IMAGE QUALITY
SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE
QUALITY MEASUREMENT METHODS PSF- POINT SPREAD FUNCTION LSF- LINE SPREAD FUNCTION CTF – CONTRAST SPREAD FUNCTION MTF- MODULATION TRANSFER FUNCTION
MTF MODULATION TRANSFER FUNCTION MOST COMMONLY USED TO DESCRIBE SPATIAL RESOLUTION IN CT
MTF MTF OF 1 – 100% TRANSFER OF OBJECT TO IMAGE MTF OF 0 – 0% TRANSFER OF OBJECT TO IMAGE
Spatial frequency Lp/cm
Spatial Frequency LARGE OBJECTS – LOW S.F. SMALL OBJECTS – HIGH S.F.
SPATIAL FREQUENCY 1 5
ABSOLUTE SIZE OF AN OBJECT IN CT IT’S EQUAL TO THE RECIPROCAL OF THE SPATIAL FREQUENCY CT SCANNER SPATIAL FREQUENCY- 15 lp/cm. WHAT SIZE OF AN OBJECT CAN THE SCANNER RESOLVE 1/15 = 0.06 cm/lp= 0.6 mm/lp 0.6 /2 = 0.3 mm object
Resolution in CT Spatial ResolutionContrast Resolution
SPATIAL RESOLUTION DEGREE OF BLURRING IN THE IMAGE MEASURE OF THE ABILITY TO DISCRIMINATE OBJECTS OF VARYING DENSITY A SMALL DISTANCE APART AGAINST A UNIFORM BACKGROUND
GEOMETRIC FACTORS F. SPOT SIZE BLURR APERTURE SIZE ( COLLIMATION) BLURR SLICE THICKNESS BLURR
PARTIAL VOLUME EFFECT 10MM SLICE THICKNESS SPREADS 2MM OBJECT OVER THE ENTIRE SLICE THICKNESS
# of projections S. resolution
ALGORITHM vs NOISE DETAIL NOISE STANDARD SMOOTHING EDGE-ENHANCEMENT
SPATIAL RESOLUTION AT HIGH CONTRAST HIGH CONTRAST RESOLUTION
DISPLAY RESOLUTION THE # OF PIXELS PER HORIZONTAL AND VERTICAL DIMENSIONS OF THE MATRIX ON THE MONITOR SCREEN OR FILM SHEET MATRIX SIZE DISPLAY RESOLUTION
DISPLAY RESOLUTION
HIGH RESOLUTION CT SLICE- 1 – 2 MM REDUCTION OF PARTIAL VOLUME ARTIFACT USE OF HIGH SPATIAL FREQ. ALGORITHM ( GOOD DETAIL) USE OF SMALL FOV ( SMALLER PIXEL)
RECONSTRUCTION ( TARGETING) vs MAGNIFICATION IN SPATIAL RESOLUTION RECONSTRUCTION - VERY LITTLE DIFFERENCE MAGNIFICATION – SIGNIFICANT DIFFERENCE, SPATIAL RESOLUTION
CONTRAST RESOLUTION SENSITIVITY OF THE SYSTEM LOW CONTRAST RESOLUTION LOW CONTRAST DETECTIBILITY
LOW CONTRAST RESOLUTION ( TISSUE RESOLUTION) THE ABILITY OF AN IMAGING SYSTEM TO DEMONSTRATE SMALL CHANGES IN TISSUE CONTRAST
FACTORS AFFECTING LOW CONTRAST RESOLUTION PHOTON FLUX SLICE THICKNESS PATIENT SIZE DETECTOR SENSITIVITY RECONSTRUCTION ALGORITHM IMAGE DISPLAY IMAGE RECORDING QUANTUM NOISE
CONTRAST RESOLUTION PHANTOM
PHOTON FLUX (P.F.) kVp P.F. mAs P.F. KVP, mAs C. RESOLUTION
PATIENT SIZE PHOTON FLUX PATIENT SIZE C. RESOLUTION
SLICE THICKNESS C. RESOLUTION SLICE THICKNESS C. RESOLUTION
DETECTOR SENSITIVITY C. RESOLUTION DETECTOR SENSITIVITY C. RESOLUTION
LOW CONTRAST RESOLUTION LOW SPATIAL FREQUENCY ALGORITHM
ALGORITHM vs NOISE DETAIL NOISE STANDARD SMOOTHING EDGE-ENHANCEMENT
QUANTUM NOISE C. RESOLUTION QUANTUM NOISE C. RESOLUTION
NOISE IN CT FLUCTUATION OF CT# BETWEEN POINTS IN THE IMAGE FOR A SCAN OF UNIFORM MATERIAL SUCH WATER.