The magnitude of H&N IMRT dose delivery errors from three possible failure modes: beam quality, symmetry, and MLC position Jackie Tonigan, M.S. Advisor:

Slides:



Advertisements
Similar presentations
بسم الله الرحمن الرحيم ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
Advertisements

Ionization Chamber Array for External Beam Radiotherapy
Commissioning an Anthropomorphic Spine and Lung Phantom for Remote Dose Verification of Institutions Participating in RTOG 0631 Douglas Caruthers, M.S.;
In the past few years the usage of conformal and IMRT treatments has been increasing rapidly. These treatments employ the use of tighter margins around.
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Slice Thickness Interpolation: The effect of interpolated slice thickness on the 2D vs 3D gamma results are shown in Table 3 for the QA data set only.
Anatomy-based MLC Field Optimization for the Treatment of Gynecologic Malignancies Myriam Bouchard M.D. Nadeau S, Germain I, Raymond P.E., Harel F, Beaulieu.
Challenges in Credentialing Institutions and Participants in Advanced Technology Clinical Trials Geoffrey Ibbott, David Followill, Andrea Molineu, Jessica.
Photon Beam Monitor-Unit Calculations
Algorithms used in heterogeneous dose calculations show systematic error as measured with the Radiological Physics Center’s anthropomorphic thorax phantom.
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID.
tomos = slice, graphein = to write
Quality Assurance: Manufacturer & Clinical Aspects  Alan Cohen, M.S. DABR  Paul Naine, MSc. MIPEM  Jim Schewe, PhD, DABMP Accuray Incorporated Elekta.
Innovation/Impact: By designing a simulated human shaped (anthropomorphic) plastic phantom with targets, organs at risk (OAR) and heterogeneities, the.
Introduction Modern radiation therapies such as intensity-modulated radiation therapy (IMRT) and volume modulated arc therapy (VMAT) demand from dose calculation.
Radiotherapy Planning for Esophageal Cancers Parag Sanghvi, MD, MSPH 9/12/07 Esophageal Cancer Tumor Board Part 1.
Impacting Radiotherapy Safety and QA with Device Improvements Benedick A Fraass, PhD, FAAPM, FASTRO, FACR Allen S. Lichter Professor of Radiation Oncology.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer.
بسم الله الرحمن الرحيم و قل رب زدنى علماً ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
Challenges for TPS Chunhua Men Elekta Software, Treatment Planning System BIRS Workshop Banff, Canada 3/12/2011.
Data Mining to Aid Beam Angle Selection for IMRT Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil- American University.
Patient Plan Results: Table 3 shows the ratio of the Pinnacle TPS calculation to the DPM recalculation for the mean dose from selected regions of interest.
Kelly Younge, Ph.DKelly Younge, Ph.D Don Roberts, Benedick Fraass, Daniel McShan, and Martha Matuszak University of Michigan, Department of Radiation Oncology,University.
Comparison of Clinical Parameters for Proton Therapy in the United States Paige Summers, MS.
IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN
The RPC Proton Therapy Approval Process
FDA Public Meeting: Device Improvements to Reduce the Number of Under-doses and Misaligned Exposures from Therapeutic Radiation.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Radiological Physics Center David Followill, Ph.D. and RPC Staff.
6. Clinical implementation and SBRT quality assurance
Use of the IC Profiler detector array for comprehensive machine QA ESTRO QA & Dosimetry Satellite Symposium Steve Morgan, Medical Physics Dept,
Dosimetric discrepancies due to positional errors in MLC movement during stereotactic lung VMAT A. Abbas, T. Karan, S. Kim, D. J. Moseley, M. M. Taremi,
The Radiological Physics Center’s Anthropomorphic Quality Assurance Phantom Program Carrie F. Amador, Nadia Hernandez, Andrea Molineu, Paola Alvarez, and.
1 A Comprehensive Study on the Heterogeneity Dose Calculation Accuracy in IMRT using an Anthropomorphic Thorax Phantom S Davidson 1, R Popple 2, G Ibbott.
1 Radiotherapy, hadrontherapy and treatment planning systems. Faiza Bourhaleb INFN-Torino University Med 1er-Morocco  Radiotherapy  Optimization techniques.
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Effectiveness of risk management process towards error reduction in the laboratory at Sakra World Hospital AUTHORS: Dr. Shabnam Roohi Mr. Deepak Agarkhed.
Automated QA at Breast Cancer Rounds – A process to improve efficiency and quality of patient care Dr. Kathy Rock Breast Cancer Clinical Fellow Rock K,
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Development of elements of 3D planning program for radiotherapy Graphical editor options  automated enclose of contour  correction of intersections 
Retroactive Calculation of TLD and Film Dose in Anthropomorphic Phantom as Assessment of Updated TPS Performance H. Alkhatib 1, S. Oves 1, B. Tsang 1,
Dr. Malhar Patel DNB (Radiation Oncology)
IAEA E-learning Program
Kasey Etreni BSc., MRT(T), RTT, CTIC
INTRODUCTION RESULTS DATA ANALYSIS DISCUSSION METHODS REFERENCES
CHAPTER11 Project Risk Management
FMEA.
FMEA PROCESS FLOW Causes/ failure mechanisms Product definition
FMEA PROCESS FLOW Determine Causes/ mechanisms failure
IAEA E-learning Program
Linac Commissioning Overview
IAEA E-learning Program
Left Posterior Superior Right Anterior Inferior
Fig. 4. Percentage of passing rate between clinical and 544 plans.
GE 6757 TOTAL QUALITY MANAGEMENT
Reducing Treatment Time and MUs by using Dynamic Conformal Arc Therapy for SBRT Breath-Hold Patients Timothy Miller, Sebastian Nunez Albermann, Besil Raju,
Insert tables Insert figure
Volumetric Modulated Arc Therapy (VMAT) versus Intensity Modulated Radiation Therapy (IMRT) for Anal Carcinoma Heather Ortega, BSRT(T), CMD, Kerry Hibbitts,
RSNA Meeting, Chicago, IL
Ch 10. A System of Dosimetric Calculations
Failure Mode and Effect Analysis
GHG meeting at ESTRO36 May, 2017
Presentation transcript:

The magnitude of H&N IMRT dose delivery errors from three possible failure modes: beam quality, symmetry, and MLC position Jackie Tonigan, M.S. Advisor: David Followill, Ph.D University of Texas M. D. Anderson AAPM Annual Meeting, Charlotte, NC. August 2, 2012

Increased Complexity and Steps Motivation Advanced RT Increased Complexity and Steps Need for Increased QM We are in a time of great advancement in radiotherapy. Our ability to treat with more conformal, higher doses is increasing and techniques for accurate target localization are improving. The motivation of this work lies in the fact that with these advances in our treatment techniques comes increased complexity in the treatment process. There are more devices, more inputs, more interfaces, and more complicated procedures. In turn, we have a greater need for comprehensive quality management and with realistic clinical limitations on personnel and time, it is important to have a strong understanding of the entire treatment process and the means by which the process may fail in order to most effectively and efficiently (manage the quality of patient treatments) prioritize quality assurance.

Sensitivity Analysis RPN=O x D x S Goal: Prioritize QM based on likelihood of error occurrence and severity of clinical impact Failure Modes and Effects Analysis Outline the process Identify failure modes and causes Determine the risk associated with each FM RPN = Risk probability number O = Probability of occurrence D = Lack of detectability S = Severity Improve QM RPN=O x D x S Sensitivity analysis is an effective way to get a good understanding of the treatment process and the risk associated with failures so that you may prioritize QM, and a number of studies in the radiotherapy community are focusing on Failure modes and effects analysis, or FMEA. An FMEA works first by outlining each step in a process, commonly done in the form of a process tree, then identifying places that errors may occur which would result in a failure of the process, known as a failure mode, and each possible cause of the failure. From there, the risk associated with each failure mode is determined by a score known as the risk probability number, which takes into account the probability of occurrence, the lack of detectability, and the severity. A scale is set and each of these values are determined by a panel of experts based on their opinion of the risk of each component. And the purpose of this work is to reduce the subjectivity of these scores by better understanding the impact of the severity of IMRT failure modes with physical measurement and calculation with the ultimate FMEA goal of identifying the most risky failure modes and using the information to improve quality management.

Couch Angle & Displacement IMRT Process Tree MU Linearity Localization Immobilization Image Transfer Positioning CT Lookup Table Simulation Imaging Treatment Planning IMRT QA Treatment Delivery Measurements Interpret Results Dose Calculation Equipment MLC Penumbra Effects MLC Positioning Beam Quality Symmetry Gantry Angle Treatment Delivery So here is a simplified depiction of the IMRT process, starting off with very basic steps, each of which then have a number of components with a number of possible failure modes and subsequent causes. My project is focusing on what I like to call physical failure modes within the treatment delivery step, and will be measuring the severity of dose delivery errors caused by failures of these listed failure modes. This specific study focuses on these three: MLC positioning, beam quality, and symmetry. Collimator Angle Contouring Prescription Plan Approval Plan Save & Transfer MLC Delivery Mechanical Delivery Localization Motion Monitoring Accessories Couch Angle & Displacement Dose Calculation

Unique Study Characteristics Extensive array of measurement capabilities (RPC) Availability of an accelerator (Clinac 2100CD) Just taken out of clinical service Modification of operating parameters Within this study are some unique characteristics. First, we had use of the RPC’s extensive measurement capabilities and second, because MDA has been replacing machines, we have access to accelerators being taken out of clinical service that we can, with the assistance of the service engineers, modify their operating parameters

Failure Mode Measurements MLC position Beam quality Symmetry Standard IMRT plan (1948 MU, 90 seg, 0.482 MCS*) Complex IMRT plan (3189 MU, 216 seg, 0.171 MCS*) Output Beam quality – ratio of depth doses Profiles (in-plane and cross-plane) Planar IMRT QA – MatriXX Phantom IMRT QA – I’mRT Body Phantom RPC IMRT H&N Phantom (end-to-end test) Here is a list of the measurements performed for each of the failure modes. We delivered both a standard H&N IMRT plan, and a more complex plan with higher MU, more segments, and a low modulation complexity score. We controlled for output, measured the beam quality with each adjustment by taking a ratio of depth doses, took profiles, performed planar and 3D phantom IMRT QA and finally delivered the plans to the RPC’s IMRT H&N phantom which as an end-to-end test provides a complete treatment assessment. A baseline was established through measurement of each of these before any alterations were made. *McNiven, et al. Med. Phys. 2010

Failure Mode Measurements MLC Position Systematic + 2 mm shift in one bank of leaves Beam Quality DD ratio increased/decreased by ~1% Symmetry Angular Steering: In-plane by ~3.5%, Cross-plane by ~3.5% Translational Steering: In-plane by ~3.5% Then each individual failure mode was simulated and the same list of measurements taken. For the MLC positional failure mode, each field was delivered with a systematic 2 mm shift outward in one bank of leaves. For beam quality, the shunt was altered to increased and subsequently decrease the depth dose ratio by 1%. The beam was steered angularly to shift the in-plane and cross-plane symmetry by 3.5% each individually, and finally it was steered translationally to shift the in-plane symmetry by 3.5%. Each of the failure modes was evaluated as an isolated event.

g-analysis difference Pilot Study Results Profiles - MatriXX No IMRT QA RPC IMRT H&N Phantom, complex plan: Range of 6 PTV TLD dose differences from baseline irradiation DTA difference from baseline irradiation (between PTV and OAR) 7%/4mm gamma index analysis average of axial and sagittal films In-plane Cross-plane Beam quality MLC (all) Alteration 5% 3% 1% (TMR2010) + 1 mm + 3 mm DTLD (%) 1 3 6 16 37 DDTA (mm) 0.2 0.9 0.7 2.5 6.8 g-analysis difference 5.2% 0.7% 23.5% 20.4% 33.1% We performed a quick pilot study on a machine going out earlier in the year with a complex treatment plan on the same RPC IMRT H&N phantom and these are the results. Here we have the difference between the baseline delivery before alterations and after alterations in the in-plane symmetry of 5%, cross-plane symmetry of 3%, beam quality of +1%, and systematic MLC shifts out in both banks of leaves by 1 and 3 mm. As expected, rather large differences were seen in deliveries with MLC positional errors. Substantial errors were also seen in each of the other failure modes, most notably the beam quality. These results encouraged us to further investigate these failure modes.

Results – MLC Offset 2% 3% 2/6 4/6 + 2 mm shift, one bank of leaves Standard Plan Complex Plan Dose difference from baseline 2% 3% # of TLD with sig. difference (p<0.05) 2/6 4/6 While analysis of the almost 100 films is ongoing, I have here the TLD dose results for the recent measurements I described. For the MLC systematic 2 mm offset in one bank, we have dose differences from the baseline delivery of up to 2% in the standard plan delivery and 3% in the complex plan delivery, with 2 and 4 of the 6 PTV TLD in the IMRT H&N phantom have significant differences from baseline.

Results – Beam Quality DD ratio increased by 1.1% DD ratio decreased by 0.6% Standard Plan Complex Plan Dose difference from baseline 2% 4% # of TLD with sig. difference (p<0.05) 0/6 6/6 For our adjustments in beam quality, there was up to 4% significant dose difference in all TLD for the complex plan here, and no significant differences in the standard plan. The decreased depth dose ratio delivery follows a similar trend with up to 2% dose difference from baseline in each plan, but with more TLD showing significance in the complex plan. Standard Plan Complex Plan Dose difference from baseline 2% # of TLD with sig. difference (p<0.05) 1/6 4/6

Results – Symmetry Angular steering, in-plane ~3.5% CROSS-PLANE Results – Symmetry Angular steering, in-plane ~3.5% Angular steering, cross-plane ~3.5% Standard Plan Complex Plan Dose difference from baseline 3% # of TLD with sig. difference (p<0.05) 2/6 3/6 For our symmetry analysis, up to 3 and 4% dose differences were seen in the angular in-plane and cross-plane deliveries, respectfully, with more TLD showing significant differences from the baseline delivery in both cross-plane deliveries. Standard Plan Complex Plan Dose difference from baseline 4% # of TLD with sig. difference (p<0.05) 6/6 5/6

Results – Symmetry Translational steering, in-plane ~3.5% 4% 4/6 2/6 CROSS-PLANE Results – Symmetry Translational steering, in-plane ~3.5% Standard Plan Complex Plan Dose difference from baseline 4% # of TLD with sig. difference (p<0.05) 4/6 2/6 And a similar difference was seen in the translational in-plane steering of up to 4%.

Summary MLC Beam Quality Symmetry More dose difference for complex plans Beam Quality Symmetry Appears to have some dose differences, film analysis required to complete assessment The impact from 3 of 9 failure modes for IMRT dose delivery have been measured that can be used to assist in the determination of the severity score for the risk probability number Stay tuned, more data to come So in summary, there were more TLD with significant dose differences in the complex plan deliveries for both the MLC positional errors and beam quality errors. We saw some significant differences in our symmetry TLD results as well, but really the trend is unclear and I think with the information from our film analysis, we will be better equipped to assess the results. Overall, the impact from 3 of our 9 physical failure modes for IMRT dose delivery have been measured, and we will be able to use this information in the determination of the severity score for the risk probability number. Stay tuned for more data to come.

Questions? Thank you! Acknowledgements Dr. David Followill This work supported by DHHS grants CA10953 and CA81647 Acknowledgements Dr. David Followill Dr. Stephen Kry Dr. Peter Balter Dr. Geoffrey Ibbott Adam Yock Austin Faught Joey Cheung Andrea Ohrt Miguel Herrera Nicholas Murray Carrie Armador Andrea Molineu Ryan Grant Questions?