Magnetic resonance imaging (MRI) safety training

Slides:



Advertisements
Similar presentations
Emergency Preparedness and Response
Advertisements

EMERGENCY MEDICAL MANAGEMENT OF RADIATION CAUSALTIES IN HOSPITAL Module XVII.
MRI Safety Training.
RESEARCH MRI Safety Training
SPILL RESPONSE Training Program.
MRI Safety This slide show will educate all healthcare providers about the safe delivery of care in the magnetic resonance (MR) environment. This slide.
State of Michigan Computed Tomography Regulations
Chemical Safety. Overview Chemical hazard classes Communication of hazards Routes of exposure Hierarchy of controls Special laboratory hazards.
25 TAC Quality Assurance in a licensed ASC
Fire Prevention and Safety Health Science Class. Rationale: Fires may occur at any time, as a result of overloading wiring, smoking, improper chemical.
MRI General Safety Rules. GENERAL RULES DO NOT ENTER the MRI Scan Room if the door is closed. Always see the MRI Technologist before entering room. Always.
MRI SAFETY FOR NON-MRI PERSONNEL.
PERMIT REQUIRED CONFINED SPACES PERMIT REQUIRED CONFINED SPACES.
MRI Safety Course Module 1 –Patient Screening Learning Objective #1 The student will be able to describe in detail the patient screening process, contraindications.
Topics of Discussion:  Introduction  Identify & describe the protocols & processes for: Lockouts Sheltering in Place Lockdown Evacuation  Active Shooter.
1 CHCOHS312A Follow safety procedures for direct care work.
Hazardous Waste and Emergency Response
Occupational health and safety
EGRESS AND FIRE PROTECTION
Safety Unit II.

The Physical Environment Tracers How Should I Prepare for the Physical Environment Tracers? Healthcare Engineering Consultants  Will limit.
Organization and Implementation of a National Regulatory Program for the Control of Radiation Sources Inspection Part II.
Merchant Shipping (Code of Safe Working Practices) Regulations Statutory Instrument 1998 No
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
Magnet Safety.
Exit Routes and Fire Protection
Mosby items and derived items © 2007, 2004 by Mosby, Inc., an affiliate of Elsevier Inc. Operations.
PROTECTIVE ACTIONS AND REENTRY. Protective Actions Promptly and effectively implemented or recommended for implementation to minimize the consequences.
Confined Spaces. Definition: Are large enough that an employee can bodily enter them Are not designed for continuous occupancy. Have limited or restricted.
Emergency Preparedness. Proposed Emergency Preparedness Rules NFR/LMC §19.326(a) deleted and moved to § for Emergency Preparedness Rules Places.
Safety and Infection Control: Reporting of Incident / Event / Irregular Occurrence / Variance & Safe Use of Equipment Lorelei Sepulveda 1.
Welcome! We’re glad you’re here
SAFETY IN THE WORKPLACE
3-MINUTE READ WORKING TOGETHER TO SAFEGUARD CHILDREN.
Handout 1: Health and safety at work
Hazard Communications Standard HS-105 (R)
MRI Safety Awareness The purpose of this healthstream presentation is to educate the hospital staff that do not routinely come into the MRI environment.
CONFINED SPACE ENTRY TRAINING PROGRAM
Methodist Hospital for Surgery
Implementation of a Surgical Safety Check List
WISHA Confined Spaces Standard
Safety Considerations
3-MINUTE READ WORKING TOGETHER TO SAFEGUARD CHILDREN.
Patient Medical Records
Jody Culham Brain and Mind Institute Department of Psychology
Goodyear Napanee Contractor Safety
Compliance Made Simple Presents:
East Carolina University Compressed Gas Safety Training Program
Reference: Diversified Health Occupations
Safety is part of our culture at Halliburton
Building Emergency Evacuation Plan (BEEP) Training for Building Emergency Coordinators and Alternates.
Fire Prevention and Safety
Injury and Illness Prevention Program (IIPP)
Magnetic field safety October 2018.
RESTRAINT & SECLUSION(R/S) for NON-NURSING
Fire Safety and Utility Controls
Safety & Health Programs
Safety & Health Programs
Human Subjects in fMRI Research
Center for Magnetic Resonance Research
Confined space.
The Emergency Action Steps
The Emergency Action Steps
The Emergency Action Steps
Safety & Health Programs
Population Monitoring
Hoglund Brain Imaging Center
NATE Tower Site Hazard Recognition Guide
Building Emergency Evacuation Plan (BEEP) Training for Building Emergency Coordinators and Alternates.
Presentation transcript:

Magnetic resonance imaging (MRI) safety training Christopher A. Lease, M.S. Medical Physicist

Tata memorial hospital Mumbai, india Two employees were pinned to machine for nearly four hours after a non-MR safe oxygen tank was brought into the room. One suffered a fractured elbow. The other suffered serious injuries to his lower body including a punctured urinary bladder and severe internal bleeding. His lower abdomen and upper thigh got crushed, severely compromising circulation to his lower body and damaging the muscles and nerves.

Westchester medical center the Michael columbini incident A six year old boy died after undergoing an MRI exam when an Oxygen tank was pulled across the room, crushing his head. The routine imaging procedure was performed following surgery for a benign brain tumor the week before. The boy was under sedation at the time the accident occurred.

Other incidents Rochester, NY The Hairpin An off-duty cop was told it was OK to walk into the MR Suite with his hand gun which immediately got pulled out of his hand, causing an accidental shot to go off. The bullet lodged in a wall and no one was injured. The Hairpin After a woman forgot to remove a hairpin, it traveled up her nose and lodged in her pharynx. This required a procedure to extract the object. At least one record of a woman dying after receiving an MR procedure with an aneurysm clip.

Dangers involved with mrI are immediate and potentially deadly The dangers of MRI are always present and can cause immediate and catastrophic damage: Projectile (Missile) hazards Medical device misplacement or disruption The three previous slides clearly demonstrate these hazards. It is the responsibility of the MR staff to understand their critical role in creating a safe MR environment for themselves, the patient, and/or any other MR personnel present. Safety should always be a first priority.

Two important references pertaining to mr safety Joint Commission Sentinel Event #38 10 explicit objectives for the institution 8 classifications of injuries that can and have occurred during the MR scanning process are described ACR Guidance Document on MR Safe Practices: 2013 14 Points

Joint commission sentinel event #38 10 Explicit Objectives 1. Restrict Access – Four Zones (Zone Posting) 2. Use trained personnel to screen all non-emergent patients twice 3. Ensure MR Tech has patient’s complete and accurate medical history and have literature readily accessible to reference the safety of any devices. 4. Have specially trained staff accompany any individuals inside the MR suite at all times who are familiar with the MR environment.

5. Education – Annually provide all medical and ancillary staff who may be expected to accompany patients into the MR Suite. Have materials available for patients and their families that explain the potential for accidents and adverse effects. 6. Precaution s to prevent burns 7. Use only equipment approved for MR scans 8. Proactively plan for managing critically ill patients who require physiological monitoring and continuous infusion of life sustaining drugs while in the suite. 9. Provide hearing protection 10. Never attempt to run a cardio-pulmonary arrest code or resuscitation within the MR Suite itself

8 Types of Injuries That Can and Have Occurred During the MR Scanning Process 1. Missile (Projectile) Effect 2. Dislodged ferromagnetic implants 3. Burns from objects that may heat (wires and leads are the most common) 4. Equipment or device malfunction due to the magnetic field (battery powered devices such as diffusion pumps and pacemakers) 5. Failure to attend patient support systems (sedation or anesthesia) - Oxygen or infusion pumps can run out if replacement is not readily available 6. Acoustic injury from machine noise 7. Adverse events from contrast agents 8. Adverse events to cryogen handling, storage, or inadvertent release in MR systems (quench)

ACR Guidance Document on MR Safe Practices: 2013 1. Establish, Implement, and Maintain Current MR Safety Policies and Procedures *2. Static Magnetic Field Issues: Site Access Restriction (Zone I-IV Postings) 3. MR Technologist 4. Pregnancy Related Issues – Patient and Worker 5. Pediatric MR Safety Concerns 6. Time Varying Gradient Magnetic Field Related Issues: Induced Voltages 7. Time Varying Gradient Magnetic Field Related Issues: Auditory Considerations 8. Time Varying Radiofrequency Magnetic Field Related Issues: Thermal 9. Drug Delivery Patches and Pads 10. Cryogen-Related Issues

ACR Guidance Document on MR Safe Practices: 2013 1. Establish, Implement, and Maintain Current MR Safety Policies and Procedures *2. Static Magnetic Field Issues: Site Access Restriction (Zone I-IV Postings) 3. MR Technologist 4. Pregnancy Related Issues – Patient and Worker 5. Pediatric MR Safety Concerns 6. Time Varying Gradient Magnetic Field Related Issues: Induced Voltages 7. Time Varying Gradient Magnetic Field Related Issues: Auditory Considerations 8. Time Varying Radiofrequency Magnetic Field Related Issues: Thermal

ACR Guidance Document on MR Safe Practices: 2013 9. Drug Delivery Patches and Pads 10. Cryogen-Related Issues 11. Claustrophobia, Anxiety, Sedation, Analgesia, and Anesthesia 12. Contrast Agent Safety 13. Patients in Whom There Are or May Be Intracranial Aneurysm Clips 14. Patients in Whom There Are or May Be Cardiac Pacemakers or Implantable Cardioverter Defibrillators

MRI SAFETY ZONES Zones are conceptual Should be clearly posted Meaning of each zone should be understood by staff

Zone I Includes all areas that are freely accessible to the general public Typically outside the MR environment itself Area through which patients, health care personnel, and other employees of the MR site access the MR environment

Zone II Interface between the publicly accessible, uncontrolled Zone I, and the strictly controlled Zones III and IV Typically where patients are greeted and are not free to move about at will, but rather under the supervision of MR personnel Answers to MR screening questions, patient history, medical insurance questions, etc., are typically obtained

Zone III Area in which free access by unscreened non-MR personnel or ferromagnetic objects or equipment can result in serious injury or death Access should be strictly restricted, with access to regions within it (including Zone IV) controlled by, and entirely under the supervision of MR Personnel Should be physically restricted from general public access (i.e. key locks, passkey, or other reliable, physically restricting method that can differentiate between MR and non-MR personnel)

Zone IV Area is synonymous with the MR scanner magnet room itself By definition will always be located within Zone III Should be demarcated and clearly marked as being potentially hazardous due to the presence of very strong magnetic fields Should be clearly marked with a red light and lighted sign stating, “The Magnet Is Always On” Non-MR personnel should be accompanied by, or under the immediate supervision of and visual contact with, one specifically identified level 2 MR person for the entirety of their duration within Zone IV (and Zone III) Level 1 & 2 MR personnel may move freely about all zones

Mr Personnel and non-mr personnel definitions All individuals working within at least Zone III of the MR environment should be documented as having successfully completed at least one of the MR Safety live lectures or prerecorded presentations approved by the MR medical director Above should be repeated annually and documented Two levels of MR personnel: Level 1 and Level 2 Level 1: Those who have passed minimal safety educational efforts to ensure their own safety as they work within Zone III Level 2: Those who have been more extensively trained and educated in the broader aspects of MR safety issues and direct neuromuscular excitation from rapidly changing gradients) MR medical director to identify necessary training and those who qualify as Level 2 Medical director shall have necessary education and experience to qualify as Level 2

Patient and non-mr personnel screening Non-MR personnel wishing to enter Zone II must pass and MR safety screening Only MR personnel are authorized to perform this Screening process and forms for patients, non-MR personnel, and MR personnel should be essentially the same Ferromagnetic detection devices can be a supplement to the screening process but not used in place of the screening process Non-emergent patients should be screened on site by a minimum of two separate individuals At least one of these should be performed by level 2 personnel and at least one should be performed interactively

Patient and non-mr personnel screening Remove all readily removable metallic personnel belongings and devices on or in them (i.e. watches, jewelry, pagers, cell phones, body piercings, contraceptive diaphragms, metallic drug delivery patches, cosmetics containing metallic particles (such as eye makeup), and clothing items) It is recommended that patients wear a site-supplied gown with no metal fasteners when feasible Anyone with a history of potential ferromagnetic foreign object penetration must undergo further investigation before being permitted entrance to Zone III Once positive identification has been made, best effort assessments should be made to identify to MR compatibility or MR safety of the implant or object Final determination of whether or not to scan any given patient with any given implant, foreign body, etc. is to be made by the level 2 designated attending MR Radiologist, the MR Medical Director, or specifically designate Level 2 MR Personnel

Patient and non-mr personnel screening All non-MR personnel with implanted electrically conductive devices upon which they are dependent should be precluded from Zone IV and physically restrained from the 5-Gauss line unless specifically cleared in writing by a level 2 designated attending Radiologist or Medical Director Arrange to prospectively educate local fire marshals, fire-fighters, and police or security personnel about the potential hazards of responding to emergencies in the MR suite Even in the presence of a true fire (or other emergency) in Zone III or IV, magnetic fields may be present and fully operational Air tanks, axes, crowbars, other firefighting equipment, guns, etc might prove catastrophic or even lethal to those responding or others in the vicinity As part of Zone III and IV restrictions, all MR suites must have clearly labeled and readily accessible MR Conditional or MR Safe fire extinguishing equipment physically stored within Zone III or IV cardiac pacemakers, implantable cardioverter defibrillators (ICDs), diaphragmatic pacemakers, electromechanically activated devices, or other

MR Personnel Screening Must undergo an MR screening process as part of their employment interview process to ensure their safety in the MR environment. For your own protection and for the protection of the non-MR personnel under their supervision, all MR personnel must immediately report any trauma, procedure, or surgery they experience or undergo where a ferromagnetic object or device may have become introduced within or on them. This will permit appropriate screening to be performed on the employee to determine the safety of permitting that employee into Zone III.

Device and object screening Ferrous objects should be restricted from entering Zone III, whenever practical (including equipment brought by contractors) Never assume MR compatibility or safety information about a device if it is not clearly documented in writing External devices or objects demonstrated to be ferromagnetic and MR Unsafe or incompatible may still, under specific circumstances, be brought into Zone III if they are under the direct supervision of specifically designated level 1 or 2 MR personnel and deemed by MR personnel to be necessary and appropriate for patient care. The designated level 1 or 2 MR personnel must be familiar with the device, its function, and the reason supporting its introduction to Zone III.

Device and object screening All portable metallic or partially metallic objects must be properly identified and labeled using the current FDA labeling criteria developed by ASTM International in standard ASTM F2503 MR Safe (Green square) Wholly, nonmetallic Unsafe (Red circle) Clearly ferromagnetic MR Conditional (Yellow triangle) Objects with an MR conditional rating should be affixed with the yellow triangular label before being brought into the scan room/Zone IV.

MR Technologist – Qualifications and responsibilities MR technologists should be in compliance with the technologist qualifications listed in the MR Accreditation Program Requirements. Except for emergent coverage, there will be a minimum of 2 MR technologists or one MR technologist and one other individual with the designation of MR personnel in the immediate Zone II through Zone IV MR environment. For emergent coverage, the MR technologist can scan with no other individuals in their Zone II through Zone IV environment as long as there is in-house, ready emergent coverage by designated department of radiology MR personnel (e.g., radiology house staff or radiology attending).

PREGNANCY RELATED ISSUES Worker Pregnant health care practitioners are permitted to work in and around the MR environment throughout all stages of their pregnancy.

PREGNANCY RELATED ISSUES Patient Present data have not conclusively documented any deleterious effects of MR imaging exposure on the developing fetus. Therefore no special consideration is recommended for the first, versus any other, trimester in pregnancy. Pregnant patients can be accepted to undergo MR scans at any stage of pregnancy if, in the determination of a level 2 MR personnel-designated attending radiologist. The risk–benefit ratio to the patient warrants that the study be performed. MR contrast agents should not be routinely provided to pregnant patients. This decision too, is on that must be made on a case-by-case basis by the covering level 2 MR personnel-designated attending radiologist who will assess the risk–benefit ratio for that particular patient. Studies have demonstrated that at least some of the gadolinium-based MR contrast agents readily pass through the placental barrier and enter the fetal circulation.  The risk to the fetus of gadolinium based MR contrast agent administration remains unknown and may be harmful.

Pediatric MR Safety Concerns Sedation and Monitoring Issues Children form the largest group requiring sedation for MRI, largely because of their inability to remain motionless during scans. Sedation protocols may vary from institution to institution according to procedures performed (diagnostic vs. interventional), the complexity of the patient population (healthy preschoolers vs. premature infants), the method of sedation (mild sedation vs. general anesthesia) and the qualifications of the sedation provider.

Pediatric MR Safety Concerns Pediatric Screening Issues Children may not be reliable historians and, especially for older children and teenagers, should be questioned both in the presence of parents or guardians and separately to maximize the possibility that all potential dangers are disclosed. It is recommended that they be gowned before entering Zone IV to help ensure that no metallic objects, toys, etc. inadvertently find their way into Zone IV. Pillows, stuffed animals, and other comfort items brought from home represent real risks and should be discouraged from entering Zone IV. If unavoidable, each should be carefully checked with a powerful handheld magnet and/or ferromagnetic detector and perhaps again in the MR scanner before permitting the patient to enter Zone IV with them to ensure that they do not contain any objectionable metallic components.

Pediatric MR Safety Concerns MR Safety of Accompanying Family or Patient Although any age patient might request that others accompany them for their MR examination, this is far more common in the pediatric population. Those accompanying or remaining with the patient should be screened using the same criteria as anyone else entering Zone IV. In general, it would be prudent to limit accompanying adults to a single individual.

Time Varying Radiofrequency Magnetic Field Related Issues: Thermal SAR (Specific Absorption Rate): Measures the power absorbed per unit mass of tissue (W/kg). Absorption of RF power increases tissue temp. In a way, this could be thought of as a measure of dose for RF fields. FDA SAR limits are as follows: Whole body - <4 W/kg Head - <3.2 W/kg Extremities (or any 1 g of tissue) - <8 W/kg

Cryogen related issues - QUENCH The greatest risk of exposure to cryogens is when the magnet has an unintentional (emergency)shut down, this is known as a quench. During a quench, the cryogen evaporates rapidly causing a loss of superconductivity in the magnet. Emergency ventilating systems direct the cryogens through a quench pipe out of the building. If the expanding helium cannot be dissipated through an external vent it may be released into the scanning room which will cause the displacement of oxygen and present a risk of asphyxiation. In the event of a quench: Evacuate all patients quickly & safely The entire MR suite should be evacuated and everyone should be moved outside the room The MR technologist must confirm the magnetic field has dissipated prior to the safe entry of response teams (i.e., Fire Department)

Additional items: Time Varying Gradient Magnetic Field Related Issues: Induced Voltages Time Varying Gradient Magnetic Field Related Issues: Auditory Considerations Drug Delivery Patches and Pads Claustrophobia, Anxiety, Sedation, Analgesia and Anesthesia Contrast Agent Safety Patients in Whom There Are or May Be Intracranial Aneurysm Clips Patients in Whom There are or May Be Cardiac Pacemakers or Implantable Cardioverter Defibrillators

References: http://www.aapm.org/meetings/amos2/pdf/59-17207-59975-979.pdf MRI Safety, R. Jason Stafford https://www.aapm.org/pubs/reports/rpt_20.pdf AAPM Report No. 20 http://onlinelibrary.wiley.com/doi/10.1002/jmri.24011/pdf ACR Guidance on Safe Practices http://www.acr.org/~/media/ACR/Documents/PDF/QualitySafety/MR%20Safety/InterrelatingSentinelEven tAlert38.pdf Interrelating Sentinel Event #23 with ACR Guidance http://www.jointcommission.org/assets/1/18/SEA_38.pdf Joint Commission Issue 38 Kanal E, Gillen J, Evans JA,Savitz DA, Shellock FG. Survey of reproductive health among female MR workers. Radiology 1993;187: 395–399.