EVALUATION OF HEMOCHROMA POC Hb READER

Slides:



Advertisements
Similar presentations
QUALITY CONTROL IN HEMATOLOGY
Advertisements

Instrumental Analysis
Learning objectives You should be able to: –Identify the requirements for the Data Collection and Processing section of the Internal Assessment –Collect.
MLAB 1415: Hematology Micros 60.
Hemoglobin concentration
Quantitation of Methemoglobin
Electrical Impedance and Colorimetric Measurements Joanna Ellis, MLS(ASCP)
World Health Organization
Laboratory Quality Control
Summary 1 l The Analytical Problem l Data Handling.
Quality Assurance.
Automation in hematology Until Slide 23
In the name of Allah. Development and psychometric Testing of a new Instrument to Measure Affecting Factors on Women’s Behaviors to Breast Cancer Prevention:
1 Seventh Lecture Error Analysis Instrumentation and Product Testing.
Quality Assurance in Haematology
Decision analysis and Risk Management course in Kuopio
QUALITY CONTROL OF PHYSICO-Chemical METHODS Introduction :Validation توثيق المصدوقية.
Chemometrics Method comparison
Method Comparison A method comparison is done when: A lab is considering performing an assay they have not performed previously or Performing an assay.
Quality Assurance.
Quality Assessment 2 Quality Control.
Validation of Analytical Method
Performance Characteristics of the FARRZYME™ High Avidity Anti-dsDNA IgG Antibody ELISA Brenda B. Suh-Lailam 1, Tyson R. Chiaro 1, K. Wayne Davis 1, Andrew.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Analytical considerations
CLS 541-Advanced Medical Haematology M. sc. Program Lab-2-
Automation in haematology
Automated CBC Parameters
Automated CBC Parameters
Quality Control Lecture 5
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
Practical Hematology Lab
Evaluation of the Presage™ ST2 ELISA Jun Lu 1, David G. Grenache 1,2 1 ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT 2 Department.
REVIEW ON PROJECT WORK Measurement… Prof. Andras Fekete Department of Physics and Control Corvinus University of Budapest.
B. Neidhart, W. Wegscheider (Eds.): Quality in Chemical Measurements © Springer-Verlag Berlin Heidelberg 2000 H. Albus Basic Course Experiments to Demonstrate.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
USE OF UNCERTAINTY OF MEASUREMENT IN TESTING ROHAN PERERA MSc ( UK ), ISO/IEC Technical Assessor, Metrology Consultant.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
Paper Writing and Abstract Writing Prof. Peih-ying Lu School of Medicine Kaohsiung Medical University.
HAEMATOLOGY LINE A.C.KAROBIA
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
EQUIPMENT and METHOD VALIDATION
Diagnostic clinical chemistry
이 장 우. 1. Introduction  HPLC-MS/MS methodology achieved its preferred status -Highly selective and effectively eliminated interference -Without.
Electrical Impedance and Colorimetric Measurements
Automatic Electronic Blood Cell Counting
Other Hematological Procedures
World Health Organization
Micros 60 & Quality Control
Manual Red Blood Cell and White Blood Cell Counts
Laboratory Quality Control
IB Chemistry Internal Assessment.
Practical clinical chemistry
Analytical Method Validation
Chapter 5 Quality Assurance and Calibration Methods
Choice of Methods and Instruments
ImmunoWELL Zika Virus Serology.
Francesco Riccobono, Barouch Giechaskiel, Pierre Bonnel
QUANTITATION OF METHEMOGLOBIN
ANALYTICAL METHOD VALIDATION
World Health Organization
Quantitative tests for proteins
Introduction To Medical Technology
Quality Control Lecture 3
Quantitative tests for proteins
▪Internal quality control:
EXTENT OF CHANGES IN PRE AND POSTDONATION DONOR VARIABLES IN SINGLE AND DOUBLE DOSE PLATELETPHERESIS AND ITS IMPLICATIONS ON DONOR SAFETY Dr. R. Sreedevi.
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Presentation transcript:

EVALUATION OF HEMOCHROMA POC Hb READER Investigators: Mr. Francis M Wanyama 1, Prof Christine S. Kigondu1 and Dr. Peter Maturi2   Department of Human Pathology, School of Medicine, University of Nairobi, 1Clinical Chemistry Section, 2Hematology and BTU Section, P.O Box 19676 – 00202, Nairobi, Kenya.

INTRODUCTION Boditech’s Hemochroma TM is a portable hemoglobin meter based on robust and reliable detection mechanism with proper handling. It is capable of delivering excellent accurate measurements comparable to high end reference hemoglobin analyzers.

RATIONALE GLP requires that all equipment upon installation should demonstrate the capability of achieving the performance required as well as complying with the specifications relevant to the tests concerned. All methods have to be validated extensively in order to confirm that they are suitable for the intended use. Laboratories are expected to carry out thorough investigations of their analyzer as part of method validation before they are put into use.

STUDY OBJECTIVE To assess the performance of Boditec’s Hemochroma TM Point of care (POC) hemogloblin reader with respect to its level of accuracy.

MATERIAL AND METHODS Test Equipment: Boditec’s hemochroma Hemochroma TM POC hemoglobin reader is developed for the quantitative assay of hemoglobin in human blood by standard cyanmethemoglobin method.

Comparative Equipment: aᶜ.T 5 differential CP analyzer Beckman Principle: It is a fully automated hematology analyzer providing a complete WBC five-part differential which is determined simultaneously by the ACV(absorbance, cytochemistry and volume technology ) and is focused flow impedance

Beckman HB measurement It is based on transmittance of light through the optical part of the first dilution/Hb bath using spectrophotometric technique at wavelength of 550nm. The transmittance of the sample dilution is compared to the transmittance of a reagent blank. The system calculates the Hb using the blank and sample readings.

The cyanmethemoglobin method The Hb released by lysis of RBC combines with the potassium cyanide to form a stable cyanmethemoglobin compound which is measured through optical part of the first dilution/Hb bath using spectrophotometric technique at a wavelength of 550nm.

Table: Technical characteristics for measurement of Hb Dilution characteristic  Volume of whole blood - 10µl Volume of Aᶜ.T differential diluent - 1700µl Preliminary dilution ratio - 1:170 Volume of the 1:170 dilution removed(for making the RBC/platelet dilution) - 42.5µl Volume of Aᶜ.T 5 differential Hb lyse - 400µl Additional volume of Aᶜ.T 5 differential diluent - 400µl Final dilution for Hb determination - 1:250 Reaction temperature - 35⁰C(95⁰F) Methods of analysis - spectrophotometry Wavelength - 550nm 

RELATIVE ACCURACY The comparative equipment used was Beckman coulter Ac TTM Haematology analyser PN 4237650c from Beckman coulter Inc, Fullerton, CA 92 835.

STATISTICAL ANALYSIS The data was analyzed statistically by use of SD, the CV and finally the results were subjected to the students ‘t’ test to test for variance and correlation. t = X1- X/SE critical values also known as ‘t’ tabulated, which tests significance [P] was obtained from the standard statistical tables

  RESULTS The Hb (g/dl) levels for 52 specimen as read on Beckman and Hemochroma TM POC and the resultant means, SD and CV% Comparative Hb levels were obtained using factor -1.7 between the Beckman and Hemochroma. Intra sample standard deviation obtained between the two machines were within 1 SD with mean SD of 0.4

Coefficient of variation (C. V%) ranged between 0 and 11 Coefficient of variation (C.V%) ranged between 0 and 11.0 per cent with only 8 points out of 52 giving above 6%. The mean C.V was 3.2% which is within the acceptable of up to 6% The student ‘t’ test gave P value of <0.0005, this shows that there is no significant statistical difference between the two methods

CONCLUSION The mean levels of Hb (g/dl) for Beckman and Hemochroma POC was 12.1 and 11.6 respectively with mean of both two level at 11.9 and 1SD of 0.4. The student ‘t’ test showed P. value of < 0.0005 showing that the two means were not statistically different when a factor of -1.7 was used. The correlation coefficiency [r] was 0.9 indicating that the two methods are highly correlated.

RECOMMENDATIONS Periodic calibration should be inbuilt based on an internal quality control and/or reference methods. The identified places where the machine can be of greater use includes Casualty Wards, Antenatal clinic, Medical clinics eg for Heamatological clinic, Research, Screening camps, Mobile clinics and Laboratories in the rural settings etc.

Thank you !!! Thank You!!