Home sphygmomanometers

Slides:



Advertisements
Similar presentations
Blood pressure measurement in primary care
Advertisements

ICT in Healthcare Topic 6.
BLOOD PRESSURE OBJECTIVES DEFINE BLOOD PRESSURE –SYSTOLIC AND DIASTOLIC NORMAL RANGE OF B/P –SYSTOLIC AND DIASTOLIC SIZE AND PLACEMENT OF B/P CUFF.
ITU-T Technical Session on EMF
1 Hypertension Center, 3rd University Dept. Medicine, Sotiria Hospital, Athens, Greece 2 Microlife AG, Heerbrugg, Switzerland 3 Microlife Corporation,
1 Mark Huffman, MD, MPH Northwestern University Feinberg School of Medicine Northwestern Memorial Hospital Healthy Transitions 21 November 2013 Blood Pressure.
Measurement of CV variables. In vitro Total control of confounding variables –Vasomotion, temperature changes, autoregulation, mean BP Most accurate because.
Ακριβής και αξιόπιστη μέτρηση ΑΠ
Measuring Blood Pressure
NUR 111: SKILL 5-5: ASSESSING ARTERIAL BLOOD PRESSURE ASSESSING A BLOOD PRESSURE : I FEEL THIS IS ONE OF THE MOST IMPORTANT SKILLS, AS A NURSE, THAT YOU.
Noninvasive Blood Pressure Monitoring Issued April 2010.
CME Program for Family Physicians Ambulatory BP Monitoring Brian Gore, MD CCFP Dip Epid. CME Program for Family Physicians Ambulatory BP Monitoring Brian.
Blood pressure measurement PHCL 326 (1+1) Clinical skills for pharmacists.
MT Najafi Oct 2014 Introduction  Self-monitoring of blood pressure by patients at home (home blood pressure monitoring (HBPM)) is being increasingly.
Copyright © 2008 Delmar Learning. All rights reserved. Unit 20 Blood Pressure.
Calibration Measuring to the Standard Blood Pressure Measurement Workshop Morehouse School of Medicine Presented by: Sandra Owen Georgia State University.
A guide for healthcare professionals Measuring Blood Pressure at Home Michigan Department of Community Health Heart Disease and Stroke Prevention Unit.
Critical Appraisal of Clinical Practice Guidelines
HOME AND AMBULATORY BLOOD PRESSURE MONITORING
Home sphygmomanometers Tekin AKPOLAT, MD Professor of Internal Medicine Ondokuz Mayis University School of Medicine Department of Nephrology SAMSUN-TURKEY.
Quality Control Project Management Unit Credit Value : 4 Essential
ACCURACY IN BLOOD PRESSURE MEASUREMENT Importance of site and accuracy.
Rasvahapot valtimotaudin vaaran arvioinnissa Blood pressure measurement - EHES September 21, 2010 Antti Jula.
Introduction In medicine, business, sports, science, and other fields, important decisions are based on statistical information drawn from samples. A sample.
G U I D E L I N E SG U I D E L I N E SG U I D E L I N E SG U I D E L I N E S JNC ESH 2003 BHS 2004 CHEP 2004 JSH 2004 HSSH 2005 HOME BP MONITORING.
Managing Blood Pressure: It Takes a Team
Copyright 2010, The World Bank Group. All Rights Reserved. Principles, criteria and methods Part 2 Quality management Produced in Collaboration between.
Section I. Accurate Measurement of Blood Pressure
SINGING FROM THE SAME HYMN SHEET Address to SATS Study Day 29 June 2013 Dr Sue Armstrong.
2015 Canadian Hypertension Education Program Recommendations
Section VII. Home BP Measurement 2015 Canadian Hypertension Education Program Recommendations.
Hypertension What is Blood Pressure?
BLOOD PRESSURE MEASUREMENT OF THE PRESSURE THAT THE BLOOD EXERTS ON THE WALLS OF THE ARTERIES DURING THE VARIOUS STAGES OF HEART ACTIVITY. –AS THE HEART.
BLOOD PRESSURE MEASUREMENT Sharon Jones Chapter 17.
Blood pressure measurement Presented by -Deepika Jain -Divya Jain
Measuring Blood Pressure. Diagnostic Criteria for HTN Two or more separate occasions – BP > 140/90 –Average of Seated Standing for > 2 min Seated with.
Please get your assigned computer. Do not turn on until instructed.
Measuring Blood Pressure. Diagnostic Criteria for HTN Two or more separate occasions – BP > 140/90 –Average of Seated Standing for > 2 min Seated with.
© 2016 Cengage Learning ®. All Rights Reserved. May not be scanned, copied or duplicated, or posted to a publicly accessible website, in whole or in part.
Lifestyles, Fitness and Rehabilitation Hypertension.
Copyright 2010, The World Bank Group. All Rights Reserved. Agricultural Census Sampling Frames and Sampling Section B 1.
© 2016 Global Market Insights, Inc. USA. All Rights Reserved Fuel Cell Market size worth $25.5bn by 2024Low Power Wide Area Network.
Clinical Quality Improvement: Achieving BP Control
Carditech, California, USA University Of West Bohemia, Czech Republic
G U I D E L I N E S HOME BP MONITORING JNC ESH 2003 CHEP 2004
Screening System for Hypertension and Diabetes at Primary Care Level
World Kidney Day 2016: Kidney Disease & Children
Dr. Kęstutis Adamonis, Dr. Romanas Zykus,
Diagnosis and initial management of hypertension in primary care
Chapter 6 Vital Signs.
World Health Organization
BIOELECTRONICS 1 Lec10: ch4 Blood Measurements By
First-in-Man, First In The USA: What’s The Difference?
Hiding In Plain Sight (HIPS) Strategy
What’s New in the 2013 ESC/ESH Hypertension Guideline
Deputy Director, Division of Biostatistics No Conflict of Interest
Clinical Engineering Lecture (3).
THE FEASIBILTY STUDY LECTURE-5.
Research amongst Physical Therapists in the State of Kuwait: Participation, Perception, Attitude and Barriers Presented by Sameera Aljadi, PT, PhD Assistant.
SAMPLING (Zikmund, Chapter 12.
CHAPTER 2 Testing Throughout the Software Life Cycle
Blood Pressure August 2015 Blood Pressure.
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Self-measured blood pressure monitoring
Diagnosis of disease M2/D2
Measurement of Blood Pressure: Issues and Challenges
Coding and processing time use
Read more STRIDE BP - Accurate BP Monitors
Levels of involvement Consultation Collaboration User control
Presentation transcript:

Home sphygmomanometers Tekin AKPOLAT, MD Professor of Internal Medicine Ondokuz Mayis University School of Medicine Department of Nephrology SAMSUN-TURKEY November 21, 2009 ANTALYA

Plan Introduction Home sphygmomanometers and features Validation protocols Limitations of the validation protocols Accuracy of sphygmomanometers Potential role of nephrologists Conclusions/Summary References

Purpose of this presentation Summarize basic features of automated home BP measurement devices Emphasize the distinction between validation, calibration and accuracy of automated home sphygmomanometers Discuss some practical points for improvement of BP control in CKD for nephrologists.

Other topics Aneroid and mercury sphygmomanometers Proper patient preparation Patient training BP measurement techniques Potential advantages of SMBP and 24h ABPM Automated devices used for 24 hours ABPM or in hospitals will not be discussed

Current situation 1 The introduction of inexpensive, easy-to-use, and automated BP measuring devices, lead to a widespread use of SMBP at home and SMBP at home became a part of clinical practice including CKD.

Current situation 2 According to a nation-wide survey in 2004–2005 in Japan, 90% of clinicians recommended home blood pressure measurement and 77% of hypertensive patients have a sphygmomanometer at home. The proportion of patients owning a monitor has increased from 49% in 2000 to 64% in 2005 (USA).

Home sphygmomanometers Mercury, aneroid, automatic, and semi-automatic Mercury: the gold standard Aneroid: practical difficulties Automated: preferred Measurement of BP with an automated device is a simple procedure.

Features of automated home sphygmomanometers 1 Measurement site (upper arm, wrist, finger) Appropriate cuff-size Availability of smaller or larger sized cuffs One device-two cuffs Comfort cuff (22-42 cm)

Features of automated home sphygmomanometers 2 Validation status in general population Validation status in special groups Model/type Accuracy

Features of automated home sphygmomanometers 3 Arrhythmia detection Consecutive 3 measurements Avoidance of pumping to too high levels Multi-user facility Easy to use XL size display Infrared interface

Features of automated home sphygmomanometers 4 Quick measurement Capacity of memory Modern style Small dimensions (suitable for traveling, handbags…) Equipped with carrying bag Soft case Easy wrap cuff

Features of automated home sphygmomanometers 5 Washable cuff cover Batteries included Minimum number of batteries Low battery indication Rechargeable battery pack Enabling PC connectability via USB cable

Features of automated home sphygmomanometers 6 Software for use with a PC Links to PC separate printer With printer Bluetooth output to telemedicine service providers Blood pressure classification indicator Hypertension Indicator

Features of automated home sphygmomanometers 7 Hide display function Advanced Positioning Sensor Wrist position sensor Voice speaks readings Voice speaks orders Automatic switch off

Features of automated home sphygmomanometers 8 Fully automated deflation Tracking of morning hypertension Time and date of measurement Pulse rate Provide trend plots

Important/essential features Measurement site (upper arm, wrist, finger) Appropriate cuff-size Validation status in general population/ special groups Accuracy

Wrist/upper arm Wrist devices are popular Wrist diameter is little affected by obesity Strict attention should be paid to having wrist at heart level while operating the device. Three specific guidelines and two relevant websites recommend upper arm devices A validated finger device is not available

Appropriate cuff-size Obesity is an increasing problem Cuff bladder must encircle at least 80% of the arm. The manufacturer's specifications can be followed Recommendations: AHA and BSH Smaller or larger cuff-sizes: Extra payment

Validation/accuracy Validation protocols are objective guides Two useful websites All models/types of any trademark or company present in the market are not validated The validation protocols do not guarantee accuracy of a particular device for an individual patient

Validation protocols 1 AAMI BHS International protocol Japan ISO-WG

Validation protocols 2 The basis of the validation tests is the comparison of BP measured by the device being tested with measurements made by trained observers, using a mercury sphygmomanometer under control of one or two experts. BP should be measured with the arm supported at heart level after 10-15 min rest.

Validation protocols 3 Indirect measurement of BP using a mercury sphygmomanometer is preferred, direct (intra-arterial) or simultaneous measurements have some disadvantages. A sequential same-arm method was used for comparison of tested automated and mercury sphygmomanometers. The devices are tested over a wide range of BP categories with a certain number of subjects in each category.

Validation in special groups 1 The BHS protocol described validation procedures for special groups such as pregnant women, children and elderly people The basis of additional validation testing for elderly population is increased arterial stiffness with aging which is a manifestation of CKD including predialysis period

Validation in special groups 2 Arterial stiffness can influence the correspondence between readings taken by mercury sphygmomanometers and oscillometric devices. Up to now, only one home sphygmomanometer was validated for dialysis patients (Blood Press Monit 2007; 12: 227-232)

Validation in special groups 3 Arterial stiffness is also increased in diabetes mellitus which is the most common cause of CKD in most of the countries. Special patient groups such as early stages of CKD, diabetic nephropathy or patients having extraosseus/vascular calcifications may require specific validation tests.

Blood Press Monit 2002; 7: 313-8.

Blood Press Monit 2002; 7: 313-8. Gerin et al have addressed the clinically relevant issue of device accuracy in individual patients for the first time. They planned a theoretical study and made an empirical test to estimate the proportion of persons for whom a BP monitor validated according to existing BHS and AAMI standards would be inaccurate.

Blood Press Monit 2002; 7: 313-8. They have shown that errors do tend to cluster within persons, and by concentrating solely on the population mean error, the BHS and AAMI protocols allow the approval of monitors that are inaccurate for a substantial proportion of people. As a result, they concluded that under these validation criteria, it is possible that more than half of patients will have an average error greater than 5 mmHg, and more than one in four will have an average error greater than 10mmHg.

They proposed two stages for validation. The model of the monitor in question should be validated at the population level. The particular monitor unit should be validated in the physician’s office for the intended user.

International protocol In order to decrease individual inaccuracy, the IP introduced a tertiary phase whereby the device was assessed according to the number of subjects in whom it gives accurate measurements in addition to its overall accuracy.

Blood Press Monit 2008; 13: 187-91 The statistical power of three validation protocols (AAMI, BHS and International) have been compared by Friedman et al and they concluded that the decrease of participants from 85 to 33 in the IP reduced its statistical power from 98 % to 70 %.

Accuracy of sphygmomanometers 1 Validation, accuracy and calibration are different and confusing concepts. Calibration is a procedure to control accurate BP measurement and it assesses a sphygmomanometer under in-vitro conditions. For automated sphygmomanometers, calibration is the assessment of the accuracy of the pressure transducer, which requires specialized equipment.

Accuracy of sphygmomanometers 2 Although calibration of the pressure transducer is essential for an automated device, it does not address the accuracy of BP determination. Accuracy can only be determined by clinical testing.

Accuracy of sphygmomanometers 3 The American Heart Association, American Society of Hypertension, Preventive Cardiovascular Nurses Association and European Society of Hypertension guidelines on SMBP emphasized the importance of checking monitors for accuracy in 2008.

Samsun experience 1 We planned a campaign to determine the accuracy of home sphygomanometers in 2006. The findings have been published in two articles [Blood Press 2008; 17: 34–41 and Blood Press Monit 2009; 14: 26–31.].

Samsun experience 2 We realized that a significant proportion of the devices had individual accuracy problems. Devices having a difference greater than 4mmHg were considered inaccurate. After learning that their newly purchased devices were inaccurate, most of the patients returned them to the retailer, who in turn forwarded them to the importer.

Samsun experience 3 The technical service of the importer found that the calibration was normal in almost all of the devices. The company did not consider inaccuracy a problem and returned the device back to the patient. The patients came back, there was nothing to do.

Published studies/guidelines The devices were checked for accuracy using two methods in previous studies: assessment of calibration and sequential measurement. The percentage of inaccurate automated devices was higher than 40 % in all studies using the sequential method for the evaluation of accuracy This high inaccuracy rate is the evidence of magnitude of the individual accuracy problem.

Calibration/Accuracy Calibration and accuracy are different concepts Accuracy can only be determined by clinical testing: But HOW?

HOW WILL THE INDIVIDUAL ACCURACY BE CHECKED? The cutoff value for inaccuracy began from greater than 3mmHg; many investigators used at least 5mmHg, and Ali and Rouse [J Hum Hypertens 2002; 16:359–361] considered inaccuracy an error of greater than 10 mmHg. The relevant guidelines emphasized the importance of checking the individual accuracy, but did not mention how to do it.

Blood Press Monit 2009; 14: 208-215.

Blood Press Monit 2009; 14: 208-215. This study described a method for the assessment of individual accuracy in details. The three pivots of this method are sequential measurement, the number of BP measurements and cutoff values for assessment.

Blood Press Monit 2009; 14: 208-215. The test has three stages Sequential measurement of BP similar to validation protocols (mercury, tested device, mercury..) was used for the first time for the assessment of individual accuracy. The tested devices were categorized into 4 groups

Potential role of nephrologists Recommendation for purchase of a validated device (trademark and model/type) Checking of the device of the patient for accuracy Teaching of BP measurement technique Observation of the patient while operating the device

Practical points Two useful websites Measurement of arm circumference Supplementation with CD, video, easily readable booklets, brochures… The PA.NET International Quality Certification Protocol (Blood Press Monit 2008; 13: 285-9)

Conclusions/Summary 1 Inaccuracy of home sphygmomanometers is a common, neglected and ignored problem. Measurement site, cuff-size, validation status, and accuracy are the most important features.

Conclusions/Summary 2 Individual accuracy can be more important in special patient groups such as early stages of CKD, diabetic nephropathy or patients having extraosseus/vascular calcifications. The nephrology clinics have a great role and responsibility to encourage optimal circumstances for SMBP at home for their patients and they can be organized for training programs and check of individual accuracy.

Conclusions/Summary 3 Active guidance of nephrologists to all steps of SMBP at home will improve control of BP and increase quality of patient care in CKD.

References http://www.dableducational.org. http://www.bhsoc.org/blood_pressure_list.stm J Hypertens 2008; 26: 1505-26. J Cardiovasc Nurs 2008; 23: 299-323. J Hypertens 1993; 11 (suppl 2): S43–S63. Blood Press Monit 2009; 14: 208-215. Blood Press Monit 2002; 7: 313-8. Blood Press Monit 2002; 7: 3-17. Hypertens Res 2003; 26: 771-82.