MRI Safety Jerry Allison, Ph.D. Department of Radiology

Slides:



Advertisements
Similar presentations
MRI Safety Training.
Advertisements

YOUR GUIDE ON MRI.
MRI Phillip W Patton, Ph.D..
1 of 24 CCS&BS MR Safety SeminarCenter for Complex Systems and Brain Sciences Magnetic Resonance Safety Seminar: An Introduction to the Potential Hazards.
MRI Safety This slide show will educate all healthcare providers about the safe delivery of care in the magnetic resonance (MR) environment. This slide.
The importance of MRI, a few numbers  MRI units worldwide in 2003  75 millions scans per year performed  Constant need for over 1000 MRI technologists.
MRI in patients with pacemakers DR PRADEEP SREEKUMAR.
Handling Liquid Cryogens Safely Tony Kent School of Physics and Astronomy.
MRI Safety Jeopardy* ForcesProjectilesPotpourriHeatingScreening Because when you think MRI safety, think * Jeopardy is a trademark of Sony.
Equipment Magnetic resonance imaging (MRI) scan requires the use of a very strong magnetic field. Unlike other devices used in radiology, MR imaging.
Consent for Research Study A study for patients newly diagnosed with advanced glioblastoma (brain cancer): Learning whether a PET scan with F-fluoromisonidazole.
Overview The scanner Dangers in around the scanner In case of an emergency Emergency numbers Emergency procedure Reporting adverse events Safety checklists.
Methods: Scanner: GE 7T Excite (bore 60cm diam x 335cm length) and GE 3T LX (bore 60cm diam x 262cm length). The gradient performance of both scanners.
Biologic Effects. Biological Interactions Radiation Ultrasound MRI.
FMRI: Biological Basis and Experiment Design Lecture 18: Physical practicalities Digression: analysis ICE9: Example for WA8 Safety limits –dB/dt –SAR –Acoustic.
MRI Physics 3: Hardware Douglas C. Noll Biomedical Engineering University of Michigan.
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
THE MAGNETIC FIELD IS AT FULL POWER ALWAYS …. IT IS ALWAYS “ON”.
MRI Bioeffects and Safety Dr. Peterson. History and Background Bloch (Stanford) & Purcell (Harvard) Nobel Prize Discovery of Spectroscopy.
MRISC Organizational Chart Research and Graduate Studies Dr. James Tracy College of Medicine Fred de Beer, M.D. Director Dr. Charlie Smith Administrative.
MRI SAFETY FOR NON-MRI PERSONNEL.
MRI Safety Wendy Winkle
Chapter 15: MRI Safety Mark D. Herbst, MD, PhD. Two main sources of danger Magnetic field RF Energy Also –Gradients –Acoustic noise.
The UE directive 2004/40/EC and Magnetic Resonance Imaging a debate Thierry Metens Ph D Université Libre de Bruxelles, Hôpital Erasme Brain awareness week.
Overview of Neurostimulation
Radio Frequency Safety. Purpose Provide: Basic Technical Understanding Overview of FCC Regulations and Compliance Issues Hazard Recognition Skills Awareness.
Medical Imaging Technology
The comparison of MRI imaging In 3.0 T & 1.5T By a.r.shoaie.
Physiological Properties of Thermal Modalities (2)
By Ass.Prof.Dr. Gehan Mosad
Uniting MRI with ULTRASOUND hhholdorf. Dr. Raymond Damadian The MRI scanner was invented by Raymond Damadian. Though, Damadian did not invent the actual.
MRI DRAFT Principles of MRI ► Sectional Anatomy ► Provides anatomic and physiologic info ► Non- invasive ► No radiation  Magnetic fields and.
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
MRI Safety Course Module 5- Magnetic Fields and Radiofrequency.
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
Magnet Safety.
Pacemakers and MRI An outdated contraindication? Emily Myers Research in Allied Health East Tennessee State University.
Consent for Research Study A study for patients newly diagnosed with advanced glioblastoma (brain cancer): Learning whether a PET scan with F-fluoromisonidazole.
Magnetic Resonance Imaging
Ultrasound 10cm 2 Ultrasound (US) Therapeutic ultrasound –Uses acoustic energy, delivered at high frequencies for therapeutic purposes Electrical current.
 A medium that provides a means for transferring energy from one place to another  It is not expensive or difficult to control the risk of an electrical.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
Beam Measurements.
Medical Resonance Imaging
Methodist Hospital for Surgery
پرتوهای غیریونساز دکتر میرشهرام حسینی پناه
Lesson Plan MRI Scan Experience
Classroom and Laboratory Safety
Hazards of Electricity Electrical Hazards Include Electrical Hazards Include Electrical Shock Electrical Shock Electrical Explosions Electrical Explosions.
Jody Culham Brain and Mind Institute Department of Psychology
Medical Imaging Imagining Modalities.
SURGICAL DIATHERMY.
Classroom & Laboratory Safety
SURGICAL DIATHERMY.
Iontophoresis Vineela.U 08B21A0538 CSE.
Classroom and Laboratory Safety
Magnetic field safety October 2018.
MEDICAL IMAGING TECHNOLOGIES
Basics of MRI By Dr. Mohamed Samieh.
Human Subjects in fMRI Research
Lesson Plan MRI Scan Experience
Center for Magnetic Resonance Research
Lesson Plan MRI Scan Experience
Lesson Plan MRI Scan Experience
FMRI vs. MRI By: Kathleen Shaffer.
Electrical Hazards By Kory Hildebrand.
Residential exposure to EMFs
Hoglund Brain Imaging Center
Superconducting Magnets
Presentation transcript:

MRI Safety Jerry Allison, Ph.D. Department of Radiology Medical College of Georgia Augusta University

Projectile Accidents Foreign Materials in the Body Other Stuff MRI Safety Projectile Accidents Foreign Materials in the Body Other Stuff

Projectile Accidents The MRI magnets are ALWAYS on (24 hours/day, 365 days/year) There is a STRONG fringe magnetic field around the magnets The fringe magnetic field is confined to the scan room 3

Projectile Accidents The fringe magnetic field can cause ferrous metal objects to fly into the bore of the magnet with great force and speed Ferrous metal objects contain iron (steel) The closer you are to the bore of the magnet, the stronger the fringe field 4

As you approach the magnet, the fringe magnetic field gets STRONGER 5

http://simplyphysics.com/flying_objects.html# 6

Projectile Accidents The fringe magnetic field can be deadly Valhalla, N.Y. Westchester County Medical Center July 27, 2001 Six year old Michael Colombini was fatally injured “when an oxygen canister was turned into a guided missile by the powerful MRI magnet” 7

Foreign Materials in the Body Foreign material in the body is a concern in MRI Foreign material is any material in the body other than the flesh and bones that we are born with Wide variety of foreign material is intentionally and unintentionally present in patients, health care providers and others 8

Foreign Materials in the Body The presence of certain foreign materials in the body can be fatal in an MRI magnet Deaths have occurred due to the presence of aneurysm clips and cardiac pacemakers in patients undergoing MRI Sight has been lost by movement of ferrous metal in the eye resulting in vitreous hemorrhage (2.0 mm x 3.5 mm iron fragment @ 0.35 T) Prostheses are usually securely anchored to bone and are usually non-ferrous. 9

Foreign Materials in the Body Fringe magnetic fields stronger than 5 Gauss can affect cardiac pacemakers. To protect patients and others having pacemakers there is a 5 Gauss Exclusion Zone around MRI magnets. At AU Health, the Exclusion Zones are the MRI scan rooms 10

Fringe Diagram of Active Shield 1.5 T* 26’ 4” 5 Gauss 16’ 6” *Siemens Symphony 11

Safety (continued) -At least two deaths have resulted from MRI Scans on patients with pacemakers. 12

Foreign Materials in the Body Heating of metals and implants has not been significant (even for large prostheses) There are resources for assessing safety of foreign materials in MRI www.mrisafety.com F.G. Shellock. Reference manual for Magnetic Resonance Safety: 2017 Edition (© 2017 730 pages; ISBN 978-0-9891632-4-8) 13

Aneurysm Clips www.mrisafety.com 14

Signage FDA Guidance for the Submission Of Premarket Notifications for Magnetic Resonance Diagnostic Devices states: "The controlled access area should be labeled "Danger - High Magnetic Field" at all entries." The term "warning" does not convey the importance of a situation that may not only be potentially hazardous, but has been responsible for serious injuries and deaths. 15

16

Other Safety Issues Static Magnetic Fields M R I Physics Course Other Safety Issues Static Magnetic Fields Time Varying Magnetic Fields RF Magnetic Fields Auditory Safety FDA Safety Guidelines Contrast Agents 17

Other Safety Issues Psychological Distress Cryogen Safety M R I Physics Course Other Safety Issues Psychological Distress Cryogen Safety Experiences at Higher Magnetic Fields Pregnancy 18

Safety (continued) Biological effects of static magnetic fields -Augmentation of the T wave during peak flow in the aortic arch -No evidence of mutagenic effects (in vitro or in vivo) -No evidence of an effect on organ development in vivo -No pronounced threshold -No trends suggestive of a dose response curve 19

Safety (continued) Biological effects of static magnetic fields “In general, there is no conclusive evidence for irreversible or significant biological effects related to acute, short-term exposure of humans to static magnetic field strengths up to 2.0 T.” MRI systems up to 8T are in use for human imaging. 20

Safety (continued) Biological effects of RF magnetic fields -RF magnetic fields are oscillating magnetic fields -These fields produce heating of tissue -Energy deposited (degree of heating) is measured by the Specific Absorption Rate (SAR) in units of Watts/kilogram 21

Safety (continued) Biological effects of RF magnetic fields - SAR depends upon flip angle, TR, tissue density, RF pulse duration and frequency, RF coil radius, tissue conductivity and static magnetic field. -Guidelines restrict rise in core temperature to <1 oC 22

Safety (continued) Biological effects of RF magnetic fields -Similar to ultrasound since as long as the energy deposited is below established limits, minimal heating of tissues occurs. -Extra precautions are necessary for patients with compromised thermoregulation: fever, diabetes, obesity, cardiovascular disease, old age, medicated (  and Ca channel blockers), anesthesia 23

Safety (continued) Biological effects of RF magnetic fields -Thermal burns -Electrical conductors can absorb energy and become extremely hot -Avoid having any cable wire or lead in contact with the patient -Surface coil cables can be placed in sleeves to prevent burns 24

Safety (continued) -FDA Safety Guidance for MR Devices http://www.fda.gov/cdrh/ode/guidance/793.html 25

Safety (continued) Biological effects of time varying (gradient) magnetic fields -Faraday’s Law: Induction of electric fields and currents in tissues (depends on tissue conductivity, gradient strength, risetime, waveshape, polarity, repetition time TR and duration of the scan) -Eddy current formation 26

Safety (continued) Biological effects of time varying (gradient) magnetic fields -Nerve stimulation: electrical stimulation of peripheral and cutaneous nerves should precede any stimulation of ventricular fibrillation. Guidelines for time varying magnetic fields are designed to prevent any perceptible nerve stimulation ( with a safety factor of times three) -Magnetophosphenes: perceptible light flashes caused by electrical stimulation of the retina (not experienced in imagers having field strengths less than 4.0 T) 27

Safety (continued) Biological effects of time varying (gradient) magnetic fields “…guidelines are believed to provide a wide margin of safety with respect to exposure to gradient magnetic fields during MRI…newer “fast scanning” techniques that may require more complex applications of gradient magnetic fields may easily exceed these recommended levels and must be thoroughly evaluated for potential health hazards.” -Time varying magnetic fields are used to stimulate bone healing 28

Stimulation Threshold Levels Mean Cardiac Stimulation Threshold Mean Respiratory Mean Painful Nerve Mean Peripheral Nerve 3600 T/s T/s = Tesla per second 0 - 20 T/s = clinical mode >20 T/s = First or Second Controlled Modes* * Limited by IRB or Human Ethical Committee 900 T/s 90 T/s 60 T/s 45 T/s Typical Operating Range 20 T/s 29 0 T/s MR Safety Guide © 2003 by General Electric Company, Inc.

Due to the rapid rate of change of the magnetic fields (dB/dt) used during some EPI scans, a percentage of patients may experience a tingling or touch sensation. This means a percentage of patients may experience Peripheral Nerve Stimulation (PNS) at 45 T/s. If this sensation is bothersome or uncomfortable to the patient, stop the scan. Change to a non-EPI pulse sequence to continue scanning the patient. MR Safety Guide © 2003 by General Electric Company, Inc. 30

Safety (continued) Auditory safety Activation of gradient magnetic fields produces significant vibrations in the gradient coils. MRI acoustical noise has been shown to produce reversible hearing impairment and could potentially produce permanent damage. Hearing protection is recommended for all patients undergoing an MRI procedure on a high-field MRI system (1.5T and 3.0T). Noise attenuating ear-plugs or head phones are routinely used in MRI 31

Safety (continued) -FDA Safety Guidance for MR Devices http://www.fda.gov/cdrh/ode/guidance/793.html 32

Safety (continued) -MRI manufacturers often use SAR level control (based on IEC guidance?): Normal mode (or clinical mode): admissible for all patients First level: increased SAR permissible under “medical supervision” Second level: use of higher levels of SAR under IRB approval (research) 33

Safety (continued) -GE SAR levels 34

Safety (continued) -GE dB/dt levels 35

Safety (continued) -Contrast agents -Gd-DTPA is significantly safer than iodine based contrast agents used in X-ray procedures (x 10) -Severe reactions do occur -Pregnancy: Gd-DTPA may have poor clearance from amniotic fluid -Gadolinium may cause Nephrogenic Systemic Fibrosis (NSF) 36

Safety (continued) -Contrast agents FDA Requests Boxed Warning for Contrast Agents Used to Improve MRI … patients with severe kidney insufficiency who receive gadolinium-based agents are at risk for developing a debilitating, and a potentially fatal disease known as nephrogenic systemic fibrosis (NSF). In addition, it would state that patients just before or just after liver transplantation, or those with chronic liver disease, are also at risk for developing NSF if they are experiencing kidney insufficiency of any severity. 37

Safety (continued) -Psychological Distress (before, during and after MRI) Claustrophobia Anxiety Depression Panic Disorders -RF coils need to be near the body for high signal to noise -Usually transient (2 cases of claustrophobia persisted requiring psychiatric treatment) 38

Safety (continued) -Psychological Distress -Countermeasures Briefing Friend or relative in scan room Physical or verbal contact with patient Prone Mirrors, prism glasses, video goggles Blindfold Relaxation strategies Psychological desensitization techniques 39

Cryogen Safety Superconducting MRI magnets use cryogens (1970 liters of liquid helium) Magnets can QUENCH resulting in conversion of the liquid helium to gaseous helium 1970 liters of liquid helium becomes 53,000 cubic feet of gaseous helium in about 1 minute 40

Cryogen Safety Magnets are designed to dump the gaseous helium directly outdoors through a quench pipe If the quench pipe fails, gaseous helium could be dumped into the scan room, displacing the oxygen Some AU Health MRI scan rooms have oxygen monitors that automatically activate emergency exhaust systems in the event the oxygen level drops below 18% (An Alarm Sounds) 41

Cryogen Safety There has been one fatality by asphyxiation related to a liquid nitrogen leak in MRI 42

Safety (continued) -Experiences with higher magnetic fields (4 T) Several 7 – 8 T systems Several 4 T systems Numerous 3T systems Bioeffects: magnetophosphenes vertigo nausea headaches metallic taste 43

Pregnancy MRI is not considered to be hazardous to the fetus Limited studies have examined teratogenic effects Cells that are reproducing rapidly, as in the first trimester, are susceptible to damage from a variety of physical agents 44

Pregnancy FDA: “the safety of MR imaging when used to image the fetus and infant has not been established” British National Radiological Protection Board: “It might be prudent to exclude pregnant women during the first three months of pregnancy” “Particular caution should be exercised with the use of MRI during the first trimester because there is a high spontaneous abortion rate (30%) with associated potential medicolegal implications.” May be preferred over studies requiring ionizing radiation Should only be performed if the result is expected to be an important factor in the management of the patient 45

ACR Guidance http://onlinelibrary.wiley.com/doi/10.1002/jmri.24011/pdf 46

AU Health MRI Safety Policy 47

Vagal Nerve Stimulators Cyberonics VNS Therapy System Procedures Several preparatory steps Pulse generator to 0 mA Transmit/receive head coil only < 2 Tesla SAR: <1.3 for a 70kg patient Time-varying intensity: < 10 Tesla/sec 48

Medtronic Deep Brain Stimulators Procedures Several preparatory steps Pulse generator to 0 mA Transmit/receive head coil only Only 1.5 Tesla systems Use correct patient weight SAR: <0.1 Gradient dB/dT: < 20 Tesla/sec 49

Projectile Accidents Foreign Materials in the Body Other Stuff MRI Safety Projectile Accidents Foreign Materials in the Body Other Stuff