X-Ray Generators and Ancillary Radiographic Equipment

Slides:



Advertisements
Similar presentations
Producing an X-ray Exposure
Advertisements

Technique Guidance Systems
Chapter 11 Prime Factors.
CHAPTER 6 THE X-RAY BEAM SPECIFIC OBJECTIVES NOTED IN THIS POWER POINT BEGIN ON PAGE
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Rad Tech 110 Circuits.
Radiographic Quality Chapter 5.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
Resident Physics Lectures
The Grid Kyle Thornton DMI 50B.
Analyzing the Image Density. Density Overall blackening of the image.
Chapter 8 The X-ray Imaging System
The X-Ray Circuit.
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
TIMERS.
Basic quality control tests in medical radiography
X-Ray Production & Emission
Exposure Timers By Prof. Stelmark.
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
The X-ray Imaging System
X-ray Generators and Transformers
X-Ray Production & Emission
Conventional and Computed Tomography
Quality Control Rad T 110.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Quality Control.
Ch. 2 – Anatomy of the X-ray Machine
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
CHAPTER 3 EQUIPMENT OPERATION AND QUALITY CONTROL
Rad T 290 Generators. Generator Components control console  kVp adjust  mA adjust  time adjust transformer  high voltage (step up)  filament »low.
Equipment Operation and Maintenance. Grid-controlled tube Designed to turn on and off rapidly, providing short precise exposures Use in fluoro and in.
RAD 254 Chapter 15 Radiographic Technique 4 primary exposure factors are: kVp, mA, time, distance.
CT Instrumentation and X-ray system
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Radiographic Quality Visibility and Sharpness
Radiographic Equipment
RAD 354 Chap. 13 Radiographic Technique
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Part No...., Module No....Lesson No
Part No...., Module No....Lesson No
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
RAD 354 Chapt. 16 Image QUALITY
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Chapter 18 Quality Control
Optimizing The Image RVT: Chapter 7
Stacy Kopso, M.Ed.,RT(R)(M). Quality Control (QC)-A program that specifically addresses the safe and reliable operation of equipment Required by the Joint.
Technique Guidance Systems By Prof. Stelmark. Anatomic Programming Anatomic programming, or anatomically programmed radiography (APR), refers to a radiographic.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Radiation Protection RTMR 284 CHAPTER 20 PART II.
Diagnostic Equipment Quality Control
Ch 36 Radiation Protection Design
Chapters 21 & 22.
Introduction to Radiography
Resident Physics Lectures (year 1)
X-RAY PRODUCTION AND EXPOSURE FACTORS
Rad Tech 110 Circuits.
Factors Effecting the Production
Exposure Factors Chapter 4
Quality Control Testing of Screen Speed
Equipment Operation and Maintenance
Resident Physics Lectures (year 1)
Artifacts and Errors.
Generation of X-rays Q&A
Presentation transcript:

X-Ray Generators and Ancillary Radiographic Equipment Chapter 7 Quality Control of X-Ray Generators and Ancillary Radiographic Equipment

X-ray Generators Single Phase Generator – uses single phase of alternating current

X-ray Generators (cont’d) Single Phase Generator – half-wave rectified

X-ray Generators (cont’d) Single Phase Generator – full-wave rectified

X-ray Generators (cont’d) Three Phase Generators – utilize 3 sources of alternating current

X-ray Generators (cont’d) Three Phase Generators – 3-phase, 6-pulse

X-ray Generators (cont’d) Three Phase Generators – 3-phase, 12-pulse

X-ray Generators (cont’d) High Frequency Generator – increases frequency of incoming alternating current

X-ray Generators (cont’d) Three main components: Control or Operating Console High Voltage Generator X-ray Tube, Tube Accessories and X-ray Table

Quality Control for Radiographic Units Three Components: Visual Inspection Environmental Inspection Performance Testing

Radiation Measurement Most performance testing performed with gas-filled chamber detectors Three different type of these detectors: Ion Chamber – most often used Proportional counter Geiger-Muller Counter

Performance Testing X-ray Generator Reproducibility or Exposure Same technical factors should always produce the same radiation output Testing requires measuring output with detector at same technical factors Any variation found during testing cannot exceed ± 5%

Performance Testing X-ray Generator (cont’d) Radiation Output Measures amount of radiation exposure per mAs of technique Units are either mR/mAs or μCkg-1/mAs Determined using radiation detector and making exposure at 10 mAs Current and previous values are compared All values should remain within ± 10%

Performance Testing X-ray Generator (cont’d) Radiation Quality and Filtration Measured by determining Half Value Layer (HVL) Determined by using a radiation detector and making exposures at 80 kVp adding aluminum filters in 1 mm increments Minimum HVL at 80 kVp is 2.3 mm aluminum equivalent

Performance Testing X-ray Generator (cont’d) Kilovolt peak (kVp) Accuracy Can be determined by using a special test cassette or a digital kVp meter Determines whether or not the kVp value on the control console corresponds with the maximum energy of the actual x-ray beam Variations between the stated kVp and the actual x-ray beam quality must be within ± 5%

Performance Testing X-ray Generator (cont’d) Timer Accuracy Testing can be performed by using either the spinning top test or a digital timer The variability allowed for timer accuracy is ± 5% for exposure times greater than 10 milliseconds (ms) and ± 20% for exposures less than 10 milliseconds (ms)

Performance Testing X-ray Generator (cont’d) Voltage Waveform Visualizes voltage waveform through x-ray tube

Performance Testing X-ray Generator (cont’d) Reciprocity The same mAs should produce the same radiation output regardless of mA and time selected Testing utilizes a radiation detector and exposures using same mAs but different mA and time stations Any variation between exposures cannot exceed ± 10%

Performance Testing X-ray Generator (cont’d) Linearity Any sequential change in mAs should produce the same sequential change in radiation output Testing involves making exposures of a radiation detector at mAs values that are double that of previous exposure Any variation between exposure cannot exceed ± 10%

Performance Testing X-ray Tube and Tube Accessories Focal Spot Size Evaluated to determine if focal spot blooming has occurred Level of blooming must not exceed NEMA standards

Performance Testing X-ray Tube and Tube Accessories (cont’d) Focal Spot Size Testing can be performed by one of three devices: Pinhole camera Focal Spot Test Tool Star or slit resolution patterns

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Restriction System Light Field-Radiation Field Congruence Assures that light field of collimator is the same area struck by x-rays The light field must be congruent to within ± 2% of the SID Evaluation can be performed by the use of either a collimator test tool or 8-penny test

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Restriction System Image Receptor-Radiation Field Alignment Determined with collimators equipped with Positive Beam Limitation (PBL) Variation between the x-ray field and image receptor size must be within ± 3% of the SID

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Restriction System Accuracy of the X-Y Scales Size of field selected by collimator adjustment knobs must correspond to the actual field size to within ± 2% of SID Can be tested with collimator test tool or 8-penny test

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Restriction System Illuminator Bulb Brightness Determined using a photometer placed tabletop at a 100 cm SID The illumination of the light source must be at least 15 foot-candles (ft-cd) or 160 lux when measured at a 100-cm (40”) distance

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Alignment X-ray beam must be mounted properly in its metal housing and aligned to the center of the bucky tray Items to evaluate include perpendicularity and x-ray beam/bucky tray alignment

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Alignment Perpendicularity The x-ray tube must be mounted in its housing so that the x-ray beam is within 1 degree of perpendicular Testing is performed with a commercially available beam alignment tool

Performance Testing X-ray Tube and Tube Accessories (cont’d) Beam Alignment X-ray beam/Bucky tray alignment Also known as central ray congruency The x-ray beam/Bucky tray alignment must be within ± 1% of SID Testing can be performed using either a beam alignment tool or the washer or coin method

Performance Testing X-ray Tube and Tube Accessories (cont’d) SID Indicator The SID indicator must be installed so that it is accurate to within ± 2% of the SID Testing can be performed with a tape measure or the triangular method

Performance Testing Overload Protection The overload protection mechanism should not permit an exposure that exceeds 80% of the tube capacity for a single exposure Testing involves selecting technical factors that would exceed limit and then depress rotor/prep button on control panel If rotor engages and ready light appears, system fails

Performance Testing (cont’d) X-ray Tube Heat Sensors Usually found on 3-phase equipment Heat sensors should provide a warning when anode heat reaches 75% of the maximum

Automatic Exposure Control Consists of two main components: Detectors – also known as sensors, cells or chambers. Actual devices can be either photodetectors, ion chambers or solid state detectors Comparator – receives signal sent by detectors and terminates exposure when a specific amount of radiation has been measured

Automatic Exposure Control Testing (cont’d) Backup, or Maximum Exposure Time The backup timer should terminate the exposure within 6 seconds or 600 mAs, whichever comes first Testing involves placing a lead apron tabletop over sensors and seeing if exposure terminates within specified values

Automatic Exposure Control Testing (cont’d) Consistency of Exposure with Varying mA System should maintain exposure level with any changes in mA Testing involves making a series of exposures of a homogenous phantom at different mA stations. Images should all be the same. Any variation cannot exceed ± 10%

Automatic Exposure Control Testing (cont’d) Consistency of Exposure with Varying kVp System should maintain exposure with any changes in kVp Testing involves making a series of exposures of a homogenous phantom at different kVp values All images should be similar

Automatic Exposure Control Testing (cont’d) Consistency of Exposure with Varying Part Thickness System should be able to compensate for changes in part thickness Testing involves a series of exposures of a homogenous phantom of varying thickness values Images should be similar

Automatic Exposure Control Testing (cont’d) Consistency of Exposure with Varying Field Sizes System should be able to compensate for any changes in the x-ray field size Testing involves a series of exposures of a homogenous phantom at different field sizes Images should be similar

Automatic Exposure Control Testing (cont’d) Consistency of AEC Detectors Each detector or sensor should function identically to all other detectors Testing involves exposures of a homogenous phantom at each of the sensors available on unit Images should be similar

Automatic Exposure Control Testing (cont’d) Reproducibility Exposures made at the same kVp and mA stations of the same phantom thickness should produce the same exposure Testing involves a series of exposures of a homogenous phantom at the same kVp and mAs Images should be similar

Automatic Exposure Control Testing (cont’d) Density Control Function Density selector should allow for changes in radiation exposure of 25% to 30% for each increment Testing involves a series of exposures of a homogenous phantom at each of the density selector options Each setting should produce an image 25% to 30% greater than the previous setting

Conventional Tomographic Systems Used to create “slices” of the body to remove superimposition and improve contrast in area of interest Two basic types are linear (sometimes called rectilinear) and pluridirectional (sometimes called multidirectional)

Quality Control of Tomographic Systems (cont’d) Section Level Requires specialized test tool The level of the section indicated on the equipment and the actual level should be the same or within ± 5 mm (some manufacturers suggest ± 1 mm for pluridirectional units)

Quality Control of Tomographic Systems (cont’d) Section Thickness Evaluation is made with special test tool

Quality Control of Tomographic Systems (cont’d) Level Incrementation Requires testing with specialized test tool A ruler should be constructed so that changing from one tomographic section to the next is accurate to within ± 2 mm

Quality Control of Tomographic Systems (cont’d) Exposure Angle Testing involves specialized test tool The value indicated on the equipment and the actual angle are the same or within ± 5 degrees for units operating at angles greater than 30 degrees and within ± 2 degrees for units operating less than 30 degrees

Quality Control of Tomographic Systems (cont’d) Spatial Resolution Testing involves imaging a test tool with a mesh pattern containing different size holes Units should be able to resolve at least 40 holes per inch

Quality Control of Tomographic Systems (cont’d) Section Uniformity and Beam Path Radiation output should be consistent throughout exposure Any variation for uniformity should be within optical density value of ± 0.3 Path closure in pluridirectional units should be within ± 10% of the path length for beam path

Quality Control of Tomographic Systems (cont’d) Patient Exposure Patient exposure for tomographic image should not exceed two times the non-tomographic exposure for the same part in the same position

Grid Testing Grid Uniformity Lead strips in grid must be uniformly spaced Image of homogenous phantom is made and optical density of the center and center of each quadrant is made Readings should be within ± 0.10 for proper uniformity

Grid Testing (cont’d) Grid Alignment Center of grid must be aligned with the detent lock position of the x-ray tube Testing requires imaging a special test tool

Portable and Mobile X-ray Generators Portable Generator – small enough to be carried by one person Mobile Generator – pushed on wheels Direct Power Unit Capacitor Discharge Cordless High Frequency