Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University.

Slides:



Advertisements
Similar presentations
Partition Coefficients Lecture 26. The Partition Coefficient Geochemists find it convenient to define a partition or distribution coefficient of element.
Advertisements

Thermobarometry Lecture 12. We now have enough thermodynamics to put it to some real use: calculating the temperatures and pressures at which mineral.
X-ray Diffraction. X-ray Generation X-ray tube (sealed) Pure metal target (Cu) Electrons remover inner-shell electrons from target. Other electrons “fall”
Trace Element Abundances in Single Presolar SiC Stardust Grains by Synchrotron X-Ray Fluorescence (SXRF) Zhonghu Cai (XOR) Barry Lai (XOR) Steve Sutton.
Using FLUKA to study Radiation Fields in ERL Components Jason E. Andrews, University of Washington Vaclav Kostroun, Mentor.
Quantitative Analysis: Intensities to Concentrations
ED and WD X-ray Analysis
etc… Analysing samples with complex geometries Particles Inclusions
Thin Film Quantitation of Chemistry and Thickness Using EPMA John Donovan Micro Analytical Facility CAMCOR (Characterization of Advanced Materials in Oregon)
X-Ray Fluorescence Analysis by Gergana Hristozova Project supervisor: s. eng. M. Gustova FLNR.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron Probe Microanalysis EPMA
Electron-impact inner shell ionization cross section measurements for heavy element impurities in fusion reactors Jingjun Zhu Institute of Nuclear Science.
John Fournelle Eugene Cameron Electron Microprobe Lab Department of Geology and Geophysics University of Wisconsin - Madison Monte Carlo Simulations and.
Where are all the PGEs in the Platreef? The use of Laser Ablation ICP-MS in revealing trace element mineralogy David Holwell 1,2 and Iain McDonald 1 1.
Quantitative Analysis Quantitative analysis using the electron microprobe involves measuring the intensities of X-ray lines generated from your unknown.
Al, Mg, Si and Na Ka Peak Shifts in Common Silicate and Oxide Minerals: Relevance to Achieving the Goal of 1% Accuracy in EPMA John H. Fournelle Eugene.
Adventures of an Igneous Petrologist in a Probe Lab John Fournelle University of Wisconsin Madison, Wisconsin.
Update to End to End LSST Science Simulation Garrett Jernigan and John Peterson December, 2004 Status of the Science End-to-End Simulator: 1. Sky Models.
Version: November 16, 2009 EPMA of Particles Electron probe microanalysis EPMA.
Thermobarometry Lecture 12. We now have enough thermodynamics to put it to some real use: calculating the temperatures and pressures at which mineral.
IEEE NSS 2012 IEEE NSS 2007 Honolulu, HI Best Student Paper (A. Lechner) IEEE TNS April 2009 Same geometry, primary generator and energy deposition scoring.
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA Monte Carlo Simulations.
Introduction to Mineralogy Dr. Tark Hamilton Chapter 3: Lecture 7 The Chemical Basis of Minerals (sizes, shapes & directions) Camosun College GEOS 250.
SEM- Schematic Overview. Electron Detection Tungsten Filament Electron Source.
Electron probe microanalysis Low Voltage SEM Operation Modified 9/23/10.
The Significance of Refractive Index “K” Values LSMs, or, Why Long Wavelength Peak Markers Don’t Line Up Correctly John Fournelle* and K. Bolger*, J. Cook*,
Analysis of Fluorine in Fe-bearing phases: Problems and solutions John Fournelle Eugene Cameron Electron Microprobe Lab Department of Geology & Geophysics.
Al and Si Quantitation in Routine Silicate EPMA: Maybe not so routine John H. Fournelle Eugene Cameron Electron Microprobe Lab Department of Geology and.
2015 Cameca Users Meeting KαKα K L M Fluorescence and absorption… A sample of olivine has Fe, Mg and Si. Fe K α = 6.4 keV Binding energies… Mg K = 1.30.
Spatial Resolution and minimum detection
A Study of Reverse MC and Space Charge Effect Simulation with Geant4
by students Rozhkov G.V. Khalikov E.V. scientific adviser Iyudin A.F.
Quantitative EPMA Mapping of Minor and Trace Elements in Zircons
Laser wakefield accelerated electrons
Induced-activity experiment:
Electron probe microanalysis EPMA
The Problem of Secondary Fluorescence in EPMA in the Application of the Ti-in-Zircon Geothermometer And the Utility of PENEPMA.
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
What is XPS? XPS (x-ray photoelectron spectroscopy) is also known as ESCA (electron spectroscopy for chemical analysis). XPS provides chemical information.
International Workshop on radiosensitization
SEM and EPMA Some Advanced Topics October 7, 2009.
Christopher R. McGann, Ph.D. Student University of Washington
Electron probe microanalysis EPMA
Preface: What’s EPMA all about? How does Geology 777 work?
Electron probe microanalysis
Electron probe microanalysis
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron Probe Microanalysis EPMA
Electron Probe Microanalysis EPMA
The Materials Computation Center. Duane D. Johnson and Richard M
Bi-plasma interactions on femtosecond time-scales
Pamela Aaron, Dr. Charles Shearer, and Paul Burger
Electron probe microanalysis EPMA
Electron Probe Microanalysis EPMA
Electron probe microanalysis
Electron probe microanalysis EPMA
Geochemistry: Exploration, Environment, Analysis
Analysis of Optical Beam Propagation Through Turbulence
Electron Probe Microanalysis EPMA
Electron Probe Microanalysis EPMA
A Short Course on Geant4 Simulation Toolkit Introduction
Electron probe microanalysis EPMA
Nanocharacterization (II)
Electron probe microanalysis - Scanning Electron Microscopy EPMA - SEM
Electron probe microanalysis
Electron probe microanalysis EPMA
SEM and EPMA Some Advanced Topics October 7, 2009.
Electron probe microanalysis
Electron probe microanalysis EPMA
Presentation transcript:

Lecture 4a Common Error in EPMA: Secondary Fluorescence from Outside Primary Excitation Volume John Fournelle, Ph.D. Department of Geoscience University of Wisconsin Madison, Wisconsin, USA December 1, 2015 Nanjing

Common EPMA needs/wants Need to measure low ppm of some element Need to analyze small crystals in volcanic glass) Need to analyze small inclusions in minerals Need to determine if there is diffusion profile between two phases But simple EPMA can produce errors if SECONDARY FLUORESCENCE is not understood

“Simple EPMA” assumes…. That all X-rays detected come from the electron beam interaction volume

Fluorescence Problems UW- Madison Geology 777 Fluorescence Problems Alloy: 90 at% Ni 10 at% Fe Goldstein et al. p. 142 Secondary fluorescence is an important issue that must be appreciated. Generated X-rays are not scattered nearly as much as incident electrons, and thus the generated X-rays can travel relatively long distances (50 um in Fig 3.49) within the specimen and produce a second generation of X-rays. If the specimen (and standards) are relatively large (=homogeneous), this is not a problem.

Fluorescence Across Phase Boundaries UW- Madison Geology 777 Fluorescence Across Phase Boundaries Reed 1993, p. 258 However, if minor or trace elements are being analyzed in small grains (Phase 1) and the host phase (2) has high element abundance, an error can be made in the EPMA analysis.

Fluorescence across boundaries UW- Madison Geology 777 Fluorescence across boundaries “false Co” “false Cu” Cu Co

Secondary Fluorescence Correction UW- Madison Geology 777 Secondary Fluorescence Correction An American Mineralogist 2003 article (Llovet & Galan) reports an innovative approach to correcting the secondary fluorescence (SF) in diffusion couples and from adjacent phases. This utilizes a complex Monte Carlo program called PENELOPE that permits complicated geometric models of electron and X-ray behavior in materials. SF can be simulated in a model that represents the actual specimen.

Secondary Fluorescence Correction UW- Madison Geology 777 Secondary Fluorescence Correction SF can be simulated with Monte Carlo program PENELOPE in a model that represents the actual specimen (e.g. Fig 1).

Secondary Fluorescence Correction UW- Madison Geology 777 Secondary Fluorescence Correction Correction for SF in Ca in olivine using PENELOPE. The corrected Ca-in-olivine temperatures matched closer the 2 pyroxene thermometer temperatures than when uncorrected. Llovet and Galan, 2003.

Ti in Quartz Geothermometer Quartz fill of rutile needles Sato and Santosh, 2007, Mineralogical Magazine, 71, 143. This published study used the Ti measured in the quartz at ~10 microns from the rutile as the ‘true’ Ti content of the quartz. It gave minimum temperatures of 1190°C.

Ti in Quartz Geothermometer Sato and Santosh, 2007, Mineralogical Magazine, 71, 143. I simulated the quartz with a rutile, at various distances from the interface in the quartz. The simulation matches the published experimental data; my conclusion is that virtually of the Ti measured by EPMA is due to Secondary Fluorescence from the rutile.

UW- Madison Geology 777 An SF real story Another lab reported 10 wt% Nb (below) in what should have been Nb-free phase (by EDS at 30 kV). The issue was small grain size and nearby Nb (right image) When I ran WDS (18 kV) I found essentially zero Nb -- what is the problem? The original researchers used Nb Ka because, with EDS, it is impossible to resolve Nb La (it sits between Al Ka, Hf Ma and Pd La). Above: EDS spectrum on Pd2HfAl 5 um away from Nb. Pd Ka is very efficient at traveling across the border and exiting the Nb Fournelle, Kim and Perepezko (2005)

And I could reproduce their error by WDS – at 28 kV! Using Nb Ka And we see that also at 28 kV, there is minimal Secondary Fluorescence for Nb La line

Experimental Validation of Nb Ka Secondary Fluorescence in Nb-free Pd2HfAl

PENELOPE/PENEPMA Rigorous Monte Carlo program that tracks both electrons and photons Complex geometries possible Steep learning curve on original software Newer interfaces easier to use: PyPenelope (Philippe Pinard)- full app (go to SourceForge to download) Probe for EPMA – simplified app for secondary fluorescence (included in CALCZAF/Standard). A tutorial is at the ProbeSoftware Forum.

A GUI for the calculation program can be downloaded from http://epmalab.uoregon.edu/calczaf.htm)