Fast SiPM readout for PET

Slides:



Advertisements
Similar presentations
PET Design: Simulation Studies using GEANT4 and GATE - Status Report - Martin Göttlich DESY.
Advertisements

1 A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout Heejong Kim 1, Chien-Min Kao 1, Chin-Tu Chen 1, Jean-Francois Genat.
Performance Evaluation of SiPM Arrays under Strong Magnetic Fields
June/5/20081 Electronics development for fast-timing PET detectors: The multi-threshold discriminator Time of Flight PET system Contents 1. Introduction.
A High-speed Adaptively-biased Current- to-current Front-end for SSPM Arrays Bob Zheng, Jean-Pierre Walder, Henrik von der Lippe, William Moses, Martin.
1 SciFi electronics meeting – CERN– June 20 th 2011 Some ideas about a FE for a SciFi tracker based on SiPM A. Comerma, D. Gascón Universitat de Barcelona.
Report on SiPM Tests SiPM as a alternative photo detector to replace PMT. Qauntify basic characteristics Measure Energy, Timing resolution Develop simulation.
Readout ASIC for SiPM detector of the CTA new generation camera (ALPS) N.Fouque, R. Hermel, F. Mehrez, Sylvie Rosier-Lees LAPP (Laboratoire d’Annecy le.
FLC Group Test-beam Studies of the Laser-Wire Detector 13 September 2006 Maximilian Micheler Supervisor: Freddy Poirier.
Experimental set-up for on the bench tests Abstract Modeling of processes in the MCP PMT Timing and Cross-Talk Properties of BURLE/Photonis Multi-Channel.
Development of a gamma-ray imager using a large area monolithic 4x4 MPPC array for a future PET scanner Takeshi Nakamori T. Kato, J. Kataoka, T. Miura,
Development of Multi-pixel photon counters(2) M.Taguchi, T.Nakaya, M.Yokoyama, S.Gomi(kyoto) T.Nakadaira, K.Yoshimura(KEK)
1 SiPM Gain Test SiPM Multiple APD pixels operating at Geiger mode. Output is the sum of the outputs from all APD pixels. Advantages Compact size High.
Optimization of Detectors for Time of Flight PET Marek Moszyński, Tomasz Szczęśniak, Soltan Institute for Nuclear Studies, Otwock-Świerk, Poland.
MPPC status M.Taguchi(kyoto) T2K ND /7/7.
Development of Multi-Pixel Photon Counters(MPPC) Makoto Taguchi Kyoto University.
Design for Wide FOV Cherenkov telescope upgrading THE 2 nd WORKSHOP OF IHEP Shoushan Zhang Institute of High Energy Physics.
Timing Studies of Hamamatsu MPPCs and MEPhI SiPM Samples Bob Wagner, Gary Drake, Patrick DeLurgio Argonne National Laboratory Qingguo Xie Department of.
Slide 1Turisini M. Frontend Electronics M.Turisini, E. Cisbani, P. Musico CLAS12 RICH Technical Review, 2013 June Requirements 2.Description of.
An ASIC for TOF and DOI Measurement with SiPM Matrices Cristoforo Marzocca, Francesco Licciulli DEI - Politecnico di Bari and INFN - Sezione di Bari, Italy.
Peter Dendooven LaBr 3 and LYSO monolithic crystals coupled to photosensor arrays for TOF-PET Physics for Health in Europe Workshop February 2-4, 2010,
A Multi-Threshold Method for TOF-PET Signal Processing Heejong Kim 1, Chien-Min Kao 1, Qingguo Xie 1, Chin-Tu Chen 1, Octavia Biris 2, Jialin Lin 2, Fukun.
CERN PH MIC group P. Jarron 07 November 06 GIGATRACKER Meeting Gigatracker Front end based on ultra fast NINO circuit P. Jarron, G. Anelli, F. Anghinolfi,
Jornadas CPAN – 14 th October 2014 – Sevilla SiUB Electronics Instrumentation Development and ASIC Design Lluís Garrido Universitat de Barcelona.
Update on works with SiPMs at Pisa Matteo Morrocchi.
Development of characterization facilities and front-end electronics for SiPM R. A. Shukla, For SiPM Development Group.
SiPM Readout ASIC LAPP Contribution JM Dubois, N. Fouque, R. Hermel, G. Lamanna, F. Mehrez, J.L. Panazol, J. Prast, S. Rosier.
Ongoing work on ASIC development at Politecnico-Bari F. Corsi, F. Ciciriello, F. Licciulli, C. Marzocca, G. Matarrese DEE - Politecnico di Bari and INFN.
KM2A PMT testing platform FENG Cunfeng, LI Chaoju, SUN Yansheng Shandong University THE 2 nd WORKSHOP OF AIR SHOWER DETECTION AT HIGH ALTITUDES IHEP, Beijing,
P. Lecoq CERN2 February ENVISION WP2 Meeting CERN Group contribution to ENVISION WP2 Paul Lecoq CERN, Geneva.
Silicon Photomultiplier (SiPM) Readout Application Specific Integrated Chip (ASIC) for Timing Huangshan Chen.
Page Detector and Electronics R&D for picosecond resolution, single photon detection and imaging J.S. MilnesPhotek Ltd T.M. ConneelyUniversity.
1 The Scintillation Tile Hodoscope (SciTil) ● Motivation ● Event timing/ event building/ software trigger ● Conversion detection ● Charged particle TOF.
 13 Readout Electronics A First Look 28-Jan-2004.
1 MPPC Timing Resolution Akito Kobayashi Shinshu University Table of contents ・ Timing Resolution ・ Set up ・ data treatment ・ result(1) ・ result(2) ・ result(3)
Activities at SMI/Vienna in testing the performance of SiPMs
Performance of scintillation pixel detectors with MPPC read-out and digital signal processing Mihael Makek with D. Bosnar, V. Gačić, L. Pavelić, P. Šenjug.
DAQ ACQUISITION FOR THE dE/dX DETECTOR
Performance of LYSO and CeBr3 crystals readout by SiPM
A 2 Gsps Waveform Digitizer ASIC in CMOS 180 nm Technology
Commissioning of the Crystal evaluation setup
Journées VLSI-FPGA-PCB Juin 2010 Xiaochao Fang
Preliminary data analysis on 16x16 modules
Results achieved so far to improve the RPC rate capability
CTA-LST meeting February 2015
Update of time measurement results with the USB WaveCatcher board & Electronics for the DIRC-like TOF prototype at SLAC D.Breton , L.Burmistov,
Ongoing R&D in Orsay/Saclay on ps time measurement: a USB-powered 2-channel 3.2GS/s 12-bit digitizer D.Breton (LAL Orsay), E.Delagnes (CEA/IRFU) Séminaire.
TDC at OMEGA I will talk about SPACIROC asic
A First Look J. Pilcher 12-Mar-2004
Christophe Beigbeder/ ETD PID meeting
Next generation 3D digital SiPM for precise timing resolution
Conceptual design of TOF and beam test results
LHCb calorimeter main features
SiPM Readout ASIC LAPP Contribution
Department of Physics and Astronomy,
The New Readout Electronics for the SLAC Focusing DIRC Prototype (SLAC experiment T-492 ) L. L. Ruckman, G. S. Varner Instrumentation Development Laboratory.
X. Zhu1, 3, Z. Deng1, 3, A. Lan2, X. Sun2, Y. Liu1, 3, Y. Shao2
BESIII EMC electronics
3x3-MPPC-SMDx16ch SiPM tileable modules from Hamamatsu
A Multi-Threshold Method for TOF-PET Signal Processing
MCP PET Simulation (7) – Pixelated X-tal
Pre-installation Tests of the LHCb Muon Chambers
HE instrument and in-orbit performance
md-NUV PET project meeting
ASPID (Application of Silicon Photomultipliers to Imaging Detectors)
Test Setup SiPM Hamamatsu S C Pixel size( 25um)
A Time Driven Readout Scheme For PET and CT… …using APDs and SiPMs
Ongoing R&D in Orsay/Saclay on ps time measurement: status of the USB-powered 2-channel 3.2GS/s 12-bit digitizer D.Breton & J.Maalmi (LAL Orsay), E.Delagnes.
Status Report on MCP PET Simulation
The MPPC Study for the GLD Calorimeter Readout
Presentation transcript:

Fast SiPM readout for PET David Gascon1, Ricardo Graciani1, David Sánchez1, John Barrio2, Gabriela Llosá2, Josep F. Oliver2, Jesús Marín3, José Manuel Perez3, Pedro Rato3, Iciar Sarasola4 1Universitat de Barcelona 2IFIC/IFIMED (CSIC-UVEG) 3CIEMAT 4CERN

Outline The FlexToT ASIC FlexToT jitter Results with 1 channel MPPC Energy Resolution FlexToT measurements at CERN DOI First results with monolithic and MPPC matrix CRT MPPC characterization Future steps

The FlexToT ASIC Joint project with CIEMAT to develop a time-over- threshold ASIC for SiPM based PET ICCUB: electronics and microelectronics design CIEMAT: PET and medical imaging instrumentation Optimized for segmented crystals with adjustable dynamic range FlexToT 16 channel SiGe BiCMOS 0.35um austriamicrosystem 10 mm2 3.3 V (10 mW/ch) QFN 64

The FlexToT ASIC FlexToT ASIC main features (I) linear time-over-threshold (ToT) design output signal width proportional to input current peak value 16 input analog channels (current) 16 digital ToT output channels (energy) 1 fast CML trigger output (time) current amplifiers with DC coupling for low input impedance (34 ) and 250 MHz bandwidth (> 400 MHz for FlexToT v2) bias voltage of each SiPM pixel individually adjustable for correcting temperature and gain variations magnetic field (MR) tolerant

The FlexToT ASIC FlexToT ASIC main features (II) low power consumption of ~10 mW/channel sub-nanosecond timing with a trigger jitter better than 30 ps (< 10 ps for FlexToT v2) limited by measuring system simple readout by an FPGA including a time-to-digital conversion (TDC) linearity better than 5% for a dynamic range in current peak values from 0.7 mA to 18 mA (55 pC to 2 nC input charge) internal gain saturation control and pile-up detection fabricated by AMS (Austria) in 0.35 μm HBT BiCMOS technology die size 2.93 mm x 2.54 mm, mounted on a 9 mm x 9 mm QFN64 package

Outline The FlexToT ASIC FlexToT jitter Results with 1 channel MPPC Energy Resolution FlexToT measurements at CERN DOI First results with monolithic and MPPC matrix CRT MPPC characterization Future steps

FlexToT v2 jitter measurements Time difference between calibration input signal and ASIC trigger Measurement setup: Agilent 81110A (330 MHz) pulse generator 4 GHz / 20 GSPs DPO scope Threshold set to detect 1 photon equivalent Jitter better than 8 ps (rms) for signals larger than 3 photoelectrons Best result: 7.2 ps (rms) for ~10 photoelectrons signal

Outline The FlexToT ASIC FlexToT jitter Results with 1 channel MPPC Energy Resolution FlexToT measurements at CERN DOI First results with monolithic and MPPC matrix CRT MPPC characterization Future steps

Energy Resolution: pixelated Energy resolution using Cs 137 and Na 22 for calibration: < 8 % 20x3x3 mm3 LFS crystals HPKK 3x3 mm2 - 50 um sensors

The FlexToT ASIC The timing performance of the FlexToT v2 ASIC was measured at CERN and compared with NINO readout electronics STSM of Iciar Sarasola (CIEMAT), COST Action TD1401 “FAST” A portable setup with two FlexToT readout cards was carried to CERN during August 2015 to evaluate: Single Photon Timing Resolution (SPTR) Coincidence Time Resolution (CTR) Measurements were performed using both NINO and FlexToT readout electronics under similar operational conditions

FlexToT v2 timing measurements at CERN Single Photon Time Resolution (SPTR) Pilas picosecond laser at 420 nm, 28 ps (FWHM) pulses at 100 kHz rate LeCroy WaveRunner 104 Xi 1 GHz oscilloscope (10 Gs/s) Laser intensity reduced by neutral density (ND) filters Measurements performed in a light-tight box at 18.5 C Hamamatsu S13360-3050CS (LCT5) 3x3 mm2 MPPC 80ps 64ps - Best results (sigma): 64 ps for NINO and 80 ps for FlexToT

FlexToT v2 timing measurements at CERN Coincidence Time Resolution (CTR) 2x2x5 mm3 LSO:Ce,Ca crystals on Hamamatsu S13360-3050CS (LCT5) 3x3 mm2 MPPCs LeCroy DDA 735Zi-A 3.5 GHz Analyzer (40 Gs/s) Measurements performed in a light-tight box at 15 C Coincidences corresponding to 511 keV photopeak (+/- 3 sigma) in both detectors  10 mW/ch  30 mW/ch Best results (FWHM): 97 ps for NINO and 128 ps for FlexToT readout

DOI in phoswich configuration The FlexToT allows time discrimination from two type of crystals with different decay times. GSO 511 keV peak Time LYSO 511 keV peak Energy - MPPC array S12572, 4x4 with 3x3mm2/ch with 50um - Stack configuration: GSO and LYSO, 1.35x1.35mm2 with 7mm and 8mm long respectively.

Outline The FlexToT ASIC FlexToT jitter Results with 1 channel MPPC Energy Resolution FlexToT measurements at CERN DOI First results with monolithic and MPPC matrix CRT MPPC characterization Future steps

Energy Resolution: monolithic Energy resolution and linearity with LYSO monolithic scintillator blocks Resolution at 511 keV: 15.4% FWHM Linearity (511-662-1274 keV): R2 = 0.99975 Raw spectra obtained by summing the ToT outputs of all channels 15 % @ 511 KeV (monolithic) 8 % @ 511 KeV (segmented)

BEST CTR with large LaBr3 crystal - MPPC array S12572, 4x4 with 3x3m/ch with 50um - LaBr3 crystal of 16x18x5mm3 MPPC array with monolithic LaBr3 in coincidence with single channel MPPC. CTR < 450ps 5mm 20mm 16mm 18mm

BEST CTR with large LYSO crystal - MPPC array S12572, 4x4 with 3x3mm2/ch with 50um - LYSO crystal of 16x18x10mm3 MPPC array with monolithic LYSO in coincidence with single channel MPPC. CTR < 650ps 10mm 20mm 16mm 18mm

Outline The FlexToT ASIC FlexToT jitter Results with 1 channel MPPC Energy Resolution FlexToT measurements at CERN DOI First results with monolithic and MPPC matrix CRT MPPC characterization Future steps

MPPC characterization: DCR 0.5pe 1.5pe 2.5pe 3.5pe - MPPC LCT2-50um - Number of pulses on a 10𝜇𝑠 windows

MPPC characterization using PACTA - MPPC single channel S13360, 6x6mm2/ch with 50um - Picosecond (50 ps FWHM) laser - Breakdown voltage 51.2V

Summary and conclusions UB and CIEMAT have designed and implemented an ASIC for the readout of SiPMs coupled to scintillators in (TOF)PET applications Good linearity (better than 5%) from 0.7 mA to >18 mA peak input current (FlexToT v2) Jitter better than 8 ps for signals larger than 3 photoelectrons Good results in SPTR, CTR and DOI capability: 80 ps 𝜎 for single photoelectron timing resolution 128 ps FWHM coincidence time resolution with LSO:Ce,Ca scintillators (FlexToT v2) DOI discrimination with stacked GSO + LYSO crystals <450 ps FWHM coincidence time resolution with LaBr3 and <650 ps FWHM for LYSO monolithic crystals (FlexToT v2) PACTA really works for MPPC characterization

FUTURE STEPS Detailed PENELOPE simulations. New ASIC version for monolithic crystal (180nm CMOS, 3.5 mW/ch) DOI corrections and channel equalization. Optimization of MR-compatible demonstrator with 64 channel DAQ board.

Acknowledgements Fast MEDAMI organizers Hamamatsu

Thanks for your attention!

FlexToT Time Properties: CTR Coincidence time resolution measurements for Classical photopeak selection Results @ ±2 20x3x3 mm3 LFS crystals LCT4 3x3mm 50um

FlexToT Time Properties: CTR CTR results with 20mm crystals: S12572 ~ 250ps FHWM LCT2 ~ 260ps FHWM LCT4 ~ 220ps FHWM

FlexToT Time Properties: SPTR Single Photon Time Resolution Picosecond (50 ps FWHM) laser Jitter floor < 10 ps rms For very large signals

FlexToT ASIC Good linearity and uniformity With only comparator threshold offset equalization Different operating ranges can be covered 28 28

MPPC characterization photon distribution charge

MPPC characterization gain plot

MPPC characterization Vb and Cd+Cq