Zinc and Copper Analysis by Colourimetry By Carolyn Fahm TEQAS ~ Trace Elements External Quality Assessment Scheme S: External Quality Assessment.

Slides:



Advertisements
Similar presentations
Assessing Risk of Heavy Metals Input in Food Chain. 1. Introduction Among the many elements that occur in the natural environment about 60% of them are.
Advertisements

Day to Day Management of Quality Control
Elements of Report Writing. Section E, Page 27 Section E, Page 27 In course website, lab handouts section will be an example manuscript In course website,
Mehlich 3 Evaluation Robert O. Miller ALP Technical Director Colorado State University Fort Collins, CO SERA-6 Meeting, Raleigh, NC, June 21, 2011 Miller,
Review of Mobile Laboratoriein Turkey for On Site Analysis An Overview Review of Mobile Laboratoriein Turkey for On Site Analysis An Overview Michael Jackman,
Summary of Key Results from the 2012/2013 Survey of Visa Applicants Who Used a Licensed Adviser Undertaken by Premium Research Prepared: July 2013.
World Health Organization
The Importance of Commutability of Reference Materials Used as Calibrators: The Example of Ceruloplasmin I. Zegers, R. Beetham, T. Keller, J. Sheldon,
L MACKAY, N CRINIS, Q LAM, R SCOTT
Laboratory Quality Control
Cal State Northridge  320 Ainsworth Sampling Distributions and Hypothesis Testing.
Quality Assurance.
Utilization of Assay Performance Characteristics to Estimate Hemoglobin A 1c Result Reliability A. Woodworth, N. Korpi-Steiner, J.J. Miller, L.V. Rao,
The reform of A level qualifications in the sciences Dennis Opposs SCORE seminar on grading of practical work in A level sciences, 17 October 2014, London.
BPS - 3rd Ed. Chapter 131 Confidence intervals: the basics.
Internal Quality Control (QC) for Medical Laboratories: An introduction Dr. Otto Panagiotakis and Dr. Alexander Haliassos ESEAP – Greek Proficiency Testing.
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
CALIBRATION METHODS.
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Unit #7 - Basic Quality Control for the Clinical Laboratory
Simulation Walk Through Seeing how a simulation could work on your course.
Quality Assessment 2 Quality Control.
Anura L. Jayewardene, PhD Drug Research Unit at the General SFGH, 1001 Potero Ave. Bldg 100, Rm 157 San Francisco, CA Phone: (415) FAX:
B. Neidhart, W. Wegscheider (Eds.): Quality in Chemical Measurements © Springer-Verlag Berlin Heidelberg 2000 P. HoulgateAssessment of Test Kits in Terms.
ABC of Quality Control A problem based approach RT ERASMUS NHLS / FACULTY OF HEALTH SCIENCES, UNIVERSITY OF STELLENBOSCH 20th July, 2007, Bela Bela.
Introduction to ISO New and modified requirements.
Copyright © 2010 Pearson Education, Inc. Chapter 22 Comparing Two Proportions.
Summary of Key Results from the 2013/2014 Survey of Visa Applicants Who Used a Licensed Adviser Survey undertaken by: Premium Research Report prepared:
AGA 2009 Tracking Survey Perceptions of Governmental Financial Management Prepared for the Association of Government Accountants December 29, 2009 © Harris.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Analytical Goals. Valid data are essential in making medical decisions, the most important concepts used in judging analytical performance :- 1)Analytical.
Basic Business Statistics, 11e © 2009 Prentice-Hall, Inc. Chap 8-1 Chapter 8 Confidence Interval Estimation Basic Business Statistics 11 th Edition.
Confidence Interval Estimation
Automotive Test & Measurement Case Studies SAAMF Roadshow Durban CSIR NML Eddie Tarnow Metrologist: Torque & Automotive 14 June 2006.
TRICS Performance and Usage Trends 2014/15 MIKE GREEN, SURREY COUNTY COUNCIL.
Lecture 4 Basic Statistics Dr. A.K.M. Shafiqul Islam School of Bioprocess Engineering University Malaysia Perlis
Quality Control Lecture 5
RADIOACTIVITY IN VIRGIN SOILS AND SOILS FROM SOME AREAS WITH CLOSED URANIUM MINING FACILITIES IN BULGARIA Ivanka Yordanova, Lidia Misheva, Martin Banov,
Monitoring the Performance of Nucleic Acid Tests using Data Generated from EDCNet and DigitalPT Wayne Dimech, Darren Jardine, Thu-Anh Pham and the staff.
Laboratory QA/QC An Overview.
BNPCV APCCB BIOLOGICAL VARIATION FOR N-TERMINAL PRO-BNP Jones GRD 1, Bennett A 2,3, Boscato L 1, Macdonald P 2, Brien J 3. Departments of Chemical.
Adv. Inst. Techs.  flame emission (eg flame photometer) known as low temperature emission ( K) ◦ first form of spectroscopy ◦ used in commercial.
CALIBRATION METHODS. For many analytical techniques, we need to evaluate the response of the unknown sample against the responses of a set of standards.
The USP Performance Test Dissolution Systems Suitability Studies Walter W. Hauck, Ph.D. USP Consultant Presentation to Advisory Committee for Pharmaceutical.
0 1 1.Key Performance Indicator Results ( ) KPI Survey Statistics Student Distribution by Year in Program KPI Overall Results Student Satisfaction.
Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function.
Recruitment Statistics Overview of Recruitment 2007/08 Of the 38 vacancies advertised this year, 27 roles were filled and 7 roles were not.
Example 10.2 Measuring Student Reaction to a New Textbook Hypothesis Tests for a Population Mean.
1 Exercise 7: Accuracy and precision. 2 Origin of the error : Accuracy and precision Systematic (not random) –bias –impossible to be corrected  accuracy.
Quality Assessment for Blood Screening Laboratories in Australia and the Asian Pacific. Darren Jardine and Staff of the National Serology Reference Laboratory,
Wenclawiak, B.: Fit for Purpose – A Customers View© Springer-Verlag Berlin Heidelberg 2003 In: Wenclawiak, Koch, Hadjicostas (eds.) Quality Assurance in.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
ARGUMENTS Chapter 15. INTRODUCTION All research projects require some argumentation An argument simply ‘combines’ existing facts to derive new facts,
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
George E. Detsis Manager, Analytical Services Program United States Department of Energy Office of Health, Safety and Security Office of Sustainability.
Patient Safety Monitoring in International Laboratories (SMILE) Mark Swartz, MT(ASCP), SMILE QA/QC Coordinator Improving the Sensitivity of QC Monitoring:
Chapter 1: The Nature of Analytical Chemistry
Table 1. The Five Medical Laboratory Case Studies Jan Martin et al. Transnational Evaluation of a New Continuing Professional Development Activity for.
TESTS OF SIGNIFICANCE. It is a test to compare the results of the method with those accepted method.
Data Screening. What is it? Data screening is very important to make sure you’ve met all your assumptions, outliers, and error problems. Each type of.
Diagnostic clinical chemistry
Trialling Copyright Virtual Management Simulations ENGINUITY.
Pakistan Society Of Chemical Pathologists Zoom Series of Lectures ZT 24. Quality Managent 1 Brig Aamir Ijaz MCPS, FCPS, FRCP (Edin), MCPS-HPE HOD and.
MARK RIDGWELL, DIASUKE TANAKA & CCG WOOD Within run Imprecision
Laboratory Quality Control
Practical clinical chemistry
Introduction To Medical Technology
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
Quality Control Lecture 3
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

Zinc and Copper Analysis by Colourimetry By Carolyn Fahm TEQAS ~ Trace Elements External Quality Assessment Scheme S: External Quality Assessment at Surrey S University of Surrey Guildford

Why We Undertook this Study The majority of laboratories returning Zinc and/or Copper results to TEQAS analyse using Flame Atomic AbsorptionThe majority of laboratories returning Zinc and/or Copper results to TEQAS analyse using Flame Atomic Absorption But there are a growing number general chemistry laboratories measuring these analytes by colourimetry, since it does not require special equipment and can be easily integrated into the regular laboratory workload.But there are a growing number general chemistry laboratories measuring these analytes by colourimetry, since it does not require special equipment and can be easily integrated into the regular laboratory workload.

Why We Undertook this Study - 2 However over the course of time we observed that the number of times that colorimetric results had performance problems that exceeded those seen by labs running Zinc and Copper by FAA.However over the course of time we observed that the number of times that colorimetric results had performance problems that exceeded those seen by labs running Zinc and Copper by FAA. We decided to increase our understanding of what was going on by querying our Zinc and Copper participants about their current or past usage of colourimetric methods.We decided to increase our understanding of what was going on by querying our Zinc and Copper participants about their current or past usage of colourimetric methods.

Questions We asked each respondent to address: What is your current method? Have you ever used or evaluated a colourimetric method? If so ~ ~ Colourimetric Assay Questions Which Method? Method?Automated Or Manual? Type and Source of Calibrant Calibrant? TraceableStandard?EvaluationResults? ObservedPerformanceProblems?

Zinc Results 51 Labs Responded51 Labs Responded 19 Labs reported that they had evaluated colourimetry.19 Labs reported that they had evaluated colourimetry. Currently X labs report Zinc byCurrently X labs report Zinc by this method. this method. 9 used Wako 4 used Randox 1 used Boehringer Mannheim or Sentinel 2 used Audit

Detailed Zinc Information Analysers in use:Analysers in use: Deproteinisation used?Deproteinisation used? Aqueous or Serum Calibrant?Aqueous or Serum Calibrant? Olympus 600,640 Roche Modular, 917 Cobas Fara, Mira yes; 6 no Aqueous; 3 Serum

Detailed Zinc Information - 2 Source of Calibrant:Source of Calibrant: Is calibrant traceable to defined standard?Is calibrant traceable to defined standard? from kit 3 3 Seronorm or other serum-based material 1 1 Merck standard solution no or not known 4 4 yes

Zinc Comments “Precision and accuracy have improved since new analysers came into use in... We appear to over- estimate zinc by about 2  mol/L. Thus our accuracy looks worse at low Zinc levels…”“Precision and accuracy have improved since new analysers came into use in... We appear to over- estimate zinc by about 2  mol/L. Thus our accuracy looks worse at low Zinc levels…” “Few "flyers" when analyses moved from old to new analyser.” “Supplier of kit admitted that this lot of reagents we had been using was un- stable and replaced it.” “Occasional odd results in UK NEQAS Scheme.”

Copper Results 37 Labs Responded37 Labs Responded 6 Labs reported that they had used colourimetry, either currently or in the past6 Labs reported that they had used colourimetry, either currently or in the past Only 4 labs were currently using a colourimetric method.Only 4 labs were currently using a colourimetric method. 1 used Spinreact 1 used Roche 1used Randox 1 used Boehringer Mannheim

Detailed Copper Information Analysers in use:Analysers in use: Deproteinisation used?Deproteinisation used? Aqueous or Serum Calibrant?Aqueous or Serum Calibrant? Cobas Fara Roche Modular, 917 Cobas Fara, Mira yes; 1 no; 1 only for plasma samples 31 3 Aqueous; 1 Serum

Detailed Copper Information - 2 Source of Calibrant:Source of Calibrant: Is calibrant traceable to defined standard?Is calibrant traceable to defined standard? 3 3 from kit 1 1 Serum-based reference material solution 2 2 no 2 2 did not state

Copper Comments “Haemolysed or lipemic samples give high results.”“Haemolysed or lipemic samples give high results.” “Occasional odd result in UK NEQAS Trace Elements Scheme.”“Occasional odd result in UK NEQAS Trace Elements Scheme.”

Zinc and Copper Precision Results Varies with Concentration ~ Several Labs gave two Values. Zinc Within Run Zinc Between Runs 0.4 – 5% Copper Within Run Copper Between Runs 5.2 – 7% 5.2 – 6.7%

The Next Step When this study was reported to the UK NEQAS Trace Elements Specialist Advisory Group, further analysis was suggested.When this study was reported to the UK NEQAS Trace Elements Specialist Advisory Group, further analysis was suggested. Since we have a database of TEQAS results going back for many years, a comparison of the TEQAS performance of colourimetric assays as opposed to other methods of analysis was undertaken.Since we have a database of TEQAS results going back for many years, a comparison of the TEQAS performance of colourimetric assays as opposed to other methods of analysis was undertaken.

Zinc Performance Profile Colourimetry typically yields higher reported concentrations across the concentration ranges.Colourimetry typically yields higher reported concentrations across the concentration ranges. However the statistics are not robust because of widely discrepant N’s (from 48 for FAA to 19 for colourimetry down to 4 for ICP-MS.) Hence the ALTM is most influenced by FAA and least influenced by ICP and Not Specified.However the statistics are not robust because of widely discrepant N’s (from 48 for FAA to 19 for colourimetry down to 4 for ICP-MS.) Hence the ALTM is most influenced by FAA and least influenced by ICP and Not Specified. CV’s (in increasing order) are: ICP, Flame AA, Colourimetry, and Not Specified.CV’s (in increasing order) are: ICP, Flame AA, Colourimetry, and Not Specified. Although there are just four in the ICP category (most precise), the least precise, not specified category has only 7!Although there are just four in the ICP category (most precise), the least precise, not specified category has only 7!

Zinc Concentration Chart

Zinc CV Chart

Zinc Concentration Chart - 2

Zinc CV Chart - 2

Zinc Concentration Chart -3

Zinc CV Chart - 3

Zinc Chart - 4

Zinc CV Chart - 4

Zinc Concentration Summary Slide: 21 Samples, over 7 Months

Zinc CV Summary Slide: 21 Samples, over 7 Months

Copper Performance Profile Colourimetry (for the two reporting participants) tended to be about the same (or slightly lower) than for other methods.Colourimetry (for the two reporting participants) tended to be about the same (or slightly lower) than for other methods. The CV’s were substantially lower – but again, with a mere two participants, few if any conclusions can be drawn.The CV’s were substantially lower – but again, with a mere two participants, few if any conclusions can be drawn. With the withdrawal of Roche’s copper reagents, it is unlikely that colourimetric copper analysis will play a significant role in the future.With the withdrawal of Roche’s copper reagents, it is unlikely that colourimetric copper analysis will play a significant role in the future.

Copper Concentration Chart

Copper CV Chart

Summary Zinc Conc. = Higher Common CV = Higher Copper Conc. =Same or Lower Extremely rare. CV = too few to assess

The entire study of TEQAS samples is available as an attachment.