Gas Chromatography. Presented By - Mr. Shaise Jacob Faculty Nirmala College of Pharmacy Muvattupuzha, Kerala India –

Slides:



Advertisements
Similar presentations
Introduction to Chromatography
Advertisements

Gas Chromatography Introduction 1.) Gas Chromatography
Gas Chromatography Lecture 36.
Gas Chromatography Vaporization of sample Gas-solid
Gas Chromatography 427 PHC.
Lecture 8b Gas Chromatography.
Gas Chromatography & Gas-Liquid Chromatography
GAS CHROMATOGRAPHY ENVE 202 Dr. Aslıhan Kerç.
GAS CHROMATOGRAPHY.
Gas Chromatography There an be many parts to a gas chromatography system but the basic components include: An injection system. A column (controllable.
Lab Methods Day June 25, 2014 Gas Chromatography
1 Gas Chromatography Lecture a. Thermal Conductivity Detector (TCD) This is a nondestructive detector which is used for the separation and collection.
In carbon-13 NMR, what do the number of peaks represent?
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
Gas Chromatography. Mobile phase: Inert gas such as N2 or He. Mobile phase: Inert gas such as N2 or He. Stationary phase: May be solid (GSC) or Stationary.
Gas chromatography is used in many research labs, industrial labs (quality control), forensic (arson and drug analysis, toxicology, etc.), environmental.
Chromatography Chapter 4.
Gas Liquid Chromatography
Gas Chromatography. Gas Chromatography Basics Gas Liquid Chromatography (GLC) Gas Solid Chromatography (GSC) Mobile phase does not interact with analyte.
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Principles of Chromatography. Chromatography is the most powerful tool for separating & measuring the components of a complex mixture. Quantitative &
LECTURE 4: CHROMATOGRAPHY Content: - Chromatographic separation - classifying analytical separations; column chromatography, planar chromatography - gas.
Chapter 6 - Chromatography
Chemical Ideas 7.6 Chromatography. The general principle. Use – to separate and identify components of mixtures. Several different types - paper, thin.
GAS – LIQUID CHROMATOGRAPHY BY M. GLORY HEPSIBAH M.PHARM (PH. ANALYSIS) BHARAT COLLEGE OF PHARMACY.
474 PHG 5.
Chapter 27 Gas Chromatography 1. Principles
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Chromatography Chapter 6.
CHROMATOGRAPHY. The general name given to methods by which two or more compounds in a mixture are physically separated by distributing themselves between.
Gas Chromatography A separation science.... What is Chromatography? The separation of a mixture of two or more compounds or ions by distribution between.
Separation Science Differences in IMFs can be used to separate chemical substances for further analysis. Differences in IMFs can be used to separate chemical.
animation.php.
Chromatography vObjective  To understand the principles of chromatography and know the specific types of Chromatograph used in the analysis of environmental.
GAS LIQUID CHROMATOGRAPHY
CHROMATOGRAPHY Chromatography basically involves the separation of mixtures due to differences in the distribution coefficient.
Introduction to Chromatography Definition Chromatography is a separation technique based on the different interactions of compounds with two phases, a.
By: Arafath and Nick. What is it  Chromatography is a technique that is used to separate the substances present in a mixture.  It is widely used to.
Analytical Separations
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
Introduction to Chromatography. Introduction Chromatography permit the scientist to separate closely related components of complex mixtures. In all chromatographic.
HPLC.
Separation Techniques
GAS CHROMATOGRAPHY In gas chromatography (GC), the sample is injected onto the head of a chromatographic column and immediately vaporized. The components.
GAS LIQUID CHROMATOGRAPHY Principles Partition of molecules between gas (mobile phase) and liquid (stationary phase).
Ch 21 – Principles of Chromatography and Mass Spectrometry Ch 22 – Gas and Liquid Chromatography.
Prof. aza Basic gas chromatography prof. aza Department of Pharmacy, Andalas University STIFI Perintis, Padang STIFAR, Pekan Baru STIFI Bhakti Pertiwi,
Chapter 32 Gas Chromatography. In gas chromatography, the components of a vaporized sample are separated by being distributed between a mobile gaseous.
Dr. Ashraf M. M. Mahmoud, Associate professor
Gas Chromatography. In Gas Chromatography (GC), a gaseous mobile phase transports a gaseous solute through a long, thin column containing solid or liquid.
Instrumental Analysis (I)  HPLC Tutorial #7 PHCMt561 – 5 th Sem. Pharm.
1 Principles of Chromatography Chap Analytical Separations and Chemical Problem Solving If you, a researcher of a food company are asked to find.
Chromatography.
M.PRASAD NAIDU Msc Medical Biochemistry, Ph.D Research scholar.
M.PRASAD NAIDU Msc Medical Biochemistry, Ph.D Research scholar.
CHROMATOGRAPHY  A laboratory technique in which the components of a sample are separated based on how they distribute between two chemical or physical.
Experiments in Analytical Chemistry
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
Overview Determining Identity Quantitative Analysis
Chemical Ideas 7.6 Chromatography.
A guide for A level students KNOCKHARDY PUBLISHING
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Chapter 27 Gas Chromatography
An Introduction to Chromatographic Separations
VAPOUR PHASE CHROMATOGRAPHY
Chromatography Daheeya Alenazi.
Principle of separation of different components:
Gas Chromatography.
GAS CHROMATOGRAPHY.
CHROMATOGRAPHY.
Presentation transcript:

Gas Chromatography

Presented By - Mr. Shaise Jacob Faculty Nirmala College of Pharmacy Muvattupuzha, Kerala India –

What is Gas Chromatography? It is also known as…It is also known as… –Gas-Liquid Chromatography (GLC)

GAS CHROMATOGRAPHY  Separation of gaseous & volatile substances  Simple & efficient in regard to separation GC consists of GSC (gas solid chromatography) GLC (gas liquid chromatography Gas → M.P Solid / Liquid → S.P GSC not used because of limited no. of S.P GSC principle is ADSORPTION GLC principle is PARTITION

Sample to be separated is converted into vapour And mixed with gaseous M.P Component more soluble in the S.P → travels slower Component less soluble in the S.P → travels faster Components are separated according to their Partition Co-efficient Criteria for compounds to be analyzed by G.C 1.VOLATILITY: 2.THERMOSTABILITY:

What is Gas Chromatography? The father of modern gas chromatography is Nobel Prize winner John Porter Martin, who also developed the first liquid-gas chromatograph. (1950)The father of modern gas chromatography is Nobel Prize winner John Porter Martin, who also developed the first liquid-gas chromatograph. (1950)

The Next Generation in Gas Chromatography

How a Gas Chromatography Machine Works –First, a vaporized sample is injected onto the chromatographic column. –Second, the sample moves through the column through the flow of inert gas. –Third, the components are recorded as a sequence of peaks as they leave the column.

Chromatographic Separation –Deals with both the stationary phase and the mobile phase. Mobile – inert gas used as carrier.Mobile – inert gas used as carrier. Stationary – liquid coated on a solid or a solid within a column.Stationary – liquid coated on a solid or a solid within a column.

Chromatographic Separation Chromatographic SeparationChromatographic Separation –In the mobile phase, components of the sample are uniquely drawn to the stationary phase and thus, enter this phase at different times. –The parts of the sample are separated within the column. –Compounds used at the stationary phase reach the detector at unique times and produce a series of peaks along a time sequence.

Chromatographic Separation (continued) –The peaks can then be read and analyzed by a forensic scientist to determine the exact components of the mixture. –Retention time is determined by each component reaching the detector at a characteristic time.

Chromatographic Analysis –The number of components in a sample is determined by the number of peaks. –The amount of a given component in a sample is determined by the area under the peaks. –The identity of components can be determined by the given retention times.

Peaks and Data

PRACTICAL REQUIREMENTS Carrier gas Flow regulators & Flow meters Injection devices Columns Temperature control devices Detectors Recorders & Integrators

CARRIER GAS » Hydrogen better thermal conductivity disadvantage: it reacts with unsaturated compounds & inflammable » Helium excellent thermal conductivity it is expensive » Nitrogen reduced sensitivity it is inexpensive

Requirements of a carrier gas Inertness Suitable for the detector High purity Easily available Cheap Should not cause the risk of fire Should give best column performance

Flow regulators & Flow meters  deliver the gas with uniform pressure/flow rate  flow meters:- Rota meter & Soap bubble flow meter Rota meter  placed before column inlet it has a glass tube with a float held on to a spring. the level of the float is determined by the flow rate of carrier gas

Soap Bubble Meter ◊ Similar to Rota meter & instead of a float, soap bubble formed indicates the flow rate

Injection Devices  Gases can be introduced into the column by valve devices  liquids can be injected through loop or septum devices

COLUMNS Important part of GC Made up of glass or stainless steel Glass column- inert, highly fragile COLUMNS can be classified  Depending on its use 1. Analytical column meters length & 3-6 mm d.m 2. Preparative column 3-6 meters length, 6-9mm d.m

 Depending on its nature 1.Packed column: columns are available in a packed manner S.P for GLC: polyethylene glycol, esters, amides, hydrocarbons, polysiloxanes… 2.Open tubular or Capillary column or Golay column  Long capillary tubing M in length  Uniform & narrow d.m of cm  Made up of stainless steel & form of a coil  Disadvantage: more sample cannot loaded

3.SCOT columns (Support coated open tubular column  Improved version of Golay / Capillary columns, have small sample capacity  Made by depositing a micron size porous layer of supporting material on the inner wall of the capillary column  Then coated with a thin film of liquid phase

Columns Packed Capillary

Equilibration of the column  Before introduction of the sample  Column is attached to instrument & desired flow rate by flow regulators  Set desired temp.  Conditioning is achieved by passing carrier gas for 24 hours

Temperature Control Devices  Preheaters: convert sample into its vapour form, present along with injecting devices  Thermostatically controlled oven: temperature maintenance in a column is highly essential for efficient separation. Two types of operations  Isothermal programming:-  Linear programming:- this method is efficient for separation of complex mixtures

Temperature Control Isothermal Gradient Instrumentation - Oven

DETECTORS  Heart of the apparatus The requirements of an ideal detector are-  Applicability to wide range of samples  Rapidity  High sensitivity  Linearity  Response should be unaffected by temperature, flow rate…  Non destructive  Simple & inexpensive

Measures the changes of thermal conductivity due to the sample (  g). Sample can be recovered. 1.Thermal Conductivity Detector (Katharometer, Hot Wire Detector)

Thermal Conductivity Basics When the carrier gas is contaminated by sample, the cooling effect of the gas changes. The difference in cooling is used to generate the detector signal. The TCD is a nondestructive, concentration sensing detector. A heated filament is cooled by the flow of carrier gas. Flow

When a separated compound elutes from the column, the thermal conductivity of the mixture of carrier gas and compound gas is lowered. The filament in the sample column becomes hotter than the control column. The imbalance between control and sample filament temperature is measured by a simple gadget and a signal is recorded Thermal Conductivity Detector

Measures heat loss from a hot filament – filament heated to const T when only carrier gas flows heat loss to metal block is constant, filament T remains constant. when an analyte species flows past the filament generally thermal conductivity goes down, T of filament will rise. (resistance of the filament will rise).

Relative Thermal Conductivity Compound Relative Thermal Conductivity Carbon Tetrachloride0.05 Benzene0.11 Hexane0.12 Argon0.12 Methanol0.13 Nitrogen0.17 Helium1.00 Hydrogen1.28

Advantages of Katharometer  Linearity is good  Applicable to most compounds  Non destructive  Simple & inexpensive Disadvantages Low sensitivity Affected by fluctuations in temperature and flow rate Biological samples cannot be analyzed

Flame Ionization Detector  Destructive detector  The effluent from the column is mixed with H & air, and ignited.  Organic compounds burning in the flame produce ions and electrons, which can conduct electricity through the flame.  A large electrical potential is applied at the burner tip  The ions collected on collector or electrode and were recorded on recorder due to electric current.

FIDs are mass sensitive rather than conc. sensitive ADVANTAGES: µg quantities of the solute can be detected Stable Responds to most of the organic compounds Linearity is excellent DA: destroy the sample

FID

Argon ionization detector  Depends on the excitation of argon atoms to a metastable state, by using radioactive energy. Argon → irradiation Argon + e - → collision Metastable Argon → collision of sub. → Ionization → ↑ Current ADVANTAGES 1.Responds to organic compounds 2.High sensitivity DISADVANTAGES 1.Response is not absolute 2.Linearity is poor 3. Sensitivity is affected by water

ELECTRON CAPTURE DETECTOR The detector consists of a cavity that contains two electrodes and a radiation source that emits  - radiation (e.g. 63 Ni, 3 H) The collision between electrons and the carrier gas (methane plus an inert gas) produces a plasma containing electrons and positive ions.

If a compound is present that contains electronegative atoms, those electrons are captured and negative ions are formed, and rate of electron collection decreases The detector selective for compounds with atoms of high electron affinity. This detector is frequently used in the analysis of chlorinated compounds e.g. – pesticides, polychlorinated biphenyls

ADVANTAGE Highly sensitive DISADVANTAGE Used only for compounds with electron affinity

RECORDERS & INTEGRATORS Record the baseline and all the peaks obtained INTEGRATORS Record the individual peaks with Rt, height….

Derivatisation of sample  Treat sample to improve the process of separation by column or detection by detector.  They are 2 types  Precolumn derivatisation Components are converted to volatile & thermo stable derivative. Conditions - Pre column derivatisation  Component ↓ volatile  Compounds are thermo labile  ↓ tailing & improve separation

Post column derivatisation  Improve response shown by detector  Components ionization / affinity towards electrons is increased Pretreatment of solid support  To overcome tailing  Generally doing separation of non polar components like esters, ethers… Techniques: 1. use more polar liquid S.P 2. Increasing amt. of liquid phase 3.Pretreatment of solid support to remove active sites.

Parameters used in GC Retention time (Rt) It is the difference in time b/w the point of injection & appearance of peak maxima. Rt measured in minutes or seconds (or) It is the time required for 50% of a component to be eluted from a column Retention volume (Vr) It is the volume of carrier gas which is required to elute 50% of the component from the column. Retention volume = Retention time ˣ Flow rate

Separation factor (S) Ratio of partition co-efficient of the two components to be separated. If more difference in partition co-efficient b/w two compounds, the peaks are far apart & S Is more. If partition co-efficient of two compounds are similar, then peaks are closer Resolution (R) The true separation of 2 consecutive peaks on a chromatogram is measured by resolution It is the measure of both column & solvent efficiencies R= 2 d W 1 +W 2

Retention time

Separation factor

Resolution

THEORETICAL PLATE  An imaginary unit of the column where equilibrium has been established between S.P & M.P  It can also be called as a functional unit of the column HETP – Height Equivalent to a Theoretical Plate  Efficiency of a column is expressed by the number of theoretical plates in the column or HETP  If HETP is less, the column is ↑ efficient.  If HETP is more, the column is ↓ efficient

HETP= L (length of the column) N (no of theoretical plates) HETP is given by Van Deemter equation HETP= A + B +Cu u A = Eddy diffusion term or multiple path diffusion which arises due to packing of the column B = Molecular diffusion, depends on flow rate C = Effect of mass transfer,depends on flow rate u = Flow rate

Efficiency ( No. of Theoretical plates) It can be determined by using the formula n = 16 Rt 2 w 2 N = no. of theoretical plates Rt = retention time W = peak width at base  The no. of theoretical plates is high, the column is highly efficient  For G.C the value of 600/ meter

Asymmetry Factor  Chromatographic peak should be symmetrical about its centre  If peak is not symmetrical- shows Fronting or Tailing  FRONTING Due to saturation of S.P & can be avoided by using less quantity of sample  TAILING Due to more active adsorption sites & can be eliminated by support pretreatment,

Asymmetry factor ( ) can be calculated by using the formula AF=b/a b & a calculated at 5% or 10% of the peak height

ADVANTAGES OF G.C  Very high resolution power, complex mixtures can be resolved into its components by this method.  Very high sensitivity with TCD, detect down to 100 ppm  It is a micro method, small sample size is required  Fast analysis is possible, gas as moving phase- rapid equilibrium  Relatively good precision & accuracy  Qualitative & quantitative analysis is possible

Gas Chromatography vials caps

Chromatographic Analysis –The number of components in a sample is determined by the number of peaks. –The amount of a given component in a sample is determined by the area under the peaks. –The identity of components can be determined by the given retention times.

Applications of G.C G.C is capable of separating, detecting & partially characterizing the organic compounds, particularly when present in small quantities. 1, Qualitative analysis Rt & RV are used for the identification & separation 2, Checking the purity of a compound Compare the chromatogram of the std. & that of the sample

3, Quantitative analysis It is necessary to measure the peak area or peak height of each component 4, used for analysis of drugs & their metabolites.

Semi-Quantitative Analysis of Fatty Acids C C C Detector Response Retention Time Peak Area Sample Concentration (mg/ml)

Tentative Identification of Unknown Compounds Response GC Retention Time on Carbowax-20 (min) Mixture of known compounds Hexane Octane Decane 1.6 min = RT Response Unknown compound may be Hexane 1.6 min = RT Retention Time on Carbowax-20 (min)

Response GC Retention Time on SE-30 Unknown compound RT= 4 min on SE-30 Response GC Retention Time on SE-30 Hexane RT= 4.0 min on SE-30 Retention Times

Advantages of Gas Chromatography Very good separation Time (analysis is short) Small sample is needed -  l Good detection system Quantitatively analyzed

How a Gas Chromatography Machine Works –First, a vaporized sample is injected onto the chromatographic column. –Second, the sample moves through the column through the flow of inert gas. –Third, the components are recorded as a sequence of peaks as they leave the column.