Testing of Highly Accurate Blackbodies Harri Latvakoski Mike Watson Shane Topham Mike Wojcik SDL/10-087.

Slides:



Advertisements
Similar presentations
Ground-based FTIR Update CAVIAR Meeting NPL, 29 th September 2010 Tom Gardiner, Marc Coleman.
Advertisements

Modeling and Sizing a Thermoelectric Cooler Within a Thermal Analyzer Jane Baumann C&R Technologies, Inc. Littleton, Colorado.
CAVITY RING DOWN SPECTROSCOPY
Development of an External Cavity Quantum Cascade Laser for High- Resolution Spectroscopy of Molecular Ions JACOB T. STEWART, BRADLEY M. GIBSON, BENJAMIN.
1 NIST LBIR Capabilities for Absolute Radiometric Calibrations Dr. Raju Datla NIST Optical Technology Division Gaithersburg, MD PIs: Dr. Adriaan.
Pre-launch Characterization of the CERES Flight Model 5 (FM5) Instrument on NPP S. Thomas a, K. J. Priestley b, M. Shankar a, N. P. Smith a, M. G. Timcoe.
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Remote sensing in meteorology
Test Cryostat, OGSE and MGSE PACS IHDR: MPE 12/13 Nov 2003 AIV1 PACS Test Cryostat, OGSE and MGSE Gerd Jakob MPE.
Uncertainty considerations for the calibration of transfer standard radiation thermometers Graham Machin, NPL.
Calibration and Status of MOBY Dennis Clark, NOAA/NESDIS Carol Johnson, NIST Steve Brown, NIST Mark Yarbrough, MLML Stephanie Flora, MLML Mike Feinholz,
14/06/20151 MORE Requirements seen from ESA Pedro Pablos 1 st MORE Team Meeting 27 Febrero 2007.
ASIC3 WorkshopLandsdowne, VA May 16-18, 2006 J. Harder Page 1 Calibration Status of the Solar Irradiance Monitor (SIM) : The Present and the Future Jerald.
Senior Design II Film Thickness Measurement Julian Peters Joe Fitzmyer Brad Demers Coordinator: Dr. Wayne Walter Advisor: Dr. Dale Ewbank Sponsor: Dr.
PERFORMANCE OF THE DELPHI REFRACTOMETER IN MONITORING THE RICH RADIATORS A. Filippas 1, E. Fokitis 1, S. Maltezos 1, K. Patrinos 1, and M. Davenport 2.
E Usadi, NPL Low Temperature Blackbody for CAVIAR: Characterisation & Calibration Eric Usadi Thanks to Chris Price Theo Theocharous Title E Usadi 02Jun08.
Millimetron mision sensitivities and instrumentation concept
Overview of Temperature Measurement ME 115 Figures are from “Practical Guidelines for Temperature Measurement” unless otherwise notedwww.omega.com.
Are you getting the concept? Calculate D a, D l, R d and s g for 1 st order diffraction under optimal conditions for the indicated 0.5 m grating with 100.
Hyperspectral Satellite Imaging Planning a Mission Victor Gardner University of Maryland 2007 AIAA Region 1 Mid-Atlantic Student Conference National Institute.
Improving Uncertainties of Non-Contact Thermometry Measurements Mark Finch Fluke Calibration.
SIMULTANEOUS MEASUREMENT OF TEMPERATURE AND PRESSURE SENSOR USING BRAGG GRATINGS.
and 4181 Precision Infrared Calibrators Sales Presentation Measure With Confidence New 4180 Series IR Calibrators.
9 March 2004 GIFTS Blackbody Subsystem Critical Design Review Emissivity Modeling and Uncertainty Budget 09 March 2004.
CLARREO FTS Design: Achieving Robust On-Orbit SI Traceability for IR - Principles Distilled from NIST John A. Dykema, Joe Demusz, Chris Tuozzolo, Norton.
Atomic Emission Spectroscopy
2010 CEOS Field Reflectance Intercomparisons Lessons Learned K. Thome 1, N. Fox 2 1 NASA/GSFC, 2 National Physical Laboratory.
Sergey Mekhontsev National Institute of Standards and Technology Optical Technology Division, Gaithersburg, MD Infrared Spectral Radiance Scale.
NA62 Gigatracker Working Group Meeting 2 February 2010 Massimiliano Fiorini CERN.
Quick Review of Remote Sensing Basic Theory Paolo Antonelli CIMSS University of Wisconsin-Madison Benevento, June 2007.
Infrared Spectroscopy using Quantum Cascade Lasers Peng Wang and Tom Tague Bruker Optics, Billerica, MA Laurent Diehl, Christian Pflügl and Federico.
Performance characteristics for measurement and instrumentation system
Slide 1 Progress towards Achieving On-Orbit SI Traceability for the CLARREO IR Payload Hampton, VA, April 10, 2012 Recent results towards verification.
HD target.
9 March 2004 GIFTS Blackbody Subsystem Critical Design Review Action Items From PDR Fred Best 9 March 2004.
Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric.
DECam Daily Flatfield Calibration DECam calibration workshop, TAMU April 20 th, 2009 Jean-Philippe Rheault, Texas A&M University.
Jörn Helbert Planetary Emissivity Laboratory Facing the heat – Obtaining near infrared real emissivity spectra at Venus surface temperatures.
1 Reflected Solar Calibration Demonstration System - SOLARIS K. Thome, D. Jennings, B. McAndrew, J. McCorkel, P. Thompson NASA/GSFC.
Thoughts on the Design of a WVR for Alan Roy (MPIfR) the Twin Telescope at Wettzell.
A Method for Correcting for Telescope Spectral Transmission in the Geosynchronous Imaging Fourier Transform Spectrometer (GIFTS) John D. Elwell, Deron.
Extended Detector Cutoff Considerations WFIRST Project Office May
Scisat Test Readiness Review MAESTRO Evaluation and Verification Tests University of Toronto 29 January 2003.
PACS IBDR MPE 27/28 Feb 2002 AIV 1 PACS IBDR Test Cryostat and OGSE Gerd Jakob MPE.
Laser heating in the diamond anvil cell. mineral properties at high P,T constrain phases, temperature, and composition of earth & planets’ interiors understand.
Initial Results from the Scintillator Fast Lost Ion Probe D. Darrow NSTX Physics Meeting February 28, 2005.
Increased Thermal Background for the post-NCS NICMOS TIPS – June 19, 2003 Megan Sosey NICMOS.
An Experimental Approach to the Prediction of Complete Millimeter and Submillimeter Spectra at Astrophysical Temperatures Ivan Medvedev and Frank C. De.
Beam on Target Diagnostics Beam on Target Meeting 2013 March Tom Shea.
Date of download: 6/1/2016 Copyright © 2016 SPIE. All rights reserved. (a) Vision of the Brillouin lidar operated from a helicopter. The center ray represents.
 FT-IR stands for Fourier Transform Infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR radiation is passed through.
Damian Luna Yetziel Sandoval – Alberto Gonzales – 80546
Use of Air Showers to Reduce Soaking Time for High Precision Dimensional Measurements Ted Doiron, Donna McLaughlin, Ann Schneider Engineering Metrology.
1 CLARREO Advances in Reflected Solar Spectra Calibration Accuracy K. Thome 1, N. Fox 2, G. Kopp 3, J. McCorkel 1, P. Pilewskie 3 1 NASA/Goddard Space.
2015 GSICS Annual Meeting, Deli India March 16~20, 2015 Xiuqing Hu National Satellite Meteorological Center, CMA Yupeng Wang, Wei Fang Changchun Institute.
CLARREO Pathfinder - Overview
Digital Light Sources First introduced in 2001.
The “Monitor to measure the Integral TRAnsmittance” (MITRA)
Paper under review for JGR-Atmospheres …
Electromagnetic Spectrum
Global Perspective, Local Commitment
Overview of Temperature Measurement ME 115
Instrument Considerations
Detective Quantum Efficiency Preliminary Design Review
GIFTS Blackbody Subsystem Critical Design Review Blackbody Requirements Flowdown Fred Best 9 March 2004.
Early calibration results of FY-4A/GIIRS during in-orbit testing
OL 750 Measurement Systems
REMOTE SENSING.
Shanghai Institute of Technical Physics , Chinese Academy of Science
Remote sensing in meteorology
Presentation transcript:

Testing of Highly Accurate Blackbodies Harri Latvakoski Mike Watson Shane Topham Mike Wojcik SDL/10-087

Outline High-accuracy blackbodies CORSAIR blackbody Blackbody testing needs and testing types Blackbody testing examples Conclusions

High accuracy blackbody needs There is increasing need for high accuracy blackbodies emissivity to Temperature uncertainties ~100 mK or less Driven by increasing accuracy needs of modern Earth-observing infrared instruments (e.g. FIRST and CLARREO) FIRST accuracy goals: K (1σ) cm K (1σ) cm -1 CLARREO accuracy requirements: ~ K (3σ) ~200 to 2000 cm -1 Maintain accuracy over 5 years on orbit CLARREO has strictest requirements but other instruments are close Both FIRST and CLARREO require long wave performance (unusual)

Blackbodies SDL has built several high accuracy blackbodies in the past E.g. for FIRST SDL has built a prototype (the CORSAIR blackbody) to show ability to meet CLARREO requirements As part of the NASA Langley IIP program CORSAIR Design is based on the FIRST blackbodies Easily scalable for different aperture sizes or requirements Simple to build LWIRCS

CORSAIR Blackbody Specifications Blackbody is designed to be a useful ground unit Wavelength range2 to 50  m Controlled temperature range:200 to 350 K Operating environment:Ambient or inside larger vacuum chamber Coolant:Liquid nitrogen Environment temperature:77 K to ambient Aperture:1.75 inches Beam divergence:6° full angle Emissivity  Temperature uncertainties30 mK under CLARREO like conditions Large operating range adds significant complexity

CORSAIR Blackbody Overall Design Vacuum shell, LN 2 and heated shroud are needed for wide range of environments

CORSAIR Blackbody Cavity Design Aluminum, steel, and fiberglass Specular trap design, works to long wavelength Numerous features keep gradients low

Blackbody testing CORSAIR emissivity  , temperature uncertainty 30 mK Can we believe it? …want proof for these high performance levels In general, blackbody performance is calculated from physical models and temperature sensor calibration. These results are quoted as performance Blackbodies usually NOT tested for performance or compared to other blackbodies Blackbody end-to-end testing is a significant expense Earth observing mission, such as CLARREO, attempting to contribute to a long record of observation need documented SI traceability. This requires testing Historically, when equipment or experimental results have been compared, they are often found to disagree by more then the claimed uncertainties

Types of blackbody testing Blackbody testing usual considered to mean end-to-end performance testing Compare blackbody to another blackbody Requires stable, highly accurate radiometer to observe both blackbodies Spectral test Can’t prove new design claiming higher accuracy than previous designs meets its requirement Observe with absolute radiometer based on electrical substitution standard Broadband test Requires very low background, otherwise similar to above It is also possible to test various blackbody characteristics one at a time 3 blackbody error sources; each can be tested to some extent Temperature gradients Temperature sensor errors Non-zero reflectivity Disadvantage: not a complete performance test Advantages: Increases confidence in model results Simpler then end-to-end test Blackbody can be designed for simple testing in some cases

Blackbody testing CORSAIR blackbody was designed for simple testing of some characteristics SDL has performed several type of blackbody testing with several blackbodies Preliminary results of testing performed on the CORSAIR blackbody follow Temperature gradient test Temperature sensor test Emissivity test End-to-end test

Testing examples – Temperature gradients Gradients across surfaces viewed in blackbody introduce error Can test for temperature gradients with sufficiently accurate temperature sensors at critical locations in blackbody Plot of sensors on CORSAIR blackbody cylinder and cone shows no gradients at -39 C No gradients to few mK level at other temps

Testing examples – sensor uncertainties Blackbody temperature sensors can be calibrated to 10 mK or better. Need to ensure sensors maintain calibration Can drift over time Can change reading from mounting or handling (mechanical stress) 1. In CORSAIR blackbody, multiple sensors agree when in blackbody so calibration is unlikely to have changed 2. CORSAIR blackbody contains phase change cells Cell with water, gallium, and mercury Cell is independently temperature controlled from blackbody Allows for calibration of temperature sensors: Take cell through melt point Calibrate sensor in cell Turn off cell, it equilibrates with blackbody Transfer calibration to blackbody sensors Blackbody TEC Container with temperature sensor Hg, H 2 O, Ga

Testing examples – sensor uncertainties Take cell to melt point and hold Absolute standard allows check of temperature sensor calibration Early results show CORSAIR cell temperature sensors consistent with known melt points to within ~20 mK. Temperature sensors were calibrated to ~15 mK (1σ) prior to placement in blackbody Can identify melt point by going through melt point and observing melt curve Absolute accuracy limited: melt curve is never completely flat, by definition melting requires gradients across cell Works well for trending Current cell design monitors material expansion to unambiguously identify melt point

Testing examples – Emissivity Emissivity must be high or at least well-known for an accurate blackbody The CORSAIR blackbody includes a emissivity monitor based on a quantum cascade laser (QCL) CLARREO will require emissivity monitor, CORSAIR includes one for demonstration Single 9.6  m 150 mW QCL Instrument viewing blackbody is detector Broad-beam Laser shines into blackbody with no optics

Testing examples – Emissivity Simple monitor design mimics observation geometry, so it allows a measure of blackbody emissivity Monitor used during test Spectrometer did not detect laser Emissivity  0.999,992 This is consistent with model and paint data at 9.6  m

Testing examples – End-to-End test We used the SDL Transfer Radiometer (SXR) to compare the CORSAIR blackbody with our existing Long-wave Infrared Calibration Source (LWIRCS) SXR (left) CORSAIR blackbody (right) The SXR is a combination spectrometer and radiometer in a single vacuum shell

SXR SXR basic operating specifications Wavelength range2 to 30  m DetectorsSi:As (11 K, single element used at a time) CoolantLiquid Helium (20 K optical bench) Filters28 (4 wheels, 8 slots) singles Window (removable)KBr, heated to 310 K ModesSpectrometer, radiometer Spectrometer resolution0.5, 1, 2, 4 cm -1 Spectrometer samplingOnce per HeNe laser fringe, Nyquist 7900 cm -1 Aperture70 mm (2.76”) Beam divergence1 mrad All SXR options software selectable while chamber is cold

CORSAIR SXR Test For testing on SXR, CORSAIR blackbody aperture was increased to 3” LWIRCS observed at 8 temps ~100, 179, 233, 249, 273, 290, 310, 335 K 8 filters in radiometer mode, 6 in spectrometer mode CORSAIR observed at 5 temps 233, 249, 273, 290, 310 K Same mode, filter set as for LWIRCS Data not yet analyzed Quick look analysis of 8.4 and 12  m narrow bands show LWIRCS, CORSAIR at same temperature consistent to better than 1%

Conclusions / Future Work Blackbody testing is useful We have performed several types of blackbody tests Test data will be processed May need to consider how to improve tests if accuracy is not at desired level