How to ensure Safety? Calibration vs Functional (Bump)Test Differentiate between calibration and functional (bump) test functional test –the sensors.

Slides:



Advertisements
Similar presentations
Gas Detection Instruments
Advertisements

Monitoring Equipment and Detection Devices Chapter 14.
Chapter 2 The Process of Experimentation
SECTION 11 LEVEL B SUIT OUT. SECTION 12 AIR MONITORING.
Sensor Operating Theory
Dynamic Performance Class 4.
Effectiveness Evaluation for Production Drugs Crystal Groesbeck, Ph.D Division of Production Drugs.
Hazmat Industry Technician. Types of Equipment Monitoring equipment Detection devices Air sampling systems.
Chapter 2 Analyzing the Business Case.
Chapter 29 Product Control and Air Monitoring. Introduction Product control techniques can provide quick reduction in damage Reduction of surface area.
Basic Air Monitoring Concepts And Use Of The Altair 4 Gas Meter
Presented on Behalf of The: Barbados Association of Professional Engineers by: Eng.: Dev Maharaj.
Air Monitoring Presented By: Etech Environmental & Safety Solutions, Inc.
Gas Sensor and Transmitter
PERMIT REQUIRED CONFINED SPACES PERMIT REQUIRED CONFINED SPACES.
Confined Space Ventilation Are We Really Moving Air, or Just Making Ourselves Feel Good?
ITX 4-Gas Detector Quick Reference Guide. Case:Type 304 Stainless Steel,.024" (.61mm) thick Dimensions:4.75" X 3.2" X 1.68" (121mm X 81mm X 43mm) Weight:18.5.
Altair® 5X MultiGas Detector Altair® 5X MultiGas Detector The ALTAIR 5X MultiGas Detector is to be used when performing a hazard assessment.
Seafood HACCP Alliance for Training and Education Chapter 10 Principle 6: Establish Verification Procedures.
Static Pressure Control Loop The purpose of the static pressure control loop is to maintain an optimal static pressure in the ductwork. The control loop.
WELCOME TO Gas Tester Training O2, LEL Monitoring
Florida Operations Level Hazardous Materials Training Unit 5.5 MISSION SPECIFIC: Air Monitoring and Sampling.
Reducing risk with Exposure Monitoring Equipment.
Monitoring and Evaluation in MCH Programs and Projects MCH in Developing Countries Feb 10, 2011.
H2S Gas Meter/Detector. Introduction Air monitoring data is useful for: *Assessing the health risks to the public and response workers. *Selecting personal.
Measurements Measurements and errors : - Here, the goal is to have some understanding of the operation and behavior of electrical test instruments. Also,
Gas sensing Panca Mudji Rahardjo, ST.MT Electrical Engineering - UB.
Balances The balance is an instrument which measures the mass of a body or substance using the gravity force which acts on that body At present, there.
? Training removes doubt, instills confidence, and lays the foundation for everyone’s skill and experience level.
Chemical Risk Assessment & Exposure Monitoring Quantitative Risk Assessment Revision December Information provided subject to the 'Conditions for.
WORKSHOP ON ACCREDITATION OF BODIES CERTIFYING MEDICAL DEVICES INT MARKET TOPIC 6 CH 5 ISO MANAGEMENT RESPONSIBILITY Philippe Bauwin Medical.
Bi-Weekly Safety Meeting November 29, The days of using canaries in mines as primitive gas detectors are long gone..they've been replaced by fixed.
1 Emergency and Disaster Response to Chemical Releases Monitoring Module 4.
Misuse of Combustible Gas Meters Supervisor Training 1/11/08.
National Highway Institute 5-1 REV-2, JAN 2006 EQUIPMENT FACTORS AFFECTING INERTIAL PROFILER MEASUREMENTS BLOCK 5.
Overview of Instrument Calibration Presents by NCQC, India.
Excavation & Trenching Awareness Plus
Atmospheric Monitoring. ATMOSPHERIC MONITORING INSTRUCTIONAL GOAL The participant will understand the need for atmospheric monitoring and the types, uses,
E VERY L IFE H AS A P URPOSE… Instrument Selection.
Rev. 1 © Chevron 2009 Hot Work Awareness Plus. Rev. 1 © Chevron Training Objectives  To explain the type of work which may constitute Hot Work.
Bi-Weekly Safety Meeting November 29, The days of using canaries in mines as primitive gas detectors are long gone…they've been replaced by fixed.
QUALITY ASSURANCE MODULE 3. Definitions Quality Assurance - A process/effort that ensures that processes are followed, that the processes of doing right.
AIA and Exposure Data Collection By: Jaco van Rensburg
Chapter 10 Verification Procedures.
SKRM Launching New Technology In Cupric Chloride Regeneration
Jim Vermeulen MSA Territory Manager
Confined Space (Permit-required) Rescue/ Emergency Services
Gas Detection in Water Treatment
Battery Basics for The Radio Operator J&K Communications, Inc.
Integrated Consulting Solutions
OH 3-1 Receiving and Storing to Maintain Quality Food Production 3 OH 3-1.
DIGITAL BREAK THROUGH IN WEIGHING
Instrument Maintenance
(Additional materials)
Lesson 8: Epic Appointment Scheduling Advanced Appointments
Electronic Control Systems Week 4 – Signaling and Calibration
Chapter 10 Verification Procedures.
Confined Space (Permit-required)
Deyuan Kong and Sara Mcmillen, Chevron Energy Technology Company USA
Ventilation Measurement Tracer Gas Decay Testing
Calibration and Validation
Measurements Measurements and errors :
This presentation document has been prepared by Vault Intelligence Limited (“Vault") and is intended for off line demonstration, presentation and educational.
How to conduct Effective Stage-1 Audit
Quality Control Lecture 3
H2S Monitors.
Biological Science Applications in Agriculture
Manure Gas Safety Monitor Types
Everyone is talking about
Radiation Detectors for the PSS1 at ESS
Presentation transcript:

How to ensure Safety?

Calibration vs Functional (Bump)Test Differentiate between calibration and functional (bump) test functional test –the sensors are simply exposed to known concentration test gas. –The readings and performance of the alarms are verified, but not adjusted. –As long as the functional test indicates readings are accurate, there is no obligation to adjust the sensitivity of the sensors. Calibration –the readings of the sensors are actually adjusted (if necessary) to match the concentrations in the calibration gas.

FUNCTIONAL (BUMP) TEST Many companies take the approach of doing a daily functional test. –This is a great approach IF it is logistically feasible. Other companies with a long experience of using confined space gas detectors (like New England Grid) find that they are safely able to extend the period between calibration checks. Some perform functional (bump) tests on a monthly basis. Other companies do a full calibration on instruments only once every 180 days. BW personnel are not the ones using the instruments in the field –not in a position to dictate the frequency of testing.

CALIBRATION We base our recommendations on the very conservative International Safety Equipment Association (ISEA) protocol for lengthening the time between testing the calibration of sensors in instruments used to monitor confined space atmosphere. –The ISEA protocol suggests performing a functional test at least once every 30 days. –There is no set interval for performing a calibration –Customers are free to calibrate their instruments once every 180 days, or choose an even longer period between scheduled calibrations IF they are able to verify through experience that this is a safe and prudent interval. It's up to the instrument user to verify that the interval specified is safe and prudent. If there is any doubt, a shorter interval between calibration checks should be specified.

RE-VERIFICATION Any conditions, incidents, experiences, or exposure to contaminants that might adversely affect the calibration –should trigger re-verification of calibration by performing a functional (bump) test before further use. –If there is any doubt about the calibration of the sensors, they should be exposed to known concentration test gas before further use.

BUMP TEST FOR CL2 AND HCL Sensor Typical Response Time –Catalytic sensor T90 <= 30 seconds (methane) T90 <= 10 seconds (hydrogen) –Infra-red sensor T90 <= 10 seconds –Electrochemical sensors CO/ H2S –T90 <= 30 seconds Cl2/ HCl –T50 <= 30 seconds –T90 <= 70 seconds –CL2/ HCL sensor Bump gas and wait for 4 minutes for sensor to stabilise Bump gas and check for sensor response at 50% response after 30 seconds

OXYGEN SENSOR Oxygen Sensor has a finite life the length of their operating life will be determined by the total exposure to Oxygen. This is because the sensor has a metal anode which is consumed by the reaction mechanism. The characteristic of Oxygen Sensor is that they have a stable signal output, which experiences minimal decline, right up until the end of the sensors life. At the end of the sensor's operating life there is a sudden and rapid drop-off in output as the sensor's anode is finally consumed.

BUMP TEST FOR O2 SENSOR A ’bump’ test is a means of verifying that an instrument is working within acceptable limits by briefly exposing to a known gas mixture formulated to change the output of all the sensors present. This is different from a calibration where the instrument is also exposed to a known gas mixture but is allowed to settle to a steady figure and the reading adjusted to the stated gas concentration of the test mixture. For oxygen monitors a level of confidence that the unit is working adequately may be gained by exhaling over the sensor inlet and viewing the reduction in reading obtained. In many cases the reduction in Oxygen concentration obtained is sufficient to trigger the low oxygen alarm.

Questions