Exercises 4.

Slides:



Advertisements
Similar presentations
Introduction to Chromatography
Advertisements

INTRODUCTION TO CHROMATOGRAPY
Introduction to Chromatographic Separations Due to lack of analytical specificity, separations are often necessary Chromatography is about separations.
Chem. 31 – 4/15 Lecture. Announcements I Exam 2 – Hope to have graded by next Monday Lab Reports –AA report now due Monday, 4/20 (best to try to complete.
The geometry of capillary columns is fairly simple, consisting of length, internal diameter, and stationary phase thickness. Nevertheless, there are endless.
Chapter 26 An Introduction to Chromatographic Separations
Chem. 133 – 5/14 Lecture. Announcements I Lab Stuff Pass out Peer Review Assignments See Term Project Handout for Format of Poster Today should be check.
Chem. 133 – 5/5 Lecture. Announcements Lab Report 2.4 due Thursday – can turn in today for reduction of late penalties Term Project Progress Report –
Chem. 133 – 5/7 Lecture. Announcements I Exam 3 on Tuesday (will give summary of material to know later) Format will be similar to other exams I will.
Lecture 35 12/2/05. Mobile phase Volume flow rate mL solvent/minute travle Linear flow rate cm of column length/min.
Lecture 15 Chromatography Introduction and Plumbing Harris Ch. 23.
Chromatography General
Intro to Chromatographic Separations Chap 26. Originally based on separation and identification by color Originally based on separation and identification.
Effect of Packing Particle Size on Plate Height. Resolution between 2 adjacent peaks.
Principles of Chromatography. Chromatography is the most powerful tool for separating & measuring the components of a complex mixture. Quantitative &
Created with MindGenius Business 2005® Performance parameters (Advanced) Theoretical Plate Model Performance parameters (Advanced) Theoretical Plate Model.
Basics of Chromatography Separation of Analytes Type of Equilibrium.
Chromatography Separates components in mixture: Based on - polarity
Chromatographic separations Chapter 26 The “stuff” you do before you analyze a “complex” sample.
1.1 General description - Sample dissolved in and transported by a mobile phase - Some components in sample interact more strongly with stationary phase.
LECTURE 9 CHROMATOGRAPHIC SEPARATIONS The “stuff” you do before you analyze a “complex” sample.
Chapter 21 Principles of Chromatography. Chromatography is the most powerful tool for separating & measuring the components of a complex mixture. Quantitative.
CHROMATOGRAPHY. The general name given to methods by which two or more compounds in a mixture are physically separated by distributing themselves between.
Mass transfer in Gas Chromatography
Chem. 230 – 9/23 Lecture.
INTRODUCTION TO CHROMATOGRAPY
Lecture 8 Van Deemter Equation!. Resolution Describes how well 2 compounds are separated Rs = 1 4 N 1/2 (  -1) k’ 1+k’ ( ) efficiency selectivity retention.
CHROMATOGRAPHY Chromatography basically involves the separation of mixtures due to differences in the distribution coefficient.
Introduction to Chromatography Definition Chromatography is a separation technique based on the different interactions of compounds with two phases, a.
Alternative Approach: Rate Theory Based on the principal of the summation of variances. σ T 2 = σ σ σ … So all contributions to the broadening.
1 An Introduction to Chromatographic Separations Lecture 34.
Gas Chromatography Lecture 38.
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC). HIGH PERFORMANCE LIQUID CHROMATOGRAPHY High Performance Liquid Chromatography (HPLC) is one of the most.
Introduction to Chromatography. Introduction Chromatography permit the scientist to separate closely related components of complex mixtures. In all chromatographic.
Separation Techniques
Ch 21 – Principles of Chromatography and Mass Spectrometry Ch 22 – Gas and Liquid Chromatography.
Prof. aza Basic gas chromatography prof. aza Department of Pharmacy, Andalas University STIFI Perintis, Padang STIFAR, Pekan Baru STIFI Bhakti Pertiwi,
HPLC 1. Introduction 1.Introduction  INSTUMENTAL ANALYSIS  PRACTICAL 213 PHC  HPLC.
 The word “chromatography” originated from two Greek words, chroma which means “color” and graphy which means “writing”.  Chromatography was founded.
Chem. 133 – 5/3 Lecture. Announcements Lab – Term Project Progress Report Due Today – Last Assignments: Term Project Poster and Peer Review Grading (Friday,
Instrumental Analysis (I) HPLC Tutorial 8. Graded presentation Students in groups of 4-5 individuals are asked to prepare a presentation (weight=5% of.
Instrumental Analysis (I)  HPLC Tutorial #7 PHCMt561 – 5 th Sem. Pharm.
1 Principles of Chromatography Chap Analytical Separations and Chemical Problem Solving If you, a researcher of a food company are asked to find.
Downloaded from کروماتوگرافی CHROMATOGRAPHY Downloaded from
Experiments in Analytical Chemistry
1.1 General description - Sample dissolved in and transported by a mobile phase - Some components in sample interact more strongly with stationary phase.
Chem. 133 – 5/2 Lecture.
Chem. 133 – 5/9 Lecture.
Chemical Ideas 7.6 Chromatography.
Lecture 5 Non-ideal conditions.
Chem. 133 – 5/11 Lecture.
An Introduction to Chromatographic Separations
An Introduction to Chromatographic Separations
Exercises 5.
Optimization & Monitoring.
Lecture 4 Band broadening.
Chem. 31 – 11/13 Lecture.
Efficiency, selectivity and resolution
Intro to Chromatography CHM 480 Suroviec
Chem. 133 – 5/4 Lecture.
Chromatography Introduction and Plumbing Harris Ch. 23
An Introduction to Chromatographic Separations
Chromatographic performance
Separations.
Retention and phase distribution
Principle of separation of different components:
Chromatography is a technique in which the components of a mixture are separated based on The differences in rates at which they are carried through a.
Chromatography is a technique in which the components of a mixture are separated based on The differences in rates at which they are carried through a.
Theory of Band Broadening
Presentation transcript:

Exercises 4

Exercise 4.1 x) List the variables that lead to zone broadening in chromatography

The van Deemter equation for plate height, H: Exercise 4.1 x) List the variables that lead to zone broadening in chromatography The van Deemter equation for plate height, H: Multiple paths, A, (”eddy diffusion”)

The van Deemter equation for plate height, H: Exercise 4.1 x) List the variables that lead to zone broadening in chromatography The van Deemter equation for plate height, H: Multiple paths, A, (”eddy diffusion”) Longitudinal diffusion, B

The van Deemter equation for plate height, H: Exercise 4.1 x) List the variables that lead to zone broadening in chromatography The van Deemter equation for plate height, H: Multiple paths, A, (”eddy diffusion”) Longitudinal diffusion, B The mass transfer term (in mobile phase, CM, and stationary phase, CS)

The van Deemter equation for plate height, H: Exercise 4.1 x) List the variables that lead to zone broadening in chromatography The van Deemter equation for plate height, H: Multiple paths, A, (”eddy diffusion”) Longitudinal diffusion, B The mass transfer term (in mobile phase, CM, and stationary phase, CS) +++ Extra column effects

Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC?

The van Deemter equation for plate height, H: Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC? The van Deemter equation for plate height, H:

The van Deemter equation for plate height, H: Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC? The van Deemter equation for plate height, H: Mass transfer in the mobile phase: r is the column radius k is the retention factor DM is the diffusion coefficient in the mobile phase C M = 1 + 6k + 11 k 2 24( + 1) r D B is i direct function of the diffusion coefficient of the analyte in the mobile phase, DM: B = 2·DM

The van Deemter equation for plate height, H: Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC? The van Deemter equation for plate height, H: Mass transfer in the mobile phase: r is the column radius k is the retention factor DM is the diffusion coefficient in the mobile phase C M = 1 + 6k + 11 k 2 24( + 1) r D B is i direct function of the diffusion coefficient of the analyte in the mobile phase, DM: B = 2·DM B  DM CM  1/DM

The van Deemter equation for plate height, H: Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC? The van Deemter equation for plate height, H: Mass transfer in the mobile phase: r is the column radius k is the retention factor DM is the diffusion coefficient in the mobile phase C M = 1 + 6k + 11 k 2 24( + 1) r D B is i direct function of the diffusion coefficient of the analyte in the mobile phase, DM: B = 2·DM Both the B-term and the C-terms are functions of the diffusion coefficients in the mobile phase. Diffusion in gases is much higher than diffusion in liquids. lower diffusion in liquids gives lower B and higher C, which will shift the minimum in the curve to lower mobile phase velocities.

Exercise 4.2 x) Why does the minimum in a plot of plate height versus mobile phase velocity occur at lower velocity in LC than in GC? A=1, B=1, C=1  uopt = √B/C = 1 A=1, B=0.5, C=2  uopt = √B/C = 0.5 Both the B-term and the C-terms are functions of the diffusion coefficients in the mobile phase. Diffusion in gases is much higher than diffusion in liquids. lower diffusion in liquids gives lower B and higher C, which will shift the minimum in the curve to lower mobile phase velocities.

Exercise 4.3 The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why.

Exercise 4.3 The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why.

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The purnell equation The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) The purnell equation

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) k is in this case very low, making the last factor in the Purnell equation significant

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) We can calculate k for tR around 2 min: tm = 1500 cm / 38 cm/s = 39 s = 0.66 min t′R = 2.00 min – 0.66 min = 1.34 min k = t′R / tM = 1.34 / 0.66 = 2.03 (for 1 um phase)

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) k = 2.03 (for 1 μm phase) All other factors kept constant, if the phase thickness is decreased to 25%, k decreases to 25% (because Kc is the same) k = 2.03 / 4 = 0.51 for (for 0.25 μm phase)

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) k = 2.03 (for 1 μm phase) 2.03 / (1+2.03) = 0.67 k = 2.03 / 4 = 0.51 for (for 0.25 μm phase) 0.51 / (1+0.51) = 0.34

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) What would be the most obvious choice for increasing the resolution with the column with 0.25 μm film?

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) What would be the most obvious choice for increasing the resolution with the column with 0.25 μm film? Decrease the temperature so that k increases to the level it is with the 1 μm film. (But the method would then take more time)

Exercise 4.3 Rs = N 4 α–1 α k(B) 1+k(B) The two chromatograms show a group of solvents separated by GC on two capillary columns with different stationary phase thickness: 0.25 μm and 1.0 μm (left and right plot respectively). Both columns are 15 m long, have internal diameter of 0.32 mm and are coated with the same stationary phase. Temperature and carrier gas settings were also identical. Helium was used as carrier gas in both cases with a linear velocity of 38 cm/s and the oven temperature was 40C. According to the van Deemter equation and the relationship CS  df one should expect that the column with 0.25 μm phase has higher separation efficiency than the column with 1.0 μm phase. But the resolution is obviously much better on the chromatogram to the right. Explain why. Rs = N 4 α–1 α k(B) 1+k(B) What would be the most obvious choice for increasing the resolution with the column with 0.25 μm film? Reducing the column length and keeping k at the same level would in this case be a better way to reduce retention time. Reducing the length to half would decrease retention times to half, but the resolution would only be decreased to approximately 71% (√0.5=0.71)

Exercise 4.4 C B Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C B 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L B 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L The Purnell equation says that Rs  N (α and k is unaffected) B Rs = N 4 α–1 α k(B) 1+k(B) N  L and Rs  N  Rs  L 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L The Purnell equation says that Rs  N (α and k is unaffected) B Rs = N 4 α–1 α k(B) 1+k(B) N  L and Rs  N  Rs  L  L  Rs2 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L The Purnell equation says that Rs  N (α and k is unaffected) B Rs = N 4 α–1 α k(B) 1+k(B) N  L and Rs  N  Rs  L  L  Rs2  (L(new) / L(old)) = (Rs(new)/ Rs(old))2 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L The Purnell equation says that Rs  N (α and k is unaffected) B Rs = N 4 α–1 α k(B) 1+k(B) N  L and Rs  N  Rs  L  L  Rs2  (L(new) / L(old)) = (Rs(new)/ Rs(old))2  L(new) = (Rs(new)/ Rs(old))2 · L(old) 1.16 1.07 14.1 13.3

Exercise 4.4 C B Column length, L = 24.7 cm None of the terms in the van Deemter equation are affected by change in column length, so plate height, H, is the same. Column length, L = 24.7 cm Rs = 0.72 (from exercise 3.3) C L N L H H =  N =  N  L The Purnell equation says that Rs  N (α and k is unaffected) B Rs = N 4 α–1 α k(B) 1+k(B) N  L and Rs  N  Rs  L  L  Rs2  (L(new) / L(old)) = (Rs(new)/ Rs(old))2  L(new) = (Rs(new)/ Rs(old))2 · L(old) = (1.5 / 0.72)2 · 24.7 cm = 107 cm 1.16 1.07 14.1 13.3

Exercise 4.4 2 4 6 8 10 12 14 16 18 20 22 min tR=5.4 min wb=0.41 min 2 4 6 8 10 12 14 16 18 20 22 min

Exercise 4.4 Both tM and t’R will increase proportionally with the column length  tR increase proportionally with the column length tR=5.4 min wb=0.41 min tR=14.1 min wb=1.16 min tR=13.3 min wb=1.07 min tR=21.6 min wb=1.72 min 2 4 6 8 10 12 14 16 18 20 22 min

Exercise 4.4 Both tM and t’R will increase proportionally with the column length  tR increase proportionally with the column length tR(old) = 21.6 min tR(new) = tR(old) · L(new) / L(old) tR(new) = 21.6 min · (107 / 24.7) = 93.8 min tR=5.4 min wb=0.41 min tR=14.1 min wb=1.16 min tR=13.3 min wb=1.07 min tR=21.6 min wb=1.72 min 2 4 6 8 10 12 14 16 18 20 22 min

Exercise 4.5

Exercise 4.5

Exercise 4.5 The optimal mobile phase velocity is found where A + B/u + C∙u has a local minimum, meaning that the derivative is 0.  uopt = √B/C = 0.0014 / 0.0031 = 0.67 mm/s

Exercise 4.5 The optimal mobile phase velocity is found where A + B/u + C∙u has a local minimum, meaning that the derivative is 0.  uopt = √B/C = 0.0014 / 0.0031 = 0.67 mm/s Note the difference to GC where optimal velocities are typically around 20-35 cm/s

Exercise 4.5 B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u H (mm) 0.020 H = A + + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u H (mm) 0.020 H = A + + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u H (mm) 0.020 H = A + + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u H (mm) 0.020 H = A + + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u H (mm) 0.020 H = A + + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 uopt B H = A + u + C∙u 0.0014 H = 0.0074 + u + 0.0031∙u 0.020 H = A + uopt + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0074 + + 0.0031∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 A term: Eq (14) CM term in packed columns Eq (16) A(new) = A(old) · (dp(new) / dp(old)) = 0.0074 · 3/5 = 0.0044 mm Needed for 4.5d C(new) = C(old) · (dp(new) / dp(old))2 = 0.0031 · (3/5)2 = 0.0011 mm uopt = B/C = 0.0014 / 0.0011 = 1.1 mm/s

Exercise 4.5 uopt B H = A + u + C∙u 0.0014 H = 0.0044 + u + 0.0011∙u 0.020 H = A + uopt + C∙u 0.018 0.016 0.0014 u 0.014 H = 0.0044 + + 0.0011∙u H (mm) 0.012 0.010 0.008 0.006 0.004 0.002 0.000 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 Velocity, u (mm/s)

Exercise 4.5 5 μm 3 μm H (mm) Velocity, u (mm/s) 0.000 0.002 0.004 0.006 0.008 0.010 0.012 0.014 0.016 0.018 0.020 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3 μm 5 μm H (mm) Velocity, u (mm/s)

Exercise 4.6 2 4 6 8 10 12 14 16 18 20 22 24 tR=4.11 min wb=1.95 min 2 4 6 8 10 12 14 16 18 20 22 24

Exercise 4.6 Eq (8) N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 tR=4.11 min wb=1.95 min tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min 2 4 6 8 10 12 14 16 18 20 22 24 Eq (8)

Exercise 4.6 N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 16 18 20 22 24 tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min tR=4.11 min wb=1.95 min The peak widths are almost identical. Normally they would increase with retention under isocratic conditions if the peak broadening were caused by the column and could be explained by the van Deemter equation.

Exercise 4.6 N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 16 18 20 22 24 tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min tR=4.11 min wb=1.95 min The peak widths are almost identical. Normally they would increase with retention under isocratic conditions if the peak broadening were caused by the column and could be explained by the van Deemter equation.

Exercise 4.6 N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 16 18 20 22 24 tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min tR=4.11 min wb=1.95 min The peak widths are almost identical. Normally they would increase with retention under isocratic conditions if the peak broadening were caused by the column and could be explained by the van Deemter equation. The contrubution from extra-column effects are independent of retention time.

Exercise 4.6 N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 16 18 20 22 24 tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min tR=4.11 min wb=1.95 min The peak widths are almost identical. Normally they would increase with retention under isocratic conditions if the peak broadening were caused by the column and could be explained by the van Deemter equation. The contrubution from extra-column effects are independent of retention time. The pattern therefore indicates a large extra-column effect.

Exercise 4.6 N = 71 N = 219 N = 673 N = 1790 2 4 6 8 10 12 14 16 18 20 22 24 tR=7.51 min wb=2.03 min tR=13.30 min wb=2.05 min tR=22.21 min wb=2.10 min tR=4.11 min wb=1.95 min Note that if you use solvent programming (LC) or temperature programming (GC) it is normal to have equal peak widths – but with isocratic/isothermal conditions

Exercise 4.6 C = 1 + 6k + 11 k 24( + 1) r D 2k d C = 3( k + 1) D Note that the van Deemter equation is not completely independent from retention time since CM and CS is dependent on k C M = 1 + 6k + 11 k 2 24( + 1) r D 2k d 2 C = f S 3( k + 1) 2 D S Retention factor, k