Automatic Pipettes Use And Calibration

Slides:



Advertisements
Similar presentations
Improving Pipetting Techniques For better accuracy and performance
Advertisements

Measuring Volume Biotech I pg. 301 to 323 in Basic Laboratory Methods.
Use and Maintenance of Micropipets
An Introduction to Microvolumetrics and Pipetting
Use and Maintenance of Micro-pipettes
Measuring Small Amounts Part 1: Using pipets and micropipets.
INTRODUCTION TO PIPETTES
Laboratory Techniques I: Dilution Go to browse and set to full screen.
Cautions! 1. Set pipet volume only within the range specified for that micropipet 2. Do not attempt to set a volume beyond the pipet’s minimum or maximum.
Oct. 22, 2012 Generosity Ludicrous: causing or deserving laughter because of absurdity Do Now: What is a microliter? How many microliter are in one liter?
Basic Laboratory Skills
Guide to Pipetting Pipette terms: Adjustment - altering the pipette so that the dispensed volume is within the specifications. Air Displacement Pipettes.
Use and Functional Verification of Automatic and Micro-Pipettes Module 7: Quality Assurance of Analytical Phase.
Micropipette Tutorial
Pipette Calibration Procedure Pipette, water and balance Test the maximum volume or minimum volume Pre-wet tips Test 5 volumes per setting.
Glucose test Ms. Ibtisam alaswad Ms. Nour A. taim.
Common Laboratory Glassware and Volumetric Glassware Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit.
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
Introduction of CLS 333.  Course Description:  This is a required course in biochemistry for preparing a generalist in medical technology program. The.
Module 11: Use and Care of Equipment At the HIV Rapid Testing Site.
Laboratory equipment.
Introduction to Equipment
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
Biochemistry Practical Dept.of Biochemistry Zhihong Li ( 李志红 )
Experimental Procedure. Overview The supernatant from a saturated calcium hydroxide solution is titrated with a standardized hydrochloric acid solution.
MEASUREMENT OF VOLUME. OVERVIEW First some general principles Then, more detail about micropipettes.
Biochemistry Practical Dept.of Biochemistry Zhihong Li ( 李志红 ) PhD.
Mastering the micropipette. What is a micropipette? How do you use a micropipette? How can I determine if I’m pipetting correctly?
Micropipettes Step-by-Step to use Micropipettes 1) Check the volume 2) Attach disposable tip (different tips for each size micropipette) 3) Depress the.
CHAPTER 1: Some Tools of the Trade Lab Purpose of Lab 1.1 Become familiar with the small volumes of solutions used in molecular biology Introduce.
How to use a micropipette. When is a micropipette needed? Micropipettes are precision instruments designed to measure and transfer small volumes of liquids.
 Accurately measure and transfer small amounts of liquid  Volumes commonly range from 0.5 – 2000  L (1/millionth of a liter aka )  2-20  L.
1. Which tool would you use for picking up 15 microliter? Which tool for 25 ml? Why? 2. Why do the little test tubes look this way? 3. Answer “Micropipette.
Micropipettes Why they are useful How they work How to care for them.
Experiment 1 Basic Laboratory Techniques Parts A, D, E, and an Unknown
Automatic pipettes use and calibration
Introduction to Titration Go to browse and set to full screen.
Proper Micropipette Technique and Slide Preparation
Use and Maintenance of Micro-pipettes
枣庄学院化学化工与材料科学学院 Preparation of Solution. 枣庄学院化学化工与材料科学学院 1. Grasp the basic Method and Operation of Preparing Solution 2. Study the Method of Use about.
Bangladesh BEST Programme
1 REGIONAL BIOTECH AT THE UNIVERSITY OF TEXAS-PAN AMERICAN BIOSCIENCE EDUCATION FOR THE RIO GRANDE VALLEY.
An Introduction to Microvolumetrics and Pipetting
Biochemistry LABORATORY NOTEBOOK AND EXPERIMENT REPORTS
Micropipette Tutorial
Fundamental Techniques and Measurements
Techniques in Molecular Biology 2017 Fall Lecture -2-
Measuring Volume.
MLAB 1335 Immunology Serology Terry Kotrla, MS, MT(ASCP)BB
مهارتك الأساسية في معامل التقنية الحيوية
Techniques in Molecular Biology 2017 Fall Lecture -2-
Measurement of Volume Chapter 20.
Metric System and Lab Skills
Use and Maintenance of Micro-pipettes
أجهزة قياس حيوي المحاضرة التاسعة.
Introduction To Medical Technology
Diluting Aqueous Solutions (page )
Introduction to Titration
An Introduction to Microvolumetrics and Pipetting
Micropipette Tutorial
How to Micropipette 1) Never adjust the volume below 5 uL or above 50uL 2) Never leave pipetter lying on bench with a tip attatched Basic Methods - Set.
Guide to Pipetting.
Pipette Technique.
Common Laboratory Glassware
Experiment 1 Basic Laboratory Techniques Parts A, D, E, and an Unknown
Measurement & Micropipetting
Laboratory Glassware.
Pipette Calibration.
Micropipette Tutorial
Presentation transcript:

Automatic Pipettes Use And Calibration Introduction To Medical Technology Lecture - 8 Automatic Pipettes Use And Calibration

Introduction Automatic pipettes are used in the biological, biochemical, and microbiological teaching and research laboratories to accurately transfer small liquid volumes. Glass pipettes which are used in chemical laboratories are not highly accurate for volumes less than 1 milliliter (1 ml), but the automatic pipettes are both accurate and precise. Each pipette can be set to transfer any volume within its own volume range. Although the complete set of seven Pipetman models covers the volume range from 0.1 l to 5000l (5 ml), you will most often use the P-20, P-200, and P-1000 models which are recommended for volume transfers of from 2 to 20 l, 50 to 200 l, and 100 to1000l (1 ml), respectively.

Pipetman Model Available

.

. Using a Micropipette Never exceed the upper or lower limits of these pipettes. Set the desired volume by turning the centrally located rings clockwise to increase volume or counterclockwise to decrease volume. The examples below show how to read the micropipette:

Place a tip on the discharge end of the pipette. The plunger will stop at two different positions when it is depressed. The first of these stopping points is the point of initial resistance and is the level of depression that will result in the desired volume of solution being transferred. The second stopping point can be found when the plunger is depressed beyond the initial resistance until it is in contact with the body of the pipette. At this point the plunger cannot be further depressed. This second stopping point is used for the complete discharging of solutions from the plastic tip. You should not reach this second stop when drawing liquid into the pipette, only when expelling the last drop.

Methods Of Pipetting Forward pipetting Reverse technique This is the most frequently used technique. In forward pipetting, an exactly set volume of liquid is aspired to the tip and then it is delivered to a new vessel. This technique is recommended for pipetting of diluted aqueous solutions, buffers, diluted acids and bases. Reverse technique  In reverse technique, a larger volume of solution is aspired to the tip. Then, an exact volume is delivered (and some liquid remains in the tip). This technique is suitable for very viscous or volatile liquids, biologic fluids, foaming solutions and for measuring very small volumes.

Forward Method Depress the plunger until you feel the initial resistance and insert tip into the solution, just barely below the surface of the liquid and not as deep as possible. Carefully and slowly release plunger Discharge the solution into the appropriate container by depressing plunger. This time, depress the plunger to the point of initial resistance, wait one second, and then continue pressing the plunger as far as it will go in order to discharge the entire volume of solution. Remove tip by pressing down on the tip discarder

Reverse Method Push the button to the second position. Dip the tip 2–5 mm bellow the level of the solution. Slowly aspire the solution to the tip. Remove the tip slowly from the solution. If necessary, wipe droplets from the external surface of the tip. Deliver the liquid to a new vessel, pushing the button to the first position only. Hold the button in the first position and remove the tip from the vessel. The liquid remaining in the tip can be returned to the stock vessel or thrown away.

NOTES Never point a pipette up. This may cause liquid to run down into the pipette and destroying it. When withdrawing liquids with the pipette, always release the plunger slowly. This prevents liquid from rushing into the end of the pipette and clogging it up. This is especially important with large volume pipettes (200-1000 μl). Be sure you use the proper size tip for each pipette. Always use a new tip for each different liquid. Use the correct pipette for the volume that is to be dispensed. Never use the 200-1000 μl pipette to dispense volumes below 200 μl. going below or above the range of the micropipette may damage the instrument.

NOTES cont. withdraw the pipette from the receiving vessel carefully, sliding the tip along the wall of the vessel. Holding the tip against the side of vessel is especially important when transferring small volumes of liquid. Remove the tip from the sample liquid. No liquid should remain on the Outside of the tip. Wipe away any droplets on the outside of the tip with a lint-free tissue, such as a gauze, but only wipe droplets from the side of the tip. Never Touch The Tip Orifice (opening) or you may absorb part of your sample.

NOTES cont. Pre-rinsing minimizes inter-tip or intra-tip variances due to any coating of the inner tip wall. When pipetting some solution, such as serum or other protein-containing solutions and organic solvents, a significant film may be retained on the inside of the wall of the disposable tips, resulting in an excessive volume error. Since this film remains relatively constant in successive pipetting with the same tip, excellent precision may be obtained by discarding the first filling of the tip and refilling the tip a second time or more times to use as the sample(s). This procedure is recommended when critical reproducibility (precision) is required.

NOTES cont. When pipetting concentrated acids of highly corrosive solutions such as halide salts, it is recommended that the pipette be disassembled after use to inspect the piston, shaft, and seal assemblies and clean them, if necessary. Air bubbles in the tip can greatly reduce pipetting accuracy. If an air bubble is trapped within the tip during intake, dispense the sample into the original vessel, check the tip immersion depth, and pipet more slowly. If an air bubble appears a second time, discard the tip and use a new one.

Pipette Calibration Principle Under a constant temperature and atmospheric pressure, the density of distilled water is constant. The volume of water can be determined by weighting dispensed water. The calibration of pipette is carried out determining the volume of water, the accuracy of measurements is effected by ambient temperature, atmospheric pressure and relative humidity. These factors are usually combined to give the Z factor, used in calculation of volume of water. Then the calculated volume of water is compared with the theoretical volume to determine the accuracy and precision of the pipette.

Material & Equipment Pipette and tips 50 ml beaker and plastic medicine cup Distilled water Temperature meter ( ±0.1℃ ) Analytical balance ( ±1.0 mg ) Atmospheric pressure meter

Procedure Determine the water temperature and record it. Place a beaker filled with distilled water into analytic balance and close the door of balance waiting for equilibrium of inner vapor. Place a plastic medicine cup on the pan and adjust the weight to zero. Put a tip onto the pipette and set the volume which is to be tested. Pre-rinse the tip: aspirate and dispense the setting volume three times and press the push – button on the second stop to remove any remaining liquid. Press the push-button to the first positive stop. Hold the pipette vertically, immerse the tip in the liquid and release the push button slowly and smoothly to aspirate the liquid. Wait one second and withdraw the tip from the liquid. Wipe any droplets away from the outside of the tip using a gauze.

Procedure cont. Place the end of the tip against the inside wall of the plastic cup at an angle of 10-40 °. Press the push-button smoothly to the first stop. Wait one second, change new site and press the push-button on the second stop. Keeping the push-button press to the end, remove the pipette by drawing the tip along the inside surface of the plastic cup and release the push button. Close the door of balance and record the value on the balance display after it has stabilized. Repeat step (6) to (12) 14 times. Eject the tip. Pipette should be calibrated every three months routinely.

Calculation Calculate the difference in weight between one reading and the next. Convert the weight unit of measured value into the volume unit of measured value using the following formula: Volume ( ml or ul ) = Weight ( mg or ug ) x Z Z value : conversion factor, which is conversion of density Calculate the average ( Mean ), accuracy, standard deviation (S.D.) and imprecision ( C.V. ) using the following formula:     n              n    2        Σ Xi           Σ ( Xi -X )         i=1               i=1      1 / 2 (Mean) X = ---------------- ;   S.D. = (----------------- )         N              N     S.D.             Measured value C.V. = ----------- X 100% ; Accuracy = --------------------- X 100%     Mean             Theoretical value Accuracy value must be 99-101 % and C.V. value must be less than 1 %.