Creating a Technique Chart

Slides:



Advertisements
Similar presentations
Producing an X-ray Exposure
Advertisements

Technical Aspects: the machine the image
EVERYTHING YOU WANTED TO KNOW BUT WERE AFRAID TO ASK!!!!
[Radiography] Technique - Exposure Factors
Formulating Technique Charts
Technique Guidance Systems
Ch. 14 Four Radiographic Qualities Ch
Chapter 11 Prime Factors.
Kilovolt Peak (kVp) kVp determine the quality of the beam.
4 Producing Quality Radiographs.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Primary Exposure Factors IV
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
AUTOMATIC EXPOSURE CONTROL DMI 63 3/10/2012 online.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Resident Physics Lectures
Chapter 19 & 20 Image Quality & Techniques
Analyzing the Image Density. Density Overall blackening of the image.
Prime Exposure Factors II By Professor Stelmark. Primary Factors The primary exposure technique factors the radiographer selects on the control panel.
Technical Artifacts and Errors
Chapter 9 Film & Film Holders.
Veterinary Radiology Safety and Procedures. What is a radiograph? When an xray beam (a form of electromagnetic radiation) penetrates tissue to form a.
TIMERS.
Basic quality control tests in medical radiography
Radiographic Technique Evaluation
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
Quality Control.
Exposure Factors or Prime Factors
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Random Sampling. Introduction Scientists cannot possibly count every organism in a population. One way to estimate the size of a population is to collect.
PRINCIPLES OF TECHNIQUE AND EXPOSURE
Radiographic Quality Visibility and Sharpness
Think about the effect each technique might achieve.
More digital reading explaining LUT RT 244 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of.
Chapter 10 Exposure Mathematics. Exposure Mathematics u We have seen in Chapter Seven that there are several exposure factors that are predetermined before.
More digital 244 wk 12 Perry Sprawls, Ph.D. Professor Emeritus Department of Radiology Emory University School of Medicine Atlanta, GA,
Lavin: Chapter 7 CTVT: pp
Tube Exposure Factors Math Technique Contrast and Density
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
TECHNIQUE CHARTS SERGEO GUILBAUD CLINICAL COORDINATOR.
Radiology Steps in radiation production Kilovoltage and voltage.
 Make and type of x-ray machine  Incoming line voltage  Kilovoltage  Milliamperage and exposure time  Focal film distance/source-imagine distance.
Week Seven.
Radiology. The Cathode… Provides a source of electrons and directs the electrons towards to anode. The cathode has a coiled wire filament that emits electrons.
Exposure Factors or Prime Factors
Producing An Image RVT: Chapter 6. Learning Objectives: Chapter 6 Understand the 4 factors of radiographic exposure and how each impacts the production.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Exposure Calculations
RADIATION PROTECTION 04/12/2016.
EVERYTHING YOU WANTED TO KNOW BUT WERE AFRAID TO ASK!!!!
Chapters 21 & 22.
Avian Radiology Birds are very well suited to radiographic study.
Optimisation of Patient Protection for Radiography
Radiographic Contrast
Producing An Image Lavin: Chapter 6 CTVT:
Factors Effecting the Production
Exposure Factors Chapter 4
Quality Control Testing of Screen Speed
Resident Physics Lectures (Year 1)
Developing a technique Chart Chapter 9
Comparing Exposure Systems
Representation of Data
Resident Physics Lectures (Year 1)
What makes a Quality Image
RECORDED DETAIL.
Artifacts and Errors.
RECORDED DETAIL.
Brightness Control Systems
Creating a Veterinary Radiographic Technique Chart
Presentation transcript:

Creating a Technique Chart The Cheat Sheet of Radiology!

What is a Technique Chart? Provides suggested technical factors for each type of anatomy. Unique to each x-ray unit Unique to each anatomical part Requires: Medium sized dog with ideal body condition One perfectly exposed radiograph Screens of the same speed and manufacturer Machine to be serviced/calibrated before beginning

Why is it Important? A quick reference for machine settings Without it: New calculations would have to be taken before every radiograph. Increased number of retakes = increased exposure to radiation.

Creating a Technique Chart Step 1: Prepare Service your processor and ensure screens and film match. Select a medium-sized dog of average weight. Step 2: Select your mAs Use the following for an average speed intensifying screen: Extremity 2.5 mAs Thorax 5 mAs Abdomen 7.5 mAs Spine 10 mAs Remember: mA x secs = mAs

Creating a Technique Chart Step 2: Select your mAs (cont.) To achieve this mAs, the mA and time need to be set separately, and the machine will calculate the mAs for you. Use these as standard mA settings: Extremity 150 mA Other 300 mA Calculate the time as follows: Extremity: 150mA X 1/60 secs= 2.5 mAs Thorax: 300 mA X 1/60 secs = 5 mAs Abdomen: 300 mA X 1/40 secs = 7.5 mAs Spine: 300 mA X 1/30 secs = 10 mAs

Creating a Technique Chart Step 3: Select your initial kVp Calculate initial kVp using Sante’s Rule Formula: 2 x thickness + 40 + grid factor *If a grid is used, add 10 to total Using a grid means to place the cassette in the tray under the radiology table Grid is usually indicated for measurements greater than or equal to 10 cms Example: Extremity = 7 cm (2 X 7) + 40 = 54 kVp No grid necessary since < 10 cms Book states >12 cm Article states >10 ** Note: At VTI, we always use a grid for cats, and sometimes will use a grid as a way to achieve a clearer image, even if the animal is small **

Creating a Technique Chart Step 4: Expose the perfect film Use the exposure factors calculated above as a starting point If you can’t see details of internal structures because the image is too dark, the radiograph has been over-penetrated (too much contrast), so decrease kVp by 15% If you can’t see details of internal structures because the image is too light, the radiograph has been under-penetrated, so increase kVp by 15% Once the radiograph is close to perfect, reduce the changes to 5% increments until you’re satisfied with the result. mAs can be adjusted up or down 50% as needed to alter density This may require a lot of exposures!

Creating a Technique Chart Step 5: Make the technique chart Complete kVp’s above and below the “perfect entry” as follows: Subtract 2 kVp from the original kVp for each cm decrease from the original Add 2 kVp to the original kVp for each cm increase from the original kVp up to 80 kVp. Add 3 kVp for each cm increase that places the kVp above 80 and up to 100 Add 4 kVp for each cm increase that places the kVp above 100 Step 6: Create a Technique Chart for each different study (abdomen, thorax, extremity, and spine) Create an additional chart for body parts that may need to be radiographed using a grid (all but extremities)

For example… Cm. kVp mAs 7 64 7.5 8 66 9 68 10 70 11 72 12 74 13 76 14 78 15 80 16 83 17 86 18 89 19 92 In this example, the perfect combination of kVp and mAs for a sample 15 cm abdomen is shown in blue…The entries above and below are filled in based on the previous rules.