Occupational Radiation Protection during High Exposure Operations

Slides:



Advertisements
Similar presentations
TRAINING COURSE ON X-RAY FOR GP
Advertisements

Radiation Safety Training Lab Survey Program Washington State University Radiation Safety Office.
1 Radiological protection, dosimetry & environmental monitoring National Training Course on Radiopharmacy, Radiochemistry and Radioisotope Production Site.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Radiation Safety Course Radiation Monitoring Heath de la Giroday Dispensing Chemist Radiation Safety Officer.
Radiation Detection Systems
Detecting and Measuring Ionizing Radiation -2
4/2003 Rev 2 II.3.6 – slide 1 of 18 Part IIQuantities and Measurements Module 3Principles of Radiation Detection and Measurement Session 6Photographic.
Radiation Detectors / Particle Detectors
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO INTAKES OF RADIONUCLIDES Interpretation of Measurement Results.
Nuclear Radiation Basics. Copyright © 2011International Medcom Contents What is Radiation? What is Radiation? Instruments that Measure Nuclear.
Familiarization and Operation
Industrial Radiography
International Atomic Energy Agency Criteria for the Need for Monitoring ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO INTAKES OF RADIONUCLIDES.
Establishment of a Training Dosimetry Laboratory in Cairo University Dr. Hassan Fathy for NSPA.
Estimation of the effects of a lead vest on dose reduction for NPP workers using Monte Carlo calculations KIM JEONG-IN.
Environmental Health XIV. Standards and Monitoring Shu-Chi Chang, Ph.D., P.E., P.A. Assistant Professor 1 and Division Chief 2 1 Department of Environmental.
IAEA International Atomic Energy Agency Emergency Response Protective Actions Day 10 – Lecture 3.
Gauges and well logging
Alexander Brandl ERHS 630 Effective and Committed Effective Dose Environmental and Radiological Health Sciences.
Intervention for Chronic and Emergency Exposure Situations Basic Concepts of Emergency Preparedness Personal Protective Equipment, Devices and Procedures.
PART IX: EMERGENCY EXPOSURE SITUATIONS Module IX.1: Generic requirements for emergency exposure situations Lesson IX.1-2: General Requirements Lecture.
Lecture 6 Review of Units used in Radiation Safety and Radiation Biology Assessment of Public Dose Unauthorized Use/Removal of Radioactive Materials Misadministrations.
EMERGENCY PLAN AND PROCEDURE IN INDUSTRY INVOLVING NORM/TENORM
Principles of Radiation Detection
Protection Against Occupational Exposure
IAEA International Atomic Energy Agency Assessment of External and Internal Exposures Michael Hajek Radiation Safety and Monitoring Section Division of.
 Justification is the answer.  Dose limits are not applicable only recommened.
IAEA International Atomic Energy Agency PGEC Part VI Planned Exposure Situations - Generic Requirements Module VI.2 Requirements for occupational exposure.
Organization and Implementation of a National Regulatory Program for the Control of Radiation Sources Inspection Part II.
1. 2 Radiation Safety 3 What is Radiation? Radiation is a form of energy. It is emitted by either the nucleus of an atom or an orbital electron. It.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Records and Record.
Intervention for Chronic and Emergency Exposure Situations Assessment and Response during Radiological Emergency Dose Assessment Overview Lecture IAEA.
International Atomic Energy Agency Assessment of Internal Exposure following Accidents or Incidents ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO INTAKE OF.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Workplace Monitoring.
Ferris State University & Michigan Department of Career Development 1 Radiation Safety Answer Key.
Introduction to nuclear medicine technology NMT 231 Aya Ahmed Saeed.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Dosimetry Services.
ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO INTAKE OF RADIONUCLIDES
Organization and Implementation of a National Regulatory Program for the Control of Radiation Sources Technical Services.
HP SURVEY INSTRUMENT CALIBRATION AND SELECTION PRINCIPLES OF RADIATION DETECTION AND QUANTIFICATION January 13 – 15, 2016 TECHNICAL MANAGEMENT SERVICES.
Radiological Monitoring and Measurement 1 OBJECTIVES Allow the participants to understand the concepts of measuring dose received by radiation workers.
IAEA International Atomic Energy Agency Occupational Radiation Protection during High Exposure Operations Lecture 4 Medical Management of Radiological.
IAEA International Atomic Energy Agency Lecture 5 – Internal dose assessment and interpretation of measurement results Postgraduate Educational Course.
IAEA International Atomic Energy Agency PART V: ASSESSMENT OF EXTERNAL AND INTERNAL EXPOSURES (OTHER THAN MEDICAL) Module V.1 - Assessment of occupational.
Technical Services. Objectives To identify the technical services needed within the infrastructure for an effective implementation of regulatory programme.
Radiation Protection RTMR 284 CHAPTER 20 PART II.
Radiological impacts from nuclear industrial facilities on the public and the environment : Their magnitude and the next 50 years forecast Sylvain Saint-Pierre.
Louisiana State University Radiation Safety Office
RADIATION PROTECTION 04/12/2016.
Lessons learned from the surveillance: Measuring methods and monitoring strategies T. R. Beck, E. Ettenhuber Federal Office for Radiation Protection, Germany.
Occupational Radiation Protection during High Exposure Operations
Topic 6 : Atomic and Nuclear Physics.
Industrial Radiography
Transposition of Requirements set out in the Basic Safety Standards for Nuclear Facilities in Lithuania Gintautas KLEVINSKAS Albinas MASTAUSKAS Radiation.
6th European ALARA Network Workshop Madrid, Spain, October 2002
Occupational Radiation Protection during High Exposure Operations
Catalina Chitu, Vasile Simionov, “CNE-PROD Cernavoda” NPP, Romania
Radiopharmaceutical Production
ADEQUACY OF EXISTING AEROSOL SAMPLERS FOR MONITORING NORM EXPOSURES European ALARA Network 9th Workshop on Occupational Exposure to Natural Radiation.
IONIZING RADIATION 1.Introduce self
Principles of Radiation Protection
HEALTH PHYSICS INSTRUMENTATION
1 Principles of Radiation Detection Operational Radiological Safety Course U.S. Army Chemical Biological Radiological & Nuclear School Edwin R. Bradley.
There are two kinds of radiation monitors used for medical purposes: survey monitors personal monitors.
Occupational Radiation Protection during High Exposure Operations
Estimation of the effects of a lead vest on dose reduction for NPP workers using Monte Carlo calculations KIM JEONG-IN.
Film badge: Measure and record radiation exposure due to gamma rays, X-rays, and Beta particles. Permanent record Cheaper Energy dependence Fading film.
 To provide an overview of external and internal dosimeters used for accupationl radiation protection 
 To provide an overview of external and internal dosimeters used for accupationl radiation protection 
Presentation transcript:

Occupational Radiation Protection during High Exposure Operations IAEA Training Material on Occupational Radiation Protection Occupational Radiation Protection during High Exposure Operations Lecture 10 Practical Radiation Monitoring and Dose Evaluation

Contents Monitoring Dose evaluation Workplace monitoring Individual monitoring Dose evaluation External exposure Internal exposure Total dose

Introduction Working in high exposure situations requires both individual and workplace monitoring. Workplace monitoring is used to avoid high doses to the exposed workers and individual monitoring is used to see what happened after the work is completed (passive monitors) and to control the work during operations (active monitors).

1. Monitoring Workplace monitoring General principles Dose rate measurement Airborne contamination measurement Surface contamination measurement 4

Workplace monitoring General principles Source related monitoring In nuclear or radiological emergency situations, the dose rate, airborne contamination and surface contamination in the workplace of emergency workers should be monitored. Furthermore it is important to continuously check for changes in the radiation levels. Workplace monitoring can be divided into two types: Source related monitoring Task related monitoring 5

Workplace monitoring Source related monitoring Monitoring performed continuously and/or periodically at fixed monitoring points in order to supervise the situation of the source and the facilities. Task related monitoring Monitoring performed beside emergency workers during a task in order to control the individual dose of emergency worker adequately. 6

Workplace monitoring In facilities where fixed monitoring equipment is installed, the radiological situation in the facilities can be checked by indicator panel or computer terminal in the control room of the facility. In the facilities without fixed monitors, radiation measurement should be performed by using portable radiation monitors. 7

Dose rate measurement A sufficient quantity of suitable dose rate meters capable of providing direct readings of the ambient dose equivalent rate (H*(10)/h) with ranges from the μSv/h to 1 Sv/h should be available. Where high levels of surface contamination with beta/gamma emitters is foreseen, equipment calibrated in H’(0.07)/h should also be available. The equipment should be type tested and appropriate for work in field conditions – for example some detectors with liquid crystal screens are hard to read in broad daylight. The screen should have back lighting for night work. 8

Dose rate measurement The energy range and ruggedness, battery life and availability, temperature range and other considerations are important criteria. The detector should “fail to safe” if it goes off scale. 9

Dose rate measurement As the dose rate meters are expected to read high dose rates, and that the correct measurement of the dose rates is fundamental to the safety of the exposed workers, it is important that: Before making the reading, the worker should assure that he or she is in a safe position or situation. When entering a high dose rate area keep the meter turned on and check the alarm settings beforehand. The worker using the dose rate meters should be familiar with the equipment through use during routine work. The dose rate meter passes through a functional test every day before use: battery, cables, background and if possible a check source. In high dose rate areas, do not take the measurement too quickly, take the minimum time necessary to make a correct reading. Record the measurement. The detector can be wrapped in plastic to avoid contamination. 10

Detection principles – gas based detectors

G-M counters G-M counters connected to a count rate meter (scaler) are the most widely used instrument due to their low cost, simplicity, reliability, and high sensitivity. The G-M counter does not measure dose rate. It registers “counts” produced by the incident radiation which produces an ionization in the sensitive volume. The G-M survey meter can be made into a fairly accurate indicator of dose rate over a wide range of photon energies. This is done through the use of filters around the detector (energy compensation) The G-M can saturate at high very dose rates. In this case, ionization chamber survey instruments should be used.

Scintillator based dose rate measurement

Scintillators The total current from the photomultiplier is a measure of the energy deposited, and therefore the dose. Plastic scintillators are tissue equivalent. They are fragile, temperature dependant and high cost. They also have high sensitivity which means they are not appropriate for high dose rate work. Scintillators an be used to make accurate dose rate measurements relevant for public protection, around or below 1 µSv/h.

Neutron survey instruments High neutron dose rates are very rare in emergency work, the exception being the continued criticality of nuclear material. The suppliers provide detectors based on boron trifluoride (BF3) or helium-3 (3He) gas. A simplified drawing of a neutron detector is shown below. D: display P: proportional counter incorporating BF3 or 3He M: cylindrical or spherical polyethylene moderator to thermalize incident fast neutrons for detection S: perforated cadmium or boron plastic sheath which modifies the energy response according to neutron quality factors to enable the direct measurement of dose equivalent rates 15

Measurement of airborne contamination Airborne contamination meters indicate the potential internal exposure when a radioactive material is distributed within the atmosphere. Grab samples may be made and then counted using pancake type surface contamination detectors, the problem being the presence of naturally occurring radionuclides. For better protection in areas where high levels of air contamination are suspected, an alarming air sampler is recommended. These samplers contain battery powered pumps and implanted silicon detectors with energy discrimination that allows the natural background to be subtracted. 16

Measurement of airborne contamination For work inside buildings, especially with high specific activity alpha emitters, semi-portable continuous air monitors for alpha and beta emitters with alarms can be used. 17

Surface contamination monitors Surface contamination monitors are usually designed to measure alpha and beta emitters deposited on a surface. Most surface contamination monitors indicate in counts/s (s-1) and the instrument needs to be calibrated for the particular radiation being detected to enable the indicated reading to be converted into meaningful units such as Bq/cm2. Alpha contamination monitors have a very low background count rate and can be used in high gamma radiation fields. Beta contamination monitors will have their reading altered by the gamma background, unless adequately shielded. In this case an indirect measurement (wipe) will be indicated. Operational limits for surface contamination in Bq/cm2 should be established during the planning phase. 18

Surface contamination monitors G-M pancake type surface contamination monitors are the most used due to their ruggedness and ease of use. As with all pancake detectors with mica window, the detection efficiency of alpha surface contamination is quite low. 19

Individual Monitoring – External exposure – An individual dosimeter is used for measurement of an external dose: Passive dosimeter Active dosimeter A passive dosimeter measures an accumulated dose. Since the battery is unnecessary, the passive dosimeter is smaller and lighter, and it cannot run out of power. However, it does not have a direct readout and cannot measure a change of dose rate, or have preset alarms to provide a warning of a change of working conditions. Typical passive dosimeter are TLD, OSL, RPL, Film etc. An active dosimeter is an individual dosimeter which integrates the counts of radiation and measures dose rate and the accumulated dose. Since the active dosimeter needs the power of a battery it generally is not suitable for continuous use over several days, and it may not be useable if there is a power blackout. 20

Individual Monitoring – Internal exposure – Typical methods of individual monitoring for intakes are whole body counting, organ counting (such as thyroid or lung monitoring) and analysis of samples of excreta. Sampling of the breathing zone with personal air samplers can also be used. In many circumstances involving exposure due to radionuclides, workplace monitoring will be needed. Monitoring for the estimation of doses from intakes of radionuclides may include one or more of the techniques: Sequential measurements of radionuclides in the whole body or in specific organs Measurements of radionuclides in biological samples such as excreta or breath Measurement of radionuclides in physical samples such as filters from personal or fixed air samples, or surface smears 21

2. Dose evaluation After an accident has occurred, the radiological consequences may be complicated by trauma or other health effects incurred by the workers. Medical treatment of injuries, especially those that are potentially life threatening, generally takes priority over biological operations, including exposure assessment. In such case, post-accident dose assessment should be conducted when the medical situation has been brought under control. 22

External exposure – calculation of external dose – An external dose is calculated based on the measurement result of a individual dosimeter. For radiation protection purposes the measured operational quantities Hp(10) and Hp(0.07) are interpreted in terms of the protection quantities effective dose E and equivalent dose to the skin and extremities HT. Realistic assumptions have to be made with respect to the type and uniformity of the radiation field and the orientation of the worker within the field. Under this conditions, the dosimeter reading gives a good estimate of the worker’s exposure without underestimating or severely overestimating the relevant protection quantity. 23

External exposure – calculation of external dose – In cases where worker moves about the workplace, 4 types of multidirectional field should generally be considered: with radiation incident predominantly from the front half space (AP) from the rear half space (PA) with radiation incident symmetrically from all directions perpendicular to the body (ROT) with radiation incident isotropically from all directions including above and below (ISO) If the radiation is expected to come from the rear, the dosimeter should be worn on the back. If the dose approaches the relevant limit or reference level, an appropriate correction factor should be applied for AP, PA, ROT or ISO geometry, based on a knowledge of the radiation and the conditions of exposure. 24

External exposure – dose reconstruction – There may be cases where the personal dosimeter has not been worn. Also there can be cases where the measurement result of a individual dosimeter may be inappropriate as a substitution of an effective dose or equivalent dose because of non-uniform exposure. In these cases, it is necessary to perform evaluation of an effective dose or an equivalent dose by an experiment or calculation. In order to verify the measurement result of a personal dosimeter the reconstruction experiment of the exposure situation is conducted if needed. 25

External exposure – whole body and extremity dose – When worker is exposed to the radiation to the whole body uniformly, the value of a personal dosimeter can be seen to be equivalent to the effective dose. In case of non-uniform exposure, equivalent dose of each tissues or organs should be evaluated based on the measured value by individual dosimeter and the result of the dose reconstruction. An effective dose can be evaluated by multiplying those equivalent doses by the corresponding tissue weighting factors. Equivalent dose to the skin of a hand can be evaluated by using the value of Hp(0.07) measured by a finger ring dosimeter. 26

Internal exposure Accidents may result in releases of radioactive materials into the working environment with the potential for high doses to the workers. Details of implementation of dose assessment of exposure due to intakes of radionuclides are described in the IAEA Safety Guide No. RS-G-1.2. 27

Internal exposure – calculation of internal dose – Once assessment of internal exposure has commenced, as much information as practicable should be gathered. Because intakes associated with accidents or incidents can result in committed effective doses which approach or exceed dose limits or reference levels, individual and specific data on the material involved are normally needed for exposure assessment. An adequate assessment of dose can be made only after considering all of the data, resolving the sources of inconsistency as far as possible, and determining the most likely and worst possible scenarios for the exposure and the magnitude of any intake. 28

Internal exposure – dose calculation code – In general, the predicted value of retention, the predicted value of excretion and the dose coefficients depend on the many parameters such as nuclide, physical and chemical form, path of intake, intake pattern (acute or chronic), time after intake, etc. For the purpose of considering these parameters, various codes for calculating a dose of internal exposure have been developed [PHE, NIRS, JAERI, etc.]. 29

Internal exposure – whole body and organ dose – The whole body dose is expressed with an effective dose. An effective dose is calculated by multiplying an intake by the corresponding dose coefficient. The tissue dose and organ dose are expressed in terms of the tissue equivalent dose. An equivalent dose is calculated by multiplying an intake by an equivalent dose coefficient. The effective dose coefficient and the equivalent dose coefficients are given in the ICRP Publication 103. 30

Total dose Effective dose Equivalent dose Lens of eyes The effective dose from external exposure and the committed effective dose from internal exposure should be totalled. Equivalent dose Lens of eyes The equivalent dose of lens of eyes should be evaluated based on Hp(3) measured by the personal dosimeter worn on a head or a collar, etc. When mixed fields of photon, beta-rays, and neutrons exist in a workplace, the Hp(3) should be totalled. 31

Total dose Equivalent dose Skin In a workplace where a worker receives almost uniform exposure to medium or high energy photon radiation, it can be considered that Hp(10) as measured by the personal dosimeter is an equivalent dose of the skin. When it is assumed that the dose of the skin becomes high locally, an extremity dosimeter should be set on the part of body and it can be considered that Hp(0.07) measured in the part is the equivalent dose of the skin. When skin contamination or work clothing contamination is discovered, the skin dose should be evaluated based on the measurement result of surface contamination on the skin or clothes. 32

Total dose Equivalent dose Thyroid The external dose of the thyroid can be estimated from Hp(10) measured with the personal dosimeter. In uniform exposure, Hp(10) can be considered to be an equivalent dose of the thyroid. In non- uniform exposure, the discrepancy in Hp(10) brought about by the difference between the location of the thyroid and the wear position of individual dosimeter should be corrected if necessary. The internal dose of the thyroid is calculated by multiplying an intake by a dose coefficient to obtain the equivalent dose of the thyroid. 33

Total dose Equivalent dose Other tissues and organs In uniform external exposure, Hp(10) can be considered to be an equivalent dose of the thyroid. In non-uniform exposure, the equivalent dose of the organ aimed is estimated by suitable method, such as dose reconstruction. The equivalent dose of each tissue and organs is defined as the sum of the equivalent dose by external exposure and the equivalent dose by internal exposure. 34