Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Herschel Calibration Workshop PACS Photometer Flux Calibration: Update Markus Nielbock (MPIA.

Slides:



Advertisements
Similar presentations
The PACS Status Herschel PV Mid-Term Review ESAC, September 1, Herschel Performance Verification Mid-Term Review PACS Status: Overview Markus Nielbock.
Advertisements

Stellar Linearity Test Jason Surace (Spitzer Science Center)
350  m Observations of Asteroids, Secondary Calibrators, and Stellar Standard(s) C. Darren Dowell Jet Propulsion Laboratory, California Institute of Technology,USA.
Calibration Plan During BASIC Science Göran Sandell Calibration Scientist.
Bruno Altieri September # 1 PACS photometer pipeline overview Bruno Altieri (ESA/HSC)
PACS SOVT-2 (SIMS) and CAPs development MPIA PACS ICC Meeting #32 1 st – 3 rd April 2009, MPE Garching Markus Nielbock (MPIA)
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th - 10 th Oct 2014 SPIRE Photometer Data Reprocessing Bernhard Schulz (NHSC/IPAC) on.
ASI Review Padova ICC Contribution 1 WP3000 : Padova ICC Contribution Staff Alberto Franceschini (AF) [Professore Ordinario] Giulia Rodighiero (GR) [Ricercatore.
PACS NHSC SPIRE Point Source Spectroscopy Webinar 21 March 2012 David Shupe, Bernhard Schulz, Kevin Xu on behalf of the SPIRE ICC Extracting Photometry.
PACS Page 1 NHSC Data Processing Workshop – Pasadena Sept 10-14, 2012 SPIRE Spectrometer Data Reduction: Mapping Observations Nanyao Lu NHSC/IPAC (On behalf.
PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching.
PACS NHSC Data Processing Workshop – Pasadena 26 th - 30 th Aug 2013 Photometer Extended Source Photometry Bernhard Schulz NHSC/IPAC on behalf of the SPIRE.
FM-ILT Results: Mechanisms FM1 Chopper and Calibration Sources Markus Nielbock (MPIA) Babar Ali (IPAC) Jeroen Bouwman (MPIA) Helmut Dannerbauer (MPIA)
1 NHSC PACS NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-202 (for Hipe 9.0) Level 0 to Level 2 processing: From raw data to calibrated.
Page 1 PACS NHSC Data Processing Workshop – Pasadena 26 th - 30 th Aug 2013 Overview of SPIRE Photometer Pipeline Kevin Xu NHSC/IPAC on behalf of the SPIRE.
PACS SOVT-2 and CAPs development MPIA PACS ICC Meeting #31 12 th – 13 th January 2009, Vienna (A) Markus Nielbock (MPIA)
Consortium Meeting La Palma October PV-Phase & Calibration Plans Sarah Leeks 1 SPIRE Consortium Meeting La Palma, Oct. 1 – PV Phase and.
1 NHSC PACS NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-202 (for Hipe 8.0) Level 0 to Level 2 processing: From raw data to calibrated.
PACS Spectrometer Spatial Calibration plan in PV phase A.Contursi D. Lutz and U. Klaas.
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th - 10 th Oct 2014 Roberta Paladini NHSC/IPAC Photometry Guidelines for PACS data.
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th - 10 th Oct 2014 The PACS Photometer: Overview and Products Roberta Paladini NHSC/IPAC.
1 NHSC PACS NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-402 (for Hipe 12.0) Level 1 to Level 2 processing: The JScanam pipeline Prepared.
PACS NHSC Data Processing Workshop – Pasadena 10 th - 14 th Sep 2012 Measuring Photometry from SPIRE Observations Presenter: David Shupe (NHSC/IPAC) on.
PACS NHSC Data Processing Workshop – Pasadena 10 th - 14 th Sep 2012 The SPIRE Destriper Bernhard Schulz NHSC/IPAC on behalf of the SPIRE ICC 1.
Page 1 PACS NHSC Webinar: New SPIRE Features in HIPE 10 6 th March 2013, Pasadena What’s New in SPIRE Photometer Pipeline in HIPE 10 Kevin Xu NHSC/IPAC.
SNAP Calibration Program Steps to Spectrophotometric Calibration The SNAP (Supernova / Acceleration Probe) mission’s primary science.
PACS Page 1 NHSC Workshop on HSA Data Oct 6-10, 2014 SPIRE Spectrometer Data Reduction: Special Cases Nanyao Lu NHSC/IPAC (On behalf of the SPIRE ICC,
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th - 10 th Oct 2014 SPIRE Broad-Band Photometry Extraction Bernhard Schulz (NHSC/IPAC)
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th - 10 th Oct 2014 The PACS Spectrometer: Overview and Products Roberta Paladini NHSC/IPAC.
Page 1 PACS NHSC Webinar: New SPIRE Features in HIPE th Nov 2012, Pasadena Update on SPIRE Photometer Pipeline Kevin Xu NHSC/IPAC on behalf of the.
PVPhotFlux PACS Photometer photometric calibration MPIA PACS Commissioning and PV Phase Plan Review 21 st – 22 nd January 2009, MPE Garching Markus Nielbock.
- page 1 NHSC – DP workshop – Feb – N. Billot PACS Deglitching PACS Photometer Data Presentation based on tutorial PACS-402.
Jan 21/22, 2009PACS PV Review Photometer Spatial Cal1 PACS Photometer Spatial Calibration PVPhotSpatial D.Lutz PACS PV Review Jan 21/ , MPE.
PACS ICC Readiness Review MPE, July 3/ PACS Photometer PV Phase Plan 1 Status Report M. Nielbock: PACS PHOT PV Phase Plan Markus Nielbock (MPIA)
CEA DSM Dapnia SAp Flux calibration of the Photometer Koryo Okumura, Marc Sauvage, Nicolas Billot, Bertrand Morin DSM/DAPNIA/Sap.
Overview, Spectrometer Products and Processing Philosophy Phil Appleton on Behalf of PACS Team PACS IFU Spectrometer.
Data Processing Workshop NHSC, Pasadena, CA 02 – 04 February 2011 Stephan Ott VG #1 PACS Overview and new developments in.
 SPIRE/PACS guaranteed time programme.  Parallel Mode Observations at 100, 160, 250, 350 and 500µm simultaneously.  Each.
NHSC PACS PACS Photometry and Errors Estimate Roberta Paladini (NSHC/Caltech) R. Paladini NHSC Archive Data Processing Workshop – August 26 th – 30 th.
PACS ICC Meeting #41 ESAC, 6/7 November 2012POP Calibration GoalsE. Sturm Post-Operations Calibration Goals E. Sturm.
PACS NHSC Data Processing Workshop Aug 26-30, 2013 Page 1 SPIRE Spectrometer Data: Calibration Updates, User Data Reprocessing, and Other Issues Nanyao.
Page 1 PACS NHSC Data Processing Workshop – Pasadena 26 th - 30 th Aug 2013 Issues with Photometer Data & How to Resolve them with HIPE Tools Kevin Xu.
PACS Page 1 NHSC Data Processing Workshop – Pasadena Aug 26-30, 2013 SPIRE Spectrometer Data Reduction: Mapping Observations Nanyao Lu NHSC/IPAC (On behalf.
1 NHSC PACS NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-402 (for Hipe 13.0) Level 1 to Level 2.5 processing: The JScanam pipeline Prepared.
PACS page 1 NHSC SPIRE Data Processing Webinars 8 th Feb 2012 PACS page 1 Overview of SPIRE Photometer Pipeline C. Kevin Xu (NHSC/IPAC)
PACS page 1 NHSC SPIRE Data Processing Webinars 7 th March 2012 PACS page 1 SPIRE Photometer Map Making C. Kevin Xu (NHSC/IPAC)
New SPIRE features in HIPE 9.1 NHSC; Nov 28, 2012 PACS Page 1 What’s New in HIPE 9.1 ( SPIRE FTS) Nanyao Lu NHSC/IPAC (on behalf of the SPIRE ICC)
PACS Page 1 NHSC Workshop on HSA Data Oct 6-10, 2014 SPIRE Spectrometer: Pipeline Calibration Nanyao Lu NHSC/IPAC (on behalf of the SPIRE ICC, HSC & NHSC)
- page 1 PACS Phil Appleton on behalf of the NHSC/HSC and the PACS ICC ; especially Bart Vandenbussche and Pierre Royer (KUL Belgium) Instrument Performance.
A. Ealet Berkeley, december Spectrograph calibration Determination of specifications Calibration strategy Note in
1 NHSC PACS NHSC/PACS Web Tutorials Running PACS photometer pipelines PACS-403 (for Hipe 13.0) Level 1 to Level 2.5 processing: The Unimap pipeline Prepared.
Page 1 PACS NHSC Data Processing Workshop – Pasadena 10 th - 14 th Sep 2012 Reprocessing Scan Maps Kevin Xu NHSC/IPAC on behalf of the SPIRE ICC.
Markus Nielbock (MPIA) – Herschel Pointing PACS ICC Meeting #38 Herschel Pointing Summary and Recent Developments Markus Nielbock (MPIA Heidelberg) with.
PACS Hitchhiker’s Guide to Herschel Archive Workshop – Pasadena 6 th – 10 th Oct 2014 Overview of SPIRE Photometer Data Reduction Pipeline Kevin Xu NHSC/IPAC.
Markus Nielbock (MPIA) – PACS Calibration Status Report Herschel Calibration Steering Group Meeting #24 PACS Calibration Status Report Markus Nielbock.
Markus Nielbock (MPIA) – PACS Photometer Flux Calibration EPOS Meeting PACS Photometer Flux Calibration Current Plans Markus Nielbock MPIA, 22 th February.
Markus Nielbock (MPIA) – Herschel/PACS DP Workshop Herschel DP Workshop PACS Photometer Flux Calibration and Point Source Photometry Markus Nielbock (MPIA.
Markus Nielbock (MPIA) – PACS Point Source AOT Pipeline Herschel Science Demonstration Phase Data Processing Workshop PACS Point Source AOT Pipeline Tutorial.
HCalSG meeting #37 Calibration updates: Asteroids T. G. Müller (MPE Garching) Publication: Far-infrared photometric observations of the outer planets and.
SPIRE Flux Calibration: Implementation
Herschel Calibration Workshop
Validation status overivew
photometry and spectroscopy with PACS
Validation status overivew
PACS Calibration Status Report
PACS / SPIRE cross-calibration on prime fidicial standard stars
PACS ICC Meeting #41 PACS Photometer
Markus Nielbock (MPIA Heidelberg)
Overview and new developments in Herschel Data Processing
What’s New in HIPE 10.0 (SPIRE FTS)
Presentation transcript:

Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Herschel Calibration Workshop PACS Photometer Flux Calibration: Update Markus Nielbock (MPIA Heidelberg) Thomas Müller (MPE Garching) on behalf of the PACS ICC and Calibration Working Group ESAC, 19 th January 2012

PACS Photometer Flux Calibration Overview Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● current status ● non-linearity correction ● influence on PSF templates and aperture correction ● roadmap for calibration product updates ● correlation with evaporator temperature

PACS Photometer Flux Calibration Current status: General remarks Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● flux calibration based on scan map observations of 5 prime fiducial stars ● model uncertainty amounts to 5% throughout ● chop-nod observations are underestimated by 4%, 4%, 6% at 70, 100, 160 µm ● aperture photometry with aperture correction (PSF template: Vesta, Mars) ● colour corrections: 1.016, 1.033, at 70, 100, 160 µm ● all prime standards are in the linear flux regime ● secondary targets (faint stars, asteroids, planets) ● observed for consistency and extension of flux range

PACS Photometer Flux Calibration Current status: 70 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Current status: 100 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Current status: 160 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration improved data processing

PACS Photometer Flux Calibration Current status: Results Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● flux calibration based on scan map observations of 5 prime fiducial stars ● scan-map: 160 µm fluxes might be underestimated by 2% ● chop-nod observations are underestimated by 4%, 4%, 6% at 70, 100, 160 µm ● fluxes of prime calibrators in PACS-P scan-map observations consistent with models within: ● 3%, 3%, 5% at 70, 100, 160 µm (weighted by calibrator, not by individual observation) ● extracted fluxes are influenced at a level of a few percent by: ● cross-talk, highpass filtering, source masking, deglitching, etc.

PACS Photometer Flux Calibration Non-linearity correction - Implementation Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● established by N. Billot (CEA/IPAC/IRAM) based on pre-flight ground calibration ● non-linearity corrections of the order of 10% for signals around 60 Jy/pixel ● → small effect (0-15%) for brightest calibration sources (> 100 Jy per point source) ● flux calibrators affected: ● planets: Uranus, Neptune (5-15%) ● bright asteroids: Ceres, Pallas, Vesta (0-5%) ● independent of gain setting ● is implemented in HIPE 8

PACS Photometer Flux Calibration Non-linearity correction: Verification 70 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 70 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 70 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 100 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 100 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 100 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 160 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 160 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Verification 160 µm Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e

PACS Photometer Flux Calibration Non-linearity correction: Results Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● correction brings Neptune and Uranus fluxes into the ± 5% corridor ● data processing step and corresponding task is validated and available as of HIPE 8 ● should be used for all point sources above 10 Jy and bright extended regions ● noticable effects occur above 100 Jy ● Neptune models: ● truth might be between esa2 and esa3 at 70, 100 µm and close to esa3 at 160 µm ● but wait for end of this talk … ● open issues: ● PSF models are based on Vesta observations from OD 160 without NL-correction ● → 70 µm PSF model affected, aperture correction compromised OD 160 Vesta observations at 100, 160 µm are okay

PACS Photometer Flux Calibration Non-linearity correction: Influence on model PSFs and EEF Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● deviations found: ● ~ 10% in peak flux, ~ 5% in total flux ● flux calibration at 70 µm affected ● aperture photometry of point sources is self-consistent and not affected ● (responsivity ↔ aperture correction) ● extended emission measured too high at 70 µm

PACS Photometer Flux Calibration Roadmap for flux calibration and aperture correction update Markus Nielbock (MPIA) – PACS Photometer Flux Calibration 1.re-analysis of PSF measurements with NL correction applied (Vesta, Mars) 2.evaluation of new EEF curves 3.generation of updated aperture correction calibration table (EEF values) 4.re-processing of all scan-map observations including NL correction using the new EEF 5.values for aperture correction 6.generation of updated response calibration product FM7 based on scan-map 7.observations of the five prime standards (will also correct 2% shift at 160 µm) 8.update of PhotApertureCorrectionPointSourceTask to accept new response calibration 9.validation of FM7 response calibration product on all calibrators (stars, asteroids, planets) 10.update of relevant documentation and processing scripts 11.release of new PSF models, EEF table, response calibration FM7 to be expected for HIPE 9

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● PACS 3 He cooler temperature rises towards end of cooling cycle ● affects detector responsivity (this is expected) ● flux in individual observations is modified by up to ~5% ● introduces a systematic component to calibration uncertainty ● relevant for variability programmes → avoid scheduling towards the end of cooling cycle ● new scheduling scheme prevents observations at high evaporator temperatures detector debiasing -1 hour

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration Neptun e Uranu s

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration

PACS Photometer Flux Calibration Secondary effects: Correlation with evaporator temperature Markus Nielbock (MPIA) – PACS Photometer Flux Calibration esa3 model seems to be the best choice for all bands

PACS Photometer Flux Calibration Summary Markus Nielbock (MPIA) – PACS Photometer Flux Calibration ● flux calibration is in very good shape ● scan-map: 160 µm fluxes might be underestimated by 2% ● chop-nod observations are underestimated by 4%, 4%, 6% at 70, 100, 160 µm ● prime calibrator model uncertainty: 5% ● fluxes of prime calibrators in PACS-P scan-map observations consistent with models within: ● 3%, 3%, 5% at 70, 100, 160 µm (weighted by calibrator, not by individual observation) ● non-linearity correction module is verified and in place ● non-linearity significantly affects point sources above 100 Jy ● (e.g. bright asteroids, planets, but also bright extended emission) ● PSF model at 70 µm based on uncorrected observations of Vesta (OD 160) ● roadmap for PSF model, responsivity and EEF/aperture correction table update in place ● → HIPE 9 ● reponsivity affected by evaporator temperature of 3 He cooler (~ 5% integrated flux) ● after eliminating those effects: Neptune model esa3 seems most suitable for PACS-P