Josiane Arnaud1, Jean-Philippe Weber2,

Slides:



Advertisements
Similar presentations
An Introduction to Quality Assurance in Analytical Science
Advertisements

Ninewells Hospital & Medical School
QUALITY CONTROL IN HEMATOLOGY
User verification of performance Zahra Khatami/Robert Hill BHR University Hospitals/Sheffield Teaching Hospitals Focus 2013.
SAMPLE DESIGN: HOW MANY WILL BE IN THE SAMPLE—DESCRIPTIVE STUDIES ?
Biological Variation Data: Time for Development of Standards? Dr William A Bartlett, On behalf of the Biological Variation Working Group, European Federation.
Harmonization of Measurement Results of the Alcohol Biomarker Carbohydrate- Deficient Transferrin by Use of the Toolbox of Technical Procedures of the.
EMODNet Chemistry Steering Committee January 2014 Rome Giordano Giorgi
1 Developments in Metrology Ed W.B. de Leer NMi Van Swinden Laboratorium The Netherlands UNECE WP.6 Presentation 24 november 2004.
QCL Training Seminar, Tanzania | 5-7 Dec 07 1 |1 | Proficiency Tests John H McB Miller Laboratory Department (DLab) European Department for the Quality.
L MACKAY, N CRINIS, Q LAM, R SCOTT
1998 EQAS Organization and Management - IFCC/EMD/C- AQ1 EQAS Organization and Management l Bibliography Fundamentals for External Quality Assessment, IFCC/EMD/C-AQ,
Utilization of Assay Performance Characteristics to Estimate Hemoglobin A 1c Result Reliability A. Woodworth, N. Korpi-Steiner, J.J. Miller, L.V. Rao,
Internal Quality Control (QC) for Medical Laboratories: An introduction Dr. Otto Panagiotakis and Dr. Alexander Haliassos ESEAP – Greek Proficiency Testing.
InterQC Delivering Simplicity in the Quality Control Management for your Laboratory Developed by Vitro.
Area deprivation and cultural inequalities in fetal growth: capturing trends over time Nathalie Auger, a,b,c Alison L Park, a,b Mark Daniel d a Institut.
THE ROLE OF EQA IN HARMONIZATION OF REFERENCE INTERVALS IN CROATIA LABQUALITY DAYS, 2015, HELSINKI FINLAND Department of Medical Biochemistry and Laboratory.
Guidance for water quality intercalibration
BOHB APCCB LABORATORY USE OF THE ABBOTT MEDISENSE METER FOR BETA-HYDROXYBUTYRATE MEASUREMENT. GRD Jones, A Screnci, P Graham Chemical Pathology,
IFCC EQAS for Reference Laboratories ________________________________________
Quality Assessment 2 Quality Control.
1. Dr. Adel M. Assiri Faculty of Medicine Department of Biochemistry Umm Al-Qura University.
Vall d’Hebron Servei de Microbiologia External quality assurance for drug susceptibility testing of M. tuberculosis against second line anti-tuberculosis.
08/ COMMON SNAILS By Dietmar Stöckl STT Consulting.
QUALITY ASSURANCE Reference Intervals Lecture 4. Normal range or Reference interval The term ‘normal range’ is commonly used when referring to the range.
Influence of Analytical Bias and Imprecision on Guideline-Driven Medical Decision Limits Per Hyltoft Petersen Hyltoft Petersen P, Klee GG. Influence of.
Conformity Assessment and Accreditation Mike Peet Chief Executive Officer South African National Accreditation System.
Prof. of Clinical Chemistry, Mansoura University.
The concept for a network of national Reference Laboratories for high risk IVDs – Results of the working group meeting 1.
Andrew Attia, Miriam Bishay & Carly Sterling September 15, 2015 Diagnostic Blood Testing PHM142 Fall 2015 Instructor: Dr. Jeffrey Henderson.
Quality Control and Patient Risk Curtis A. Parvin, Ph. D
Dr. Peter Bode, Associate Professor Delft University of Technology, Interfaculty Reactor Institute Mekelweg JB Delft, The Netherlands Tel
BNPCV APCCB BIOLOGICAL VARIATION FOR N-TERMINAL PRO-BNP Jones GRD 1, Bennett A 2,3, Boscato L 1, Macdonald P 2, Brien J 3. Departments of Chemical.
BMTA July 2005: 1 Valid Analytical Measurement Studies of Proficiency Testing scheme performance S Ellison LGC Limited, Teddington The work described in.
All you wanted to know about laboratory tests, but were afraid to ask By Dietmar Stöckl STT Consulting The informed patient meeting the laboratory.
Update on the P. falciparum Standard & Replacement of the HBV & HCV International Standards Sally Baylis, NIBSC SoGAT XIX.
Evaluation of the Presage™ ST2 ELISA Jun Lu 1, David G. Grenache 1,2 1 ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 2 Department.
Statistics & graphics for the laboratory 25 Biological variation Introduction Estimation (ANOVA application) Index-of-individuality Comparison of a result.
Meeting of ICAR Reference Laboratory Network, 16 June ICAR Session Niagara Falls ICAR Reference Laboratory Network - 4 th Meeting, Niagara.
For further information please contact: ERNDIM QUANTITATIVE ORGANIC ACIDS IN URINE: HIGHLIGHTS AND RECOMMENDATIONS Martens.
Verification of qualitative methods
VERIFICATION OF REFERENCE INTERVALS GIAN LUCA SALVAGNO, MD, PhD University of Verona, Italy.
Quality control & Statistics. Definition: it is the science of gathering, analyzing, interpreting and representing data. Example: introduction a new test.
Mauro Panteghini Who, what and when to do in validation/verification of methods.
PT/External Quality Assessment Principles & Concepts Dr Marjan R. Farzami
Quality Assessment of MFA’s evaluations Rita Tesselaar Policy and operations Evaluation Department Netherlands Ministry of Foreign Affairs.
Quality Control: Analysis Of Data Pawan Angra MS Division of Laboratory Systems Public Health Practice Program Office Centers for Disease Control and.
Estimating measurement uncertainty
TTB M (04/2010) Methods Criteria and Performance Patricia Nedialkova, Ph.D., TTB International Wine Technical Forum May 7, 2015 Prepared for the.
11 BMTA, Teddington, 20 th October 2015 Dr Chris Harrington Deputy Director SAS Trace Element Laboratory, Surrey Pathology Services (SPS) External Quality.
CURRENT SITUATION AND PROBLEM OF IQC AND EQC IN INDONESIA
Horace Gumba KEMRI-Wellcome Trust Research Programme
Pakistan Society Of Chemical Pathologists Zoom Series of Lectures ZT 24. Quality Managent 1 Brig Aamir Ijaz MCPS, FCPS, FRCP (Edin), MCPS-HPE HOD and.
P.J. Showell1, E.A. Lynch1, H.D. Carr-Smith1, A.R. Bradwell2
BANGLADESH UNIVERSITY OF ENGINEERING & TECHNOLOGY , BANGLADESH
Comparison of Nurse Mentor and Instructor
Robust Measurement of Branched Chain Amino Acids on the Vantera Clinical Analyzer and the Clinical Association of NMR-Measured Valine with Type 2 Diabetes.
بسم الله الرحمن الرحیم.
محدوده مرجع در آزمايشگاه تشخيص طبی Reference Interval
محدوده مرجع در آزمايشگاه تشخيص طبی در بخش بیوشیمی Reference Interval
Quality Assurance Reference Intervals.
Italian EQA study for HCV RNA, HIV RNA and HBV DNA G. M. Bisso, K
SAMPLE DESIGN: HOW MANY WILL BE IN THE SAMPLE—DESCRIPTIVE STUDIES ?
Quality Control Lecture 3
PT-WFD - The self committed Network of PT providers
Intermediate methods in observational epidemiology 2008
Philippe QUEVAUVILLER
A proposal for Derivation of Analytical Quality Goal for Precision
Bland-Altman plots for intra and inter observer variability for the assessment of total Agatston score for patients with a calcium score less than 1000.
STANDARDISATION, HARMONISATION AND TRACEABILITY
Presentation transcript:

Josiane Arnaud1, Jean-Philippe Weber2, Are desirable quality specifications for evaluating laboratory performances in external quality assessment schemes for copper, zinc and selenium in human serum or plasma attainable? Josiane Arnaud1, Jean-Philippe Weber2, Cas W Weykamp3, Patrick J Parsons4, Olav Mazarrasa5, Antonio Menditto6, Marina Patriarca6, Andrew Taylor7 1 – CHU de Grenoble, France; 2 – Institut National de Santé Publique, Québec, Canada; 3 – Queen Beatrix Hospital, Winterswijk, The Netherlands; 4 – New York State Department of Health, Albany, USA; 5 – Centro de Seguridad y Salud en el Trabajo, Santander, Spain; 6 – Istituto Superiore di Sanità, Rome, Italy; 7 – University of Surrey, Guildford, UK

Thematic Network Organizers of external quality assessment / proficiency testing schemes related to occupational and environmental laboratory medicine Conflicting conclusions at the participant level Different possibilities for organisation, statistical analyses, participant monitoring ISO 13528, EURACHEM, EQALM, ILAC, IUPAC, IFCC… TE EQAS consensus procedures Coordonnateur : Dr Andrew Taylor (University of Surrey)

Uncertainty evaluation Calculated (sample dependent)  Consensus participants’ standard deviation (after exclusion of outliers) Reference laboratory standard deviation Metrological laboratory standard deviation Defined (analyte and medium)  Defined by the law  Quality specification = QS

Quality specification criteria Must take into account:  Analytical possibilities (concentration range)  Expertise of laboratory  Biological individual variability  Overlaps between « normal » and pathological states Law recommendations (if any) EQAS role: laboratory licensing or education

Quality specification calculation Horwitz Equation : 0.02C0.8495 Fraser’s equation: 0.25(CV2intra+CV2inter)1/2 + z(0.5xCVintra) (100/4)* ref. range/mean ref range Tonk equation: Quality specification calculation

Quality specification calculation Fraser’s equation, 1999 Desirable total error % = 0.25 (CV2intra + CV2 inter)1/2 + z (0,5 CV intra) CV intra = intra-individual variability CV inter = inter individual variability z = 1.65 for a 95% probability level. Fraser CG. Scand J Clin Lab Invest 1999;59:487-90.

Individual Variability (published datas* on healthy populations) Copper Zinc Selenium intra 2.3 – 9.2 4.4 - 9.4 3.6 - 5.1 inter 12.2 - 16.9 4.1 - 10.7 13.0 - 13.9 *References Gallagher SK, et al. Clin Chem 1989;35:369-73 Gonzalez-Revalderia J, et al. Clin Chem 1990;36:2140-1. Sabban A. Thesis, University of Surrey, 2005. Lacher DA, et al. Clin Chem 2005;51:450-2.

Cu Zn Se quality specification Minimal QS*,% 12 15 Desirable QS*,% 8 10 Optimal QS*,% 4 5 *Desirable total error % = 0.25(CV2intra+CV2inter)1/2 + z(0.5xCVintra) Optimal = 0.5 Desirable. Minimal = 1.5 Desirable.

Uncertainty and concentration 0,00 10,00 20,00 30,00 40,00 50,00 60,00 70,00 Zn CV % (participants) Zn target concentration, µmol/l (consensus after exclusion of outliers) Minimal QS: 1.2 µmol/l  8 µmol/l Desirable QS: 0.8 µmol/l Optimal QS: 0.4 µmol/l Zinc

Cu Zn Se quality specification (low concentrations) Cutoff, µmol/l 7 8 0.6 Minimal QS, µmol/l 0.84 1.20 0.072 Desirable QS, µmol/l 0.56 0.80 0.048 Optimal QS, µmol/l 0.28 0.40 0.024

Quality specification evaluation

Method Sample concentration selection Cu, µmol/l 3.5 9 15 20 30 Nb participants 82 128 184 239 168 Se, µmol/l 0.25 0.5 1.0 1.5 3.0 69 101 95 143 97 Zn, µmol/l 8 144 156 228

Z-score = (xi-X)/(QS/2) Method Z-score Z-score = (xi-X)/(QS/2) where xi = participant’s result, X = target value QS = quality specification. QS is divided by two to be consistent with the ISO guide which states that performance is Satisfactory if Z-score 2 Unacceptable if Z-score > 2 Allows comparison of  samples

Cu, Zn and Se Z-score percentiles Results Cu, Zn and Se Z-score percentiles Minimal Cu Minimal Se Minimal Zn Desirable Cu Desirable Se Desirable Zn Optimal Cu Optimal Se Optimal Zn -8,00 -6,00 -4,00 -2,00 0,00 2,00 4,00 6,00 8,00 10,00 Z-score, percentiles

Unacceptable results, % Cu &Se 30 3 20 1.5 15 1.0 9 0.5 3.5 0.25 Concentrations, µmol/l Zn 20 15 8 % of Z-score>2

Discussion Participant value Target value TE EQAS Conclusion 22.0 20.0 NY Satisfactory BE Unacceptable Discussion

Discussion Fraser’s equation, 1999 Desirable total error % = 0.25 (CV2intra + CV2 inter)1/2 + z (0,5 CV intra) Limitations Literature agreement Pathological states Analytical possibilities Advantages Biological variability Sample and run independent

Discussion Imprecision, % Cu Se Zn Fraser’s QS, optimal (this work) 4 5 Fraser’s QS, desirable (this work) 8 10 Fraser’s QS, desirable (Ricos, plasma) 12 14.5 11 Fraser’s QS, desirable (Ricos, serum) 14 13.5 Analytical variability (ref. laboratories) 4.3 6.0 5.8 Analytical variability (participants) 10.1 17.4 14.7 Fraser’s QS, minimal (this work) 15 Ricos C et al. Scand J Clin Lab Invest 1999;59:491-500

Discussion : Comparison to current QS – Cu & Zn CA, FR : 0.08x + 0.4 BE, UK : 0.031x + 0.9 IT : 0.075x + 0.6 Cu NY : 0.1x Zn NY : 0.15x 0,00 1,00 2,00 3,00 4,00 5,00 6,00 7,00 8,00 10 20 30 40 50 60 Concentration, µmol/l QS, µmol/l Minimal Zn QS : 0.15x if >8 µmol/l Minimal Cu QS : 0.12x if >7 µmol/l

Discussion : Comparison to current QS - Se IT QS = 0.06x + 0.06 BE, CA, FR, UK QS = 0.06x + 0.07 NY QS = 0.2x 0,00 0,10 0,20 0,30 0,40 0,50 0,60 0,70 0,80 0,90 1,00 2,00 3,00 4,00 5,00 Se concentration, µmol/l Se QS, µmol/l Minimal QS = 0.12x if x> 0.6 µmol/l

Conclusion Patient’s diagnosis and follow-up Reliable determinations Participant laboratory TE EQAS objective quality specification Analytical performances Individual variabilty harmonisation

Thank you for your attention