30 Day Outcomes from the SOURCE XT TAVI Post Approval Study

Slides:



Advertisements
Similar presentations
Thirty-day Outcomes from the Multi-centre European Pivotal Trial for Transapical Transcatheter Aortic Valve Implantation with a Self-expanding Prosthesis.
Advertisements

STS 2015 John V. Conte, MD Professor of Surgery Johns Hopkins University School of Medicine On Behalf of the CoreValve US Investigators Transcatheter Aortic.
Three-year clinical and echocardiographic follow-up of aortic stenosis patients implanted with a self-expending bioprosthesis Sabine Bleiziffer German.
Lessons from TAVR Randomized Trials and Registries E Murat Tuzcu, MD Professor of Medicine Cleveland Clinic Financial disclosures: None PARTNER Executive.
Long-Term Outcomes Using a Self- Expanding Bioprosthesis in Patients With Severe Aortic Stenosis Deemed Extreme Risk for Surgery: Two-Year Results From.
One Year Outcomes in Real World Patients Treated with Transcatheter Aortic Valve Implantation The ADVANCE Study Axel Linke University of Leipzig Heart.
INTERNATIONAL. CAUTION: For distribution only in markets where CoreValve® is approved. Not for distribution in U.S., Canada or Japan. Medtronic, Inc
Conflicts of interests for Leif Thuesen, M.D.
University Heart Center Hamburg
Dr Martyn Thomas Director of Cardiac Services Guys and St Thomas NHS Foundation Trust A Member of Kings Health Partners London.
Corrado Tamburino, MD, PhD; Davide Capodanno, MD; Angelo Ramondo, MD; Anna Sonia Petronio, MD; Federica Ettori, MD; Gennaro Santoro, MD; Silvio Klugmann,
Prosthesis-Patient Mismatch in High Risk Patients with Severe Aortic Stenosis in a Randomized Trial of a Self-Expanding Prosthesis George L. Zorn, III.
Axel Linke University of Leipzig Heart Center, Leipzig, Germany Sabine Bleiziffer German Heart Center, Munich, Germany Johan Bosmans University Hospital.
TCT 2015 | San Francisco | October 15, 2015 Transcatheter Aortic Valve Replacement for Failed Surgical Bioprostheses Danny Dvir, MD John G. Webb, MD and.
TCT 2015 | San Francisco | October 15, 2015 Transcatheter Aortic Valve Replacement for Failed Surgical Bioprostheses Danny Dvir, MD John G. Webb, MD and.
The Impact of Prior Stroke on the Outcome of Patients with Severe Aortic Stenosis Undergoing Transcatheter Aortic Valve Replacement Romain Didier, MD;
UC c EN. Through Medtronic sponsored research, the Transcatheter Aortic Valves clinical portfolio is studying over 11,000 subjects at over 125.
INTERNATIONAL. CAUTION: For distribution only in markets where CoreValve® is approved. Not for distribution in U.S., Canada or Japan. Medtronic, Inc
Martin B. Leon, MD on behalf of the PARTNER Investigators TCT 2010; Washington, DC; September 23, 2010 Transcatheter Aortic Valve Implantation in Inoperable.
TCT 2015 | San Francisco | October 15, 2015 Howard C. Herrmann, MD on behalf of The PARTNER II Trial Investigators SAPIEN 3: Evaluation of a Balloon- Expandable.
INTERNATIONAL. CAUTION: For distribution only in markets where CoreValve® is approved. Not for distribution in U.S., Canada or Japan. Medtronic, Inc
Longest Follow-up After Implantation of a Self-Expanding Repositionable Transcatheter Aortic Valve: Final Follow-up of the Evolut R CE Study Stephen Brecker,
G. Michael Deeb, MD On Behalf of the CoreValve US Investigators
Disclosure Statement of Financial Interest
Patients at intermediate surgical risk undergoing isolated interventional or surgical aortic valve replacement for severe symptomatic aortic valve stenosis.
The OPTIMAX first-in-man study Mid-term clinical outcome of Titanium-Nitride-Oxide-coated Cobalt Chromium stents in patients with de novo coronary artery.
New Data from The PARTNER Trial
Extending the Boundaries of TAVR: Future Directions
On behalf of the FORWARD Study Investigators
Highlights From the SAPIEN 3 Experience in Intermediate-Risk Patients Vinod H. Thourani, MD on behalf of the PARTNER Trial Investigators Professor.
The Medtronic Ventor EngagerTM TAVI System
Raj R. Makkar, MD On behalf of The PARTNER Trial Investigators
Design & Product Development Larry L Wood
Are we ready to perform TAVI in Intermediate Risk Patients?
Potential conflicts of interest
Updates From NOTION: The First All-Comer TAVR Trial
30-Day Outcomes From John Webb, MD
Transcatheter or Surgical Aortic Valve Replacement in Intermediate Risk Patients with Aortic Stenosis Description: The goal of the trial was to assess.
Review of the Latest OUS Data from the Self-Expanding Valve Studies
First Report of One-Year Outcomes of the REPRISE I Feasibility Study of the Repositionable Lotus Aortic Valve Replacement System Ian T. Meredith.
30-Day Safety and Echocardiographic Outcomes Following Transcatheter Aortic Valve Replacement with the Self-Expanding Repositionable Evolut PRO System.
Trans-Apical Aortic Valve Implant:
ONE-YEAR OUTCOMES OLAF WENDLER, MD PHD FRCS
First Report of Three-Year Outcomes With the Repositionable and Fully Retrievable Lotus™ Aortic Valve Replacement System: Results From the REPRISE I.
Early Outcomes with the Evolut R Repositionable Self-Expanding Transcatheter Aortic Valve in the United States Mathew Williams, MD, For the Evolut R US.
University of Pennsylvania
Giuseppe Tarantini MD, PhD
Early Recovery of Left Ventricular Systolic Function After CoreValve Transcatheter Aortic Valve Replacement Harold L. Dauerman, MD; Michael J. Reardon,
Jodi J. Akin, MS VP Global Clinical Affairs,
Direct Flow Medical Experience with a Conformable, Repositionable Retrievable Percutaneous Aortic Valve Reginald Low MD University of California, Davis.
TAVI „Catch me if you can!“
The Impact of Live Case Transmission on Patient Outcomes during Transcatheter Aortic Valve Replacement: Results from the VERITAS Study Dr. Ron Waksman.
Transcatheter Heart Valves
Michael Mack, M.D. Dallas, TX
Vinod H. Thourani, MD on behalf of The PARTNER Trial Investigators
Successful Cox Maze Procedure During Mitral Valve Surgery Restores Patient Survival Without Increasing Operative Risk Niv Ad, MD Chief, Cardiac Surgery.
Axel Linke University of Leipzig Heart Center, Leipzig, Germany
Annual Outcomes With Transcatheter Valve Therapy
One Year Outcomes in Real World Patients Treated with Transcatheter Aortic Valve Implantation The ADVANCE Study Axel Linke University of Leipzig Heart.
Second Generation Valves: What Will Be Different?
University Heart Center Hamburg
S.G. Worthley, MB, BS, PhD., S. Redwood, MD, PhD.,
Balloon-Expandable Transcatheter Valve System : OUS Data
Annual Outcomes With Transcatheter Valve Therapy
Late Follow-Up from the PARTNER Aortic Valve-in-Valve Registry
Samir R. Kapadia, MD On behalf of The PARTNER Trial Investigators
The Synergy between Percutaneous Coronary Intervention with TAXUS and Cardiac Surgery: The SYNTAX Study One Year Results of the PCI and CABG Registries.
On behalf of all principal COMPARE II investigators:
Five-Year Outcomes after Randomization to Transcatheter or Surgical Aortic Valve Replacement: Final Results of The PARTNER 1 Trial Michael J. Mack, MD.
The Synergy between Percutaneous Coronary Intervention with TAXUS and Cardiac Surgery: The SYNTAX Study One Year Results of the PCI and CABG Registries.
Presentation transcript:

30 Day Outcomes from the SOURCE XT TAVI Post Approval Study Olaf Wendler MD PhD FRCS Professor of Cardiac Surgery King's College Hospital/King’s Health Partners, London, UK on behalf of the SOURCE XT Investigators

   Potential conflicts of interest Speaker’s name: Olaf Wendler  I have the following potential conflicts of interest to report:  Research contracts Edwards Lifesciences  Consulting Medtronic, JenaValve, Edwards Lifesciences, SJM  Employment in industry  Stockholder of a healthcare company  Owner of a healthcare company  Other(s)  I do not have any potential conflict of interest   

Introduction First TAVI performed by Alain Cribier in 2002 CE-Mark approval of the Edwards SAPIEN™ THV in 2007 The „SOURCE“ Post Approval Registry included 2300+ patients treated using the SAPIEN™ THV Data generated provided important information for new product and clinical development Next generation SAPIEN XT™ THV and its delivery systems are assessed in the „SOURCE XT“ Post Approval Study

Edwards SAPIEN XTTM Evolution Stainless Steel Frame Equine Pericardial Tissue Stainless Steel Frame Bovine Pericardial Tissue Cobalt-Chromium Frame Bovine Pericardial Tissue Semi-closed leaflets Reduced crimped profile 2004 2007 2010 Cribier-Edwards™ THV 23mm Edwards SAPIEN™ THV 23 mm and 26 mm Edwards SAPIEN XT ™ THV 23 mm, 26 mm, and 29mm 4

SOURCE - SOURCE XT Multicenter, prospective, observational, post-approval studies to Observe use of the SAPIEN/SAPIEN XT™ THV in the real world Evaluate long-term safety of SAPIEN/SAPIEN XT™ THV SOURCE was further aimed to Develope physician training Optimize patient screening and selection SOURCE XT is further aimed to Incorporate VARC definitions Include assessment of frailty Develop a TAVI risk prediction model

Design and Execution SOURCE Registry SOURCE XT Post Approval Study Patients 2307 2706 Enrollment Nov 2007 – Dec 2009 July 2010 – Oct 2011 Centers 37 94 Countries 14 17 Access Routes TF, TA TF, TA, TAO, SC THV SAPIEN™ 23mm, 26mm SAPIEN XT™ 23mm, 26mm, 29mm Adverse Event Adjudication Principal Investigators Independent Clinical Endpoint Committee, VARC Guidelines Monitored No Yes Follow-up 30 Days, annually to 2 Years 30 Days, annually to 5 Years

Executive Committee Prof. Dr. Helmut Baumgartner - Muenster, Germany Prof. Dr. Gino Gerosa - Padua, Italy Dr. Paolo Rubino - Mercogliano, Italy Prof. Dr. Patrick Nataf - Paris, France Dr. Martyn Thomas - London, United Kingdom Prof. Dr. Alec Vahanian - Paris, France Prof. Dr. Thomas Walther - Bad Nauheim, Germany Prof. Dr. Olaf Wendler-London - United Kingdom Prof. Dr. Stephan Windecker-Bern, Switzerland

Access Routes Subclavian 0.3%

Methods Focus on quality of data Centers participating had to submit consecutive group of their patient experience using the Edwards SAPIEN XT™ Decision of access route made by the individual center Cardiologists and Cardiac Surgeons as PI‘s in participating centers to support the „Heart Team“ approach „We present the interim analysis of the 30 day outcomes with a focus on TF and TA access routes“

Baseline Characteristics - Cardiac Transfemoral N = 1694 Transapical N = 906 P NYHA III/IV (%) 77.2 75.9 0.5 LV Ejection Fraction (mean ±SD) 55.3 ±17.0 53.5 ± 12.4 <.0001 Coronary Artery Disease (%) 40.2 51.8 Previous MI (%) 12.2 20.9 Previous PCI (%) 27.2 36.7 Previous CABG (%) 12.3 24.4 Atrial Fibrillation (%) 22.8 29.9 Previous Pacemaker (%) 10.0 12.7 0.03 Previous Ao Bioprosthesis (%) 1.8 4.0 0.001

Baseline Characteristics Transfemoral N = 1694 Transapical N = 906 P Age (years, mean ±SD) 81.9 ±6.6 80.2 ±6.4 <.0001 Female (%) 64.5 45.6 Previous Stroke (%) 7.3 10.6 0.005 Renal Failure (%) 24.9 35.2 Peripheral Vascular Disease (%) 14.6 33.0 Diabetes Mellitus (%) 26.8 33.7 0.0003 Porcelain Aorta (%) 4.4 10.7 Frail condition (mean±SD) 5.2 ±1.4 0.7 COPD (%) 19.6 21.9 0.2 Logistic EuroSCORE (mean±SD) 19.9 ±12.0 21.8 ±13.8 0.0004 STS Score (mean±SD) 8.4 ±7.0 8.7 ±7.2 Logistic EuroSCORE TF TA ≤40 92.3 88.5 >40 6.7 10.8 Missing 0.9 0.8

Procedural Information Characteristic Transfemoral N = 1694 Transapical N = 906 P Total Procedure Time (min, mean ±SD) 79.7 ±51.7 90.7 ±56.3 <0.0001 Fluoroscopy Time (min, mean ±SD) 15.0 ±20.8 7.4 ±19.3 Contrast Volume (ml, mean ±SD) 141.1 ±104.0 95.0 ±81.8 General Anesthesia (%) 50.1 96.0 Pre-implant BAV (%) 98.5 93.4 Annular Diameter (mm, mean ± SD) 21.7 ± 3.5 22.5 ± 3.4 Valve size implanted 23 mm (%) 26 mm (%) 29 mm (%) NA (%) 47.8 51.8 -- 0.3 33.0 41.8 23.8 1.4

Procedural Complications Adverse Events Transfemoral N = 1694 Transapical N = 906 P Conversion to Surgery (%) 0.3 0.6 Cardiac Tamponade (%) 0.7 0.1 0.04 Annular Dissection (%) 0.4 Coronary Occlusion (%) 1.0 SAPIEN-in-SAPIEN (Bailout) (%) 0.8 1.7 0.07 Valve Embolization (%) 0.9 0.2 Aborted Procedure (%) 1.1 Unable to Cross Native Valve (%) --

Clinical Outcomes at 30 Days Transfemoral N = 1694 Transapical N = 906 All-cause Mortality (%)* 4.3 9.9 Any Stroke (%)* 2.3 2.1 TIA (%) 0.6 0.7 Myocardial Infarction (%) 0.5 New Pacemaker (%) 8.0 10.9 Vascular Complication – Major (%)* 7.3 3.6 Renal Failure with Temporary Dialysis) (%) 1.2 4.0 Major Bleeding (%) 5.0 11.4 * VARC definitions /CEC Adjudicated

Subanalyses Mortality Stratified LES Transfemoral N = 1694 Transapical N=906 ≤40 4.0 8.4 >40 7.0 22.4 SAPIEN XTTM Size Transfemoral N = 1694 Transapical N=906 23mm (%) 4.3 8.9 26mm (%) 4.1 11.9 29mm (%) N/A 6.0 Missing (%) 0.6 0.06

Echo at 30 Days*: TF – TA Mean Gradient (mm Hg) EF (%) AVA(cm2) p <.0001 EF (%) p <.0001 Baseline 30 Days *Site reported, Paired observation analysis

Paravalvular Leak at 30 Days*: TF - TA *Site reported

Conclusions (I) Mortality in TF (4.3%) is one of the lowest ever reported for TF registries One third of patients were selected for a TA approach Mortality in TA has decreased but remains higher in the very high-risk patient group For 29 mm TA prostheses (not available for TF, no selection bias) mortality for TA is only 6% Compared to SOURCE, mean log EuroSCORE has decreased but stays within the label indication of 15 for the SAPIEN XTTM TA patients continue to present with more co-morbidities

Conclusions (II) Major vascular complications were more common after TF and bleeding more often seen after TA Procedural complications were rare and early hemodynamic function excellent using the SAPIEN XT™ LV ejection fraction was improved significantly using both access routes The number of patients with no/trace paravalvular leakage is high and it remains to be seen which effect this may has on the 1-year survival It remains to be seen which effect frailty has on 1-year survival