Use of Dual Energy CT to Image Valve Settings of Programmable Shunts: Ex Vivo Testing Authors: K. Jim Hsieh, MD; Alexander C. Mamourian MD Department of.

Slides:



Advertisements
Similar presentations
An algorithm for metal streaking artifact reduction in cone beam CT M. Bazalova 1,2, G. Landry 1, L. Beaulieu 3, F. Verhaegen 1,4 1-McGill University,
Advertisements

Introduction Magnetic resonance (MR) imaging is recognised as offering potential benefits in the delineation of target volumes for radiotherapy (RT). For.
EP Discordance of CTA and Digital Subtraction Angiography in Diagnosing Vasospasm Following Subarachnoid Hemorrhage ASNR 53 rd Annual Meeting – April.
Sunday Case of the Day How does the presence of the object impact patient dose? A. Increases by 20% - 50% B. Increases by 3% - 5% C. No Change D. Decreases.
New Hybrid Monochromatic CT Images Achieve High Suppression of Metallic Artifacts While Preserving Soft Tissue and Iodine Contrast Michal Kulon, Peter.
In Pediatric Patients with Hypoplastic Internal Auditory Canals, What is the Utility of Thin-Section T2-Weighted Imaging to Determine the Contents of.
Sunday Case of the Day Physics
Andrew D. Hardie, MD Carlo N. De Cecco, MD
Applications of Dual-Energy Computed Tomography for the Evaluation of Head and Neck Squamous Cell Carcinoma  Reza Forghani, MD, PhD, Hillary R. Kelly,
Color pressure maps (mm Hg) in a healthy subject in the axial (A) and coronal (D) planes as well as for a patient (C and F) with a type IIa DAVF affecting.
Lateral skull radiograph performed in our patient at age 4 years shows cystoperitoneal shunt connected to a VP shunt via a Y connector (dashed arrow, reservoir.
Patient 4.A, Axial T2-weighted scan shows a huge extra-axial cyst (white arrows) with septation (black arrow) that displaces the corpus callosum posteriorly.B,
Ruptured PICA aneurysm in a 62-year-old man presenting in poor clinical condition. Ruptured PICA aneurysm in a 62-year-old man presenting in poor clinical.
Incidental detection of recurrent cholesteatoma in a patient being evaluated for an unrelated indication. Incidental detection of recurrent cholesteatoma.
Demonstration of the creation of a patient-specific brain mold for minimizing tissue distortion during fixation. Demonstration of the creation of a patient-specific.
A 50-year-old man with MD. Axial thin-section CT image shows decreased distance between the vertical limb of the posterior semicircular canal and the posterior.
Intraosseous temporal bone meningioma in a 45-year-old woman who presented with left-sided hearing loss and tinnitus. Intraosseous temporal bone meningioma.
Two coronal views of the same case, demonstrating the aggressive appearing imaging features of mastoid segment FNSs. The left coronal T1-weighted postcontrast-enhanced.
A, MIP of 3 mm thickness of the 1.5T scan.
Case 1: 15-year-old boy with juvenile nasopharyngeal angiofibroma on the right side. Case 1: 15-year-old boy with juvenile nasopharyngeal angiofibroma.
Preoperative 3D angiogram (A) shows a very wide-neck large ICA aneurysm. Preoperative 3D angiogram (A) shows a very wide-neck large ICA aneurysm. It could.
False-positive findings at the base of the left pterygoid process on conventional CT images alone in a 63-year-old woman with nasopharyngeal carcinoma.
Comparison of different DWI techniques.
Detection of recurrent cholesteatoma when physical examination is obscured and CT is indeterminate. Detection of recurrent cholesteatoma when physical.
3D-printed model of the lumbar spine (bottom left panel) of a patient with L1 left lamina osteoblastoma, designed to replicate the patient's anatomy as.
Singular-energy magnitude and location at peak systole in aneurysm volumes (gray) of aneurysm 1, 2, 3, and 5 for 3D PC-MR imaging; CFD with inflow boundary.
Axial contrast-enhanced neck CT scans and 3D reformat performed in a 59-year-old-man who underwent fibular free flap reconstruction for osteomyelitis complicating.
42-year-old male patient with follow-up neck CT for lymphoma at 70 kVp (A) and corresponding previous CT at 120 kVp (B). 42-year-old male patient with.
An 11-year-old girl with left-face sensitivity and left-temple pain.
Case 4: 25-year-old woman with pulsatile tinnitus.
Axial CT images at the basal ganglia level show subarachnoid hemorrhage on the right and a chronic lacunar infarct on the left. Axial CT images at the.
Scout suture evaluation.
Sarcoma growing into the left maxilla.
AP (A) and lateral (B) radiographs demonstrating a discontinuous segment of the catheter, with broken catheter ends in the subcutaneous tissue of the lower.
Bony cochlear nerve canal atresia in a patient with CND
SWI data in a patient with progressive MS exquisitely demonstrate the location of iron deposition in the deep gray matter. SWI data in a patient with progressive.
Coronal reconstructed noncontrast multidetector CT image of the chest demonstrates an extensive paravertebral soft-tissue mass and bone destruction at.
An oligodendroglioma in the right frontoinsular region.
A, Axial non-enhanced CT scan (soft-tissue algorithm) shows a heterogeneous right temporal bone lesion, with cortical thinning and bone remodeling of the.
Castellvi type IIIa and IIIb LSTVs
Conventional GE magnitude images show both calcifications and hemorrhage as areas of hypointense signal intensity. Conventional GE magnitude images show.
Four more examples of missed additional aneurysms on DSA
Cystic changes within the endplates adjacent to the implants.
A,B. A,B. Contrast-enhanced CT scan of the neck demonstrates extensive soft tissue edema. At the level of the vocal cords, there is a linear attenuation.
KD involving the postauricular area and occipital scalp in an 11-year-old boy. KD involving the postauricular area and occipital scalp in an 11-year-old.
Results of aneurysm treatment with flow modification only.
Lateral radiograph during inflation of a 10-mm balloon in an S2 midline sacral body fracture (same patient as in Fig 12). Lateral radiograph during inflation.
Temporal bone CT in the Pöschl (A) and Stenvers (B) planes demonstrating a large defect (arrows) in the roof of the right superior semicircular canal.
Orbital amyloidoma in a 32-year old woman with a slowly increasing mass in the right orbit and generalized bone pain. Orbital amyloidoma in a 32-year old.
1, Axial T2 image in patient 1 demonstrates bilateral cystic spaces in the biparietal periventricular white matter. 2, Axial T2 image in patient 2 demonstrating.
Bone algorithm CT images from the same case, demonstrating focal enlargement of the right tympanic segment, in the axial (left) and coronal (right) planes.
Sagittal reconstruction of a CT scan of a 6-year-old boy (patient 3) after a fall (GCS = 15) demonstrates an REH (thick white arrow), which causes mild.
Fourth ventricular CSF pulsation artifact in four subjects.
Automatic exposure control in the head of a pediatric patient.
Axial T2-weighted image (A) demonstrates focal cortical dysplasia (arrow) centered in the left anterior temporal lobe in a right-handed patient. Axial.
Patient 10. Patient 10. A 67-year-old man after radiation therapy at an outside hospital for base-of-tongue cancer. A and B, Axial contrast-enhanced CT.
A–C, Thin-section (1-mm) coronal and axial CT images of the skull base obtained with an edge-enhancing bone algorithm show enlargement (arrows) of the.
Serial lateral radiographs of a 69-year-old patient from group 2, with severe low back pain following forward bending.A, Initial radiographs demonstrate.
Number of white streaks for the cobalt-containing alloy (upper, black lines) and titanium clips (lower, white lines) scanned in high mode (H) (open symbols,
MR images in a 69-year-old woman with cervical and thoracic back pain.
Globally increased ASL signal intensity due to artifact.
Contrast-enhanced axial CT scan through the larynx with soft-tissue reconstruction demonstrates complete ossification of the thyroid (white arrow) and.
Axial CT at level of thyroid cartilage shown on soft tissue window (width, 340 HU; center, 43 HU) in panel A and narrow window (width, 1 HU; center, 130.
51-year-old man with headache, neck pain, and loss of consciousness after a fall. 51-year-old man with headache, neck pain, and loss of consciousness after.
False-positive findings at the base of the left pterygoid process on conventional CT images alone in a 63-year-old woman with nasopharyngeal carcinoma.
Tilted head position for intracranial CT angiography in a patient who has undergone clipping of a single aneurysm. Tilted head position for intracranial.
A, Setting code for the Sophysa Sophy SM8 valve.
A 73-year-old woman with well-differentiated SCCA of the lacrimal sac and nasolacrimal duct. A 73-year-old woman with well-differentiated SCCA of the lacrimal.
Orbital infiltration. Orbital infiltration. A, Soft tissue is seen in the posterior left orbit in this patient with metastatic disease to the bone. The.
Imaging of a 65-year-old man who presented with intractable epistaxis and whose history included right nephrectomy for renal cell carcinoma 5 years earlier.
Presentation transcript:

Use of Dual Energy CT to Image Valve Settings of Programmable Shunts: Ex Vivo Testing Authors: K. Jim Hsieh, MD; Alexander C. Mamourian MD Department of Radiology University of Pennsylvania ASNR 54 th Annual Meeting and The Foundation of the ASNR Symposium May 21 st – 26 th, 2016 Washington D.C. eP-99

Disclosures The authors have no financial disclosures.

Purpose Programmable shunts are commonly used for CSF diversion because they allow modifying cerebrospinal fluid (CSF) drainage rate without the need for repeat surgery. Establishing the valve settings, however, requires a hand held device that is specific to that valve vendor or plain film x-rays oriented perpendicular to the valve. These plain films must be performed in addition to the computed tomography (CT) that is considered the standard of care in patients with suspected shunt failure.

Purpose We wondered whether dual energy CT imaging with or without metal suppression software could be used to predict shunt settings in programmable shunts since that would eliminate the need for plain films in many cases. The purpose of this scientific paper is to determine whether it is possible to use CT imaging of two common programmable shunts, i.e. Strata (Medtronic) and Hakim (Codman) ex vivo to predict the pressure settings of their programmable valves. We also investigated whether a single energy CT scanner would be sufficient since dual energy scanners are not widely available.

Photon starvation artifacts from metal is a common limitation of CT imaging. Dual energy image acquisition offers one method to reduce the magnitude of the artifact. dual energy CTvirtual high kV monoenergetic The data acquired using dual energy CT imaging (Siemens) was used here to create a virtual high energy mono energetic reconstruction (right). Photos courtesy of John Woo, M.D. University of Pennsylvania

Material and Methods A Medtronic Programmable Strata valve was applied to a dried skull and then scanned at the five settings allowed by the device. The valve changes were performed and verified using the specific hardware for that device provided by Medtronic prior to each scan. Computed tomography scans were performed on a Siemens SOMATOM Force CT scanner using dual energy with a CTDIvol of 50 mGy.

Material and Methods The source 1 mm images were used to create a set of volume rendered (VR) images using TeraRecon 3D software (Foster City, CA). These VR images were used as a reference chart of the valve appearance at each setting ( ). The skull phantom then was scanned again at the five different shunt settings in random order and this imaging data was presented as a set of unknowns to three neuroradiologists without prior knowledge of the settings for each scan and there were duplicates in the test set.

Material and Methods A Codman Hakim valve was also applied to a dried skull and scanned at three settings (50, 120, and 190) on a Siemens SOMATOM Force CT scanner with dual energy. The source 1 mm images were used to create volume rendered images using TeraRecon 3D software and evaluated using the same approach as the Strata valve.

Material and Methods We then wanted to determine if we could reproduce these results with the Medtronic Programmable Strata valve on a non-dual energy CT scanner. A Medtronic Strata Valve was again applied to a dried skull and then scanned at the five settings allowed by the device on a Siemens SOMATOM Definition Edge CT scanner using a single energy (kV 120), and source 1 mm images were used to create volume rendered images using TeraRecon 3D and analyzed.

Results The volume rendered images of the Medtronic Strata valve scanned utilizing dual energy setting with and without metal suppression had little resemblance to the appearance of the setting as seen on plain film images. For this reason the usual chart of settings was of no value. We created a “VR chart” however. We found that each of the five settings appeared differently on the volume rendered images from the others. This was validated when three neuroradiologists were able to use the table previously generated to correctly determine valve setting in each of the unknown scans.

Medtronic Strata valve chart of settings as seen on standard plain radiographs. S.S. Lollis et al. AJNR Am J Neuroradiol 2010;31: ©2010 by American Society of Neuroradiology

Reference Chart – VR images post-processed using metal suppression processing of dual energy CT data

Surface Models of Programmable Shunt Valve (Medtronic, Strata) using CT data from dual energy scanner at 1.0 and 2.5 valve settings. 1.0 valve setting 2.5 valve setting Notice the size and shape of the “shoe” (arrows). While these images do not correspond to appearance of the valve on plain films, the VR images differ at the various settings.

Programmable Shunt Valve (Medtronic, Strata), VR and plain film Notice the “bunny ears” are visible on both at this setting. 2.0 VR model from CT data Plain film image at same setting (2.0)

Reference Chart – VR images post-processed CT data using soft tissue filter (kernel) without metal suppression

Reference chart – VR images post-processed using bone filtered data and without metal suppression

Methods The imaging of the Codman device was unsuccessful since there were no unique features on the VR models that would allow us to differentiate between the different shunt settings.

Codman Hakim Programmable Valve, plain film and VR from CT. S.S. Lollis et al. AJNR Am J Neuroradiol 2010;31: ©2010 by American Society of Neuroradiology There was insufficient detail on VR image (right) to delineate the notch that shows the shunt setting that is visible on plain film image (left).

Results Since dual energy CT may not always be available at all sites we hoped to replicate the results of CT imaging of the Medtronic Strata valve on a single energy CT scanner. We found that VR image detail was not as good from this scanner as the dual energy CT scanner and for that reason proved to be insufficient to determine each of the five valve settings reliably.

Surface Models of the Strata Valve using a standard, single energy CT scanner without metal suppression. 0.5 setting 2.5 setting1.5 setting

Conclusion We demonstrated “proof of concept” ex vivo that a dual energy CT is capable of demonstrating each of the five valve settings on the Medtronic Strata programmable shunt. Further testing is necessary in vivo but if validated this capability would be beneficial since it would eliminate the need for additional imaging with plain films in cases of suspected programmable shunt malfunction. Only one of the two valves we tested offered this capability. We believe manufacturers should be aware of the benefit to patients by designing or modifying programmable valves in a fashion that would allow imaging of their settings with standard CT as well as plain films.

References Lollis SS, Mamourian AC, Vaccaro TJ, Duhaime AC. Programmable CSF Shunt Valves: Radiographic Identification and Interpretation. AJNR Am J Neuroradiol 2010;31: Mamourian AC, Pluta DJ, Eskey CJ, Merlis AL. Optimizing computed tomography to reduce artifacts from titanium aneurysm clips: an in vitro study. J Neurosurg 2007;107: