Cornell ERL Prototype Injector DC Photocathode Gun Design Review Vacuum Systems Yulin Li, January 5th– 6th, 2011 1/5-6/2011 Cornell Gun Review.

Slides:



Advertisements
Similar presentations
1 Vacuum Pumping Via Titanium- Zirconium-Vanadium Thin Films Yulin Li Laboratory for Elementary-Particles Physics, Cornell University, Ithaca, NY 14853,
Advertisements

Status of 500-kV DC gun at JAEA N. Nishimori, R. Nagai, R. Hajima Japan Atomic Energy Agency (JAEA) M. Yamamoto, T. Miyajima, Y. Honda KEK H. Iijima, M.
05/20/ High-Current Polarized Source Developments Evgeni Tsentalovich MIT.
GaAs and CsKSb Photocathodes for DC Gun
Vacuum, Surfaces & Coatings Group Technology Department Vacuum tests for CLIC module prototypes 6 November 2013 C. Garion2 Outline: Reminder: specification.
30/05-03/06/2007 LCWS2007 and ILC2007, DESY Hamburg Germany Vacuum System Specifications What needs to be specified in Vacuum Specification for ICL Damping.
1 Tetsuhiko Yorita JASRI/SPring-8 Acc. Div. SPring-8 Vacuum System Contents - Outline of Vacuum system - Pressure behavior - Some topics.
Design and operation of existing ZS B.Balhan, J.Borburgh, B.Pinget B.BalhanDesign and operation of existing ZS SPS ZS Electrostatic Septum Upgrade Review.
09/13/20111 Status of high intensity polarized electron gun project at MIT-Bates Evgeni Tsentalovich MIT.
High Current Electron Source for Cooling Jefferson Lab Internal MEIC Accelerator Design Review January 17, 2014 Riad Suleiman.
PESP 2008 JLAB 1st-3rd Oct Photocathode Preparation System for the ALICE Photoinjector Keith Middleman Vacuum Science Group ASTeC, Daresbury UK.
Cornell ERL Vacuum System as a Model for ILC Damping Vacuum System Conceptual Design Yulin Li & Joe Conway C ornell L aboratory for A ccelerator-based.
Electron Source Configuration Axel Brachmann - SLAC - Jan , KEK GDE meeting International Linear Collider at Stanford Linear Accelerator Center.
2009/1/16-18 ILD09 Seoul 1 Notes for ILD Beam Pipe (Technical Aspect) Y. Suetsugu, KEK Parasitic loss Vacuum pressure profile Some comments for beam pipe.
Operated by the Southeastern Universities Research Association for the U.S. Department Of Energy Thomas Jefferson National Accelerator Facility M. Poelker,
Preliminary calculations for the vacuum system of ELENA R. Kersevan TE/VSC-IVM – 21/6/2012 The vacuum sectors, the position of vacuum pumps, flanges, and.
500kV Gun development at KEK M. Yamamoto, Y. Honda, T. Miyajima, K. Sato, T. Muto, T. Uchiyama, Y. Saito, M. Kobayashi KEK R. Hajima, N. Nishimori, R.
09/10/2007 Polarized Source for eRHIC Evgeni Tsentalovich MIT.
High Intensity Polarized Electron Gun Studies at MIT-Bates 10/01/2008 PESP Evgeni Tsentalovich MIT.
DC photogun vacuum characterization through photocathode lifetime studies Marcy L. Stutzman, Philip Adderley, Joseph Grames, Matthew Poelker, Ken Surles-Law.
Operated by the Southeastern Universities Research Association for the U.S. Depart. Of Energy Thomas Jefferson National Accelerator Facility Lifetime Measurements.
September 17-21, 2007Workshop on ILC Interaction Region Engineering Design, SLAC IR Vacuum Systems first thoughts Oleg Malyshev ASTeC, STFC Daresbury Laboratory.
SKEKB Mini Work SKEKB Vacuum System – Arc Section – Contents Y.Suetsugu KEKB Vacuum Group 1.Beam Chambers 2.Pumps: Pump, Pressure,
2007/09/17-21 SLAC IRENG07 1 A Basic Design of IR Vacuum system Y. Suetsugu, KEK Possibility of a pumping system without in-situ baking at z < L*
Vacuum, H-C. Hseuh January 4, 2007 RHIC II Project Internal Cost Review H.-C. Hseuh Vacuum Systems January 4, 2007 Project Overview.
THIN FILMS FOR CLIC ELEMENTS Outline Motivation The role of MME-CCS DB and MB transfer lines Main beam Main beam quadrupoles Other issues conclusions CLIC.
CEBAF Polarized Electron Guns Marcy L. Stutzman for the Jefferson Lab Polarized Source Group.
Beam Loss Issues for High-Power, High-Brightness Electron Injectors Bruce Dunham Cornell University December 18, 2012.
ASTRID2 Niels Hertel Nov 2012 AARHUS UNIVERSITET ASTRID2 – Status ESLS XX Berlin 2012 Niels Hertel ISA, Aarhus University, Denmark 1.
LHC Beampipes Ray Veness / AT-VAC FP XII 08Beam Vacuum- R.Veness.
11/12/20151 Status of high intensity polarized electron gun project at MIT-Bates Evgeni Tsentalovich MIT.
A. Kraemer, OLAVIII, July 11th, Gas Loads into SIS100 Cryogenic Vacuum Sections A. Kraemer, Calculations and Transparencies courtesy of St. Wilfert.
LEReC System 1. Project Schedule 2015 shutdown COMPETE! Move beam line equipment from IP 01:00 Install cooling section beam line 2016 shutdown Install.
325 MHz Superconducting Spoke Cavity Coupler status. T. Khabiboulline Power Coupler design for Superconducting Spoke cavities. Originally.
2007/09/17 SLAC IRENG07 1 Comment on IR Vacuum system Y. Suetsugu Pumping System without in-situ baking at z < L*
LEReC System 1. LEReC Injection section 2016 CeC Beam Dump 45°dipole (CeC) Profile Monitor (ERL HE) No SRF Cavity DC e - Gun ERL Solenoid ERL Halo Monitors.
NEG pumps Used at CEBAF, FEL electron guns for >15 years Three varieties – Laminated strips: used inserted in CEBAF gun – Sintered disks – Beampipe coating.
Vacuum System P. Chiggiato TE-VSC
An Electron source for PERLE
Beam Commissioning Adam Bartnik.
CBETA Vacuum System and Beam Stop (Conceptual Design) Yulin Li, CLASSE.
Vacuum Problems in the ILC Damping Ring
CBETA Vacuum System and Beam Stop Yulin Li, CLASSE.
RTML Vacuum System RDR Summary FermiLab October 24, 2007
GaAs Photocathode Development
Challenges of vacuum chambers with adjustable gap for SC undulators
G.Bregliozzi - VSC-LBV Section Collimation working group 08/07/02013
CEPC Vacuum System Dong Haiyi 2017/11/5.
Current Research CLASSE
DC Injector and Space Charge Simulation Status
Polarized Gun R&D at Fermilab
Christian Hammill, Wayne State University
Thermal Outgassing Rate for Various Beam Line Materials
RF Absorption Measurements from Beam Pipe Coatings
-Honors Thesis Defense- In Situ Ellipsometry of Surfaces in an Ultrahigh Vacuum Thin Film Deposition Chamber Joseph Choi Department of Physics and Astronomy,
The Cornell High Brightness Injector
Cornell Injector Performance
Further Measurements of Photocathode
SCRF for cw operating XFEL
Charlie Sinclair Cornell University (retired)
Electron Source Configuration
Status of 500-kV DC gun at JAEA
Cryomodule design modifications for C75
Center for Injectors and Sources
NEG-coated gun: Black R30 arc at 227kV, white R30 arc at 164 after refurbishing and second bake Carlos Wednesday, August 10, 2018.
Hydrogen and Ultra-High Vacuum
CEPC Vacuum Systems Design Consideration
Shukui Zhang, Matt Poelker, Marcy Stutzman
LLGUN2 Biased Anode Commissioning
Residual Gases in CEBAF Warm Beam Line Sources and Evacuation
Presentation transcript:

Cornell ERL Prototype Injector DC Photocathode Gun Design Review Vacuum Systems Yulin Li, January 5th– 6th, 2011 1/5-6/2011 Cornell Gun Review

Outline Vacuum Requirements – In general Present vacuum system and performance Vacuum system Issues Possible improvements 1/5-6/2011 Cornell Gun Review

Photocathode Gun Vacuum Requirements (I) (GaAs Cathode Dark Lifetime) Dominated by photocathode lifetime – dark and beam operation Base pressure – keep reactive residual gas pressure sufficiently low, dark lifetime > days H2 O2 CO CO2 Dmax (L) ? ~ 10-2 ~ 1 ~ 0.5 Pmax (torr) ~ 10-14 ~ 10-12 ~ 10-13 *) for dark time of 106 sec 1/5-6/2011 Cornell Gun Review

Photocathode Gun Vacuum Requirements (II) (GaAs Cathode Beam Lifetime) “Sudden” death – HV arcing ? Ion back-bombardment  Ions originated from cathode-anode gap  Ions originated from trapped ions beyond anode ! Estimate ion flux to the cathode beyond anode (ion  2x10-23 m-2): (Ref. 1) For pressure of 10-8 Pa in the drift , and laser spot of Ф0.2cm, and Ld = 0.25m (ions/sec·cm2 @ 100mA) The ion flux seems not very high. But what is the mechanism of the QE damage ? Ref.1 Pozdeyev, Phys. Rev. ST Accel. Beams 10, 083501 (2007) 1/5-6/2011 Cornell Gun Review

ERL Prototype Injector Layout 1 2 3 4 5 DC Photo-cathode gun keV Beamline (A1) SC RF cavity Cryo-module 600 kW Beam Dump MeV Beamlines w/ full suite of beam instrumentation Conductance limited beampipes (Ф35mm typical) and cryo-pumping in the ICM effectively separated diagnostic beamlines and beam dump from the photocathode fun A small (2°) angle in the dump beamline shield the photocathode from radiation originated in the dump, if any 1/5-6/2011 Cornell Gun Review

Achieving XHV in Photocathode Gun Ultra-Low Outgassing in Stainless Steels by Mild Heat Treatment The heat-treated (400°C bake in air for a duration of up to 100 hr) all stainless steel materials for the gun chamber. SST outgassing rates below 10-13 torr·l/s·cm2 achieved with F0>3, a dimensionless time scale, F0. t – heating time; d –thickness D=D0 exp(-Ed /kT) – Hydrogen diffusion constant Ref. C. D. Park, et al, J. Vac. Sci. Technol. A 26 1166 (2008) 1/5-6/2011 Cornell Gun Review

Pumping in the Photocathode Gun 400 l/s Ion Pump NEG Modules The gun chamber is lined with 16 SAES WR 1650 st707 NEG modules 4 SAES WR 950 st707 NEG modules inserted into a 400-l/s noble diode ion pump Total H2 pumping speed exceeding 10,000 l/s !! However: Base pressure ~ 5x10-12 torr Outgassing from thick flanges, ceramics, cathode structures, cathode support stock ?? Limited by equilibrium H2 pressure of the NEGs and ion pump ?? 1/5-6/2011 Cornell Gun Review

A1 Section – KeV Beamline Pumps (ion pumps and NEG) 1 2 3 4 5 1 Meter 1 DC Photo-cathode gun 2 BPM/Corrector/Solenoid 3 Laser Entrance with adjustable in-vacuum mirrors 4 RF Buncher 5 RF Shielded Sliding Joint and Beam Profile Viewer Very limited installed vacuum pumping due to lack of space Very difficult for a thorough/good bakeout, due to many temperature sensitive hardware 1/5-6/2011 Cornell Gun Review

Vacuum Base Pressures in Gun & A1 1. A1 Connected to the Gun and ICM 2. A1 isolated from the Gun 3. A1 isolated from the Gun & the ICM For sufficiently low base pressure to ensure adequate GaAs cathode lifetime, improvements are needed for the A1 section Many ‘trapped or hidden’ surfaces (viewer, SLDJT, BPM, buncher tuner, etc.) Part of ICM warm beampipe that is not bakeable may be a major source of outgassing 1/5-6/2011 Cornell Gun Review

ICM Warm Pipe as a Part of A1 Unbakeable surfaces exposed to A1 during A1 bakeout 1/5-6/2011 Cornell Gun Review

Gun/A1 Vacuum in Beam Runs – I 1/5-6/2011 Cornell Gun Review

Gun/A1 Vacuum in Beam Runs – II Beam Current Clear showing beam induced pressure rises in A1 (and the Gun). Causes :  Beam halo ?  RF-induced in the buncher ? 1/5-6/2011 Cornell Gun Review

Calculated Pressure Profiles Calculated using one-dimensional finite-element method, to fit measured data, assuming ONLY thermal gas-loads 1/5-6/2011 Cornell Gun Review

Improving Vacuum in the Gun Understand outgassing from other components in the gun, such as the segmented ceramics, copper parts for the cathodes, etc. Explored other pumping methods (better combination of NEGs and SIP, cryo-pumps) NEG coating (gun chamber, gun components, ceramic guard rings, a NEG coated liner, etc.) Double-walled thin SST chamber ? Improving gun bakebility 1/5-6/2011 Cornell Gun Review

Improving Vacuum in “A1” Section Increase installed pumping (adding pumps, increase conductance to the pumps, etc.) Reducing ‘trapped’ surfaces, as much as possible Enhance bakebility of the A1 section Apply NEG thin film coating to the A1 section beam pipes and components. NEG ? 1/5-6/2011 Cornell Gun Review

Calculated Pressure Profiles w/ NEG Coating 1/5-6/2011 Cornell Gun Review

Brief Summary Present gun vacuum seem sufficiently good to maintain adequate GaAs cathode dark lifetime However, base pressure in the preparation chamber needing improvement if multiple cathodes to be stored Vacuum levels in both the gun and the A1 section needing improvement to achieve acceptable GaAs cathode beam-lifetime However, significant improvement won’t be trivial Reduce ion back-scattering using ion-rejecting biased anode/ring(s) OR, alternative photocathodes, such as multi-alkali 1/5-6/2011 Cornell Gun Review