Thin Film Technology - Student Talk Sophie Chauvin 02/02/16 Leak Detection method for Vacuum Systems 1.

Slides:



Advertisements
Similar presentations
Turbo Pumps Julia Bobak Nanofab Workshop February 11, 2014.
Advertisements

FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Overview of Vacuum Systems Tejas Deshpande 24 July, 2014.
Patricia Aguar Bartolomé Institut für Kernphysik, Universität Mainz PSTP 2013 Workshop, Charlottesville 11th September 2013.
Module 2: Analytical methods and separation techniques.
Mass spectroscopy. In a typical MS procedure:  1- a sample is loaded onto the MS instrument, and undergoes vaporization.  2- the components of the sample.
High-Vacuum Technology Course
Mass Spectrometry The substance being analyzed (solid or liquid) is injected into the mass spectrometer and vaporized at elevated temperature and reduced.
Mass Spectroscopy Mass Spectrometry ä Most useful tool for molecular structure determination if you can get it into gas phase ä Molecular weight of.
Instrumental Chemistry Chapter 11 Atomic Mass Spectrometry.
Mass Spectroscopy 1Dr. Nikhat Siddiqi. Mass spectrometry is a powerful analytical technique that is used to identify unknown compounds, to quantify known.
What are the three common states of matter? Solid, plasma, liquid Liquid, Gas, Plasma Solid, Liquid, Gas None of the above.
6. Mass Spectrometry Adv. Inst. Techs. How does it work? a very small amount of sample is bombarded by a beam of high energy (usually electron beam) produces.
Photoelectron Spectroscopy Lecture 5 – instrumental details –General spectrometer design –Vacuum generation and measurement.
Mass Spectrometer Mass spectrometere is a very powerful method to analyse the structure of organic compounds.
Properties of Fluids SPH4C. Fluids Liquids and gases are both fluids: a fluid is any substance that flows and takes the shape of its container.
Pressure 1 atmosphere ~ 1 bar ~ 760 mm Hg ~ 760 torr ~ 100,000 Pa Ion gauges read in mbar i.e. 1x mbar = 1x atm. Sometimes ion gauges read.
Chapter 8 – Mass Spectrometry. Mass Spectrometry The mass spectrometer can be used for: – Quantitative analysis – as a sophisticated and very sensitive.
GCMS: gas chromatography and mass spectrometry
Forensic Arson and Explosive Investigations. Two Main Areas of Interest: isolation and analysis of flammable residues collection and analysis of explosive.
Year 12 Chemistry Unit 3 – AOS 1 Chemical Analysis.
ASIPP In-time retention evaluation by particle balance analysis on HT-7 Y. YANG*, and HT-7 team Institute of Plasma Physics, Chinese Academy of Sciences.
Mass spectrometry and the PITZ Sven Lederer Technisches Seminar
Vacuum Fundamentals 1 atmosphere = 760 mm Hg = kPa 1 torr = 1 mm Hg vacuum range pressure range low 760 ~ 25 torr medium 25~ high ~ 10.
William Kou Jonathan Mertz. Introduction The field of Mass Spectrometry using Ambient Ionization techniques has grown exponentially since The direct.
S. Jachmich (slide 1) Vessel Conditioning SL-Training, Nov 2010 Vessel conditioning Stefan Jachmich SL-Training 2010.
Vacuum Fundamentals 1 atmosphere = 760 mm Hg = kPa 1 torr = 1 mm Hg vacuum range pressure range low 760 ~ 25 torr medium 25~ high ~ 10.
High-Vacuum Technology Course
DENSITY. Density is the measure of mass per unit volume of substance. So, density is a measure of the amount of matter (particles) in a given volume (space)
Center for Materials for Information Technology an NSF Materials Science and Engineering Center Vacuum Systems Lecture 6 G.J. Mankey
Chapter 13 States of Matter Liquids and Solids Changes of State.
PIMS IN INSTITUTE FOR NUCLEAR STUDIES Brazing Andrzej Soltan Institute for Nuclear Studies.
1.2 Atomic Structure (Time needed: 6 class periods)
26.1 Action of Electric and Magnetic Fields on Matter Chapter 26.
-Plasma can be produced when a laser ionizes gas molecules in a medium -Normally, ordinary gases are transparent to electromagnetic radiation. Why then.
Fabrication and measurement of RF components for CLIC study
SCC Science Dept F W Aston SCC Science Dept Aston’s Mass Spectrometer.
MC 13.3 Spectroscopy, Pt III 1 Introduction to Mass Spectrometry (cont) Principles of Electron-Impact Mass Spectrometry:  A mass spectrometer produces.
Chromatography and Instrumentation. Chromatography Separate Analyze Identify Purify Quantify Components Mixture Chromatography is used by scientists to:
THE MASS SPECTROMETER describes how a mass spectrum is produced using a mass spectrometer.
Mass Spectrometry Relative atomic masses and the mass of individual isotopes can be determined using a mass spectrometer. The principle behind mass spectrometry.
Applications as Residual Gas Analyzer (RGA)
Experience for vacuum plants during XFEL cavity production at companies and at LASA Paolo Michelato TTC Topical Workshop on SRF Cryomodule Clean Room Assembly,
CHROMATOGRAPHY Chromatography is used to separate and analyse small amounts of mixtures Methods involve a stationary phase and a mobile phase. There are.
MICRO-STRIP METAL DETECTOR FOR BEAM DIAGNOSTICS PRINCIPLE OF OPERATION Passing through metal strips a beam of charged particles or synchrotron radiation.
Mass Spectroscopy “You can’t make an omelette without breaking eggs” The Joker.
Atomic Structure Topic The mass spectrometer Describe and explain the operation of a mass spectrometer Describe how the mass spectrometer.
Leak Detection method for Vacuum Systems Alexander Permogorov
Welcome To PPI Pumps Pvt. Ltd. Introduction about Vacuum Pump Vacuum pump is used to pull air or gases from a confined or sealed space.
Matter Created by : Reynaldo Thomas. What is Matter?  An object which takes up space and have mass is called matter.  Everything in your environment.
Vacuum Systems for Electron Microscopy Constraints on Specimens Specimens placed in the electron microscope must be able to withstand very high vacuum.
THE MASS SPECTROMETER WHAT IS A MASS SPECTROMETER
Holders and Pumps Summary
Atomic Structure Topic 2.
Atomic structure L.O. To be able to recall definitions of relative atomic, isotopic and molecular mass. To demonstrate an understanding of the basic.
The Mass Spectrometer Honours
Continuous Flow Isotope Ratio Mass Spectrometry
Atomic structure L.O. 1. To be able to recall definitions of relative atomic, isotopic and molecular mass. 2. To learn about the key principles of mass.
Atomic Structure/ Sub-particles- Definitions Mass Spectrometer
Mass Spectroscopy. Mass Spectroscopy Mass Spectrometry Most useful tool for molecular structure determination if you can get it into gas phase Molecular.
Mass Spectroscopy (MS) Applications as Residual Gas Analyzer (RGA)
1.2 Atomic Structure (Time needed: 6 class periods)
VIII. Production of Vacuum
Atomic Structure.
A sample is injected into the spectrometer and heated to a vapour.
The Mass Spectrometer.
BASIC VACUUM PRACTICE.
Residual Gas Analyzer (RGA)
CHROMATOGRAPHY.
Presentation transcript:

Thin Film Technology - Student Talk Sophie Chauvin 02/02/16 Leak Detection method for Vacuum Systems 1

Definition of the problem Maintenance of the vacuum pressure in the vacuum chamber Need of a good isolation Tightness specifications required  Find a method to detect leaks in a vacuum system Different possible sources of a leak Leak or no leak, leak size, leak location 2

Problem solution  Method: helium leak detector Why helium? small mass small atom volume low partial pressure in air 3

Problem solution Principle: N. Hilleret, “Leak Detection”, CERN, Geneva-Switzerland pp Leak detection cell 4

Problem solution 5  Direct-flow method K. Zapfe, “Leak Detection”, Deutsches Elektronen-Synchrotron DESY, Hamburg, Germany pp Different methods to connect the vacuum system

Problem solution 6  Counter-flow method K. Zapfe, “Leak Detection”, Deutsches Elektronen-Synchrotron DESY, Hamburg, Germany pp

Conclusion 7 Advantages of the counter-flow detectors Commercial leak detectors Time-consuming process

Information The leak detection cell is the main part of the helium leak detector. The residual gas is first ionised using electron beam technology. Then the ionised molecules are accelerated with an accelerating electrode. A mass spectrometer is used to filter the ions and a magnetic field enables the separation of the helium ions from the others. The principle of the mass spectrometer is to measure the gas masses in order to separate the different gases. The magnetic field is usually generated by permanent magnets. 8

Information There are different ways to connect the leak detector to a vacuum system. In the direct-flow method, the leak detection cell and the pump are directly connected to the vacuum system. The helium leak rate is defined using the helium partial pressure in the cell and effective helium pumping speed of the high vacuum pump: 9 Therefore if we want to increase the sensitivity, the pumping speed of the tracer gas (helium) has to be reduced but without decreasing the pumping speed of the other gases (which are mostly water).

Information 10 So water needs a high pumping speed and helium a low pumping speed. A liquid-nitrogen trap is used which enables to condense water. The pressure of the cell is lowered by the high pumping speed of the trap so the leak detection can start earlier. This detector is able to detect leaks at Pa.m 3.s -1 for small systems. As the trap needs to be refilled regularly, it is necessary to have a source of liquid nitrogen.

Information 11 In the counter-flow method, the leak detection cell is not directly connected to the vacuum system; it is connected to the inlet of a high-vacuum pump. The compression ratio K of turbomolecular pumps increases quickly with the mass of the pumped gas so injecting the gas from the cell to the pump enables the obtaining of an backstreaming flux enriched for lighter gases. The helium leak rate is defined using the helium partial pressure in the cell and effective helium pumping speed of the high vacuum pump but also with the compression ratio: The helium partial pressure in the detection cell is smaller is the case of a counter-flow method due to the turbomolecular pump.

Information 12 In practice, it will be impossible to have a complete tight vacuum system and it’s not even needed but it has to be small enough to maintain the required pressure in the system. For a high vacuum system:  If leak rate < mbar.l/s : the system is very tight  If leak rate < mbar.l/s : the system is tight  If leak rate < mbar.l/s : the system is leaky

References A. Preglj, M. Drab, M. Moztic, “Leak Detection Methods and Defining the Sizes of Leaks”, NDT.net -February 1999, Vol.4, No2 N. Hilleret, “Leak Detection”, CERN, Geneva-Switzerland pp K. Zapfe, “Leak Detection”, Deutsches Elektronen-Synchrotron DESY, Hamburg, Germany pp