GC/MS A Very Useful Analytical Tool

Slides:



Advertisements
Similar presentations
Mass Spectrometry.
Advertisements

GAS CHROMATOGRAPHY ENVE 202 Dr. Aslıhan Kerç.
FC-MS from Teledyne Isco CombiFlash ® a Name You Can Rely On.
Mass Spectrometry.
Chapter 14 Mass Spectroscopy.
In carbon-13 NMR, what do the number of peaks represent?
June 10, Introduction to LC/GC Instrumentation CM3007 Introduction to Separation Science Analytical Chemistry Lecture #1.
1 Chapter 24 GC Gas Chromatography. 2 GC Mechanism of separation is primarily volatility. Difference in boiling point, vapor pressure etc. What controls.
Molecular Mass Spectrometry
Gas Liquid Chromatography
CHAPTER 29 Supercritical Fluid Chromatography The mobile phase is a supercritical fluid (a fluid above its critical T and critical pressure) Supercritical.
Mass Spectrometry Mass spectrometry (MS) is not true “spectroscopy” because it does not involve the absorption of electromagnetic radiation to form an.
Principles of Chromatography. Chromatography is the most powerful tool for separating & measuring the components of a complex mixture. Quantitative &
Introduction to Mass Spectrometry (MS)
Mass Spectrometry Chapter 14 Chapter 14.
GCMS: gas chromatography and mass spectrometry
Chemical Analysis. Analytical Techniques When chemical evidence is collected at a crime scene, it must be run through an instrument. These instruments.
Chapter 27 Gas Chromatography 1. Principles
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 22 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Magnet Analytical Chemistry Unit 4
Introduction to Mass Spectrometry (MS)
1 Chemical Analysis by Mass Spectrometry. 2 All chemical substances are combinations of atoms. Atoms of different elements have different masses (H =
Created with MindGenius Business 2005® Mass Spectrometry Mass Spectrometry.
Analytical Separations
Basic Gas Chromatography. History Separation of dyes by Runge Separation of plant pigments by Tswett Theoretical gc (Martin & Synge)
Mass Spectroscopy Introduction.
HPLC.
Chapter 28 High Performance Liquid Chromatography.
Chemistry 2412 L Dr. Sheppard
GAS LIQUID CHROMATOGRAPHY Principles Partition of molecules between gas (mobile phase) and liquid (stationary phase).
Ch 21 – Principles of Chromatography and Mass Spectrometry Ch 22 – Gas and Liquid Chromatography.
1 Chapter 12 Mass Spectrometry ( 質譜 ) Organic Chemistry, 7 th Edition L. G. Wade, Jr. 藥學系 藥物分析科 許秀蘊 教授
Instrumental Analysis
Mass Spectrometry.
Mass Spectrometry Molecular weight can be obtained from a very small sample. It does not involve the absorption or emission of light. A beam of high-energy.
1 Principles of Chromatography Chap Analytical Separations and Chemical Problem Solving If you, a researcher of a food company are asked to find.
Reading assignment: section 26E(p781) Chapter 26 # 2, 3, 14, 15, 16 Chapter 27 # 7(a,d,f), 22, 23, 24, 25.
Bioseparation II Chromatography Techniques. Chromatography Most widely used purification technique used for biomolecules. Most widely used purification.
High Performance Liquid Chromatography Presented by Dr. Kamal Modi 2 nd Year Resident.
High Performance Liquid Chromatography
Overview Determining Identity Quantitative Analysis
Determining the Structure of an Organic Compound
A guide for A level students KNOCKHARDY PUBLISHING
Components of Mass Spectrometer
GAS CHROMATOGRAPH. GAS CHROMATOGRAPH Principle: Which type of compounds can analyse? Volatile Thermostable.
Introduction to Mass Spectrometry (MS)
Mass Spectrometry Courtesy
Metabolomics Part 2 Mass Spectrometry
A guide for A level students KNOCKHARDY PUBLISHING
Introduction to Mass Spectrometry (MS)
INTRODUCTION TO SPECTROPHOTOMETRY
Determining the Structure of an Organic Compound
Chromatographic separation
Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample. The amount of light absorbed by.
Lecture 22 Introduction to Mass Spectrometry Lecture Problem 7 Due
12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy Based on McMurry’s Organic Chemistry, 7th edition.
12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy Based on McMurry’s Organic Chemistry, 7th edition.
BASICS OF SUPERCRITICAL FLUID CHROMATOGRAPHY
12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy Based on McMurry’s Organic Chemistry, 7th edition.
Chemistry - Option A : Modern Analytical Chemistry
Mass Spectrometry. 3 The GC-MS => A mixture of compounds is separated by gas chromatography, then identified by mass spectrometry.
Chromatography Daheeya Alenazi.
Principle of separation of different components:
12. Structure Determination: Mass Spectrometry and Infrared Spectroscopy Based on McMurry’s Organic Chemistry, 7th edition.
Mass Spectrometry THE MAIN USE OF MS IN ORG CHEM IS:
Department of Chemistry St. Joseph’s College (Autonomous)
A guide for A level students KNOCKHARDY PUBLISHING
GAS CHROMATOGRAPHY.
CHROMATOGRAPHY.
Determining the Structure of an Organic Compound
Presentation transcript:

GC/MS A Very Useful Analytical Tool Introduction to Gas Chromatography Introduction to Mass Spectroscopy GC/MS

Chromatography The separation of a mixture by distribution of its components between a mobile and stationary phase over time mobile phase = solvent stationary phase = column packing material

Milestones in Chromatography 1903 Tswett - plant pigments separated on chalk columns 1931 Lederer & Kuhn - LC of carotenoids 1938 TLC and ion exchange 1950 reverse phase LC 1954 Martin & Synge (Nobel Prize) 1959 Gel permeation 1965 instrumental LC (Waters)

Purpose of Chromatography Analytical - determine chemical composition of a sample, and identify compounds via comparison of the retention time of known molecules Preparative - purify and collect one or more components of a sample

Chromatography Chromatogram - Detector signal vs. retention time or volume time or volume Detector Signal 1 2

Definitions Mobile phase - phase that moves through chromatograph In GC - carrier gas is the mobile phase Stationary phase - column; phase that is stationary in chromatograph Bonded phase - reactive groups imparted to stationary phase in order to achieve selectivity

Classification of Chromatography Mobile phase Attractive forces

Types of Chromatography Based on Mobile Phase Gas - Gas chromatography (GC) 1951 Martin and James (fatty acids) Liquid - Liquid chromatography (LC) 1964 Horvath (Yale) instrument 1966 Horvath and Lipsky (nucleic acid components) Supercritical fluid - Supercritical fluid chromatography (SFC) 1958 Lovelock (Yale)

Classification based on Mobile Phase Gas Chromatography Pyrolysis GC - heat solid materials to 500 - 10000C so they decompose into gaseous products Gas - solid Gas - liquid Stationary Phase Sample MUST be volatile at temperatures BELOW 3500C

Classification based on Mobile Phase Liquid chromatography (LC) Column (gravity flow) Thin layer (adsorption) High performance (pressure flow)

Types of Chromatography Based on Attractive Forces Adsorption - for polar non-ionic compounds Ion Exchange - for ionic compounds Anion - analyte is anion; bonded phase has positive charge Cation – analyte is cation; bonded phase has negative charge Partition - based on the relative solubility of analyte in mobile and stationary phases Normal – analyte is nonpolar organic; stationary phase MORE polar than the mobile phase Reverse – analyte is polar organic; stationary phase LESS polar than the mobile phase Size Exclusion - stationary phase is a porous matrix; sieving

Detectors UV-vis Refractive Index (RI) Mass spectrometry (MS) Electrochemical (EC) amperometric NMR - novel

Gas Chromatography Mobile phase is gas, the carrier Stationary phase is liquid or solid The analyzed compounds must be volatile, and stable under the column condition, primarily stable while heated.

Principle of GC Theoretically, the GC is still a process of extraction, one phase is gas, the other is the stationary phase The separation efficiency of GC depends on the volatility of molecule and the interaction between the stationary phase and molecules

GC Liquid Phase Low volatility High boiling point Chemically unreactive Examples: 1-squalene Tetrahydroxyethylenediamine Carbowax (polyethylene glycol) Silica gel coated with silane or hydrocarbon

Phases Here are some of the commonly used phases. They range in polarity from non-polar such as low polarity DMS to the higher polarity of mixtures with DPS with any combination available. There are specialty phases with very high polarity such as cyanopropylphenyl siloxane or trifluoropropyl methyl siloxane for fluorinated compounds. Some of these columns are used for separation confirmation such as the cyanopropylphenyl siloxane column for method 551.1.

Modes of GC Separation Capillary (open tubular) Packed Inner wall modified with thin (1 m) film of liquid 0.3 - 0.5 mm ID; 10 - 50 m length Packed Solid particles either porous or non-porous coated with thin (1 m) film of liquid 1 - 8 mm ID; 1 - 10 m length

Modes of GC Separation Isothermal (GC) Programmed temperature (GC) Raising column temperature (GC) Decreases retention time Sharpens peaks

Possible Detector High sensitivity - high ratio of response/concentration Universal or selective response selectivity - ability to distinguish between species Rapid response Linearity - concentration range over which signal is proportional to concentration Stability with respect to noise (baseline noise) and time (drift)

Detectors for GC Destructive Non-destructive Mass Spectral (CI/EI) Flame Ionization (FID) Nitrogen-Phosphorus (NPD) Flame Photometric (FPD) Electrolytic Conductivity (Hall/ELCD) Non-destructive Thermal Conductivity (TCD) Electron Capture (ECD) Photo Ionization (PID) Fourier Transfer Infrared (FTIR)

Detectors for GC Electron capture (ECD) Thermal conductivity (TCD) radioactive good for X-, NO2- and conjugated Thermal conductivity (TCD) change in resistance of heated wire Flame ionization (FID) destruction of combustible sample in flame produces measurable current Fourier transform infrared (FTIR) Mass spectrometry (MS)

Instrumentation for GC Carrier gas N2, He, H2 Injector Column Detector Computer oven

Gas Chromatograph Injection Port Detector Capillary Column Data System or Recorder Block diagram of GC. Carrier Gas Supply Oven

Columns Packed Capillary Packed - As suggested by the term, it is filled with a coated inert solid support such as fire brick, alumina, and graphite with a specific mesh size. The coatings are called phases and for best results are chemically bonded to the support. Chemical bonding provides for longer column life and less bleeding (major source of background noise) contributing to lower sensitivity. Column dimensions 1/8” - 1/4” ID x up to about 6’ using glass or stainless steel. Advantages - higher capacity (higher conc). Disadvantages: low resolution and low S/N. Capillary - Here the phase (film) is coated on the inside diameter of the capillary wall with film thickness range of 0.1 to 5μ where the ticker film provides for better resolution but also allows for more bleed. Typical dimensions .25mm - .53mm ID x up to 60m made of fused silica coated with polyamide. Advantages: high resolution and better S/N. Disadvantages: low capacity and cost.

The Gas Chromatograph

Separation of a Mixture

Gas Chromatogram Highest b.p. Lowest b.p. Retention time

Triangulation of a Peak

Sample Percentage Composition Calculation

Gas Chromatography: Results In a modern gas chromatography instrument, the results are displayed and analyzed using a computerized data station. It is no longer necessary to calculate peak areas by triangulation; this determination is made electronically.

Hexanes (mixture of isomers) 68-70oC Cyclohexane 80oC Heptane 98oC Toluene 110oC Mixture separates by distillation according to the boiling point. Compounds with the lower boiling point come off first! The same is true on the gas chromatographic column; the lower boiling compound comes off first!

1) hexane has the lowest retention time 2) toluene has the highest retention time The four compounds come off in the order of increasing boiling point. Hexane is actually a mixture of three compounds. It is usually called “hexanes”

Retention Times vs Response Factors Component Retention Time (min) Response Factor Hexanes (mixture of isomers) 3.054 1.022 Cyclohexane 3.491 1.133 Heptane 3.812 1.000 Toluene 4.331 1.381 NOTE: These values are for illustration purposes. Your actual values will be different!

Examples of GC Spectra Note peaks 15, 16 17 & 18 on the DB-5 column and note the same peaks on the DB-1701 column. This shows the need for confirmatory columns (columns with different phases) so that separation of the compounds can be verified.

Separation by Different Column Note peaks 1 and 2 using the different columns. The separation is achievable because the compounds are different. Also note peaks 4 and 5 and 11 and 12. Even using different columns, positional isomers, o-, m- and p- cresol are difficult to separate. Supelco® PTE-5 Supelco® SPB-50

Usage and Limitation of GC Quantitative Analysis Qualitative Analysis (retention time) High Separation Efficiency Cannot determine the actual structure even if a standard compound is used for standard

Mass Spectroscopy Molecules after ionized, carry charge Ionized particles move in either electric or magnetic field The distance the charged particles can move depends on the strength of magnetic (or electric field) and the mass-to-charge ratio Record of number of particles detected as a function of mass-to-charge ratio is the mass spectrum

Mass Spectrometry Molecular weight can be obtained from a very small sample. It does not involve the absorption or emission of light. A beam of high-energy electrons breaks the molecule apart. The masses of the fragments and their relative abundance reveal information about the structure of the molecule.

MS Components Ionization source Analyzer Detector

Mass Spectrometer Sample Introduction Data Output Inlet Data System Source Mass Analyzer Ion Detector Block diagram of mass spectrometer. Vacuum Pumps

Electron Impact Ionization ~70 Volts Electron Collector (Trap) Positive Ions + Repeller Neutral Molecules Inlet _ _ + to Analyzer + + + + + + e- e- e- EI source _ Electrons Filament Extraction Plate

Mass Spectrometer

Quadrupole Ion Filter resonant ion non-resonant ion _ Detector + + _ Source DC and AC Voltages

Separation of Ions Only the cations are deflected by the magnetic field. Amount of deflection depends on m/z. The detector signal is proportional to the number of ions hitting it. By varying the magnetic field, ions of all masses are collected and counted.

Ionization Methods Electron capture (EC) Chemical ionization (CI) 70 eV e-  neutral molecule  energetic molecular ion hard; fragmentation Chemical ionization (CI) Reagent ion + molecule  molecular ion + reagent ion Reagent ion = He, OH- (water), CH5+ or CH3+ (CH4) soft; less fragmentation

Ionization Methods Electrospray (ESI) generation of ions by desolvation or desorption of charged liquid droplets Matrix Assisted Laser Desorption (MALDI) ionization facilitated by laser irradiation of sample dissolved in an organic matrix EX: sinapinic acid

Electron Impact Ionization A high-energy electron can dislodge an electron from a bond, creating a radical cation (a positive ion with an unpaired e-).

EI (electron ionization) Mass Spectrum of unknown 1

CI (chemical ionization) Mass Spectrum of unknown 1 Reagent gas: Methane Both 72 and 88 are less than the mass in the EI spectrum Molecular ion is not sable with methane as reagent gas.

CI (chemical ionization) Mass Spectrum of unknown 1 Reagent gas: iso-Butane if 129 is M+1 => M.W. = 128  very labile cmpound M+1?

Types of MS Analyzers Quadrupole - most common Ion trap Time of Flight (TOF)

Two Operational Modes Scan Selective ion monitoring (SIM) Collect mass data over known range Slow Selective ion monitoring (SIM) Sample mass at predetermined values Fast

Mass Spectrum X - axis: m/z mass - based on 12C  12.0000 Y - axis: relative abundance usually normalized with largest line (base peak) 0 - 100 %

Features of a Basic Spectrum Tallest signal: called the base peak Signals: fragments of original molecule Molecular ion: mass of original molecule Scale: indicates mass of the cation fragments

Molecular Ion Ion whose mass equals that calculated from the molecular formula using the masses for each element which have the highest natural abundance; often tallest peak in highest m/z group Base peak - most intense peak in spectrum; not necessarily the molecular ion peak!

How are Ions Generated in Mass Spec Experiments? Radical cations are often abbreviated as: Notice the brackets have been replaced by a corner-brace symbol

What is a Radical Cation? Radical Cations are molecules that have lost one electron As a result the molecule becomes positively charged ONLY 9 e- still 10 (+) 1 (+) charge 1 C 6 (+) 4 H 4 (+) 10 (+) 1 C 6 e- 4 H 4 e- 10 e- 1 C 6 (+) 4 H 4 (+) 10 (+)

Why is it a Radical Cation? Each bond contains two electrons 8 total valence electrons Somewhere in the molecule a “one-electron bond” exists A “one-electron bond” creates an unpaired electron --- which is a radical More protons (+) than electrons (-) creates a cation

Structure of a Radical Cation Predictable? YES!!! Let’s discuss the radical cation structures possible for our favorite molecule Consider: The stability of carbocations Question: Is the stability of carbocations related to radicals?

Radical Cation Stability Cation stability can be used to predict fragmentation of a molecule Cation stability:

Application of Cation Stability Predict the radical cations that are most likely formed from the following molecules

Common Fragmentation Patterns Mass Spectroscopy Common Fragmentation Patterns

General Trends Fragments with most stable cation are preferred Form radical stabilized cations before tertiary cations Main types of radical cation fragmentation One bond cleavage – forms cation and radical Two bond cleavage – forms neutral and a new radical cation Benzylic rearrangements McLafferty Rearrangement

One Bond Cleavage Examples

One Bond Cleavage Examples

Two Bond Cleavage Examples

Two Bond Cleavage Examples

Benzylic Rearrangments Tropylium ion formation Tropylium ion structure:

McLafferty Rearrangement the rearrangement of mono-unsaturated molecular ion or radical ion with the cleavage of a,b-bond of the unsaturated system along with the concomitant transfer of a -hydrogen in a six-membered transition state by the formation of a pair of unsaturated fragments regardless of which fragment hold the charge.

Isotopes Most abundant isotope of an element is set to 100% Abundance of other isotopes are normalized with respect to it

Three Classes of Isotopes A - only a single isotope Examples: F, P, I A+1 - two isotopes with significant relative abundance differing by 1 mass unit Examples: H, C, N A+2 - two isotopes with significant relative abundance differing by 2 mass units Examples: Br, Cl, O, S

Isotopic Abundance 81Br

Molecules with Heteroatoms Isotopes: present in their usual abundance. Hydrocarbons contain 1.1% C-13, so there will be a small M+1 peak. If Br is present, M+2 is equal to M+. If Cl is present, M+2 is one-third of M+. If iodine is present, peak at 127, large gap. If N is present, M+ will be an odd number. If S is present, M+2 will be 4% of M+.

Cl2 Revisited Two isotopes: 35Cl and 37Cl Three possible species formed: 35Cl35Cl, 37Cl35Cl, and 37Cl37Cl Relative abundance: 35Cl35Cl: 1.0 x 1.0 = 1.0 37Cl35Cl and 35Cl37Cl: 1.0 x 0.325 = 0.325 each or 0.66 37Cl37Cl: 0.325 x 0.325 = 0.106 So, answer: 3 peaks at 70, 72, and 74 with relative intensities of 100, 32.5, and 10.6 %

Major Steps in Analysis of Mass Spectra Identification of molecular ion Base peak Examination of isotopic distribution pattern Negative information Determine elemental composition Analysis of fragmentation pattern Propose possible structures Compare postulated species to available reference spectra

Masses are graphed or tabulated according to their relative abundance.

Mass Spectra of Alkanes More stable carbocations will be more abundant.

Fragmentation of Hexane Starting Point M.W. of Hexanes = 86.15 g/mol

Fragmentation of Hexane 2nd: Determine ID of most abundant fragment Molecular Ion – base peak = difference 86 – 57 = 29 Difference = CH3CH2

Fragmentation of Hexane 3rd: Determine ID of other prominent fragments Peak = 43 CH3CH2CH2

MW = 86

Mass Spectrum with Chlorine

MW = 106 M+ - HCl 35Cl 37Cl 100% 33% 108 M+ - Cl 106

Mass Spectrum with Bromine

MW = 150 M+ - Br -14 (CH2) 79Br 81Br 100% 98% -15 (CH3) 150 152

MW = 120

MW = 100

MW = 86

MW = 102

MW = 102

Mass Spectrum with Sulfur

MW = 104

Mass Spectra of Alkenes Resonance-stabilized cations favored.

Mass Spectra of Alcohols Alcohols usually lose a water molecule. M+ may not be visible.

High Resolution MS Masses measured to 1 part in 20,000. A molecule with mass of 44 could be C3H8, C2H4O, CO2, or CN2H4. If a more exact mass is 44.029, pick the correct structure from the table: C3H8 C2H4O CO2 CN2H4 44.06260 44.02620 43.98983 44.03740

The GC-MS A mixture of compounds is separated by gas chromatography, then identified by mass spectrometry.

GC/MS