GC Trouble shooting.

Slides:



Advertisements
Similar presentations
Gas Chromatography Introduction 1.) Gas Chromatography
Advertisements

Gas Chromatography Vaporization of sample Gas-solid
Gas Chromatography 427 PHC.
Lecture 8b Gas Chromatography.
CAPILLARY GC INLETS.
GAS CHROMATOGRAPHY ENVE 202 Dr. Aslıhan Kerç.
Lab Methods Day June 25, 2014 Gas Chromatography
Mass Flow Controller. Function Delivers user specified flow rate of gas. Flow rate is mass flow rate (i.e. number of molecules per sec Mol/sec), not volume.
Daikin Altherma™ Troubling Shooting
6890 GC Detectors - The Flame Ionization Detector
TROUBLESHOOTING. 5WSS TROUBLESHOOTING ERROR CODEFAULTY INDICATION E1ROOM SENSOR CONTACT LOOSE E2INDOOR COIL SENSOR CONTACT LOOSE E3WATER OUTLET SENSOR.
GC. Columns have a finite life, but they will last much longer if the gas chromatography system is operated with care. The major causes of reduced column.
The Foundations: Classical Split and Splitless Injection
Chromatography Integration and Trouble-shooting
Simdis Trouble Shooting
Mobile phase is a gas! Stationary phase could be anything but a gas
TROBLESHOOTING GAS CHROMATOGRAPHY (GC) Parto Zist Behboud Reza Daj 8/16/2014.
Oxygen Concentrators.
6890/7890 GC Hardware and Technology Overview. 2 In this Section, We Will Discuss:  The major components of the gas chromatograph.  The typical chromatogram.
Normal Operating Parameters Maintenance Schedule Common Problems Normal Operating Parameters Maintenance Schedule Common Problems ClearWater Tech, LLC.
1 April 7, 2009 C. Lu Monitoring the RPC Gas Mixture with a Gas Chromatograph C. Lu Princeton University.
Dehydration of 2-Butanol and Dehydrobromination of 2-Bromobutane;
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
Next Home Remote Controlled Cliqueur troubleshooting guide Start the slide show then click Next.
Trouble Shooting and Performance Testing
AC Analytical Controls January 2001
Gas Chromatography. Gas Chromatography Basics Gas Liquid Chromatography (GLC) Gas Solid Chromatography (GSC) Mobile phase does not interact with analyte.
Glycol Pumps Operating Pressure: 100 to 2000 psi Circulation Rate: 3 to 450 gph.
FUEL INJECTION SYSTEM OPERATION
Gas Chromatography Carrier Gas Flow Control Injector Port Column Column Oven Detector Recorder The GC system consists of gas supplies for the mobile phase.
Chapter 27 Gas Chromatography 1. Principles
PAS-Intro1 Dated 24th Aug Copyright © Siemens AG 1999 All Rights Reserved Siemens Advanced Engineering Pte Ltd Process Analytical Services Division.
Temperature Control Loop
animation.php.
Diagnostics – Quaternary Pump & Fraction Collector Analytical 920 LCSemi-Prep 940 LC.
An Introduction to Chromatography What is chromatography? The separation of a mixture by distribution of its components between a mobile and stationary.
900-LC P. van Poppel – Field Support Specialist October 2009.
Advanced Method for Renewable Ethanol by Direct Synthesis from Syngas for Renewable Fuel Applications Julia Fisher, Chemical Engineering, Arizona State.
GAS CHROMATOGRAPHY (GC). GAS CHROMATOGRAPHY Sample is injected (using a syringe) into the injection port. Sample vaporizes and is forced into the column.
F-time Reformate Mode. F-time OPNA Mode F-time Winterspec Mode.
Intensive General Chemistry Chemical separations II Isabelle Vu Trieu
Analytical Separations
Created with MindGenius Business 2005® Instrumentation (1) Instrumentation (1) Origins Preparative column chromatography (Glass columns) Internal diameter:
FAME in AVTUR using GC Heart-Cut and refocusing – IP PM EL Piet Koppen – AC Analytical Controls EI Fame in Jet workshop March 1, 2012.
9100 HPLC Trouble shooting.
Problem Solving using Error Code Saijo Denki Inverter R-410A
Gas Chromatography. In Gas Chromatography (GC), a gaseous mobile phase transports a gaseous solute through a long, thin column containing solid or liquid.
HPLC TROUBLESHOOTING. There are five major of symptoms to help you quickly identify the source of the problem(s): I.Pressure abnormalities II.Leaks III.Problems.
Lecture 9 Gas Chromatography Lecture 9 – Chromatography, Dr. Rasha Hanafi 1© Dr. Rasha Hanafi, GUC.
Experiments in Analytical Chemistry
Gas Chromatography Chap 27 Types: Gas-solid chromatography (GSC)
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
BioTector COD TOC TN analyzer
The BioTector B7000i TOC analyzer with integrated oxygen concentrator
TXV Devices.
Air Pak Drying Systems Presented By: Dry Air Systems, Inc. Innovator of Unique Regenerative Desiccant Air Dryers For Compressed Air & Fluid Power Controls,
6850/6890 分流动画 Split/Splitless 进样口 – 部件
Overview Determining Identity Quantitative Analysis
Emission Control Diagnosis and Service
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Chem. 133 – 5/11 Lecture.
T-6B On-Board Oxygen Generation System (OBOGS)
Gas Chromatography.
Gas Chromatography.
FIGURE 35-1 If the vacuum hose is removed from the fuel-pressure regulator when the engine is running, the fuel pressure should increase. If it does not.
Gas chromatography Software
32A-4 Chromatographic Detectors
GAS CHROMATOGRAPHY.
32A-2 Sample Injection system
T-6B On-Board Oxygen Generation System (OBOGS)
Presentation transcript:

GC Trouble shooting

Gas Chromatography Gas Chromatograph Detector Injector Data Column Sample injection Chromatogram Detector Injector Carrier Gas Data System Column Gas Chromatograph 1 2

Injector Maintenance Capillary INJ Packed INJ

Injector Maintenance 1. Injector Leak Contamination - Septum, Glass insert & O-ring - Ferrule & Nut Contamination - Septum, Glass insert - Tubing

Injecting whole vaporized sample Packed Injector – Leak Injecting whole vaporized sample <Symptom> Decreased pressure compared to usual Reduced peak area with shifted RT <Remedy> Septum replacement Check the leak on column connection - Flow the gas without column and block the inlet port by a finger -> If column pressure increases, Check column connection ->If not, Check the leak on parts ahead of column - Welding, Sensor or Valve trouble -> Contact an engineer

Packed Injector – Contamination <Symptom> Irregular Ghost peak observed without sample injection <Remedy> 1) Contamination : Septum or Glass insert replacement 2) Check either injector or column contamination Check the ghost peak by injector temperature off, no sample injection, and oven program mode. => If Ghost peak area is reduced by injector temperature off, injector might be contaminated. Replace Glass insert with O-ring. => If Ghost peak area is not reduced by injector temperature off, Column might be contaminated. Condition the column.

Injecting a part of sample Capillary Injector – Leak <Symptom> - Column pressure is reduced - Split vent flow rate is increased as column flow rate is reduced - Reduced peak area with shifted RT <Remedy> Septum replacement Tighten Retainer Nut Check the leak on column connection - Flow the gas without column and block the inlet port by a finger -> If column pressure increases, Check column connection ->If not, Check the leak on parts ahead of column - Welding, Sensor or Valve trouble -> Contact an engineer Retainer Nut Injecting a part of sample

Capillary Injector – Contamination <Symptom> Irregular Ghost peak observed without sample injection <Remedy> 1) Contamination : Septum or Glass insert replacement 2) Check either injector or column contamination as same as packed injector. <Other trouble> Unstable flow rate – Perform Sensor zero, Valve Calibration and check APC parts. Peak area is increased higher 3times than usual. – Split vent Line is clogged. Split Vent line and Trap replacement Abnormal temperature and flow rate display - Check RTD sensor, APC board, Main board, AC power board

Column Oven Maintenance Oven temperature control problem - Temperature doesn`t increase : Check Heater coil resistance. (22ohm) - Abnormal temperature display : Check RTD sensor resistance. (110ohm) - Temperature fluctuation : Check Power supply. (220~240V) 2. Column Performance degradation & Contamination - Poor peak resolution, RT shifted, and Baseline increased <Column conditioning> - Connect the column to injector only and run oven program mode. If new column is connected to FID without conditioning, FID jet is frequently clogged. - Set proper column flow rate as per column ID. 100℃ 1hour 2-3 ℃ /min Column max temp.-30℃

Max temperature for typical columns

Detector Maintenance

Detector Maintenance <FID> <TCD> Ignition failure -Flow rate, Ignitor, Jet Abnormal signal - Collector contamination, FID B/D failure <TCD> Sensitivity trouble - Abnormal TCD cell -TCD B/D failure - Leak

FID – Ignition failure 1. Ignitor coil isn`t brighten after GC makes ‘Tic’ sound. => Coil is broken. Ignitor Replacement. 2. Flow rate check: Air 300, H2 30-40, Mkup 10-25(Cap). If Air or Mkup flow rate is too high, ignition is generally failed. If H2 flow rate is high, ignition sound is loud. 3. Jet or gas line is clogged. Check the Jet status under Room temp. Flow check Air H2

FID – Abnormal signal Normal signal is less than 50 1. Maximum signal display: 1,000,000 1) Moisture in FID => Bake FID out at 300 ℃ or clean the collector. 2) Shortage between spring and FID => Check FID whole assembly position. 3) FID B/D failure => Board replacement 2. High signal 1) Before ignition: Collector contamination (Clean it with Acetone) 2) After ignition: - Jet, or Gas line contamination - Injector or Column contamination (Plug FID port by a blank Nut and check the signal whether it is from FID, Injector, or Column.)

TCD – Sensitivity failure *Caution : Turn on Filament after all gas flows to TCD. Turn off Filament before turn off gas flow rate. * Check TCD cell resistance under room temperature. 1) Low sensitivity 2) Abnormal signal TCD Board and cell connection

TCD – Low sensitivity 1) Check Sam/Ref flow rate by a Flowmeter => Check the Leak in Cell Tubing. => Check the Leak on column connection. 2) Set proper Sense value? 6100GC : H2,He : 6, N2 and Ar : 3 or 4 6500GC : H2: 10, He : 8 or 9, N2 and Ar : 6 or 7 3) TCD Cell resistance measurement : Difference must be within 0.5Ω. 4) TCD B/D failure : Board replacement

TCD – Abnormal signal * Normal TCD signal under same Sam/Ref flow rate -Block Ref outlet by a finger for a second and signal changes to +3000. -Block Sam outlet as same as above and signal must change to -3000. (1:1 proportion change but polarity is opposite.) 1) Maximum signal display - Difference flow rate between Sam and Ref is too high and no flow rate either Sam or Ref. - Cell contamination =>Thermal conditioning (300 ℃) 2) Signal doesn`t change from Sen 0 to 9. TCD B/D failure =>Board replacement