Basic requirements on area dosemeters

Slides:



Advertisements
Similar presentations
Paolo Gemma, Senior Expert, Huawei
Advertisements

4/2003 Rev 2 I.1.1 – slide 1 of 13 Session I.1.1 Part I Review of Fundamentals Module 1Introduction Session 1Learning Objectives, Course Table of Contents.
1 Radiological protection, dosimetry & environmental monitoring National Training Course on Radiopharmacy, Radiochemistry and Radioisotope Production Site.
Detecting and Measuring Ionizing Radiation -2
Radiation Protection at Linear Electron Accelerator LINAC-800 N.I. Balalykin, V.Ph. Minashkin, M.A. Nozdrin, G.D. Shirkov, V.Yu. Schegolev Veksler and.
Copyright Alan Rowley Associates Steps to an Accurate Result Select a method and validate it as suitable for the purpose envisaged. Establish that.
Centre for Radiation, Chemical and Environmental Hazards Radiation Protection Division formerly the National Radiological Protection Board Guidance for.
S4: Market Surveillance Physikalisch-Technische Bundesanstalt Session 4: Market surveillance Peter Ulbig, Harry Stolz Belgrade, 31 October.
Radiation Exposure, Dose and Relative Biological Effectiveness in Medicine Background Image:
IAEA International Atomic Energy Agency Regulations Part II: Basic Concepts and Definitions Day 8 – Lecture 5(2)
IAEA Quantities and Measurements - 3 Radiation Protection Quantities Day 3 – Lecture 1 1.
IAEA plans with respect to environmental protection EC PROTECT Workshop Oslo, Norway, 28–30 January 2008.
PART IX: EMERGENCY EXPOSURE SITUATIONS Module IX.1: Generic requirements for emergency exposure situations Lesson IX.1-2: General Requirements Lecture.
THE RADIOLOGICAL PROTECTION INSTITUTE OF IRELAND Ms Máirín Colmáin Senior Technician Measurement Services.
Manfred Kochsiek Wilfried Schulz Physikalisch-Technische Bundesanstalt Braunschweig und Berlin Modernisation of Legal Metrology in Germany.
S3: Module D Physikalisch-Technische Bundesanstalt Session 3: Conformity Assessment Module D Peter Ulbig, Harry Stolz Belgrade, 31 October.
1 / 38 Introduction Compatibility Modular Concept WELMEC 2.4 Buenos Aires INIT Oliver Mack; Sascha Mäuselein The Modular Concept For Load Cells Conclusion.
Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.
RADIATION DOSIMETER MKS-01СА1М Technical proposal.
Milestones M5 and M6 Mid-Term Review, October 2013M. Silari – ARDENT overview1 M5: comparison of detector technologies (WP1) M6: choice of detector technology.
Good management Processing record Qualified and experienced operators Quality Manual  A Radiation processing facility should have the following vital.
OSL Albedo Neutron Dosimeter
Lecture#16 Estimation of the system’s dependability The Bonch-Bruevich Saint-Petersburg State University of Telecommunications Series of lectures “Telecommunication.
Accurate gamma-ray spectrometry of environmental samples: a challenge O. Sima - Bucharest University D. Arnold - PTB Braunschweig C. Dovlete - ERL Bucharest.
Prime Responsibility for Radiation Safety
1 / 24 Introduction Eccentricity effects INTI Buenos Aires Oliver Mack: Recommendation For The Revision Of Test Procedures For Load Cells In Legal Metrology.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Dosimetric Quantities.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Calibration and Test.
International Atomic Energy Agency ASSESSMENT OF OCCUPATIONAL EXPOSURE DUE TO EXTERNAL RADIATION SOURCES AND INTAKES OF RADIONUCLIDES Dosimetry Services.
1 Course : بسم الله الرّحمن الرّحيم Chapter 7: BSS & International Regulations Omrane KADRI, Ph.D. Office 2021 Health Safety & Radiation.
Organization and Implementation of a National Regulatory Program for the Control of Radiation Sources Technical Services.
Centre for Radiation, Chemical and Environmental Hazards Radiation Protection Division formerly the National Radiological Protection Board Comparison of.
3/2003 Rev 1 II.2.9a – slide 1 of 25 IAEA Post Graduate Educational Course Radiation Protection and Safe Use of Radiation Sources Part IIQuantities and.
M.Moll, M.Silari, H.Vincke – 3.April Mixed field irradiation -- Who answered ?  In total 36 forms filled / 34 persons answered: 38% 62% CERN:
Informations and comments on ECE-TRANS-WP.29-GRB e 1 Louis-Ferdinand PARDO (France) Transmitted by the expert from France Informal document GRB
IAEA International Atomic Energy Agency Radiation protection of the public IAEA Regional Training Course on Radiation Protection of patients for Radiographers,
Introduction of OIML R59 Moisture Meters for Cereal Grains and Oilseeds (1984) 1 Dr. Tsuyoshi Matsumoto Chair of APLMF WG on Quality Measurements of Agricultural.
MAIN FEATURES FOR CALIBRATING DOSEMETERS ORAMED WP5 J.-M. Bordy Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France ORAMED 2011 January 2011.
Ma zhongjian Ding yadong Wang qingbin Wu qingbiao Radiation Protection Group/IHEP.
Legal Metrology in Europe: Harmonising requirements trough directives and the cooperation of authorities trough WELMEC by Knut Lindløv Director Legal.
Enhancement of Radiological Safety Culture within Ignalina NPP and Contractors Viktor PLETNIOV Ignalina Nuclear Power Plant Lithuania Gintautas KLEVINSKAS.
IAEA International Atomic Energy Agency Lecture 7 – Calibration and Testing of Radiation Protection Instruments Postgraduate Educational Course in radiation.
Technical Services. Objectives To identify the technical services needed within the infrastructure for an effective implementation of regulatory programme.
EC/IAEA/ISOE Workshop Portoroz Slovenia April 2002 Electronic Personal Dosimetry Presented by Andy Weeks.
Chapter 2 Radiation Interactions with Matter East China Institute of Technology School of Nuclear Engineering and Technology LIU Yi-Bao Wang Ling.
Radiation Protection RTMR 284 CHAPTER 20 PART II.
Training Module Preparation for “AERB norms, licenses and signage”
Louisiana State University Radiation Safety Office
From Microdosimetry to Nanodosimetry
Industrial Radiography
Austrian Research Centers Seibersdorf
Regulations Part II: Basic Concepts and Definitions
6th European ALARA Network Workshop Madrid, Spain, October 2002
AUBERT Bernard*, GUILABERT Nadine°, LAMON Annick* and RICARD Marcel*
Calibration of Weighing Instruments
Recommendations for IAEA Documentation for Industrial Radiography
Fundamentals of ISO.
Technologies and type examination
Risk Management with Minimum Weight
Polarized Positrons at Jefferson Lab
X Rays Medical Physics Notes.
CABINET X-RAY MACHINE OPERATOR RADIATION SAFETY TRAINING COURSE
GSC: Standardization Advancing Global Communications
The dangers of radioactivity
Radiation Protection Handbook
GSC: Standardization Advancing Global Communications
Hp(10) irradiations for personal dosimetry traceability
Radiation fields During 1st stage beam commissioning
Radiation Detectors for the PSS1 at ESS
Permit to Work Process​ HSE GROUP RULE (CR-GR-HSE-402)
Presentation transcript:

Basic requirements on area dosemeters Design and type test requirements Dr. Hayo Zutz

Content Introduction - IEC and PTB-requirements Requirements on area dosemeters - fundamental - dosimetric Existing problems - pulsed radiation 14.02.2017 Dr. Hayo Zutz

IEC and PTB requirements On the international level the requirements on area dosemeters are given by standards of the International Electrotechnical Commission IEC 60846-1:2009 Radiation protection instrumentation – Ambient and/or directional dose equivalent (rate) meters and/or monitors for beta, X and gamma radiation – Part 1: Portable workplace and environmental meters and monitors IEC ausschreiben 14.02.2017 Dr. Hayo Zutz

IEC and PTB requirements requirements on dosemeters reference radiation fields PTB-A sind national Österreich, Schweiz akzeptiert Internationales Qualitätssiegel active members convener of relevant WG active members convener of relevant WG National PTB type-test requirements: PTB-A 14.02.2017 Dr. Hayo Zutz

Situation in Germany Regulated by the German act concerning the placement and provision of measuring instruments on the market, their use and verification, and also on prepackages (Weights and Measures Act) Dosemeters which should by used for legally relevant measurements must have a conformity assessment type-testing by PTB examination of each single dosemeter 14.02.2017 Dr. Hayo Zutz

General IEC 60846-1:2009 covers dosemeter in the following range: The German PTB-A 23.3 are limited to photon radiation and the maximum energy range is 5 keV to 7 MeV as stated by the Measures and Verification Ordinance (MessEV) 14.02.2017 Dr. Hayo Zutz

General requirements Indication In units of dose equivalent (rate), e.g. mSv or mSv/h Changing of measurement range and read-out scale simultaneous Scales readable under normal lighting conditions Test for failure of display Alarm and alarm levels Overload indication Instruction manual Designation and inscriptions Type designation, Serial number measurand and measuring range 14.02.2017 Dr. Hayo Zutz

Software requirements The software of the dosemeter (firmware and/or PC) have to fulfill the WELMEC software guide 7.2 (risk class C). The requirements depend e.g. on: Type of system e.g. build only for this purpose or can be used for other purposes (PC,…) Storage of measurement data Transmission of measurement data Example of requirements: Software & calibration parameters secured against any changes (intended or unintended) 14.02.2017 Dr. Hayo Zutz

Metrological requirements All tests are type-tests: individual calibration of each dosemeter not considered only limits for the relative response: Calculate the response: Normalizing to response at reference condition: Compare relative response with limits 14.02.2017 Dr. Hayo Zutz

Metrological requirements Linearity & coefficient of variation PTB-A also allows for three orders of magnitude including 1 mSv and 100 mSv/h Beispiel von Dosimetern 14.02.2017 Dr. Hayo Zutz

Metrological requirements Linearity & coefficient of variation PTB-A also allows for three orders of magnitude including 1 mSv and 100 mSv/h Beispiel von Dosimetern 14.02.2017 Dr. Hayo Zutz

Metrological requirements Energy and angle dependence PTB-A 80 keV to 1,25 MeV or 20 keV to 150 keV 14.02.2017 Dr. Hayo Zutz

Metrological requirements Dose rate during dose measurement and overload PTB-A: Dose rate: Overload up to 50 times the range maximum or 10 Sv/h Dose: at least 1 Sv, else 50 times the range maximum or 10 Sv 14.02.2017 Dr. Hayo Zutz

Metrological requirements Response time τ Als Bild darstellen 14.02.2017 Dr. Hayo Zutz

Metrological requirements Response time τ Response time Als Bild darstellen 14.02.2017 Dr. Hayo Zutz

Metrological requirements Response time τ IEC 60846-1:2009 PTB-A: τ <10 s 𝐻 END < 10 mSv/h τ = 2 s 𝐻 END > 10 mSv/h Als Bild darstellen 60 s for 𝐻 END < 60 µSv/h 60 𝑠 − 𝐻 END −60 µSv h 940 µSv h ∙50 s for 60 µSv/h ≤ 𝐻 END < 1 mSv/h 10 s for 𝐻 END ≤ 1 mSv/h 14.02.2017 Dr. Hayo Zutz

EMC requirements Requirements on electromagnetic compatibility PTB-A Designed by Blossomstar / Freepik Bild: Klingeldes Smartphone 80 MHz to 2 GHz 0 to 30 V/m 14.02.2017 Dr. Hayo Zutz

EMC requirements Additional requirements in PTB-A on electromagnetic compatibility 14.02.2017 Dr. Hayo Zutz

Science and material testing Pulsed radiation Application of pulsed fields increased remarkably … can be produced as a pulsed field or can appear pulsed for the dosemeter “Pulsed”: duration less 10 s X-ray diagnostics Science and material testing X-ray scanner 14.02.2017 Dr. Hayo Zutz

Electronic dosemeters in pulsed fields Pulsed X-ray fields at PTB Let’s see what happens … EPD Mk2 DMC 2000 S Typical electronic personal dosemeters (results also valid for area dosemeters!) 14.02.2017 Dr. Hayo Zutz

Electronic dosemeters in pulsed fields Constant dose per pulse Direct beam in medical fields Type testing in continuous radiation Tests are necessary! Response cw 10 s 100 ms 2 ms Pulse duration Dr. Hayo Zutz 14.02.2017

Pulsed radiation Application of pulsed fields increased remarkably … can be produced as a pulsed field or can appear pulsed for the dosemeter “Pulsed”: duration less 10 s Dosemeters are only tested in continuous fields Characteristics of dosemeters determined in cont. fields can’t be transferred to those in pulsed fields Electronic dosemeters could measure considerably wrong, or even fail completely in pulsed fields 14.02.2017 Dr. Hayo Zutz

Pulsed radiation The reference fields for testing the dosemeter are described by: ISO/TS 18090-1:2015: “Radiological Protection – Characteristics of reference pulsed radiation – Part I Photon radiation” The testing procedure for dosemeters and the requirements are given in: IEC/TS 62743:2012: “Radiation Protection Instrumentation – Electronic counting dosemeters for pulsed fields of ionizing radiation” for counting dosemeter and IEC/TS 63050: “Radiation Protection Instrumentation – Electronic dosemeters for pulsed fields of ionizing” as a more generalized version (in preparation) 14.02.2017 Dr. Hayo Zutz

Pulsed radiation The reference fields for testing the dosemeter are described by: ISO/TS 18090-1:2015: “Radiological Protection – Characteristics of reference pulsed radiation – Part I Photon radiation” The testing procedure for dosemeters and the requirements are given in: IEC/TS 62743:2012: “Radiation Protection Instrumentation – Electronic counting dosemeters for pulsed fields of ionizing radiation” for counting dosemeter and IEC/TS 63050: “Radiation Protection Instrumentation – Electronic dosemeters for pulsed fields of ionizing” as a more generalized version (in preparation) 14.02.2017 Dr. Hayo Zutz

Pulsed radiation The testing procedure for dosemeters and the requirements are given in: IEC/TS 62743:2012: “Radiation Protection Instrumentation – Electronic counting dosemeters for pulsed fields of ionizing radiation” for counting dosemeter and IEC/TS 63050: “Radiation Protection Instrumentation – Electronic dosemeters for pulsed fields of ionizing” as a more generalized version (in preparation) The concept is similar to that used for other influence quantities. A 20 % influence of the pulsation is allowed with respect to continuous radiation and the parameter range is determined. No general “approved for pulsed radiation” 14.02.2017 Dr. Hayo Zutz

Summary IEC 60846-1 and PTB-A 23.3 include requirements on area dosemeters Indication Software Linearity and coefficient of variation Photon energy and angle dependence EMC Pulsed radiation is a problem for current dosemeters Can this be overcome by MEDIPIX/TIMEPIX? 14.02.2017 Dr. Hayo Zutz

Physikalisch-Technische Bundesanstalt Braunschweig and Berlin Bundesallee 100 38116 Braunschweig Dr. Hayo Zutz Telefon: +49 531 592-6341 E-Mail: hayo.zutz@ptb.de www.ptb.de Stand: 02/17