Head computed tomography doses in four selected hospitals in Nigeria

Slides:



Advertisements
Similar presentations
Standardizing Methods of Calculating and Reporting CT Dose.
Advertisements

NEW JERSEY CT DOSE PROJECT UPDATE New Jersey Medical Physicist Meeting March 9, 2006 Web version with summary of meeting outcomes.
CT Scanning: Dosimetry and Artefacts
IAEA International Atomic Energy Agency Strengthening justification of medical exposure in diagnostic imaging IRPA13, May 2012, Glasgow, Scotland.
Radiation Dose in CT Scanning: A Team Approach
RAD309 Patient Dose.
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
Business: Cell/Text: Presented by: Thomas J. LaRocca, M.S., D.A.B.R. Medical Physicist
RADIATION MONITORING STRATEGY IN EUROPE Guy FRIJA TORONTO, May 2013.
Standards and Guidance
PART IX: EMERGENCY EXPOSURE SITUATIONS Module IX.1: Generic requirements for emergency exposure situations Lesson IX.1-2: General Requirements Lecture.
Kalpana Kanal, Ph.D., DABR Associate Professor, Diagnostic Physics
Dose Survey in Computed Tomography DS /CM Kampala IAEA/RCA Kampala.
Radiation Protection of Patients Unit
IAEA International Atomic Energy Agency Interventional Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(3)
Knowledge of radiation exposure in common radiological examinations amongst radiology department staff AL Chang, LH Cope, DH Keane, S Wood Presented by.
Pratik Mehta, B.A. 1 ; Kate MacGregor, M.P.H. 2 ; Robert Gould, D.Sc. 2 ; Diana Miglioretti, Ph.D. 3 ; J. Anthony Seibert, Ph.D. 4 ; Rebecca Smith-Bindman,
CT physics and instrumentation
BEIR VII Implications of the Report for the Future of Medical Imaging G. Donald Frey, Ph.D. Department of Radiology Medical University of South Carolina.
Low Dose Head Computed Tomography Protocol for Evaluation of Non-traumatic Emergencies in Children Rishi Mhapsekar, MD Marilyn J. Siegel, MD Robert McKinstry.
Reduction of effective and organ dose to the eye lens in cerebral MDCT scans using iterative image reconstruction Zizka J, Jandura J, Kvasnicka T, Klzo.
JUSTIFICATION OF COMPUTERIZED TOMOGRAPHY EXAMINATIONS AND RADIATION RISKS IN EVERYDAY RADIOLOGICAL PRACTICE Darka R. Hadnađev, Olivera Nikolić, Sanja Stojanović.
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
How we set a DRL An example using CT David Sutton / Colin Martin Kampala IAEA/RCA Kampala.
A Pain in the Neck! Too Many Normal Paediatric Cervical Spine CTs in trauma? SC Shelmerdine, BN Bhaludin, WY Mok, L Woods, V Cook Epsom & St Helier University.
1 Radiation Dosimetry in X-Ray Computed Tomography (CT): Standardization and Regulation Stanley H. Stern, Ph.D. Center for Devices and Radiological Health.
MEASUREMENTS OF RADIATION DOSES IN MULTISLICES COMPUTED TOMOGRAPHY EXAMINATIONS A. ELMAHDI*, A. SULIEMAN *Presenting author 1 Sudan Atomic Energy Commission,
Diagnostic reference levels in Medical Imaging. Concept and practice
Radiation Sources in medicine diagnostic Radiology
Robyn Wright  Main health system campus located in Ann Arbor, MI  Department of Radiology established in 1913  U of M Hospital  C.S. Mott Childrens.
International Atomic Energy Agency Medical exposure in radiology: Scope and responsibilities Module VIII.1 - Part 2 : Training.
What are the dose quantities in CT ? IAEA/RCA Kampala Kampala.
Objectives Identify the image quality characteristics that apply to all medical imaging modalities Understand the concept of image optimization Review.
IS THERE A RATIONAL APPROACH TO RADIATION RISK? Philip F. Judy Ph.D. Department of Radiology Brigham and Women's Hospital New England AAPM Chapter Meeting.
Jordan Russell, Medical Student, FRCP (C), Brent Burbridge, MD, FRCP (C), Jennifer Tynan, MD.
CT Image Quality for CT Pulmonary Embolism (CT PE) and Chest CT Aorta Acquired with Power Injection via an Arm Port Ryan Verity 1, David Leswick.
Medicine, Nursing and Health Sciences Overview of Radiography and Medical Imaging Course Associate Professor Peter Wells, PhD Adjunct Professor Department.
Research Methods: 1 M.Sc. Physiotherapy/Podiatry/Pain Writing for Publication.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 12.1 : Shielding and X-ray room design Practical.
Wednesday Case of the Day History: 32-year-old female (approximately standard-sized female) undergoes CT scan of abdomen and pelvis after motor vehicle.
Global CT Scan and PET Scan Market by Manufacturers, Countries, Type and Application, Forecast to WEBSITE Date: May 2017.
Introduction of a low dose protocol for thoracic ablation under CT guidance Thea Buchan, Jason Godbold, Dr Steve Bandula Interventional Oncology Service,
Moayyad E Suleiman Prof. Patrick C Brennan Dr. Peter Kench
Materials and Methods Continued: Materials and Methods:
EQUIPMENT DOWNTIME IN THE RADIOLOGY DEPARTMENT OF TEACHING HOSPITALS IN NIGERIA 1C.C. Nzotta PhD, 2T. Adejoh M.Sc, & 3B.F. Nkubli.
Routine screening for Hepatitis B and Human Immunodeficiency virus before upper gastrointestinal endoscopy: is it necessary? By Dr Ray-Offor E FWACS,FMAS,DMAS(Ind),
Patient exposure trends and problems in implementing ALARA
Role of Adaptive Statistical Iterative Reconstruction (ASIR) in lowering radiation dose for pediatric head CT Electronic Poster-eP 140 Azadeh Tabari, MD.
European Journal of Radiology
University of Ulster, School of Life and Health Sciences.
CURRENT TRENDS AND LIMITATIONS IN ESTABLISHING DIAGNOSTIC REFERENCE LEVELS FOR RADIOLOGICAL EXAMINATIONS IN NIGERIA The 48th Annual Conference and.
How we set a DRL An example using CT
Sunday Case of the Day Physics
Optimisation of Patient Protection for Radiography
Sample General Electric CT Dose Report from a CT of the chest, abdomen, and pelvis. The dose report contains information about the patient (deleted), type.
Optimisation of Chest Computed Tomography Using a Phantom: Impact of mAs and Reconstruction Techniques on Image Quality C.S. Reis1; T. Faqir2; V. Harsaker3;
CT dose optimization software synergy – A clinical perspective
PRACTICAL APPLICATION OF Diagnostic Reference Levels
Head Computed Tomography Dose Output
Radiation Dose Management (RDM) Market to Cross $1.7 billion by 2023.
Exhibit Number: C19 Evaluation and Reduction of Head Computed Tomography Dose Because of full frame graphic use title slide sparingly because too many.
The efficacy of using CAD for detection of
Wednesday Case of the Day
Dose-Length Product to Effective Dose Conversion Factors for Common Computed Tomography Examinations Based on Canadian Clinical Experience  Idris A. Elbakri,
Yogesh Thakur, PhD, MCCPM, Thorarin A
UTILIZATION OF COMPUTED TOMOGRAPHY IN YANGON GENARAL HOSPITAL DR
Test tool for assessing lead equivalence in protective lead apparels
Roshan S Livingstone and Benedicta R Department of Radiology
Optimisation of paediatric x-ray examinations
The Radiography Student at The Johns Hopkins Hospital
Presentation transcript:

Head computed tomography doses in four selected hospitals in Nigeria Tom Adejoh (M.Sc) *Mark C. Okeji (Ph.D) **Musa Y. Dambele (M.Sc) Radiology Department NAU Teaching Hospital, Nnewi, Nigeria *Department of Radiography & Radiological Sciences, University of Nigeria **Department of Medical Radiography Bayero University, Kano, Nigeria PACORI CONFERENCE, TANZANIA 16 - 19 FEBRUARY, 2017

BACKGROUND OF THE STUDY Stochastic effects are main concerns of patient dose in CT (Ionnis, 2010). Justification, limitations and optimization are relevant principles of CT radiation protection ( ICRP, 1996). Efficient optimization → high quality images with minimal radiation dose (saravanakumar, 2014). Regular dose audits will produce efficient optimization (Huda, 2008). In Africa there is a paucity of literature on CT dose (Ogbole, 2014)

Global range for head CT LITERATURE REVIEW CT Dosimetrics CTDIvol (mGy) DLP (mGy-cm) (Foley 2012; IEC 2002) Global range for head CT CTDIvol: 32 – 77 mGy DLP: 925 -1050 mGy-cm (Saravanakumar, 2014; McCollough, 2009; Brix 2003) LITERATURE REVIEW

STATEMENT OF THE PROBLEM The region selected had the highest number of CT scanners of different models and radiation outputs. A perusal of series ‘999’ showed consistent higher dose than the 1999 European Commission recommendations of 60 mGy (CTDIvol) and 1050 mGy.cm (DLP). There was no evidence of dose survey since the scanners were installed. To audit the CT dose from the subregion as a quality control measure SIGNIFICANCE OF THE STUDY Awareness on dose output for the CT community Practical actions on dose optimization as a result of findings. AIM OF THE STUDY

Method & MATERIAL * Design: retrospective study * Place of study: Three tertiary hospitals (centres A-C) & a private diagnostic centre in Anambra State of Nigeria * Duration of study: February – August, 2016 *Ethical Considerations: approval & confidentiality settled *Inclusion criteria →for centre:highest throughput, available CT dosimetrics →for subjects: ≥ 18 years, nil scalp pathology * Scanners :GE, 4-slice; Toshiba, 16-slice (2x); Philips, 16-slice * Data analysis: SPSS version 20.0

Methodology A retrospective analyses of archived CT images of subjects aged ≥ 18 years Confidentiality of digital information was maintained by the activation of image anonymity features on the console. All the images that met the inclusion criteria were included in the study. The exposure parameters extracted from the first axial series and the on-screen CTDIvol and DLP for the subjects were recorded. The mean and 75th percentile of the CTDIvol and DLP were then calculated for each centre as well as for the four centres combined. Data were subjected to descriptive statistics and analyzed using analysis of variance.

Table 1: Anthropotechnical parameters for head CT RESULTS Table 1: Anthropotechnical parameters for head CT Variables Range Centre A Centre B Centre C Centre D M:F ratio 104:96 24:26 25:25 30:20 Age range 18 - 93 19 - 73 20 - 82 18 - 75 kVp 80 - 140 120 140 mA 10 - 350 220 250 mA modulation Yes/No No Yes Gantry rot time (s) 0.48 - 4 1 2 Scan range (mm) 180 - 320 200 Pitch 0.75 – 1.5 1.5 0.7 0.75 Scan mode Helical/axial Axial Helical Azimuth 0 - 360 90/180 0/90

Table 2: DOSE OUTPUT FROM THE CENTRES Variables European Commission 75th percentile Centre A B C D All centres Mean CTDIvol (mGy) 60 58 57 73 44 55 75th percentile CTDIvol (mGy) 59 86 46 66 Mean DLP (mGy) 1050 891 1195 1612 734 1108 75th percentile DLP (mGy-cm) 945 1414 1785 794 1444

Discussion Remedy Peer review Initial protocol Radiographer: Programmer: Initial protocol Peer review Radiographer: Remedy

DISCUSSION CONTINUED Initial programme: Default protocols that are faulty often give high radiation dose (centres B and C from table 1 & 2). → appropriate ab initio programme is a great step in dose optimization Peer review of dose output places an obligation on radiographer to optimize: output from Nigeria were higher than recommended values by EC and from literature Radiographer: The remedy for acceptable dose output lies with the radiographer’s regular dose audit.

RECOMMENDATIONS LIMITATIONS CONCLUSION * Dose output from the four centres in Nigeria were higher (66 mGy/ 1444 mGy-cm) than the recommendations of the European Commission (60 mGy/ 1050 mGy-cm). LIMITATIONS Absence of dosimetrics on some scanners decreased our sample size * A more widespread dose audit is advised * Regular dose audit of CT scanners is recommeded * Default protocol audit and remediation is also advised * Radiologists should request for dose summary to place an obligation on CT radiographers to be dose conscious RECOMMENDATIONS

REFERENCE Foley SJ, McEntee MF and Rainford LA. Establishment of CT diagnostic reference levels in Ireland. British Journal of Radiology, 2012; 85(1018):1390– 1397 Ogbole GI, Obed R. Radiation doses in computed tomography: Need for optimization and application of dose reference levels in Nigeria. West African Journal of Radiology, 2014;21(1):1-6 Huda W, Nickoloff EL, Boone JM. Overview of patient dosimetry in diagnostic radiology in the USA for the past 50 years. Medical Physics,2008;35:5713–28. International Electrotechnical Commission Medical electrical equipment—part 2–44: particular requirements for the safety of X-ray equipment for computed tomography. Geneva, Switzerland: IEC; 2002 McCollough CH, Primak AN, Braun N,Kofler J, Yu L, and Christner J, Strategies for Reducing Radiation Dose in CT. Radiological Clinics of North America, 2009; 47(1): 27–40. Saravanakumar A, Vaideki K, Govindarajan KN, and Jayakumar S. Establishment of diagnostic reference levels in computed tomography for select procedures in Pudhuchery, India. Journal of Medical Physics, 2014;39(1): 50– 55.