Patient Synchrony in bellavista Auto-Rise Pressure Rise Auto-Synch Expiration Trigger 1.

Slides:



Advertisements
Similar presentations
Neonatal Mechanical Ventilation
Advertisements

Patient – Ventilator Asynchrony
CPAP/PSV.
Bilevel Titrations: Who, What, Why, and When Gary Hamilton, BS, RRT
Trigger/Limit/Cycle/Baseline
Pressure support ventilation Dr Vincent Ioos Pulmonologist and Intensivist Medical ICU, PIMS 1st International Conference Pulmonology and Critical Care.
Interpretation of Ventilator Graphics
Ventilator Graphics: Not Just Pretty Lines
Initiation of Mechanical Ventilation
Nesreen El-Sayed Morsy Aly Thoracic Medicine Department
Patrick Gleason MS4 University of South Carolina School of Medicine 1.
SERVO-i VENTILATOR GRAPHICS
How a Breath is Delivered
By: Susan P. Pilbeam, MS, RRT, FAARC John D. Hiser, MEd, RRT, FAARC
Newer Modes of Mechanical Ventilation Timothy A. Morris, M.D. Division of Pulmonary and Critical Care Medicine Medical Director, Respiratory Care University.
Effect of different cycling off criteria and positive end-expiratory pressure during pressure support ventilation in patients with chronic obstructive.
Critical Care Ventilation Technology Perspective Fran Hegarty.
Ventilator Graphics Chapter 10. Graphics Monitor the function of the ventilator Evaluate the patient’s response to the ventilator Help the clinician adjust.
Basic Concepts in Adult Mechanical Ventilation
BASICS OF WAVEFORM INTERPRETATION Michael Haines, MPH, RRT-NPS, AE-C
Ventilator Graphics Emeritus Professor Georgia State University
Mechanical Ventilation 1
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
3 nd LECTURE VENTILATORS Part One. Ventilators One of the major life support systems. Ventilators take over the vital role of the respiratory muscles.
Mechanical Ventilation Graphical Assessment
Are You Optimizing Every Bilevel Breath? Jim Eddins, RRT.
 Understand the dual control concept  Understand the pressure regulation mechanism in PRVC  Demonstration of PRVC  Settings and adjustment with Servo.
Ventilatory Modes. Volume Controlled Mandatory Breath Gas is delivered at a constant flow until the set tidal volume is reached. Pressure rises to a.
Ventilatory Modes Graphnet Ventilator.
PRESSURE CONTROL VENTILATION
Ventilators for Interns
Mechanical Ventilation
CLINICAL UTILITY OF VENTILATOR GRAPHICS
Principles of Mechanical Ventilation Magdy M Khalil, MD, EDIC Prof. Pulmonary& Critical Care Medicine.
Pressure Regulated Volume Control Mode
Mechanical Ventilation Basic Modes
Mechanical Ventilation - Introduction
Introduction to Basic Waveforms
Ventilator waveform analysis: often ignored bedside assessment
Ventilation Sam Petty Clinical Specialist Physiotherapist
Mechanical Ventilation in newborns
Mechanical Ventilator 2
Mechanical Ventilator 1
Mechanical ventilator
Tips for putting a patient on a F-60
Criteria used to determine breath types during mechanical ventilation
Advanced Modes of Mechanical Ventilation
Basic Concepts in Adult Mechanical Ventilation
FLIGHT MEDICAL B-Lev Mode Biphasic Ventilation Confidential.
Patient-ventilator interaction: insight from waveforms
Respiratory System Elastance Monitoring during PEEP Titration
You could ventilate a patient
Mechanical ventilator
Airway pressure and flow waveforms during constant flow volume control ventilation, illustrating the effect of an end-inspiratory breath-hold. Airway pressure.
Adjusting comfort settings on Newport HT70 Series Ventilators
Example ventilator screen during nasal neurally adjusted ventilatory assist in a premature neonate (23 weeks gestational age, 560 g) with respiratory distress.
Patient–ventilator interaction
Shape-signal method of triggering combines shape signal (A) and volume (B) methods of triggering. Shape-signal method of triggering combines shape signal.
Even though this patient is undergoing positive-pressure mechanical ventilation, the first 4 breaths have a relatively negative pressure (ie, pressure.
The third breath has a negative deflection (ie, below PEEP) at the end of the mechanical breath (arrow A) associated with a flow increase (arrow B), indicating.
Asynchrony index at baseline and following optimization of pressure support (PS) level (A), and following optimization of mechanical inspiratory time (mechanical.
Characteristics of a pressure-supported breath.
The peak flows (60 L/min) and flow patterns are the same for all the breaths. The peak flows (60 L/min) and flow patterns are the same for all the breaths.
During this tracing of 30 seconds, the ventilator displays that the patient rate is 16 breaths/min. During this tracing of 30 seconds, the ventilator displays.
Capire e interpretare le curve sul ventilatore
This tracing depicts 30 seconds of information.
Effect of respiratory mechanics on cycling of pressure support from inhalation to exhalation. Effect of respiratory mechanics on cycling of pressure support.
Control circuit for an adaptive pressure targeting scheme (eg, Pressure Regulated Volume Control). Control circuit for an adaptive pressure targeting scheme.
Effects of increasing the cycling-off threshold according to prolonged (A) or short (B) time constant of the respiratory system. Effects of increasing.
Airway pressure and flow graphics illustrate delayed cycling.
Presentation transcript:

Patient Synchrony in bellavista Auto-Rise Pressure Rise Auto-Synch Expiration Trigger 1

Synchrony = Free Breathing 2 Mistakes can be made in the settings which influence free breathing. Trigger Expiratory Trigger Rise Time Auto-RiseAuto-Synch

Adaptive Rise Time, Auto-Rise From first breath  to fully adapted  in a few breaths. The rise time is reduced; as a result peak flow is optimized. Auto rise makes the rise time as fast as possible to optimize peak flow and ventilator response while at the same time minimizing overshoot

Auto-Rise Auto-Synch™ Auto-Rise combined with the Auto-Synch exhalation trigger is designed to facilitate and optimize invasive and non-invasive ventilation performance. Benefits: Simplified ventilation setup Optimal rise time at all times No false overpressure alarms Additional safety for neonates

Interactivity and Synchrony Robert Brink imtmedical 2013 Optimal Slope rise & Expiratory Threshold Adjustment Pressure Support cycles to exhalation when patient stops inhaling. Ideal expiratory threshold setting =

The Challenges Expiratory Threshold Exp Trigger The expiratory threshold setting determines the flowrate at which a Pressure Support breath will cycle-off. Slope Rise Slope rise setting impacts cycling off. Adjustment of slope rise may require re-adjustment of expiratory threshold. Interactivity and Synchrony

Time Flow Expiratory Threshold Expiratory Trigger Adjustment It is typically expressed as a percent of peak flow

Interactivity and Synchrony 30% of Peak Flow Time Flow Peak Flow Expiratory Threshold Adjustment And is based on the decay in flow during inspiration.

Interactivity and Synchrony Time Flow Pulmonary Time Constant Expiratory Threshold Adjustment The flow decay is the result of the patient’s pulmonary time constant.

Slope Rise Adjustment and Expiratory Threshold Adjustment Interactivity work together to bring synchrony back to Pressure Support and VTPS

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Electromyography Diaphragm Abdomen If Expiratory Threshold Too Low Optimal Expiratory Threshold Adjustment Challenge

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Electromyography Diaphragm Abdomen The breath may cycle-off too late Optimal Expiratory Threshold Adjustment Challenge

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Electromyography Diaphragm Abdomen This can cause the patient to exhale against continuing support Optimal Expiratory Threshold Adjustment Challenge

Interactivity and Synchrony And also cause the ventilator to miss triggering a breath the next time the patient attempts to breathe Airway Pressure Time (s) Target Pressure PEEP Flow Electromyography Diaphragm Abdomen Missed Trigger Optimal Expiratory Threshold Adjustment Challenge

Interactivity and Synchrony Time Flow Long Time Constant Optimal Expiratory Threshold Adjustment Challenge Obstructed patients’ long time constants cause delivered flow to decay more slowly. Obstructed patient

Interactivity and Synchrony Time Flow Peak Flow patient insp. effort ends (patient is ready to exhale) 50% Peak Flow Optimal Expiratory Threshold Adjustment Challenge Patient inspiratory efforts may end when inspiratory flow has decayed to about 50% of their peak-flow. At that point, the patient is ready to exhale but the ventilator prolongs the breath Obstructed patient

Interactivity and Synchrony Time Flow Peak Flow Obstructed patient Late cycling off Optimal Expiratory Threshold Adjustment Challenge It is common for the breath to cycle off too late on obstructed patients if the Expiratory Threshold setting is lower than 50%.

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Challenge Late cycling off can also happen if slope rise is adjusted to a slower value after expiratory threshold has been set to an appropriate level.

Interactivity and Synchrony Time Flow 28% of Peak Flow Slope Rise and Expiratory Threshold Interactivity Challenge

Interactivity and Synchrony Time Flow 28% of Peak Flow Slope Rise and Expiratory Threshold Interactivity Challenge The slower slope rise will result in a lower flow at the onset of the breath.

Interactivity and Synchrony Time Flow 28% of Peak Flow Fast Slope Slow Slope Slope Rise and Expiratory Threshold Interactivity And that will prolong the time it takes for flow to drop to the % of flow required to cycle the breath off. (Since this is now a lower flow.)

Interactivity and Synchrony Time Flow 28% of Peak Flow Fast Slope Slow Slope Slope Rise and Expiratory Threshold Interactivity The longer the time constant, the more breath cycling off will be prolonged when Slope Rise is adjusted to a slower value.

Interactivity and Synchrony Problems related to late cycling-off (late termination) Late cycle-off Inadequate lung emptying time Auto-PEEP Missed Triggers Delayed cycling-off Airway Pressure Time (s) Flow Electro- myography Diaphragm Abdomen Missed Trigger Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Example: Exp Threshold = 30% Slope Rise setting results in a Peak flow = 100 L/min Cycling off flow = 30 L/min (30% of 100) Breath cycles off when inspiratory flow drops to 30 L/min Slope Rise Setting is lowered and it now results in a Peak flow = 50 L/min Cycling off flow = 15 L/min (30% of 50) Breath will not cycle off until flow drops to 15 L/min

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Challenge Time Flow Short Time Constant Restrictive disorder Early cycling off may happen to patients with restrictive disorders, because they have short time constants and therefore their inspiratory flow decays more rapidly during pressure targeted breaths.

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Challenge Flow drops to the cycling off threshold more quickly

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Challenge Early cycling off can also happen with other patients if slope rise is adjusted to a faster value after the expiratory threshold has been set to an appropriate level.

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Slope Rise and Expiratory Threshold Interactivity Diaphragm Electromyography

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Airway Pressure Time (s) Target Pressure PEEP Flow Diaphragm Electromyography Slope Rise and Expiratory Threshold Interactivity Early cycling off commonly causes double triggering

Interactivity and Synchrony Then Missed Triggers Airway Pressure Time (s) Target Pressure PEEP Flow Electromyography Diaphragm Abdomen

Interactivity and Synchrony Effects of a rapid Slope/Rise on Triggering (if Exp Threshold too high) In Pressure Support, a rapid slope rise can cause early cycling off and double triggering if the Expiratory Threshold setting is set too high Flow high Reach Exp Threshold Too Early Early Cycle off Patient continues inhaling Double Trigger Decreased trigger reliability from auto PEEP Airway Pressure Flow Slope Rise and Expiratory Threshold Interactivity

Interactivity and Synchrony Slope Rise and Expiratory Threshold Interactivity Confusion Clinicians may avoid using higher pressure support levels or faster slope rise settings because they associate them with negative patient responses. In reality, higher pressure support levels and faster slope rise settings may be used successfully without missed triggers and dis-synchrony if the expiratory threshold is adjusted to a suitable level.

Interactivity and Synchrony Bibliography of Published Literature Relevant to Slope Rise Selection Chatmongkolchart S, Williams P, Hess D, Kacmarek RM, Evaluation of Inspiratory Rise Time and Inspiration Termination Criteria in New -Generation Mechanical Ventilators: A Lung Model Study. Respir Care Jul;46(7): Chatburn, RL Primiano F, A New System for Understanding, Modes of Mechanical Ventilation, Respiratory Care. Respir Care Jun;46(6): Kallet RH, Campbell AR, Alonso JA, Morabito DJ, Mackersie RC.. The effects of pressure control versus volume control assisted ventilation on patient work of breathing in acute lung injury and acute respiratory distress syndrome. Respir Care Sep;45(9): Kallet RHCampbell ARAlonso JAMorabito DJMackersie RC Bonmarchand G, Chevron V, Ménard JF, Girault C, Moritz-Berthelot F, Pasquis P, Leroy J. Effects of pressure ramp slope values on the work of breathing during pressure support ventilation in restrictive patients. Crit Care Med Apr;27(4): Bonmarchand GChevron VMénard JFGirault CMoritz-Berthelot FPasquis PLeroy J Bunburaphong T, Imanaka H, Nishimura M, Hess D, Kacmarek RM. Performance characteristics of bilevel pressure ventilators: a lung model study. Chest Apr;111(4): Bunburaphong TImanaka HNishimura MHess DKacmarek RM Croci M, Pelosi P, Chiumello D, Gattinoni L. Regulation of Pressurization Rate Reduces Inspiratory Effort during Pressure Support Ventilation: A Bench Study. Respir Care 1996 Oct;41(10):

Intelligent bellavista Control to the Rescue! Interactivity and Synchrony Auto-Rise Auto-Synch™