1 PMT Univ. of Tokyo Yasuko HISAMATSU ICEPP, MEG Collaboration meeting Feb. 10th, 2004.

Slides:



Advertisements
Similar presentations
Prototype of the Daya Bay Neutrino Detector Wang Zhimin IHEP, Daya Bay.
Advertisements

1 Satoshi Mihara for the   e  collaboration, muegamma review at PSI, Jan 2003 Photon Detector ICEPP, Univ. of Tokyo Satoshi Mihara.
Liquid Xenon Carlorimetry at the MEG Experiment Satoshi MIHARA Univ. of Tokyo.
Calibration of the 10inch PMT for IceCube Experiment 03UM1106 Kazuhiro Fujimoto A thesis submitted in partial fulfillment of the requirements of the degree.
ACD calibrations Pedestals Measured from online script Measure PHA w/ HV off, no charge injection Use cyclic triggers ~ ADC counts, very narrow.
CFT Calibration Calibration Workshop Calibration Requirements Calibration Scheme Online Calibration databases.
XEC Hardware Status MEG Review Meeting 2010 Liquid Xenon Group.
Xenon Detector Status Liquid Xenon Detector Group.
Measurement of the absolute efficiency,
Performance of the PHENIX Muon Tracking System in Run-2 Ming X. Liu Los Alamos National Lab (for the PHENIX Collaboration) –Detector Commissioning –Detector.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
Satoshi Mihara ICEPP, Univ. of Tokyo Feb MEG Review Meeting 1 CEX beam test at piE1 Satoshi Mihara.
The PEPPo e - & e + polarization measurements E. Fanchini On behalf of the PEPPo collaboration POSIPOL 2012 Zeuthen 4-6 September E. Fanchini -Posipol.
Leroy Nicolas, HESS Calibration results, 28 th ICRC Tsukuba Japan, August Calibration results of the first two H·E·S·S· telescopes Nicolas Leroy.
Performance test of STS demonstrators Anton Lymanets 15 th CBM collaboration meeting, April 12 th, 2010.
14/02/2007 Paolo Walter Cattaneo 1 1.Trigger analysis 2.Muon rate 3.Q distribution 4.Baseline 5.Pulse shape 6.Z measurement 7.Att measurement OUTLINE.
Analysis of PSI beam test R.Sawada 09/Feb/2004 MEG collaboration R.Sawada 09/Feb/2004 MEG collaboration
1 N eutrino E xtended S ubmarine T elescope with O ceanographic R esearch Operation and performance of the NESTOR test detector.
E-165 (FLASH) Status 8:00am 6/13/2004. Thick Target Program Primary goal is to measure the fluorescence light output at different stages of shower development.
Neutron Monitoring Detector in KIMS Jungwon Kwak Seoul National University 2003 October 25 th KPS meeting.
Possible calibration methods for the final LXe calorimeter A. Papa 01/20/2004.
XMASS experiment Current status 10 th ICEPP Symposium in Hakuba 16 Feb 2004 Yohei Ashie ICRR Univ.of Tokyo.
1 Report on analysis of PoGO Beam Test at Spring-8 Tsunefumi Mizuno July 15, 2003 July 21, 2003 revised August 1, 2003 updated.
1 Satoshi Mihara for the   e  collaboration, review meeting at PSI, Jul 2002 Photon Detector Satoshi Mihara ICEPP, Univ. of Tokyo 1.Large Prototype.
in Beijing August 18, 2004 Recent status of the XMASS project Physics goals at XMASS Overview of XMASS Current status of R&D Summary.
Neutron detection in LHe ( HMI run 2004) R.Golub, E. Korobkina, J. Zou M. Hayden, G. Archibold J. Boissevain, W.S.Wilburn C. Gould.
Liquid Xenon Detector and Related Topics
Status Report for KEK-PS E391a KEK IPNS G.Y.Lim 14 April 2003.
Abort Gap Monitoring Randy Thurman-Keup 6 / 8 / 2004 LARP Meeting.
1 A.Tsirigotus Hellenic Open University N eutrino E xtended S ubmarine T elescope with O ceanographic R esearch The NESTOR O.M.
Study of the MPPC for the GLD Calorimeter readout Satoru Uozumi (Shinshu University) Feb Beijing Introduction Basic performances Future.
DOM Two Dimensional Scanning using testDAQ Hiroko Miyamoto T. Noda Faculty of Science Department of Physics Chiba University.
Current status of MEG Experiment Yasuko HISAMATSU ICEPP, The Univ. of Tokyo ICEPP Symposium.
Performance of SiPM Ryuhei Nakamura (Kobe Univ.) GLC calorimeter group (KEK, Kobe, Konan, Niigata, Shinshu, Tsukuba) Contents ・ Introduction ・ Test results.
Liquid Xenon Photon Detector Feb MEG Review Meeting Liquid Xenon Detector and Related Topics CEX beam test at piE1 Oct-Dec/03 Detector calibration.
Possible calibration methods for the final LXe calorimeter A. Papa 02/11/
Design for Wide FOV Cherenkov telescope upgrading THE 2 nd WORKSHOP OF IHEP Shoushan Zhang Institute of High Energy Physics.
G. Tzanakos, MINERvA Collaboration Meeting, Jan 14, Goal: Build a Test Stand that can test and characterize 10 PMTs. Assumptions about MINERvA Electronics:
DESY BT analysis - MPPC Saturation Correction - S. Uozumi Feb Sci-ECAL meeting 1.MPPC Gain Measurement (with LED data) 2.Inter-calibration of readout.
MEG 実験用液体 Xeγ 線検出器用 光電子増倍管について I 森研究室修士2年 久松康子 東大素粒子セ, 早大理工総研 A, 高エネ研 B, BINP-Novosibirsk C, INFN-Pisa D, PSI E 岩本敏幸, 内山雄祐, 大谷航, 小曽根健嗣, 菊池順 A, 古田島拓也 A,
1 Electronics Status Trigger and DAQ run successfully in RUN2006 for the first time Trigger communication to DRS boards via trigger bus Trigger firmware.
PMT Calibration R.Sawada 7/Jan/2007. Time calibration Method was talked at the previous meeting. The problems which was shown before were because I used.
Time and amplitude calibration of the Baikal-GVD neutrino telescope Vladimir Aynutdinov, Bair Shaybonov for Baikal collaboration S Vladimir Aynutdinov,
Liquid Xe detector for m +  e + g search Kenji Ozone ( ICEPP, Univ. of Tokyo, Japan ) Introduction prototype R&D ー PMTs ー small & large type summary Outline.
Current status of XMASS experiment 11 th International Workshop on Low Temperature Detectors (LTD-11) Takeda Hall, University of Tokyo, JAPAN 8/1, 2005.
Recent Results from RENO NUFACT2014 August. 25 to 30, 2014, Glasgow, Scotland, U.K. Hyunkwan Seo on behalf of the RENO Collaboration Seoul National University.
R&D works on Liquid Xenon Photon Detector for μ  e γ experiment at PSI Satoshi Mihara ICEPP, Univ. of Tokyo Outline Introduction Prototype R&D works Summary.
A. Baldini PSI July 05 Overview of the experiment MEG is being built (Beam line, Magnet, LXe, DC, TC, Elect., Software) Delays in some items: O(months)
PMT time offset calibration (not completed) R.Sawada 27/Dec/2007.
CsI Veto Detector Performance Study He Dao DMRC. Tsinghua University For KIMS Collaboration.
Upgrade of the MEG liquid xenon calorimeter with VUV-light sensitive large area SiPMs Kei Ieki for the MEG-II collaboration 1 II.
Development of UV-sensitive MPPC for upgrade of liquid xenon detector in MEG experiment Daisuke Kaneko, on behalf of the MEG Collaboration µ γ Liquid xenon.
1 Report on analysis of PoGO Beam Test at Spring-8 Tsunefumi Mizuno July 15, 2003 July 21, 2003 revised.
1 PMT Univ. of Tokyo Yasuko HISAMATSU ICEPP, The University of Tokyo MEG VRVS meeting Jan. 20th, 2004.
Performance of 1600-pixel MPPC for the GLD Calorimeter Readout Jan. 30(Tue.) Korea-Japan Joint Shinshu Univ. Takashi Maeda ( Univ. of Tsukuba)
 13 Readout Electronics A First Look 28-Jan-2004.
g beam test of the Liquid Xe calorimeter for the MEG experiment
Liquid Xenon Detector for the MEG Experiment
A First Look J. Pilcher 12-Mar-2004
MEG Experiment at PSI R&D of Liquid Xenon Photon Detector
Upgrade of LXe gamma-ray detector in MEG experiment
Upgrade of LXe gamma-ray detector in MEG experiment
Upgrade of LXe gamma-ray detector in MEG experiment
Upgrade of LXe gamma-ray detector in MEG experiment
Upgrade of LXe gamma-ray detector in MEG experiment
MEG実験の液体Xe検出器について 東大 ICEPP  森研究室 M1 金子大輔.
B. Genolini for the IPNO detector dept.
Collaboration between Hall C and Hall D
MEG Summary T. Mori for MEG Collaboration February 9, 2005.
New Results from the MEG Experiment
Presentation transcript:

1 PMT Univ. of Tokyo Yasuko HISAMATSU ICEPP, MEG Collaboration meeting Feb. 10th, 2004

2 PMT Output Deterioration  During the pi0 beam output from some PMTs reduced by 20-30% when beam was ON. (The background was estimated to be ~ 1.5*10^6p.e. per sec)  Similar change was observed under the background simulated using LED pulses -> The Deterioration of old type PMT output due to the high rate background.  Output from Al Strip Type PMT (new type) needs to be tested under the high rate background, 10^6 ~ 10^8 p.e. per sec. Yasuko HISAMATSUMEG Collaboration PSI February 2004

3 PMT Univ. of Tokyo Set up PMT (Al Strip Type) TB0094 TB0102 TB0239 TB0268 TB0284 TB0302 TB0415 TB0439 TB0568 PMT Alpha source( 241 Am ) LED Liq. Xe 5.5MeV alpha peak ADC channel Yasuko HISAMATSUMEG Collaboration February 2004

4 PMT Test of Tokyo Xe tank Purification system Liq.Xe chamber alpha source LED PMT Chamber Inside

5 Condition and Procedure alpha source : ~200Hz, LEDpulse height:500p.e. ~ 12000p.e. per event pulse shape: ~10nsec rate: 100Hz ~ 10KHz Trigger: alpha self trigger (veto by LED driver pulse) Procedure Pedestal Run & Gain calibration using LED Alpha LED OFF Alpha LED ON (LED : high rate background) -Change LED Pulse height, rate and PMT gain Yasuko HISAMATSUMEG Collaboration February 2004

6 Condition and Procedure  DAQ: -Alpha run with LED OFF -Alpha run with LED ON  The cause of the change in PMT output  Breeder current ?  Deterioration of photocathode? Low gain run Peak/Sigma (gain independent) The change in PMT output? Yasuko HISAMATSUMEG Collaboration February 2004

7 Thermal neutron background estimation Condition of the estimation  16 n/cm^2/sec (MEG TN022)  From all direction 6*10^5 photons ( : ~9MeV gamma ) are generated in LP MC 5.2 * 10^6 p.e. Scaled to the final detector (surface area) 3.2 * 10^7 detector = 5.0 *10^-6 PMTs were tested under the background of 10^6 ~ 10^8 p.e./sec

8 For instance…TB0415 5*e^5 gain peak (LED ON)/peak (LED OFF) 1*e^6 gain peak (LED ON)/peak (LED OFF) 1*e^6 gain ON)/ OFF) 10^7p.e. per sec

9 Result 5*e5gain X: number of photoelectrons per second from LED Y: alpha peak ON) / alpha peak OFF) Yasuko HISAMATSUMEG Collaboration February 2004 TB0094TB0302 TB0268 TB0568TB0102 TB0415 TB0284 TB0239 TB0439

10 Result 1*e6gain X: number of photoelectrons per second from LED Y: alpha peak ON) / alpha peak OFF) Yasuko HISAMATSUMEG Collaboration February 2004 TB0094TB0302 TB0268 TB0568TB0102 TB0415 TB0284TB0439 TB0239

11 X: number of photoelectrons per second from LED Y: ON) / OFF) Result 5*e5gain Yasuko HISAMATSUMEG Collaboration February 2004 TB0094TB0302 TB0268 TB0568TB0102 TB0415 TB0284TB0439 TB0239

12 Result 1*e6gain X: number of photoelectrons per second from LED Y: ON) / OFF) TB0094TB0302 TB0268 TB0568 TB0102 TB0415 TB0284 TB0439 TB0239 Yasuko HISAMATSUMEG Collaboration February 2004

13 PMT Output vs Elapsed Time X: event number Y: ADC count TB0239 5*10^5gain Background: 8.32*10^7pe/sec 1*10^6gain Background 1.33*10^8pe/sec ~10sec~20sec  DAQ started 2 ~ 10 sec after the LED had been turned on.  Stable output up to 5*10^7 photoelectrons/sec Yasuko HISAMATSUMEG Collaboration February 2004

14 Summary  9 Al Strip Type PMTs (new type) have been tested under the high rate background environment at the university of Tokyo.  So far, for the 9 PMTs, the stability of PMT output is better than 10% under the background up to 10^7 p.e. per sec.  For the 8 PMTs out of 9 PMTs, no output deterioration under background of ~1.5*10^6p.e. per sec.  For a PMT, TB0239, the output deterioration due to the breeder circuit was found under the background of ~1.9 * 10 ^6p.e. per sec.  The background number of photoelectrons should be suppressed to 10*6 p.e./sec.  Necessary to suppress thermal neutrons to 1/10.  Neutron background using LP is recommended. Yasuko HISAMATSUMEG Collaboration February 2004