Light mixer proposal Arjan Heering, Anton Karneyeu

Slides:



Advertisements
Similar presentations
2 Introduction   MiniCal test-beam studies started at the beginning of March (till March 6 we only had 17 APD’s, then 33 APD’s)   A few days were.
Advertisements

Photodetector Timing Resolution Burle Micro-Channel Plate Photo Multiplier Tubes (MCP-PMTs) H. Wells Wulsin SLAC Group B Winter 2005.
Microscope Performance at elevated dark rates Richard Jones University of Connecticut collaboration GlueX collaboration meeting, Newport News, Feb. 2-4,
Tagger Electronics Part 1: tagger focal plane microscope Part 2: tagger fixed array Part 3: trigger and digitization Richard Jones, University of Connecticut.
1 A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat.
ECAL Testbeam Meeting, Rome 28 March 2007 Toyoko Orimoto Adolf Bornheim, Chris Rogan, Yong Yang California Institute of Technology Lastest Results from.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
May 11, 2006CALICE Collaboration Meeting, McGill Uni, Montréal, Québec, CA 1 UV LED quality test  All 872 UV LEDs tested  Selection is about to finish.
Status of scintillator KLM study P. Pakhlov (ITEP)
CALICE Meeting DESY ITEP&MEPhI status report on tile production and R&D activities Michael Danilov ITEP.
Uni Bergen:G. Eigen, J. Zalieckas, E. van der Kraaij FZU Prague: J. Cvach, J. Kvasnicka, I. Polák CALICE meeting, Argonne, 19/03/2014.
R&D on W-SciFi Calorimeters for EIC at Brookhaven E.Kistenev, S.Stoll, A.Sukhanov, C.Woody PHENIX Group E.Aschenauer and S.Fazio Spin and EIC Group Physics.
Scintillation hodoscope with SiPM readout for the CLAS detector S. Stepanyan (JLAB) IEEE conference, Dresden, October 21, 2008.
The Scintillator ECAL Beam Test at FNAL Adil Khan International Linear Collider Workshop 2010 LCWS10 & ILC10, Beijing, China CALICE Scintillator ECAL group.
The Scintillator ECAL Beam Test at FNAL K. Kotera, Shinshu-u, 1st October 2009 CALICE Scintillator ECAL group; Kobe University, Kyungpook University, the.
THE FORWARD PROTON DETECTOR AT DZERO Gilvan Alves Lafex/CBPF 1) MOTIVATION 2) DETECTOR OPTIONS 3) FPD R&D 4) OUTLOOK Lishep 98 Lafex/CBPF Feb 17, 1998.
Montpellier, November 15, 2003 J. Cvach, TileHCAL and APD readout1 TileHCAL- fibre readout by APD APDs and preamplifiers Energy scan with DESY beam –Energy.
Uni Bergen:G. Eigen, E. van der Kraaij, J. Zalieckas FZU Prague: J. Cvach, J. Kvasnička, I. Polák Linear Collider Workshop, Tokyo 13/11/2013.
R&D of MPPC for T2K experiment PD07 : Photosensor Workshop /6/28 (Thu) S.Gomi T.Nakaya M.Yokoyama H.Kawamuko ( Kyoto University ) T.Nakadaira.
U. Akgun, HCAL Fall Meeting, 11/11/2004, Fermilab, IL HF PMT ISSUES By Ugur Akgun The University of Iowa.
Scintillator tile-SiPM system development for CALICE Engineering AHCAL Prototype Michael Danilov, ITEP & CALICE LCWS10 Beijing March 2010 Outline New scintillator.
DAQ for 4-th DC S.Popescu. Introduction We have to define DAQ chapter of the DOD for the following detectors –Vertex detector –TPC –Calorimeter –Muon.
Motivation General rule for muon triggers: Never neglect a possible backup reduction factor. It will always come back to you. Even if RPC trigger works.
5 July 2010Ivo Polák, FZU, Prague LED notched fibre distributing system Calibration system for SiPM Ivo Polák, Ji ř í Kvasni č ka
Development of Multi-Pixel Photon Counters(MPPC) Makoto Taguchi Kyoto University.
Performance of SiPM Ryuhei Nakamura (Kobe Univ.) GLC calorimeter group (KEK, Kobe, Konan, Niigata, Shinshu, Tsukuba) Contents ・ Introduction ・ Test results.
AHCAL Electronics. SPIROC2 and HBU measurement results Mathias Reinecke CALICE main meeting Univ. HASSAN II, Casablanca, Morocco Sept. 23rd, 2010.
28 June 2007G. Pauletta: ALCPG Tests of IRST SiPMs G. Pauletta Univ. & I.N.F.N. Udine Outline 1.IRST SiPMs : baseline characteristics 2.first application.
Linearity Tests The laser induced high noise pulses on the APD rendering it useless. We tried several methods to shield and filter the noise. With a stabilized.
Multipixel Geiger mode photo-sensors (MRS APD’s) Yury Kudenko ISS meeting, KEK, 25 January 2006 INR, Moscow.
SiPM for CBM Michael Danilov ITEP(Moscow) Muon Detector and/or Preshower CBM Meeting ITEP
LC Power Distribution & Pulsing Workshop, May 2011 Super-ALTRO Demonstrator Test Results LC Power Distribution & Pulsing Workshop, May nd November.
DESY BT analysis - MPPC Saturation Correction - S. Uozumi Feb Sci-ECAL meeting 1.MPPC Gain Measurement (with LED data) 2.Inter-calibration of readout.
2007 detectors study at CERN 1.SiPM time resolution 2.MAPD S60 3.Blue sensitive MAPD 4.New 16ch board 5.New HV.
(s)T3B Update – Calibration and Temperature Corrections AHCAL meeting– December 13 th 2011 – Hamburg Christian Soldner Max-Planck-Institute for Physics.
Systematics on ScECAL 1 st prototype DESY Apr , CALICE Satoru Uozumi for the CALICE collaboration NIM A, 763 (2014) 278.
F Don Lincoln, Fermilab f Fermilab/Boeing Test Results for HiSTE-VI Don Lincoln Fermi National Accelerator Laboratory.
Short report on MPPC gain monitoring system S. Uozumi Jun ScECAL meeting At the FNAL BT, having the gain monitoring system for all the channel.
Beta-ray test for strip scintillator readout MPPC GLD Cal group KEK 06/2/28 (Tue) Niigata university Sayaka IBA.
DHCAL Jan Blaha R&D is in framework of the CALICE collaboration CLIC08 Workshop CERN, 14 – 17 October 2008.
Testbeam analysis Lesya Shchutska. 2 beam telescope ECAL trigger  Prototype: short bars (3×7.35×114 mm 3 ), W absorber, 21 layer, 18 X 0  Readout: Signal.
ILC-CAL Test status of HPK photon sensors Kobe-U Y.Tamura / K.Kawagoe Confidential: do not distribute outside.
Status of NEWCHOD E.Guschin (INR), S.Kholodenko (IHEP), Yu.Kudenko (INR), I.Mannelli (Pisa), O.Mineev (INR), V.Obraztsov (IHEP), V.Semenov(IHEP), V.Sugonyaev.
CdTe prototype detector testing Anja Schubert The University of Melbourne 9 May 2011 Updates.
Sensitivity of HO to Muons Shashi Dugad for HO group India-CMS Meeting 6-7 Oct
Silicon Photomultiplier Development at GRAPES-3 K.C.Ravindran T.I.F.R, OOTY WAPP 2010 Worshop On behalf of GRAPES-3 Collaboration.
Study of the MPPC for the GLD Calorimeter Readout Satoru Uozumi (Shinshu University) for the GLD Calorimeter Group Kobe Introduction Performance.
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK) for KEKDTP photon sensor group.
M.Taguchi and T.Nobuhara(Kyoto) HPK MPPC(Multi Pixel Photon Counter) status T2K280m meeting.
28/11/02Guy Dewhirst1 Ultra Light Weights Method Guy Dewhirst Imperial College London.
3/06/06 CALOR 06Alexandre Zabi - Imperial College1 CMS ECAL Performance: Test Beam Results Alexandre Zabi on behalf of the CMS ECAL Group CMS ECAL.
Performance of 1600-pixel MPPC for the GLD Calorimeter Readout Jan. 30(Tue.) Korea-Japan Joint Shinshu Univ. Takashi Maeda ( Univ. of Tsukuba)
IFR Detector R&D status
IFR Detector R&D status
LED notched fibre distributing system Calibration system for SiPM
SiPM linearization status update
Roman Mizuk (ITEP, Moscow)
Scintillation Detectors in High Energy Physics
The Beam Test at Fermilab:
288 channel HE RBX? A conceptual exercise, NOT a proposal
Design of Digital Filter Bank and General Purpose Digital Shaper
HCAL – SiPM upgrade requirements
Update on TB 2007 Xtal Irradiation Studies at H4
ITEP&MEPhI status report on tile production and R&D activities
Scintillator tile- SiPM systems R&D
Tagger Microscope Detector & Electronics Status
R&D of MPPC for T2K experiment
Scintillator HCal Prototype
T2Kロシアグループに向けたMPPC(SiPM)の性能評価
R&D of MPPC in kyoto M.taguchi.
Presentation transcript:

Light mixer proposal Arjan Heering, Anton Karneyeu HCAL Phase 1 Upgrades Meeting / 28 Apr 2016

The problem Original design: fiber attached directly to the surface of the input window SiPM operates best (linearity of response) if the light distributed uniformly over the input window Solution: introduce light mixer Light input fibers SiPM SiPM input window

Setup LED pulse is 25 ns APD for light reference, connected to QADC SiPM connected to QIE11 readout LED APD 4 input channels: Y11 -> 4 meter clear fiber -> ODU -> MIXER -> SiPM

Configurations Connected fibers: 1 fiber 4 fibers Mixer diameter: 2.5 mm 2.6 mm 2.7 mm 2.8 mm Mixer thickness: No mixer 1 mm 1.6 mm

All measurements / RAW

SiPM Gain calculation Low-light runs: resolution is enough to resolve single electron (pixel) peaks Average gain over all channels / runs: 52 fC / pixel

SiPM charge -> pixels

Light yield efficiency, % Diam\Thickness 1fib 1 mm 4fib 1 mm 1 fib 1.6 mm 4 fib 1.6 mm 2.5 85 77 75 2.6 78 76 2.7 88 83 73 2.8 70 61 65 Normalized to the baseline (no mixer) Extracted from low-light region (linear behavior) => Diameter 2.6 mm / thickness 1.6 mm looks best

APD Gain calculation At low light SiMP expected to demonstrate linear behavior Fit 5 low-light points with line, the slope gives qadc counts to effective photo- electrons conversion

APD qadc counts -> effective photo electrons APD effective p.e. = Input Nph * PDE * mixer_efficiency From TB: 15k p.e. ~ 300 GeV

Linear correction factor Mixer improves linearity Thickness 1.6 mm and 1 mm perform similar

Linearity of mixer t = 1 mm / d = 2.6 mm

Conclusions Mixer thickness 1 mm, diameter 2.6 mm

Backup

Correction factor (vs. real light)

Correction factor (vs. real light) of the mixer t = 1 mm / d = 2.6 mm

Shunt factor 11.5 (design)

Shunt factor 10.9