreement to PTB better than expanded uncertainty. NIST vs KCRV k=1 unc. bars

Process Measurements Division Summary of international comparisons Direct validation 79 kPa to 6.8 MPa Bilateral comparison with NPL India: –Used NPLI artifact (NPLI-4, V-607) –Confirms equivalence 0.4 to 4.0 MPa

Process Measurements Division Status of Calibration Measurement Capabilities (CMCs) CMCs are statements of the measurement capabilities of each NMI in a database maintained by the BIPM and publicly available on the Web Established as part of Mutual Recognition Arrangement Changes in CMCs must be approved by regional metrology organization (RMO), submitted to BIPM, then approved by the remaining 4 RMOs The revised NIST gas pressure scale: –Submitted for SIM review: May 2008 –Approved by SIM, submitted to BIPM: June 2008 –Currently undergoing inter-regional review

Process Measurements Division Summary NIST gas pressure scale to 17 MPa now based on large diameter piston gauges –Dimensional measurements, modeling, UIM comparisons, and crossfloats confirm results Uncertainties reduced over most of scale –Direct calibration to primary std. improves u for PGs to 1.4 MPa –Better distortion estimation improves u for 7 MPa and 17 MPa PGs –Uncertainties depend on pressure Shifts in effective area of standards –5 ppm lower for standards < 350 kPa –9 ppm or less, most other ranges Distortion coefficients included in most PG ranges Results supported by Key Comparisons New uncertainties expected in NIST cal service late 2008 early 2009